The product is intended as an X-Ray source for diagnosis. Operators of the Nano-x Cart X-Ray System are healthcare professionals familiar with and responsible for the X-Ray examinations being performed. Indication for Use – The device is designed to perform radiographic X-Ray examinations of hands, wrists, and fingers, on adult patients. Limitations for use – This device is not intended for general radiographic X-Ray examinations other than the indicated use, or for mammographic, angiographic, interventional, or fluoroscopic applications. This device is not intended for pediatric use.
Device Story
Mobile X-ray system for point-of-care radiographic imaging of hands, wrists, and fingers. System comprises mobile enclosure, cold cathode X-ray tube, pulse generator, high-voltage generator, and touchscreen interface. Operator selects energy parameters (40kVp, 0.1–2mAs) via touchscreen; microprocessor manages exposure consistency. Device operates with specific Flat Panel Digital X-ray Detector (EVS3643). Used by trained healthcare professionals in clinical settings where patient transport to fixed imaging suites is impractical. Output is radiographic image data captured by external detector. System design includes emergency stop, isolation transformer, and cooling fluid for tube assembly. Benefits include portable diagnostic capability for extremity imaging.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing and phantom imaging studies. Phantom study used anthropomorphic hand phantoms (Kyoto Kagaku) to compare output against predicate. Three experts evaluated images, concluding Nanox Cart output is equivalent to predicate. Bench testing verified electrical safety, electromagnetic compatibility, and radiation protection per IEC 60601 series standards.
Technological Characteristics
Mobile X-ray system; cold cathode X-ray tube; 40kVp fixed voltage; 0.1-2mAs variable charge; 0.3mm focal spot. Components: mobile cart, pulse generator, microprocessor, touchscreen LCD. Power: 240VAC/50Hz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 62304. Connectivity: Operates with external EVS3643 digital detector.
Indications for Use
Indicated for radiographic X-ray examinations of hands, wrists, and fingers in adult patients. Not for pediatric use, general radiography, mammography, angiography, interventional, or fluoroscopic applications.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Flat Panel Digital X-ray Detector Model EVS3643 (K162552)
Submission Summary (Full Text)
{0}------------------------------------------------
April 1, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nano-x Imaging Ltd. % Rafael Aguila Responsible Third-Party Official Accelerated Device Approval Services 6800 S.W. 40th Street. Ste. 403 LUDLUM FL 33155
Re: K203782
Trade/Device Name: Nanox Cart X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: March 1, 2021 Received: March 24, 2021
Dear Rafael Aguila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203782
Device Name Nanox Cart X-Ray System
#### Indications for Use (Describe)
The product is intended as an X-Ray source for diagnosis. Operators of the Nano-x Cart X-Ray System are healtheare professionals familiar with and responsible for the X-Ray examinations being performed.
Indication for Use – The device is designed to perform radiographic X-Ray examinations of hands, wrists, and fingers, on adult patients.
Limitations for use – This device is not intended for general radiographic X-Ray examinations other than the indicated use, or for mammographic, angiographic, interventional, or fluoroscopic applications. This device is not intended for pediatric use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | Submission File | |
|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------|
| | Nano-x Imaging Ltd.<br>Neve Ilan Communication Center, Zip 9085000, Israel<br>Telephone: +972-2-574-6300 | |
| | 510(k) Summary | K203782 |
| Applicant: | Nano-x Imaging Ltd.<br>Neve Ilan Communications Center<br>Jerusalem District<br>Israel, 9085000<br>Phone: +972 (2) 574-6300 | |
| Contact Person: | Yuval Shapiro<br>QA/RA<br>Nano-x Imaging Ltd.<br>Phone: +972 (54) 205-2524<br>Email: yuval.s@Nanox.vision | |
| Date Prepared: | Mar. 1, 2021 | |
| Device Trade Name:<br>Device Common Name:<br>Classification Name:<br>Regulation Number:<br>Regulation Class:<br>Product Code: | Nanox Cart X-Ray System<br>Mobile X-Ray Device<br>Mobile X-Ray system<br>21 CFR 892.1720<br>II<br>IZL | |
| Predicate Device:<br>Manufacturer:<br>510(k) Number:<br>Regulation Name:<br>Number:<br>Regulatory Class:<br>Product Code: | AMX-4 Mobile X-Ray System<br>GE Medical Systems<br>K021016<br>Mobile x-ray system<br>21 CFR 892.1720<br>II<br>IZL | |
| Reference Device:<br>Manufacturer:<br>510(k) Number:<br>Regulation Name:<br>Number:<br>Regulatory Class:<br>Product Code: | DRX-Revolution Nano Mobile X-ray System<br>Carestream Health Inc.<br>K173924<br>Mobile x-ray system<br>21 CFR 892.1720<br>II<br>IZL | |
## Nanox Cart X-Ray System – 510(k) Submission File
Page 1 of 8
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a series of six distinct symbols or glyphs arranged horizontally. Each symbol is rendered in black against a white background, creating a stark contrast. The symbols appear to be abstract and geometric in nature, with varying shapes and orientations. The symbols include shapes resembling a horseshoe, an upside down V, and a square.
### Nanox Cart X-Ray System - 510(k) Submission File
Nano-x Imaging Ltd. Neve Ilan Communication Center, Zip 9085000, Israel Telephone: +972-2-574-6300
#### Device Description:
The Nanox Cart X-Ray system is a mobile X-Ray system designed to perform radiographic X-Ray examinations of hands, wrists, and fingers, on adult patients. The system facilitates X-ray examinations in situations where it is not possible to transport the patient to a ward with fixed equipment.
The Nanox Cart X-Ray system consists of a mobile system enclosure; an X-Ray tube assembly consists of the X-ray Tube, Cooling Fluid, electronics, and casing; a pulse generator and a highvoltage generator; a microprocessor that provides increased exposure consistency and efficient operation; and a touch-based screen LCD which provides a simple and user-friendly interface and technique selection.
The Nanox Cart is specified and designed to operate only with a Flat Panel Digital X-ray Detector Model EVS3643, manufactured by DRTECH Inc. (K162552)
The intended operators of the Nanox Cart X-ray System are healthcare professionals familiar with and responsible for the X-ray examinations being performed.
To minimize electrical, mechanical, and radiation hazards, the Nanox Cart X-ray System adheres to recognized and established industry practices and standards.
### Intended Use/Indications for Use:
The product is intended as an X-Ray source for diagnosis. Operators of the Nano-x Cart X-Ray System are healthcare professionals familiar with and responsible for the X-Ray examinations being performed.
Indication for Use – The device is designed to perform radiographic X-Ray examinations of hands, wrists, and fingers, on adult patients.
Limitations for use – This device is not intended for general radiographic X-Ray examinations other than the indicated use, or for mammographic, interventional, or fluoroscopic applications. This device is not intended for pediatric use.
#### Summary of Technology Characteristics:
Table 1 provides a comparison of the technological characteristics for the Nanox Cart X-Ray System and the predicate device.
Table 2 provides a comparison of the technological characteristics for the Nanox Cart X-Ray System and the reference device.
{5}------------------------------------------------
## UANOX
### Nanox Cart X-Ray System - 510(k) Submission File
Nano-x Imaging Ltd. Neve Ilan Communication Center, Zip 9085000, Israel Telephone: +972-2-574-6300
### Table 1: Comparison of Technological Characteristics
| | Nanox Cart X-Ray System<br>(K203782) | AMX-4 Mobile X-Ray System<br>(K021016) | Significant<br>Difference |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Regulatory | | | |
| 510(k) | K203782 | K021016¹ | |
| Device class | II | II | Same |
| Classification | Mobile x-ray system, Class II;<br>21 CFR 892.1720 | Mobile x-ray system, Class II; 21 CFR 892.1720 | Same |
| Product code | IZL | IZL | Same |
| Indications /<br>Intended use | The device is designed to<br>perform radiographic X-Ray<br>examinations of hands, wrists,<br>and fingers, on adult patients. | The AMX-4 Plus Mobile X-ray<br>System is indicated for use in<br>generating radiographic images of<br>human anatomy in all general-<br>purpose X-ray diagnostic procedures.<br>It may be used in radiology<br>departments, emergency rooms,<br>intensive care units, operating rooms,<br>pediatrics, orthopedics, and clinics. | Similar |
| Intended<br>users | X-ray Technicians trained to<br>operate the system | X-ray Technicians trained to operate<br>the system | Same |
| Physical Characteristics | | | |
| System<br>Components | The Nanox Cart X-Ray System<br>is comprised of the following<br>components:<br><br>1) Cart – The mobile<br>enclosure of the system.<br>2) Wheels and Breaks – Set<br>of 4 (four) wheels and<br>integral breaks that are<br>placed to enable the<br>mobility of the cart and<br>have a stationary<br>positioning during<br>operation (cart will not be<br>operated during movement).<br>3) On/Off Switch – A switch<br>that switches the system on.<br>4) X-Ray Stop (Emergency | The AMX-4mobile X-ray system is<br>comprised of the following<br>components:<br><br>1) Cart<br>2) Collimator<br>3) Column<br>4) Power Connectors<br>5) Rotating Arm<br>6) Docking tube and docking<br>receiver<br>7) E-Stop (Emergency Stop)<br>knobs<br>8) Grid alignment system<br>9) PREP/EXP switch<br>10) Screen<br>11) Tube head assembly<br>12) X-Ray Tube | Similar |
| | Nanox Cart X-Ray System<br>(K203782) | AMX-4 Mobile X-Ray System<br>(K021016) | Significant<br>Difference |
| | that prevents the system<br>from operating, should the<br>operator determine to halt<br>procedure.<br>5) X-Ray Switch – A switch<br>that enables the X-ray<br>radiation energy.<br>6) Display - Operators display<br>of energy emitted by the<br>Nanox Cart. The display is<br>a touchscreen that enables<br>adjusting the energy pulse<br>(touchscreen) from 0.1mAs<br>up to 2mAs in steps of<br>0.1mAs.<br>7) Controller – the Display<br>and a micro-controller<br>(Arduino Mega 256)<br>controls the Nanox Cart X-<br>ray System's functionality<br>and GUI display.<br>8) Pulse Generator - The<br>pulse generator enables a<br>pulse of 40kVp and 0.1—<br>2mAs.<br>9) Isolation Transformer –<br>Enables protection from<br>power mains surges.<br>10) Power Supply – Power<br>mains input of<br>230VAC/50Hz/300W.<br>11) Tube Assembly – The X-<br>ray tube assembly consists<br>of the X-ray Tube, Cooling<br>Fluid, electronics, and<br>casing to enable the<br>emission of the radiation<br>energy.<br>12) X-ray Tube - Nano-x's<br>Cold Cathode tube. | | |
| Imaging<br>Radiographic<br>film cassette | The Nanox Cart is specified and<br>designed to operate only with a<br>Flat Panel Digital X-ray | Any commercially available<br>Radiographic film cassette | Different |
| | Nanox Cart X-Ray System<br>(K203782) | AMX-4 Mobile X-Ray System<br>(K021016) | Significant<br>Difference |
| | Detector Model EVS3643, manufactured by DRTECH Inc.<br>(K162552) | | |
| Technology | | | |
| Principles of<br>Operation | The positions of the Tube<br>Assembly and detector holder<br>are prefixed. The technician<br>positions the patient's hand<br>inside the indicated irradiation<br>area on the detector.<br>The technician sets the required<br>energy on the Display's<br>touchscreen. Irradiation energy<br>is set with a fixed radiographic<br>output voltage of 40kVp and a<br>variable charge of 0.1–2mAs.<br>Exposure time varies from<br>50ms up-to 1s (non-selectable).<br>Once energy is set, the<br>technician operates the system<br>by pressing the X-Ray switch. | The technician has the ability to<br>position the AMX-4's tube assembly<br>using the rotating arm. The AMX-4<br>is an all-inclusive device, where X-<br>ray images are constructed on its<br>Display.<br>The AMX-4 has a rechargeable array<br>of 12.9V batteries that has the ability<br>to operate the following methods:<br>1. Two 70 kVp, 10 mAs X-ray<br>exposures, >7 seconds of<br>preparation, > 25 seconds of field<br>light, 5 minutes of drive time, 9<br>minutes of idle time<br>2. The AMX 4 batteries will provide<br>enough capacity for 165 or more 100<br>kVp, 100 mAs X-ray exposures.<br>Each exposure includes 4 seconds of<br>preparation time and 30 seconds of<br>idle time for battery recovery. This<br>number may be reduced by<br>additional idle time required for<br>X-ray tube cooling. | Different |
| System<br>Power | 240 VAC, 50 Hz | 100/200 VAC, 50/60 Hz | Same |
| Battery<br>Powered | No | Yes | Different |
| kV Range | 40 kVp | 50-125 kVp, in 24 steps | Different |
| mAs Range | 0.1 to 2mAs, in 0.1mAs step | 0.40 to 320 mAs, in 30 steps | Different |
| Tube<br>Type/Model | Nanox Tube | GE X-Ray tube model HRT09 | Similar |
| Focal Spot | 0.3 mm | 0.6/2.0 mm | Similar |
| Target Angle | 0 degrees | 15 degrees | Similar |
| Column<br>Rotation<br>Range | Fixed | (+/-)270 degrees | Different |
| | Nanox Cart X-Ray System<br>(K203782) | AMX-4 Mobile X-Ray System<br>(K021016) | Significant<br>Difference |
| Max Speed<br>(kph) | 4 | 4.8 | Similar |
| Horizontal<br>Travel (cm) | Fixed | 40.6 | Different |
| Vertical<br>Travel (cm) | Fixed | 138.4 | Different |
| Maximum<br>Tube Voltage | 40 kV | 100 kV | Different |
| Aperture<br>shape | Rectangular | Rectangular | Same |
| Aperture<br>Control | Fixed | Operator Adjustable | Different |
<sup>1</sup> https://www.accessdata.fda.go, v/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?lD=K021016
{6}------------------------------------------------
# «OUWU
### Nanox Cart X-Ray System – 510(k) Submission File
Nano-x Imaging Ltd. Neve Ilan Communication Center, Zip 9085000, Israel Telephone: +972-2-574-6300
{7}------------------------------------------------
## «OUWU
### Nanox Cart X-Ray System – 510(k) Submission File
Nano-x Imaging Ltd. Neve Ilan Communication Center, Zip 9085000, Israel Telephone: +972-2-574-6300
{8}------------------------------------------------
«OUWU
### Nanox Cart X-Ray System - 510(k) Submission File
Nano-x Imaging Ltd. Neve Ilan Communication Center, Zip 9085000, Israel Telephone: +972-2-574-6300
### Table 2: Subject and Reference Device (DRX-Revolution Nano X-ray system) Comparison
| | Nanox Cart X-Ray System<br>(K203782) | DRX-Revolution Nano<br>X-ray system | Significant<br>Difference |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Regulatory | | | |
| 510(k) | K203782 | K173924 | |
| Device class | II | II | Same |
| Classification | Mobile x-ray system, Class II;<br>21 CFR 892.1720 | Mobile x-ray system, Class II; 21 CFR<br>892.1720 | Same |
| Product code | IZL | IZL | Same |
| Indications /<br>Intended use | The device is designed to<br>perform radiographic X-Ray<br>examinations of hands, wrists,<br>and fingers, on adult patients. | The device is designed to perform<br>radiographic x-ray examinations on<br>pediatric and adult patients, in all<br>patient treatment areas. | Similar |
| Intended<br>users | X-ray Technicians trained to<br>operate the system | X-ray Technicians trained to operate<br>the system | Same |
| Technology | | | |
| System Power | 240 VAC, 50 Hz | 100 – 240 V, 50/60 Hz | Same |
| Power output | 0.08kW | Maximum 4.8 kW @ 104 msec and<br>7.7 kW @ 13 msec of power | Similar |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.