K203800 · Syntr Health Technologies, Inc. · MUU · Jul 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203800
Device Name
SyntrFuge System
Applicant
Syntr Health Technologies, Inc.
Product Code
MUU · General, Plastic Surgery
Decision Date
Jul 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SyntrFuge™ System is a sterile medical device used in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal, urological surgery, general surgery, gynecological surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
Device Story
SyntrFuge System is a sterile, single-use disposable device for concentrating lipoaspirate. Input: autologous adipose tissue harvested via standard lipoplasty system. Operation: system uses AdipoSets (AdipoChambers and AdipoChip) in a centrifuge (SyntrFPU 360) to concentrate tissue; uses sterile saline for washing. Output: concentrated adipose tissue for reinjection. Used in sterile surgical settings by surgeons. Benefits: enables autologous fat transfer for body contouring or reconstructive needs; maintains cell viability.
Clinical Evidence
Bench testing only. Verification testing included design verification, cleaning validation, cell viability, distribution simulation, packaging integrity, usability, and sterilization/shelf-life. Biocompatibility testing per ISO 10993-1:2018 confirmed device is non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and lacks systemic toxicity. Mean nucleated cell viability of 87.8% demonstrated.
Technological Characteristics
Sterile, single-use disposable system. Components: AdipoSets (AdipoChambers and AdipoChip) made of medical-grade resin. Processing principle: centrifugation (using SyntrFPU 360). Processing volume: 5mL to 15mL. Sterilization: Ethylene-Oxide (EO) to SAL 10^-6.
Indications for Use
Indicated for patients undergoing surgical procedures requiring autologous adipose tissue transfer, including orthopedic, arthroscopic, neurosurgical, gastrointestinal, urological, general, gynecological, laparoscopic, and plastic/reconstructive surgery. Contraindicated for use with manipulated fat.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 2, 2021
Syntr Health Technologies, Inc. % Nevine Erian Regulatory Consultant BQC Consulting, LLC 24341 Barbados Dr. Dana Point, California 92629
Re: K203800
Trade/Device Name: SyntrFuge System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: May 11, 2021 Received: May 13, 2021
Dear Nevine Erian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K203800
Device Name
#### SyntrFuge™ System
Indications for Use (Describe)
The SyntrFuge™ System is a sterile medical device used in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system.
The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal, urological surgery, general surgery, gynecological surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired.
Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <b>X</b> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------|---------------------------------------------|
|-------------------------------------------------------|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203800 – 510(k) Summary
| Submitter | Syntr Health Technologies, Inc.<br>5270 California Ave., Suite #300<br>Irvine, CA 92617, U.S.<br>Establishment Reg. No. – 3017791237 | | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|--|
| Contact | Ahmed Zobi<br>Chief Executive Officer<br>Phone 949 732-4298<br>E-Mail azobi@syntrtech.com | | |
| Official Correspondent | Nevine Erian<br>Regulatory Consultant<br>BQC Consulting, LLC<br>24341 Barbados Dr.<br>Dana Point, CA 92629<br>Phone (949) 370-7155<br>Fax (949) 240-8829<br>E-mail nevine@bqcconsulting.com | | |
| Date Prepared | June 27, 2021 | | |
| ■ | Trade/Device Name | SyntrFuge™ System | |
| ■ | Common Name | Suction Lipoplasty System | |
| ■ | Classification Name | System, Suction, Lipoplasty | |
| ■ | Regulation Number | 21 CFR 878.5040 | |
| ■ | Product Code | MUU | |
## Predicate Devices
- AdiPrep™ Adipose Transfer System (Harvest Technologies Corp.) K121005— Predicate
{4}------------------------------------------------
# Device Description
The SyntrFuge System is a sterile single use disposable medical device, comprised of three AdipoSets. Each AdipoSet is comprised of two AdipoChambers and one AdipoChip, two AdipoChambers attach to the AdipoChip by luer connectors.
The SyntrFuge™ System concentrates lipoaspirate, following a syringe vacuum liposuction method.
The SyntrFuge System is utilized in medical procedures where reinjection of adipose tissue into the same individual during the same surgical procedure is needed.
## Indication for Use
The SyntrFuge™ System is a sterile medical device used in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system.
The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired.
Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
# Safety and Performance Testing
## Biocompatibility Testing
The SyntrFuge System was tested and meets the biocompatibility requirements of ISO 10993-1:2018 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process.
Test results demonstrated that the SyntrFuge System is non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic and has no systemic toxicity effects.
#### Performance Testing
The SyntrFuge™ System has undergone verification testing to ensure it meets product specification requirements. Verification testing consisted of the following:
- 트 SyntrFuge System Design Verification Testing
- Cleaning Validation
- . Cell Viability
{5}------------------------------------------------
- 트 Distribution Simulation and Packaging Integrity Testing
- . Usability Testing
- . Sterilization Validation and Shelf-Life Testing
All testing met specifications. Bench test results allowed us to conclude that the SyntrFuge System meets its intended use.
## Substantial Equivalence
The SyntrFuge System is substantially equivalent to the predicate device.
### Substantial Equivalence Comparison of the SyntrFuge System to the Predicate Device
| Attribute | SyntrFuge™ System | AdiPrep™ Adipose<br>Transfer System<br>(Predicate) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification & Regulation<br>Number | System, Suction,<br>Lipoplasty<br>21 CFR 878.5040 | System, Suction,<br>Lipoplasty<br>21 CFR 878.5040 |
| Product Code | MUU | MUU |
| Indications for use | | |
| Used in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. | Yes | Yes |
| Intended for use in the following surgical specialities when concentration of harvested tissue is desired.<br>Orthopedic surgery Arthroscopic surgery Neurosurgery Gastrointestinal surgery Urological surgery General surgery Gynecological surgery Thoracic surgery Laparoscopic surgery Plastic and reconstructive surgery | Yes | Yes |
| Attribute | SyntrFuge™ System | AdiPrep™ Adipose<br>Transfer System<br>(Predicate) |
| Technological Attribute | | |
| Processing Pack Components | 3 AdipoSets, comprised<br>each of an AdipoChip and<br>2 AdipoChambers | Harvest AdiPrep Process<br>Disposable unit,<br>centrifuge tubes with<br>filter, aspiration & fat<br>injection cannula, fat<br>injection syringes, skin<br>puncture needles, and<br>oil extraction syringe &<br>needle. |
| Material Type | AdipoChip &<br>AdipoChambers<br>composed of medical<br>grade resin. | Cannula & syringes<br>composed of medical<br>grade plastics. |
| Sterilization Method | Ethylene-Oxide (EO) | Ethylene-Oxide (EO) |
| Sterility Assurance | SAL 10-6 | SAL 10-6 |
| Design | AdipoSet/Chamber | Syringe |
| Mode of Operation | Centrifugation | Centrifugation |
| Processing Volume Capacity | 5mL to 15mL | 5 mL to 25 mL |
| Uses Laboratory Centrifuge<br>System | Yes - SyntrFPU 360 | Yes - SmartPrep2 |
| Disposable or Reusable | Single Use, Disposable | Single Use, Disposable |
| Procedural Attribute | | |
| Harvesting Method | Cannula | Cannula |
| Tissue Washing Media | Sterile Saline | Sterile Saline |
| Processing Output | Concentrated Adipose<br>Tissue | Concentrated Adipose<br>Tissue |
| Nucleated Cell Viability (%) | Mean = 87.8% | Mean = 83.5% |
| For Adipose Transfer | Yes | Yes |
| Environment of Use | Sterile Setting | Sterile Setting |
{6}------------------------------------------------
The differences in product configuration between the SyntrFuge System and the AdiPrep Adipose Transfer System do not impact safety and effectiveness, as the SyntrFuge System
{7}------------------------------------------------
shares the same indications and achieves the same fat concentration as the predicate device.
## Conclusion
Information provided in this application demonstrates that the SyntrFuge System is substantially equivalent to the predicate device. The SyntrFuge System has same indications for use and similar technological characteristics as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.