← Product Code [FXX](/productcode/FXX) · K203775

# AIC disposable medical mask (K203775)

_The Aleen International Corporation · FXX · Mar 12, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203775

## Device Facts

- **Applicant:** The Aleen International Corporation
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Mar 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The AIC Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

## Device Story

AIC Disposable Medical Mask; flat pleated design; three-layer construction (inner/outer spun-bond polypropylene; middle melt-blown polypropylene filter). Features malleable aluminum nose piece; available in tie-on or elastic ear loop configurations. Used in clinical settings for infection control; worn by healthcare personnel and patients to reduce exposure to blood, body fluids, and particulates. Provides barrier protection; single-use; non-sterile. No electronic components or software.

## Clinical Evidence

No clinical data. Bench testing only. Performance metrics: Bacterial filtration efficiency 99.3%; Differential pressure 4.40 mmH2O/cm²; Fluid resistance 96/96 pass at 160 mmHg; Particulate filtration efficiency 99.9%; Flammability Class 1.

## Technological Characteristics

Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (middle). Malleable aluminum nose piece. Ear loops (polyester/elastic) or spun-bond polypropylene ties. ASTM F2100 Level 3 performance. Non-sterile. No energy source or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Surgical Face Mask ([K153496](/device/K153496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 12, 2021

The Aleen International Corporation % Diana Lam Regulatory Affairs Consultant Duocare, LLC 370 W. Grand Blvd. Suite 110 Corona, California 92882

Re: K203775

Trade/Device Name: AIC disposable medical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 28, 2020 Received: December 28, 2020

Dear Diana Lam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203775

Device Name AIC Disposable Medical Mask

#### Indications for Use (Describe)

The AIC Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided nonsterile.

| Type of Use (Select one or both, as applicable) | Routine Use Per 31 CFR 331.3(a)(2) - Payments    Data File Comparison Under 31 CFR 331.3(a)(2) - Screening |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K203775

# Date Summary Prepared:

March 10, 2021

## Applicant:

The Aleen International Corporation

Address: Jiutian Green Science and Technology Industrial Park, Heyuan High-Tech Zone, Heyuan, Guangdong, China 517000

www.aleenic.com

## Contact:

Can Huang

The Aleen International Corporation

Email: chuang@aleenic.com

Phone: (+001) 626-620-0184

# Correspondent Contact:

Diana Lam,

DuoCare, LLC

info@@duocarepro.com

### Device Information:

Device Name: AIC Disposable Medical Mask

Trade Name: AIC Disposable Medical Mask

Common Name: Disposable Medical Mask

Device Classification Name: Surgical Apparel

Classification Regulation Number: 21 CFR 878.4040

Device Classification: Class II

Classification Product Code: FXX

{4}------------------------------------------------

### Predicate Device

| Predicate Device                    | Disposable Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K153496                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Xiantao Rayxin Medical<br>Products Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                 |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Item Name                           | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                 |
| Device name                         | AIC Disposable Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Disposable Surgical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                 |
|                                     | Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Face Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                 |
| Manufacturer                        | The Aleen International                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Xiantao Rayxin Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                 |
|                                     | Corporation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Products Co., ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                 |
| 510(K) No.                          | K203775                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K153496                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                 |
| Regulation No.                      | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
| Classification Name                 | Surgical Apparel                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Surgical Apparel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
| Regulatory Class                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
| Product Code                        | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
| Indications for use                 | The AIC Disposable<br>Medical Masks are<br>intended to be worn to<br>protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. The face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device, provided non-<br>sterile.                                                           | The Disposable Surgical<br>Face Masks are intended<br>to be worn to protect<br>both the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device(s),<br>provided non-sterile.                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
| Device Description                  | The AIC Disposable<br>Medical Masks are Blue<br>and white color, and Flat<br>Pleated type mask,<br>utilizing Tie-On or Ear<br>Loops way for wearing,<br>and they all has Nose<br>Piece design for fitting<br>the facemask around the<br>nose.<br>The blue colorant is<br>polypropylene (PP)<br>master batch.<br>The proposed device(s)<br>are manufactured with<br>three layers, the inner<br>and outer layers are<br>made of spun-bond<br>polypropylene, and the<br>middle layer is made of | The proposed device(s)<br>are Blue color, and Flat<br>Pleated type mask,<br>utilizing Tie-On or Ear<br>Loops way for wearing,<br>and they all has Nose<br>Piece design for fitting<br>the facemask around the<br>nose.<br>The blue colorant is<br>polypropylene (PP)<br>master batch.<br>The proposed device(s)<br>are manufactured with<br>three layers, the inner<br>and outer layers are<br>made of spun-bond<br>polypropylene, and the<br>middle layer is made of<br>melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                 |
|                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | melt blown polypropylene<br>filter.<br>The model of proposed<br>device, tie-on, is held in<br>place over the user's<br>mouth and nose by four<br>ties welded to the<br>facemask. The tie is<br>made of spun-bond<br>polypropylene.<br>The model of proposed<br>device, ear loops, is held<br>in place over the user's<br>mouth and nose by two<br>elastic ear loops welded<br>to the facemask. The<br>elastic ear loops are not<br>made with natural rubber<br>latex.<br>The nose piece<br>contained in the<br>proposed device(s) is in<br>the layers of facemask to<br>allow the user to fit the<br>facemask around their<br>nose, which is made of<br>malleable aluminum wire.<br>The proposed device(s)<br>are sold non-sterile and<br>are intended to be single<br>use, disposable device | The model of proposed<br>device, tie-on, is held in<br>place over the user's<br>mouth and nose by four<br>ties welded to the<br>facemask. The tie is<br>made of spun-bond<br>polypropylene.<br>The model of proposed<br>device, ear loops, is held<br>in place over the user's<br>mouth and nose by two<br>elastic ear loops welded<br>to the facemask. The<br>elastic ear loops are not<br>made with natural rubber<br>latex.<br>The nose piece<br>contained in the<br>proposed device(s) is in<br>the layers of facemask to<br>allow the user to fit the<br>facemask around their<br>nose, which is made of<br>malleable aluminum wire.<br>The proposed device(s)<br>are sold non-sterile and<br>are intended to be single<br>use, disposable device. |                                                 |
| Ear loop model and<br>tie-on model  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Ear Loops, Tie-On, Flat                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Ear Loops, Tie-On, Flat<br>Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                            |
|                                     | Outer<br>layer                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Pleated, 3 layers<br>Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                 |
|                                     | Middle<br>layer                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                 |
| Mater-<br>ial                       | Inner<br>layer                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                            |
|                                     | Ties                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                 |
|                                     | Nose piece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                 |
| Ear loops                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Polyester                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Polyester                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                 |
|                                     | Mask cover<br>width                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | $9.5 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | $9.5 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                            |
|                                     | Mask cover<br>length                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | $17.5 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | $17.5 \text{ cm }\pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                            |
| Dimension                           | Single ear loop Length for Ear Loops style                                                                                                                                                                                                                                                                                                                                                                                                                                                   | $18.0 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | $17.0 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar                                         |
|                                     | Single tie Length for Tie-On style                                                                                                                                                                                                                                                                                                                                                                                                                                                           | $38.5 \pm 5.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | $40 \pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar                                         |
|                                     | Nose piece length                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | $10.5\pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | $11.0\pm 1.0$ cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar                                         |
| Actual performance values (Average) | Bacterial filtration efficiency                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 99.3% (3 lots, 32 samples/lot)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 98.7%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar                                         |
|                                     | Differential pressure (Delta-P)                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 4.40 mmH2O/cm² (3 lots, 32 samples/lot)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 4.2 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Similar                                         |
|                                     | Fluid Resistance Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 96 out of 96 Pass at160 mmHg (3 lots, 32 samples/lot)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 32 out of 32 pass at 120 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                            |
|                                     | Particulate Filtration Efficiency                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 99.9% (3 lots, 32 samples/lot)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 98.46%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar                                         |
|                                     | Flammability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Class 1 (3 lots, 32 samples/lot)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Class 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                            |
| Biocompatibility                    | Cytotoxicity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Under the conditions of the study, the proposed device extract demonstrated evidence of potential cytotoxicity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Under the conditions of the study, not cytotoxicity effect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Different                                       |
|                                     | Skin Irritation                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Under the conditions of the study, the proposed device extract was determined to be non - irritating                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Under the conditions of the study, not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                            |
|                                     | Skin Sensitization                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Under the conditions of the study, the proposed device extract was determined to be non - sensitizing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Under conditions of the study, not a sensitizer.…

---

**Source:** [https://fda-staging.innolitics.com/device/K203775](https://fda-staging.innolitics.com/device/K203775)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).