← Product Code [HWC](/productcode/HWC) · K203773

# BRM TOOL Screws (K203773)

_Brm Extremities Srl · HWC · Jan 14, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203773

## Device Facts

- **Applicant:** Brm Extremities Srl
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jan 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The BRM TOOL Screws are indicated for use for aligned bone fracture repair and arthrodesis, osteotomy, joint fusion and bone fragment fixation appropriate with the size of the screw.

## Device Story

BRM TOOL Screws are interfragmentary compression osteosynthesis screws used in orthopedic surgery to support bone, ligament, tendon, or cartilage structures; intended to reduce fractures in skeletal bone epiphyses. Excludes spinal, rib cage, and skull applications. Screws are manufactured from titanium alloy. Used by orthopedic surgeons in clinical settings to provide mechanical fixation of bone fragments. Output is physical stabilization of bone segments, facilitating healing. Benefit is restoration of skeletal alignment and stability.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing included torsional yield strength, driving torque, and pullout strength per ASTM F543 and ASTM F1264. Biocompatibility/sterility compliance verified via LAL testing per AAMI ST-72.

## Technological Characteristics

Titanium alloy bone fixation screws. Mechanical performance validated per ASTM F543 and ASTM F1264. Sterility/endotoxin compliance per AAMI ST-72.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- OrthoSolutions System26 Cannulated Extremity Screws ([K163489](/device/K163489.md))

## Reference Devices

- OVERMED MINIARS Screw ([K143596](/device/K143596.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 14, 2022

BRM Extremities Srl % Margeaux Rogers Associate Director, Regulatory Affairs MCRA, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001

Re: K203773

Trade/Device Name: BRM TOOL Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: December 23, 2020 Received: December 23, 2020

Dear Margeaux Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203773

Device Name BRM TOOL Screws

Indications for Use (Describe)

The BRM TOOL Screws are indicated for use for aligned bone fracture repair and arthrodesis, osteotomy, joint fusion and bone fragment fixation appropriate with the size of the screw.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Manufacturer:       | BRM Extremities S.r.l.                                                                                                                                                |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registered Office:  | Via Lorenzo Mascheroni 29<br>20145 Milano (MI) - Italy                                                                                                                |
| Headquarter:        | Via Papa Giovanni XXIII, 9<br>23862 Civate (LC) – Italy                                                                                                               |
| Contact:            | Ing. Andrea De Maglio<br>Regulatory Affairs Manager<br>BRM Extremities Srl<br>Tel. +39 0341 1693087<br>Cell. +39 389 1938034<br>Email. a.demaglio@brm-extremities.com |
| Prepared By:        | MCRA, LLC<br>703 8th Street NW<br>Washington, DC 20001<br>Phone: 202.552.5800                                                                                         |
| Date Prepared:      | January 13th, 2022                                                                                                                                                    |
| Device Trade Name:  | BRM TOOL Screws                                                                                                                                                       |
| Device Common Name: | Smooth or Threaded Metallic Bone Fixation Fastener                                                                                                                    |
| Classification:     | 21 CFR 888.3040 – Screw, Fixation, Bone<br>Class II                                                                                                                   |
| Product Codes:      | HWC                                                                                                                                                                   |

## Indications for Use:

The BRM TOOL Screws are indicated for use for aligned bone fracture repair and arthrodesis, osteotomy, joint fusion and bone fragment fixation appropriate with the size of the screw.

# Device Description:

The BRM TOOL Screws are interfragmentary compression osteosynthesis screws, i.e., devices used in surgical orthopedic interventions to provide support to the bone, ligament, tendon or cartilage structure, with the aim to reduce fractures in several parts of the skeleton, particularly of bone epiphyses. The application field excludes spinal, rib cage and skull bones. The BRM TOOL Screws are available in Titanium alloy.

{4}------------------------------------------------

### Predicate Devices:

The BRM Screw Family of devices are substantially equivalent to the primary predicate OrthoSolutions System26 Cannulated Extremity Screws (K163489) and reference predicate OVERMED MINIARS Screw (K143596) with respect to intended use, indications for use, design and materials.

### Technological Comparison:

Testing included torsional yield strength, driving torque and pullout strength testing. Results were determined to meet the pre-defined acceptance criteria per ASTM F543 and ASTM F1264. Additionally, the BRM Screw Family is in compliance with LAL testing requirements for orthopedic devices per AAMI ST-72 testing.

#### Substantial Equivalence:

The BRM Screw Family and the legally marketed predicate devices have the same intended use and indications for use, similar dimensions, geometry and materials. Additionally, information presented in the Mechanical Testing section demonstrate that the screws met the pre-defined acceptance criteria per ASTM F543 and ASTM F1264.

---

**Source:** [https://fda-staging.innolitics.com/device/K203773](https://fda-staging.innolitics.com/device/K203773)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
