K203786 · Ge Medical Systems Information Technologies, Inc. · DPS · Apr 30, 2021 · Cardiovascular
Device Facts
Record ID
K203786
Device Name
MAC 7 - Resting ECG Analysis System
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Apr 30, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MAC 7 Resting ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information from adult or pediatric populations. Basic system delivers 3, 6, or 12 lead ECG's and interpretive analysis. Transmission and reception of ECG data and other clinical data to and from a central clinical information system is optional. The MAC 7 Resting ECG Analysis System is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital, medical professional's facility or wherever ECG testing is performed.
Device Story
Mobile electrocardiograph; acquires 3, 6, or 12-lead ECG signals via surface electrodes; performs interpretive analysis; displays and prints reports. Used in clinical settings (hospitals, medical facilities) by physicians or trained personnel. Optional trolley for mobility. Connects to network via wired LAN or WiFi; interfaces with cardiology information systems (e.g., GEHC MUSE) for order/demographic download and report transmission. Provides data for physician review; supports clinical decision-making by providing automated interpretation; requires physician overread. Benefits include efficient ECG acquisition, standardized analysis, and integration with clinical information systems.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including risk analysis, design reviews, software development lifecycle, unit-level testing, usability testing, connectivity bench testing, integration testing, and safety/performance verification.
Technological Characteristics
Mobile electrocardiograph; 3, 6, or 12-lead acquisition; wired LAN/WiFi connectivity; optional trolley. Complies with IEC 60601-1 Ed. 3.1. Software-based interpretive analysis. No sterilization required.
Indications for Use
Indicated for adult and pediatric (birth through 21 years) patients to acquire, analyze, display, and print ECGs. Provides data interpretation for physician consideration. Not intended as sole diagnostic means; requires physician overread and consideration of other patient data. Used in clinical settings by physicians or trained personnel under physician orders.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 30, 2021
GE Medical Systems Information Technologies, Inc. Lee Bush Director Regulatory Affairs 9900 Innovation Drive Wauwatosa, Wisconsin 53226
Re: K203786
Trade/Device Name: MAC 7 - Resting ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK, DXH Dated: April 2, 2021 Received: April 7, 2021
Dear Lee Bush:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K203786
Device Name
MAC 7 Resting ECG Analysis System
#### Indications for Use (Describe)
The MAC 7 Resting ECG Analysis System is a non-invasive prescription device.
· The device is indicated for use to acquire, analyze, display and print electrocardiograms.
- · The device is indicated for use to provide interpretation of the data for consideration by a physician.
· The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.
· It is not intended as a sole means of diagnosis. The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data.
· The device is indicated for use on adult and pediatric (birth through 21 years of age) populations
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
***_****_* and a lot of lines
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510k Summary
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
DATE: December 22, 2020
### SUBMITTER
GE MEDICAL SYSTEMS Information Technologies, Inc. 9900 Innovation Drive Milwaukee County Research Park Wauwatosa, Wisconsin 53226-4856
Primary Contact Person:
Lee Bush Regulatory Affairs Manager GE Medical Systems Information Technologies, Inc. 9900 Innovation Drive Milwaukee County Research Park Wauwatosa, Wisconsin 53226-4856 Phone: +1 (262) 309-9429 Fax: +1 (414) 721-3864 E-mail: Lee.Bush(@ge.com
Secondary Contact Person:
MANJUNATHA K N Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. Telephone: +91 80 40887193 Fax at +919901233442 E-mail:Manjunatha.kn(@med.ge.com
## PRODUCT IDENTIFICATION
| Device Trade Name: | MAC 7 Resting ECG Analysis System |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name: | Electrocardiograph |
| Regulation Number -<br>Classification Name(s) | 21 CFR 870.2340 - Electrocardiograph<br>21 CFR 870.1425 – Programmable Diagnostic Computer<br>21 CFR 870.2920 – Telephone electrocardiograph<br>Transmitter and Receiver |
| Device Classification: | Class II |
| Product Code(s): | DPS, DQK, DXH |
| Predicate Device(s): | MAC VU360 Resting ECG Analysis System (K173830) |
| Device Description: | The MAC 7 Resting ECG Analysis System is a mobile<br>electrocardiograph designed to acquire, analyze, display, and<br>record ECG signals from surface ECG electrodes. |
| | The MAC 7 can capture 3, 6 or 12 lead electrocardiograms,<br>provide interpretive analysis, and print reports. |
| | The device can connect to a network, either through a wired<br>LAN connection or via wireless WiFi access points. Once on<br>the network, the device can optionally interface with the<br>cardiology information systems such as the GEHC MUSE®<br> |
| | The device provides state-of-the-art information technology<br>security features and a contemporary user interface.<br>Mobility is provided via an optional trolley. |
| Intended Use: | The MAC 7 Resting ECG Analysis System is intended to<br>acquire, analyze, display, and record electrocardiographic<br>information from adult or pediatric populations. Basic<br>system delivers 3, 6, or 12 lead ECG's and interpretive<br>analysis. Transmission and reception of ECG data and other<br>clinical data to and from a central clinical information<br>system is optional. |
| | The MAC 7 Resting ECG Analysis System is intended to be<br>used under the direct supervision of a licensed healthcare<br>practitioner, by trained operators in a hospital, medical<br>professional's facility or wherever ECG testing is performed. |
| Indications for Use | The MAC 7 Resting ECG Analysis System is a non-invasive<br>prescription device.<br>• The device is indicated for use to acquire, analyze, display<br>and print electrocardiograms.<br>• The device is indicated for use to provide interpretation<br>of the data for consideration by a physician.<br>• The device is indicated for use in a clinical setting, by a<br>physician or by trained personnel who are acting on the<br>orders of a licensed physician. It is not intended as a sole<br>means of diagnosis |
{4}------------------------------------------------
{5}------------------------------------------------
| | <ul><li>The interpretations of ECG offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.</li></ul> |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology: | The MAC 7 employs the same fundamental scientific technology, basic design, construction, materials, energy source, control mechanism, operating principles as the predicate device MAC VU360 in acquiring, analyzing, recording, displaying and printing ECG data for both adult and pediatric populations. |
| | The basic system prints 3, 6 or 12 leads of ECG and provides optional transmission and reception of ECG data to and from a central ECG cardiovascular information system. The system can be upgraded with software options, similar to the predicate device. |
| | The MAC 7 is similar to the MAC VU360 Resting ECG Analysis System, K173830, in the technology of downloading orders and patient demographics from a central ECG cardiovascular information system (e.g. MUSE) as well as supporting ECG reports in PDF. Both are able to use WiFi communication. |
| Performance Standards: | The MAC 7 Resting ECG Analysis System has completed testing and demonstrated compliance with IEC 60601-1 Ed. 3.1 and it associated collateral and particular standards. |
| Sterilization: | The MAC 7 Resting ECG Analysis System does not require sterilization. |
| Determination of Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The MAC 7 Resting ECG Analysis System and its applications comply with applicable voluntary standards. It was designed and manufactured under the Quality System Regulations of 21CFR820 and ISO 13485. The following quality assurance measures were applied to the development of the system: <ul><li>Risk Analysis</li><li>Required Reviews</li><li>Design Reviews</li><li>Software Development Lifecycle</li><li>Testing on unit level</li><li>Usability Testing</li><li>Connectivity Bench Testing</li></ul> |
- Connectivity Bench Testing
{6}------------------------------------------------
- Integration testing (System verification)
- Performance testing (Verification)
- Safety testing (Verification)
- Summary of Clinical Tests: The subject of this premarket submission, MAC 7 Resting ECG Analysis System, did not require clinical studies to support substantial equivalence.
- GE Healthcare considers the MAC 7 Resting ECG Analysis Conclusion: System to be as safe, and as effective, and performance is substantially equivalent to the predicate device, MAC VU360 Resting ECG Analysis System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.