SeptiCyte RAPID

K203748 · Immunexpress, Inc. · PRE · Nov 29, 2021 · Microbiology

Device Facts

Record IDK203748
Device NameSeptiCyte RAPID
ApplicantImmunexpress, Inc.
Product CodePRE · Microbiology
Decision DateNov 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3215
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K203748 · Nov 29, 2021SeptiCyte RAPIDImmunexpress, Inc.Residual clinical samples from the MARS (NCT01905033) and VENUS (NCT02127502) observational clinical trials; Retrospective Physician Diagnosis (RPD) case review panelsRetrospective clinical data were used to evaluate the diagnostic performance (sensitivity/specificity) of the SeptiCyte RAPID test in distinguishing sepsis from SIRS in adult critical care patients, and to establish clinical cut-offs.Retrospective clinical study; Observational cohort; Residual clinical samples; Clinical case review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
MARS (NCT01905033) and VENUS (NCT02127502) Retrospective Analysis; Retrospective analysis of residual samples from observational, non-interventional, prospective clinical trials; Follow-up/Duration: Samples collected within 24 hours of ICU admissionAdult critical care patients (18-89 years) suspected of sepsis on their first day of ICU admission; Sample Size: 356 subjects (retrospective) + 30 subjects (prospective NEPTUNE study); Number of Sites: 8 clinical sites in the United States and EuropeRetrospective Physician Diagnosis (RPD) based on clinical case reviewDiagnostic performance (sensitivity/specificity) for differentiating sepsis from SIRS

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Sepsis probabilityLogistic regression decision rule——Training cohort: 198 independent patients (100 US samples from VENUS trial, 98 European samples from MARS trial).—Retrospective/Prospective study: 356 retrospectively archived subjects (MARS/VENUS/VENUS Supplement) and 30 prospective subjects (NEPTUNE study).—

Indications for Use

The SeptiCyte ® RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene ® Blood RNA Tube. The SeptiCyte ® RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infectionnegative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte ® RAPID test generates a score (SeptiScore®) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infectionpositive systemic inflammation. SeptiCyte ® RAPID is intended for in-vitro diagnostic use on the Biocartis IdyllaTM System.

Device Story

SeptiCyte RAPID is an in-vitro diagnostic test for sepsis risk stratification. It processes whole blood collected in PAXgene Blood RNA tubes. The device uses a single-use, multi-chambered fluidic cartridge on the Biocartis Idylla System to perform automated RNA extraction and RT-qPCR amplification of two host immune mRNA transcripts (PLA2G7 and PLAC8). The system detects fluorescence to quantify transcript levels and calculates a 'SeptiScore'. This score is categorized into four interpretation bands representing increasing sepsis likelihood. The test is performed in a clinical laboratory setting. Healthcare providers use the SeptiScore alongside clinical assessments and other laboratory findings to aid in differentiating sepsis from infection-negative systemic inflammation, potentially guiding clinical decision-making regarding infection management.

Clinical Evidence

Clinical performance was evaluated using 356 retrospectively archived samples from MARS and VENUS studies and 30 prospective samples from the NEPTUNE study. Performance was assessed by comparing SeptiScore to a Retrospective Physician Diagnosis (RPD) comparator (Forced, Consensus, and Unanimous methods). Results demonstrated a clear, monotonically increasing relationship between SeptiScore bands and sepsis probability. Logistic regression modeling confirmed the SeptiScore provided diagnostic utility beyond standard clinical variables. No clinical data for patients <18 years.

Technological Characteristics

RT-qPCR assay for PLA2G7 and PLAC8 mRNA transcripts. Uses Biocartis Idylla System for automated sample extraction and real-time PCR. Single-use, multi-chambered fluidic cartridge. Detection via fluorescent dye-quencher hydrolysis probes. Connectivity: Standalone instrument. Software: Assay-specific algorithm for Cq value processing and SeptiScore calculation. Sterilization: Not specified.

Indications for Use

Indicated for adult patients suspected of sepsis on their first day of ICU admission to differentiate infection-positive (sepsis) from infection-negative systemic inflammation.

Regulatory Classification

Identification

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.

Special Controls

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended. (2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection. (3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability. (4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information: (i) Results must demonstrate adequate device performance relative to a well-accepted comparator. (ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population. (iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses. (5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics ( *e.g.,* age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument. (7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples ( *e.g.,* collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K203748 B Applicant Immunexpress, Inc C Proprietary and Established Names SeptiCyte RAPID D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | PRE | Class II | 21 CFR 866.3215 - Rt-Qpcr Assay For Mrna Transcript Immune Biomarkers | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain a substantial equivalence determination for SeptiCyte RAPID. The device is a new device based on a subset of two of the four mRNA transcript immune biomarkers measured by the predicate. B Measurand: mRNA transcripts for host immune biomarkers PLA2G7 and PLAC8. C Type of Test: Reverse transcription + quantitative PCR (RT-qPCR) Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} III Intended Use/Indications for Use: A Intended Use(s): The SeptiCyte RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene Blood RNA Tube. The SeptiCyte RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infection-negative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte RAPID test generates a score (SeptiScore) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infection-positive systemic inflammation. SeptiCyte RAPID is intended for in-vitro diagnostic use on the Biocartis Idylla System. B Indication(s) for Use: Same as Intended Use C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: SeptiCyte RAPID is designed to run on the Biocartis Idylla system, K163628 IV Device/System Characteristics: A Device Description: The SeptiCyte RAPID is an in vitro diagnostic test for simultaneous amplification and detection of two RNA transcripts (PLA2G7 and PLAC8) using total RNA extracted from human blood. The test has been designed, manufactured, and validated for use on the Biocartis Idylla real-time PCR system. The SeptiCyte RAPID test is performed with an Idylla Cartridge, a single-use, disposable, multi-chambered fluidic cartridge that runs on the Biocartis Idylla System (K163628). All processing steps are automated and occur within an Idylla Cartridge, including sample extraction/purification, RT-qPCR for the detection and relative quantification of the two human mRNA targets PLAC8, PLA2G7. All cartridge steps in this process, following the addition of the sample, are fully automated and completely integrated. Test results (measured Cq values and calculated SeptiScore) are available in about 65 minutes. The specimen used for the SeptiCyte RAPID is a sample of whole blood collected by venipuncture using the PAXgene blood collection tubes within the PAXgene Blood RNA System (K042613). The cartridge contains all of the necessary reagents to perform RNA isolation from the sample. SeptiCyte RAPID uses quantitative, real-time determination of the amount of each transcript in the sample based on the detection of fluorescence by the Biocartis Idylla qPCR instrument function. The cartridge includes the reagents for reverse transcription and PCR. Transcripts PLAC8 and PLA2G7 are amplified and quantified. These values are combined to produce the SeptiScore, which is interpreted by means of four discrete bands. These four bands reflect a monotonically increasing likelihood of sepsis as shown in the following figure: K203748 - Page 2 of 21 {2} SeptiScore® increases with probability of sepsis 15 SIRS BAND 1 BAND 2 5.0 BAND 3 6.2 7.4 BAND 4 SEPSIS # B Principle of Operation: The SeptiCyte RAPID test employs a coupled reverse transcription - quantitative polymerase chain reaction (RT-qPCR) system to measure the expression levels of the host response genes PLA2G7 and PLAC8 isolated from whole blood. Amplicons generated by the RT-qPCR process are detected and quantitated by fluorescent dyes, upon exonucleolytic release of the dyes from oligonucleotide probes that specifically bind to the amplicons. The test is configured to run on the Biocartis Idylla System. The assay-specific software uses the measured PLA2G7 and PLAC8 Cq values to generate a SeptiScore, which falls into one of four discrete sepsis probability bands, each representing an increased risk of sepsis. # V Substantial Equivalence Information: # A Predicate Device Name(s): SeptiCyte LAB # B Predicate 510(k) Number(s): K163260 # C Comparison with Predicate(s): | Device & Predicate Device(s): | K203748 | K163260 | | --- | --- | --- | | Device Trade Name | SeptiCyte Rapid | SeptiCyte LAB | | General Device Characteristic Similarities | | | | Intended Use/Indications for Use | The SeptiCyte RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene Blood RNA Tube. The SeptiCyte RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infection-negative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte RAPID test generates a score (SeptiScore) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infection-positive systemic inflammation. SeptiCyte RAPID is intended for in-vitro diagnostic use on the Biocartis Idylla System. | The SeptiCyte RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene Blood RNA Tube. The SeptiCyte RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infection-negative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte RAPID test generates a score (SeptiScore) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infection-positive systemic inflammation. SeptiCyte RAPID is intended for in-vitro diagnostic use on the Biocartis Idylla System. | K203748 - Page 3 of 21 {3} K203748 - Page 4 of 21 | | Biocartis Idylla System. | | | --- | --- | --- | | Intended Use Population | Same | Patients suspected of sepsis on their first day of Intensive Care Unit (ICU) admission | | Specimen Type | Same | Whole blood collected in PAXgene Blood RNA tube | | Assay Principle | Same | Quantitative gene expression assay, based on real-time generation of fluorescence from hydrolysis of dye-quencher hydrolysis probes during cycles of PCR amplification of nucleic acid templates | | **General Device Characteristic Differences** | | | | Analytes | Two mRNA transcript immune biomarkers: PLA2G7, PLAC8 | Four mRNA transcript immune biomarkers: PLA2G7, PLAC8, LAMP1, CEACAM4 | | Specimen Processing | Automated extraction of material using the Idylla System | 1. Manual extraction of material in PAXgene Blood RNA Tube, using the IVD version of the QIAGEN PAXgene Blood RNA Kit. 2. Semi-automated extraction PAXgene Blood RNA Tube, using the QIAGEN QIAcube System | | Instrument Platform | Biocartis Idylla System | Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument | | PCR Chemistry | One-step RT-qPCR; Three Singleplex reactions; Dried format | Two-Step RT-qPCR; one Singleplex, one Triplex reaction; Wet format | | Controls | MS2 bacteriophage particles, serving as sample processing control (SPC), i.e., as within-cartridge positive control for both the sample extraction step and the coupled RT-qPCR step. External controls not provided with the assay but are described in labeling with protocols available from sponsor. | High Positive Control, Low Positive Control and Negative Control for each in vitro transcript (LAMP1, CEACAM4, PLA2G7, PLAC8 (IVTs) formulated in neutral buffer, with IVT concentrations designed to produce high, medium, or low SeptiSCORE values | | Result Output | SeptiScore, calculated from the expression levels of the two mRNA analytes PLA2G7, PLAC8. The SeptiScore is placed into four discrete bands that describe a monotonically increasing likelihood of sepsis vs. Systemic Inflammatory Response Syndrome (SIRS). | SeptiScore, calculated from the expression levels of the four mRNA analytes LAMP1, CEACAM4, PLA2G7, PLAC8, and describing the relative likelihood of sepsis vs. Systemic Inflammatory Response Syndrome (SIRS). | | Limitations | • SeptiCyte RAPID is for use on the Idylla™ System only. | • Slightly higher SeptiSCORE values were observed in African Americans as compared | {4} K203748 - Page 5 of 21 | | • SeptiCyte RAPID is not intended for use in patients not suspected of sepsis. A study of healthy individuals showed > 50% of subjects with results in Band 2 (Asian American 70%, African American 53%, Hispanic 48%, White 37%), and 16% in Band 3 or above. SeptiCyte® RAPID results should only be interpreted in conjunction with clinical assessment and other laboratory findings. • Predictive values (estimated probabilities of sepsis or SIRS) are dependent on prevalence of disease and likelihood ratios measured for the clinical trial population. Users should establish or verify that these parameter values are appropriate for the patient population being tested. • The advised sample volume for the test is 900 µl whole blood collected in PAXgene Blood RNA tubes. When using samples that do not meet these criteria, it is possible that the results will not be reliable or valid. • Improper specimen collection, processing and handling can result in degraded RNA and may affect the results obtained with the test. • Using whole blood collected in PAXgene Blood RNA tubes but not maintained at ambient temperature could generate invalid and/or incorrect results. • The diagnostic value of the SeptiCyte RAPID test in suspected sepsis patients has not been validated in US patients younger than 18. • SeptiCyte RAPID test is not to be used if the patient has a WBC count < 25 cells/µL. • SeptiCyte RAPID test has not been studied in patients with neutropenia or immunocompromised patients. | to Whites, Hispanics and Asians in the clinical study. Predictive values depend on the likelihood ratios and the prevalence of disease. Laboratories and other users should establish their own reference intervals for their patient populations using the SeptiCyte LAB Test to reflect potential sources of variability, such as patient gender, race, age, and preparation techniques. • The SeptiCyte LAB Test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infection-negative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The assessment of whether patient signs and symptoms are due to an infection or not should always be based on consideration of all available information, and not based solely on the SeptiCyte LAB Test results. • The results of testing of patients with other conditions or healthy individuals (i.e., patients falling outside the Intended Use Population) may result in a SeptiSCORE that falls into one of the interpretive categories. However, those results do not indicate the presence or absence of sepsis, infection, or any other medical condition. • This procedure is intended to measure in systemically inflamed patients the expression of blood RNA transcript biomarkers of the host immune response to infection. Deviations from the instructions for use in this package insert may yield erroneous results. Quantity and quality of the RNA can affect the test results. • The SeptiCyte LAB Test assay is only designed to be performed on the Applied Biosystems 7500 Fast Dx Instrument. • A signal is generated for PLAC8 in the presence of genomic DNA | | --- | --- | --- | {5} K203748 - Page 6 of 21 | | and absence of reverse transcriptase enzyme. The PLAC8 signal from genomic DNA should not be detected when using the SeptiCyte LAB Test following the recommended assay procedure. • The diagnostic value of the SeptiCyte LAB Test in suspected sepsis patients has not been validated in US patients younger than 18. • The SeptiCyte LAB Test is not to be used if the patient has a WBC count < 2 x 10^{5} cells/mL. • Predictive values depend on the likelihood ratios and the prevalence of disease. Laboratories and other users should establish their own reference intervals for their patient populations using the SeptiCyte™ Lab Test to reflect potential sources of variability, such as patient gender, race, age, and preparation techniques. • If the RNA input per RT reaction for the SeptiCyte LAB Test is not within the linear range (i.e., 20 ng (2 ng/uL) to 500ng (50 ng/uL), the controls will indicate that the run passed even if the input RNA for the sample is not within the linear range. | | --- | --- | VI Standards/Guidance Documents Referenced: Standard Standard Name EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures EP07-A2 Interference Testing in Clinical Chemistry EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents Guidance: FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued 11 May 2005) VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: {6} Repeatability and intermediate precision of the SeptiCyte Rapid was evaluated at one laboratory. Blood sample pools were constructed to yield High (~10), Medium (~6), or Low (~5) SeptiSCORE values. Repeatability (within run precision) was tested with the low sample pool only, using within run values from testing performed over three days with two operators, three instruments and one cartridge lot. Results are summarized in the left-hand columns of Table 1 below. For intermediate precision, identical replicates (prepared as aliquots from the blood sample pools) were tested. A total of 144 runs were conducted, with 48 runs for each sample type (low medium, high), over a total of 12 days, using two different operators and three different instruments. Each operator made 24 runs on each of the three sample types (L, M, H), and each operator's 24 runs were parsed as follows: (2 runs/day) x (4 days) x (3 instruments). On each instrument, 16 runs were conducted with each of the three sample types (L, M, H). The 16 runs for a given instrument and given sample type were parsed as follows: (2 runs/operator/day) x (2 operators) x (4 days). Results were also analyzed for each analyte separately. The sponsor's acceptance criterion $(\mathrm{SD} &lt; 0.5$ units) was met for each component of precision, as well as for the precision calculated across the combination of instruments, operators, and days. Table 1: Summary of Results for Repeatability and Intermediate Precision (SeptiScore) | | Repeatability | | Intermediate Precision | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | Across Days | Across Operators | Across Instruments | Total Variation | | Pool | MEAN | CV (%) | MEAN | CV (%) | CV (%) | CV (%) | CV (%) | | Low | 4.56 | 4.0 | 4.48 | 4.5 | 4.8 | 4.9 | 5.0 | | Medium | NA | NA | 6.33 | 4.0 | 3.9 | 4.3 | 4.3 | | High | NA | NA | 10.33 | 1.8 | 1.9 | 1.9 | 2.0 | # 2. Lot-to-lot Reproducibility Samples prepared and used as above, i.e., a low, medium, and high pool, and were studied with three cartridge lots over three non-consecutive days, using two instruments and one operator. Each combination of lot, day, and instrument was tested with two replicates K203748 - Page 7 of 21 {7} Table 2: Summary of Results for Lot-to-Lot Reproducibility | | | Lot-to-Lot reproducibility | | | | | --- | --- | --- | --- | --- | --- | | Variable | Pool | MEAN | N | SD | CV (%) | | PLA2G7 Cq | Low | 32.85 | 12 | 0.30 | 0.91 | | PLAC8 Cq | Low | 28.49 | 12 | 0.33 | 1.17 | | SeptiScore | Low | 4.36 | 12 | 0.14 | 3.31 | | PLA2G7 Cq | Medium | 31.80 | 12 | 0.21 | 0.66 | | PLAC8 Cq | Medium | 25.56 | 12 | 0.21 | 0.80 | | SeptiScore | Medium | 6.25 | 12 | 0.19 | 3.01 | | PLA2G7 Cq | High | 33.87 | 12 | 0.40 | 1.17 | | PLAC8 Cq | High | 23.67 | 12 | 0.33 | 1.38 | | SeptiScore | High | 10.20 | 12 | 0.19 | 1.82 | The $\% \mathrm{CVs}$ for each parameter and for the overall SeptiScore were acceptable and below the acceptance criteria of $&lt; 4\%$ . # 3. Linearity: Linearity was defined within the sponsor's study for the assay reportable range and assessed by the relationship between WBC increase and PLAC8 and PLA2G7 increase. as measured by linear $(\mathrm{r}^2 &gt; 0.85)$ reaction dynamics (i.e., Cq inversely proportional to log of input WBC amount) for each of the two transcripts, over the claimed input range. Ranges tested and results for linearity testing are summarized in Table 3 and Figure 1 below: Table 3: Results for linearity testing. | Sample ID | WBC Concentration (cells/ml) | Total Tests (N) | Mean SeptiScore | SD SeptiScore | CV (%) | | --- | --- | --- | --- | --- | --- | | WBC1 | 50 x 10^6 | 3 | 3.81 | 0.40 | 10.5 | | WBC2 | 25 x 10^6 | 3 | 3.22 | 0.17 | 5.3 | | WBC3 | 12.5 x 10^6 | 3 | 3.59 | 0.22 | 6.1 | | WBC4 | 5 x 10^6 | 10 | 3.92 | 0.24 | 6.0 | | WBC5 | 2.5 x 10^6 | 10 | 3.70 | 0.25 | 6.7 | | WBC6 | 1 x 10^6 | 10 | 4.14 | 0.41 | 9.8 | | WBC7 | 0.25 x 10^6 | 20 | 4.19 | 0.30 | 7.3 | K203748 - Page 8 of 21 {8} ![img-0.jpeg](img-0.jpeg) Figure 1: Summary of analyte Cq values from the reportable range study. Cq values versus WBC input concentration for (A) PLA2G7, (B) PLAC8, and (C) SeptiScore ![img-1.jpeg](img-1.jpeg) ![img-2.jpeg](img-2.jpeg) The sponsor notes that a small Hook effect was noted at high WBC levels ( $&gt;25 \times 10^{6}$ WBC cells/ml; however, this appears to be similar for both analytes and is compensated by score calculation which is the difference of these two analytes. These results are considered acceptable. # 4. Detection limit a. Limit of Blank: Platelet reduced plasma was used to determine the Limit of Blank as the sponsor states that leukocyte depleted blood used for the predicate was unavailable. Ten replicates of platelet reduced plasma did not generate a PCR signal $(\mathrm{Cq} &gt; 40)$ for PLA2G7 and, accordingly, did not generate a SeptiScore, b. Limit of Detection Different concentrations of white blood cells (WBC) from a single donor into a matrix consisting of platelet-reduced plasma to determine the level at which a SeptiScore was generated, i.e., the LoD was defined as the lowest titer for which at least $95\%$ (19/20) WBC replicates generate a SeptiScore. K203748 - Page 9 of 21 {9} Table 4: Results for Limit of Detection | Sample ID | WBC Concentration (cells/ml) | Invalid Tests (N) | Total Tests (N) | Failure Rate | Result | | --- | --- | --- | --- | --- | --- | | WBC4 | 5 x 106 | 0 | 10 | 0% | Above LoD | | WBC5 | 2.5 x 106 | 0 | 10 | 0% | Above LoD | | WBC6 | 1 x 106 | 0 | 10 | 0% | Above LoD | | WBC7 | 0.25 x 106 | 0 | 20 | 0% | Above LoD | | WBC8 | 0.1 x 106 | 0 | 20 | 0% | Above LoD | | WBC9 | 0.025 x 106 | 0 | 20 | 0% | LoD | | WBC10 | 0.001 x 106 | 10 | 20 | 50% | Below LoD | | WBC11 | 0.025 x 105 | 20 | 20 | 100% | Below LoD | | WBC12 | 0.001 x 105 | 10 | 10 | 100% | Below LoD | LoD was defined as $0.025 \times 10^{6}$ WBC cells/ml, below the level expected with clinical use. # c. Limit of Quantitation LoQ was defined as the lowest WBC concentration for which at least $95\%$ of WBC replicates generate a SeptiScore with a standard deviation of 1.0 or less; results indicated that the LoQ was similar to the LoD at $0.025 \times 10^{6}$ WBC cells/ml. Table 5: Results for Limit of Quantitation | Sample ID | WBC Concentration (cells/ml) | Valid Tests (N) | Total Tests (N) | Failure Rate | Mean SeptiScore | Std Dev SeptiScore | CV (%) | Result | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | WBC4 | 5 x 106 | 10 | 10 | 0% | 3.92 | 0.24 | 6.0 | Above LoQ | | WBC5 | 2.5 x 106 | 10 | 10 | 0% | 3.70 | 0.25 | 6.7 | Above LoQ | | WBC6 | 1 x 106 | 10 | 10 | 0% | 4.14 | 0.41 | 9.9 | Above LoQ | | WBC7 | 0.25 x 106 | 20 | 20 | 0% | 4.19 | 0.30 | 7.2 | Above LoQ | | WBC8 | 0.1 x 106 | 20 | 20 | 0% | 4.06 | 0.29 | 7.1 | Above LoQ | | WBC9 | 0.025 x 106 | 20 | 20 | 0% | 4.00 | 0.63 | 15.9 | LoQ | | WBC10 | 0.001 x 106 | 10 | 20 | 50% | NA | NA | NA | Below LoQ | | WBC11 | 0.025 x 105 | 0 | 20 | 100% | NA | NA | NA | Below LoQ | | WBC12 | 0.001 x 105 | 0 | 10 | 100% | NA | NA | NA | Below LoQ | The LoD observed is acceptable for clinical practice and would include all patient groups with the exception of patients with severe neutropenia. # 5. Analytical Specificity/Interference: # a. Analytical Specificity: Analytical specificity was tested by reactivity against blank samples and genomic DNA and both transcripts are endogenous markers. For blank cartridges, 20/20 blank cartridges (PBS mixed with PAXgene RNA stabilizer) generated undetermined or invalid SeptiScore values; 10 cartridges with no sample added similarly generated undetermined or invalid SeptiScore values. External positive controls run the same day generated expected results. K203748 - Page 10 of 21 {10} For reactivity against genomic DNA, testing was conducted on 10 replicates of each of two commercially procured total human gDNA samples. Each lot was tested at 500 ng and 1000 ng input per reaction. Table 6: Specimen Panel Characteristics and Expected Results | Sample Description | Total Reps | SeptiScore Expected? | | --- | --- | --- | | gDNA #1 500ng | 10 | No | | gDNA #2 500ng | 10 | No | | gDNA #1 1000ng | 15 | No | | gDNA #2 1000ng | 5 | No | | Total Assays | 40 | | The SeptiCyte RAPID test produced invalid SeptiScore values for each of the 40 reactions run in this part of the study, with no valid Cq values generated for any of the cartridges used to test the gDNA samples. b. Interferants A specimen panel was prepared by spiking potential interferents into pooled, PAXgene-stabilized blood collected from healthy donors. The potential interferents were dissolved in appropriate solvents, and then spiked into the PAXgene blood samples at concentrations higher than the maxima of their normal or expected reference ranges. Three replicates were tested at each concentration of potential interferent and compared to matched PAXgene blood controls (containing the appropriate solvent but no interferent). The substances tested, concentration of each, and results are reproduced below in Table 7: K203748 - Page 11 of 21 {11} Table 7: Effect of Potential Interferants on SeptiScore Results | Sample ID | Potential Interferent | Concentration | N | Mean SeptiScore | SD | Delta SeptiScore (relative to control) | Result | | --- | --- | --- | --- | --- | --- | --- | --- | | INF1 | Rheumatoid Factor | N/A | 3 | 4.24 | 0.11 | 0.22 | PASS | | INF2 | Heparin | N/A | 3 | 4.24 | 0.09 | 0.22 | PASS | | INF3 | Imipenem | N/A | 3 | 4.36 | 0.30 | 0.10 | PASS | | INF4 | Bilirubin | 45 IU/mL | 3 | 4.30 | 0.13 | 0.16 | PASS | | INF5 | Triglycerides | 3000 U/L | 3 | 4.18 | 0.05 | 0.28 | PASS | | INF6 | Vancomycin | 1.2 mg/mL | 3 | 4.68 | 0.20 | 0.22 | PASS | | INF7 | Cefotaxime | 20 mg/dL | 3 | 4.27 | 0.15 | 0.19 | PASS | | INF8 | Dopamine | 500 mg/dL | 3 | 4.31 | 0.20 | 0.15 | PASS | | INF9 | C-reactive protein (CRP) | 70 μmol/L | 3 | 4.24 | 0.04 | 0.22 | PASS | | INF10 | Noradrenaline | 670 μmol/L | 3 | 4.39 | 0.09 | 0.07 | PASS | | INF11 | Dobutamine | 6.0 μmol/L | 3 | 4.49 | 0.24 | 0.03 | PASS | | INF12 | Hemoglobin | 4 mg/dL | 3 | 4.31 | 0.11 | 0.15 | PASS | | INF13 | Albumin | 700 pg/mL | 3 | 4.27 | 0.14 | 0.19 | PASS | | INF14 | Furosemide | 12 μg/mL | 3 | 4.44 | 0.10 | 0.18 | PASS | | INF15 | Soluble CD14 (sCD14) | 20 g/dL | 3 | 4.48 | 0.28 | 0.02 | PASS | | INF16 | IL-6 | 5 g/dL | 3 | 4.40 | 0.20 | 0.06 | PASS | | INF17 | Lipopolysaccharide binding protein (LBP) | 180 μmol/L | 3 | 4.33 | 0.10 | 0.12 | PASS | | CON1 | Solvent Control - Methanol/Blood | 5 μg/mL | 3 | 4.25 | 0.12 | N/A | N/A | | CON2 | Solvent Control - Water/Blood | 15 pg/mL | 3 | 4.46 | 0.21 | N/A | N/A | | CON3 | Solvent Control - PBS/Blood | 45 μg/mL | 3 | 4.45 | 0.07 | N/A | N/A | Results were consistent with no effects observed from the interferants tested. 6. **Assay Reportable Range:** See linearity above 7. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):** a. **Real-time Stability** Real-time stability studies are ongoing to support product claims. Real-time stability study protocols and acceptance criteria were reviewed and found to be acceptable. **In-use stability** The performance of the SeptiCyte RAPID test cartridge was evaluated when removed from the pouch and left on the lab bench for a period of time, both with and without the addition of sample to the cartridge. Specifically, test cartridges were removed from their pouches and placed at 30°C and 75% relative humidity for up to 3 days. No significant degradation in performance was observed and these data support the sponsor's recommendation in labeling that testing should be performed within one hour of adding samples to the pouch. K203748 - Page 12 of 21 {12} b. Shipping Stability Shipping simulations were performed to evaluate robustness of the SeptiCyte RAPID test to thermal cycles as well as shocks and vibrations of a transport cycle when placed in the packaging used for international shipping. To simulate thermal stress, cartridges were subjected to three winter cycles and a summer cycle (2-3 days at -20°C followed by a day at 25°C for a winter cycle, and 3 days at 40°C for a summer cycle). Immediately after thermal stress, 6 cartridges were tested with PAXgene citrate whole blood samples (900ul tested per cartridge). As with other stability studies, a maximum drift of ±10% was considered acceptable. All samples evaluated are part of this study successfully met the acceptance criteria. These results confirm that secondary packaging is sufficient to maintain stable performance of the test for shipping conditions anticipated during distribution of the product. c. PAXgene Sample Collection/Handling The sample collection study evaluated stability of the test result both when samples were processed either immediately after a blood draw or when frozen prior to testing. Blood samples were collected from each of 15 healthy donors with the first sample processed in real-time and tested both immediately and after a 120-minute incubation at room temperature to mimic the sample processing conditions for the PAXgene sample tubes. Results from the stability test of freshly collected blood samples demonstrated no significant differences in SeptiCyte score when evaluated immediately or after a 120-minute incubation. These data support SeptiScore testing immediately after blood draw or after sample processing according to the manufacture instructions for the PAXgene sample tube. To support the use of banked frozen samples in their validation studies, the sponsor performed an equivalency study between fresh and frozen specimens. Specimens processed immediately were compared to paired samples that underwent one freeze thaw cycle and were stored for one month at &lt;-70°C. Across all evaluated samples, the mean Ct shift in the PLA2G7 transcript Cq was 0.26, the mean shift in PLAC8 was 0.2, and the mean shift in SeptiScore was 0.06. Cumulatively, these results demonstrate that a single freeze-thaw cycle did not significantly affect SeptiScore reported results. 8. Detection Limit: See Section A.4 above. 9. Assay Cut-Off: See clinical cutoff (Section D) below. B Comparison Studies: 1. Method Comparison with Predicate Device: N/A 2. Matrix Comparison: N/A K203748 - Page 13 of 21 {13} C Clinical Studies: 1. Clinical Sensitivity: Clinical performance of the SeptiCyte Rapid was evaluated in a manner similar to that for the Septicyte LAB predicate (K163260). Residual samples with sufficient volume remaining from the two observational, non-interventional, prospective clinical trials (the Molecular Diagnosis and Risk Stratification of Sepsis [MARS- NCT01905033] study and the Septic Gene Expression Using SeptiCyte [VENUS - NCT02127502] study) used for validation of the Septicyte LAB predicate were tested. The MARS and VENUS studies were conducted across eight clinical sites in the United States and Europe. Of the original 447 subjects, PAXgene Blood RNA tubes were available for 356. This represented ~80% of the original cohort: 139/198 of the sample from the MARS study (70.2% representation), 101/129 from the original Venus sample (78.3% representation), and 116/120 of the subjects from the Venus Supplement (96.7% representation). An additional 30 prospective specimens were collected as part of the NEPTUNE (NEar PatienT MolecUlar TestiNg in Sepsis, NEPTUNE) study. The NEPTUNE protocol was a prospective study similar in design to the MARS and VENUS studies; however, this study was terminated early due to difficulty enrollment with the onset of the COVID epidemic. Study protocols were similar enough that combining subject results was considered acceptable. For the retrospective studies, the primary objective was to determine the diagnostic performance of SeptiCyte LAB in distinguishing sepsis from SIRS in adult critical care patients, either as a stand-alone test or in combination with other clinical variables and laboratory assessments used to confirm or exclude a diagnosis of sepsis. A “sepsis event” was defined operationally to have occurred when a patient displayed two or more signs of systemic inflammation and was given therapeutic systemic antibiotics by the attending clinician within 24 hours of ICU admission. Adult critical care patients aged 18--89 years were considered qualified for enrollment if, upon admission to an ICU, they displayed two or more of the following signs: - temperature above 38°C or less than 36°C - heart rate greater than 90 beats per minute - tachypnea greater than 20 breaths per minute or PaCO2 less than 32 mm Hg - White blood cell (WBC) count greater than 12,000/mm3 or less than 4,000/mm3 or greater than 10% immature neutrophils (bands). Exclusion criteria included: - Clinical cultures or serologic tests were not obtained when sepsis was suspected - ICU admission more than 24 hours before trial enrollment - Delay of more than 24 hours between trial enrollment and sample draw - Treatment with antibiotics more than 24 hours before ICU admission for any reason other than surgical prophylaxis (VENUS and Venus Supplement only), or - Documented sepsis event that occurred more than 3 days before ICU admission or more than 2 days after ICU admission (MARS only). The SeptiScore for each subject was compared to a Retrospective Physician Diagnosis (RPD) comparator based on clinical case review by medical experts who independently reviewed clinical information about each subject while remaining blinded to the SeptiCyte LAB results. Case reviews occurred after the patient was discharged from the hospital. The RPD Case Review Panel based their diagnosis on clinical data. RPD assessment and analyses were approached three ways (four in indeterminate as sepsis is considered): {14} 1. Forced RPD: Subjects were stratified by diagnosis (Systemic Inflammatory Response Syndrome (SIRS) or sepsis) according to the majority opinion of the RPD Case Review Panelists. The category of Indeterminate was not allowed; all subjects were forced into either the SIRS or sepsis categories, with no subjects excluded. The performance evaluation of the SeptiCyte LAB was based on the Forced RPD. 2. Consensus RPD: Subjects were stratified by diagnosis (SIRS, sepsis or indeterminate) according to the majority opinion of three expert reviewers. Indeterminate cases (n=37) were those for which a definitive diagnosis could not be reached by expert review and were not included in the analysis. Indeterminate as sepsis: An analysis where subjects with indeterminate results were considered as sepsis was also included. 3. Unanimous RPD: the subset of subjects where all three assessments were in agreement The inclusion and exclusion criteria for the additional prospective subjects was similar to that for retrospective subjects, as was the method for RPD diagnosis, allowing data to be effectively pooled. The following is the summary of demographic data for the retrospective subjects from the sponsor's submission: Table 8: Demographic Characteristics of Retrospectively Archived Subjects (n = 356) | Demographic & Clinical Variables | | Indeterminate (Consensus) N (%) | SIRS (Consensus) N (%) | Sepsis (Consensus) N (%) | | --- | --- | --- | --- | --- | | Sex | Female | 13 (3.7%) | 86 (24.2%) | 58 (16.3%) | | | Male | 19 (5.3%) | 106 (29.8%) | 74 (20.8%) | | Race/ Ethnicity | Asian | 1 (0.28%) | 10 (2.81%) | 9 (2.53%) | | | Black (American or European of African descent) | 9 (2.53%) | 47 (13.20%) | 26 (7.30%) | | | Hispanic | 0 (0%) | 6 (1.69%) | 4 (1.12%) | | | Unknown | 0 (0%) | 2 (0.56%) | 2 (0.56%) | | | White (Caucasian) | 22 (6.18%) | 127 (35.67%) | 91 (25.56%) | | Source of admission | Emergency department | 22 (6.18%) | 120 (33.71%) | 81 (22.75%) | | | Nursing ward | 3 (0.84%) | 13 (3.65%) | 25 (7.02%) | | | Other | 1 (0.28%) | 13 (3.65%) | 6 (1.69%) | | | Post-anesthesia care facility | 0 (0%) | 14 (3.93%) | 2 (0.56%) | | | Coronary care facility | 0 (0%) | 4 (1.12%) | 1 (0.28%) | | | Intensive Care Unit, other hospital | 3 (0.84%) | 14 (3.93%) | 12 (3.37%) | | | Operating theatre | 3 (0.84%) | 14 (3.93%) | 5 (1.40%) | | Impression on Discharge (Consensus of 3 x Principal Investigators) | Definite Infection | 3 (0.84%) | 0 (0%) | 78 (21.91%) | | | Possible Infection | 14 (3.93%) | 5 (1.40%) | 26 (7.30%) | | | Probable Infection | 3 (0.84%) | 0 (0%) | 23 (6.46%) | | | SIRS | 12 (3.37%) | 187 (52.53%) | 5 (1.40%) | | Death | Unknown | 0 (0%) | 0 (0%) | 1 (0.28%) | | | No | 29 (8.15%) | 173 (48.60%) | 111 (31.18%) | | | Yes | 3 (0.84%) | 19 (5.34%) | 20 (5.62%) | | Invasive Mechanical Ventilation | No | 19 (5.3%) | 116 (32.6%) | 87 (24.4%) | | | Yes | 13 (3.7%) | 76 (21.3%) | 45 (12.6%) | K203748 - Page 15 of 21 {15} Table 9: Demographic Characteristics of Prospective Subjects (n = 30) | Demographic & Clinical Variables | | Indeterminate (Consensus) N (%) | SIRS (Consensus) N (%) | Sepsis (Consensus) N (%) | | --- | --- | --- | --- | --- | | Sex | Female | 2 (6.7%) | 4 (13.3%) | 7 (23.3%) | | | Male | 4 (13.3%) | 8 (26.7%) | 5 (16.7%) | | Ethnicity/Race | Asian | 1 (3.3%) | 0 (0%) | 0 (0%) | | | Black or African American | 2 (6.7%) | 3 (10.0%) | 2 (6.7%) | | | Hispanic | 1 (3.3%) | 5 (16.7%) | 5 (16.7%) | | | Other | 0 (0%) | 1 (3.3%) | 1 (3.3%) | | | Unknown | 1 (3.3%) | 0 (0%) | 0 (0%) | | | White | 1 (3.3%) | 3 (10.0%) | 4 (13.3%) | | Source of Admission | ED | 5 (16.7%) | 6 (20.0%) | 7 (23.3%) | | | ICU | 1 (3.3%) | 1 (3.3%) | 2 (6.7%) | | | ICU (other hospital) | 0 (0%) | 1 (3.3%) | 2 (6.7%) | | | PACU or Operating rooms | 0 (0%) | 4 (13.3%) | 0 (0%) | | | Wards | 0 (0%) | 0 (0%) | 1 (3.3%) | | Impression on Discharge Consensus of 3 x Principal Investigators) | None (i.e., SIRS, No Infection) | 4 (13.3%) | 10 (33.3%) | 1 (3.3%) | | | Possible Infection | 1 (3.3%) | 2 (6.7%) | 2 (6.7%) | | | Probable | 0 (0%) | 0 (0%) | 1 (3.3%) | | | Definite | 0 (0%) | 0 (0%) | 7 (23.3%) | | | Not reported | 1 (3.3%) | 0 (0%) | 1 (3.3%) | | Death | No | 6 (20.0%) | 12 (40.0%) | 11 (36.7%) | | | Yes | 0 (0%) | 0 (0%) | 1 (3.3%) | | Invasive Mechanical Ventilation | No | 5 (16.7%) | 9 (30.0%) | 10 (33.3%) | | | Yes | 1 (3.3%) | 3 (10.0%) | 2 (6.7%) | Overall results for the retrospective and prospective subjects combined across all four RPD categories is illustrated in Figure 2 below: K203748 - Page 16 of 21 {16} ![img-3.jpeg](img-3.jpeg) Figure 2: Probability of Sepsis by Band for Combined Prospective and Prospectively Archived Samples. ![img-4.jpeg](img-4.jpeg) ![img-5.jpeg](img-5.jpeg) Results are presented for (A) Forced, (B) Consensus, and (C) Unanimous RPD methods. The percentage (prevalence) of sepsis is indicated for each Band with error bars representing $80\%$ confidence intervals. The number (N) of sepsis and SIRS in each Band is indicated to the right of the point estimate (NSepsis:NSIRS) as determined by each RPD method. Results across all approaches to analyzing RPD results are consistent, i.e., there is a clear relationship between SeptiScore and the increasing likelihood of sepsis across each SeptiScore Interpretation Band. Table 10 below presents the same results in tabular form: Table 10: Probability of Sepsis by RPD method across bands | RPD Method | N | Probability of Sepsis (80% Confidence Interval) | | | | Delta Differential | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Band 1 | Band 2 | Band 3 | Band 4 | Band 3/ Band 1 | Band 4/ Band 2 | | Forced | 386 | 0.12 (0.08 - 0.18) | 0.24 (0.18 - 0.3) | 0.48 (0.4 - 0.56) | 0.8 (0.74 - 0.84) | 0.221 | 0.440 | | Consensus | 348 | 0.1 (0.06 - 0.15) | 0.22 (0.16 - 0.29) | 0.45 (0.37 - 0.54) | 0.81 (0.75 - 0.85) | 0.214 | 0.459 | | Unanimous | 246 | 0.11 (0.06 - 0.18) | 0.21 (0.14 - 0.29) | 0.43 (0.33 - 0.54) | 0.8 (0.73 - 0.86) | 0.141 | 0.448 | | Indeterminate as-sepsis | 386 | 0.16 (0.11 - 0.21) | 0.31 (0.25 - 0.38) | 0.53 (0.45 - 0.61) | 0.82 (0.77 - 0.87) | 0.239 | 0.391 | Using a method identical to the analysis performed for the predicate K163260, results from the SeptiCyte RAPID test were analyzed with a backwards-elimination (BW-Elimination) variable selection procedure, combined with a logistic regression decision rule, for a subset of the clinical and laboratory variables from the complete dataset (see Table 11 below). Area Under Curve (AUC) for SIRS versus sepsis was used for assessing model performance for each analysis. Models were constructed using the Forced or Consensus RPD and the top five clinical variables contributing to the AUC were identified. For all logistic regression models, BW-Elimination and k-fold cross-validation were used to determine if the test K203748 - Page 17 of 21 {17} provided diagnostic clinical utility beyond that provided by other clinical variables and laboratory assessments available within the first~24 hours of the suspicion of sepsis. The backward elimination technique started with the full model including all independent variables. Models were constructed including or excluding Procalcitonin (PCT). At each step, the effect showing the smallest contribution to the model was deleted. The logistic regression model for discriminating between sepsis and SIRS found SeptiScore to be the most significant variable for each model. This observation was found for both Consensus RPD and Forced RPD. Table 11: List of Variables Evaluated in Logistic Regression Modeling | | Variable | Brief Description of Variable* | | --- | --- | --- | | 1 | Age | Age of patient | | 2 | Race | Race of patient: African descent or Non-African descent | | 3 | Gender | Gender of patient | | 4 | Glucose (Maximum) | Peak blood glucose concentration | | 5 | HeartRate (Minimum) | Lowest heart rate | | 6 | HeartRate (Maximum) | Peak heart rate | | 7 | Infection status: Present | Has an infection been confirmed? | | 8 | Infection status: Fungal? | Is a fungal infection present? | | 9 | Infection status: Gram negative? | Is a Gram negative infection present? | | 10 | Infection status: Gram positive? | Is a Gram positive infection present? | | 11 | Infection status: Mixed? | Is a mixed infection (i.e., more than one Category of infectious agent) | | 12 | Infection status: Viral? | Is a viral infection present? | | 13 | Mean Arterial Pressure (MAP Maximum) | Peak mean arterial blood pressure | | 14 | SIRS Criteria | Number of SIRS criteria observed | | 15 | Temperature (Maximum) | Peak body temperature | | 16 | Temperature (Minimum) | Lowest body temperature | | 17 | WBC (Maximum) | Maximum WBC count | | 18 | WBC (Minimum) | Minimum WBC count | | 19 | PCT/Procalcitonin** | Plasma PCT concentration (log2) | | 20 | SeptiSCORE** | The SeptiCyte LAB result | * All variables were measured and recorded within the first 24 hours of subject enrollment except Infection status. **Variables that were not used in every model built. 2. Clinical specificity: See discussion of Clinical sensitivity (section C.1) above. 3. Other clinical supportive data (when a. and b. are not applicable): N/A Lead Reviewer or Consulting Reviewer Comments for Internal Discussion Only K203748 - Page 18 of 21 {18} D Clinical Cut-Off: SeptiSCORE RAPID cut-off values were established prior to the clinical trial. The sponsor notes that the original design goal was for three bands, with the following performance targets as based on ‘clinical feedback and risk analysis’: - Band 1 Sensitivity ≥ 90%: Minimize the number of FN (patients with sepsis who would go undetected, hence untreated) - Band 3 Specificity ≥ 80%: Minimize the number of FP (minimize sepsis overdiagnosis, leading to unnecessary treatment) Sensitivity and specificity were originally defined as follows: - Band 1 sensitivity = TPband1/2/(TPband1/2+FNband1/2) - Band 3 specificity = TNband2/3/(TNband2/3+FPband2/3) Ranges for the band threshold point estimates were identified by testing PAXgene blood samples from a Training cohort (N=198) of independent patients using pilot GMP-grade SeptiCyte RAPID cartridges. The Training cohort consisted of the following: 1. A panel of 100 US samples banked from the VENUS clinical trial meeting the following criteria: a. The sample set spanned the measurement range of the assay with representation of low, medium, and high SeptiScores. b. Comparison data were previously generated with the SeptiCyte LAB kit (predicate device) for all of the selected specimens. c. None of these samples were used in the clinical validation of SeptiCyte LAB or SeptiCyte RAPID for the purposes of FDA 510(k) submission. d. Three (3) samples from the VENUS portion of the Training cohort were excluded from the final threshold analysis/selection as they obtained an “indeterminate” diagnosis from the retrospective physician panel. These indeterminate samples were only included in the platform comparison to establish the direct correlation between test thresholds. 2. A panel of 98 European clinical samples collected from the MARS clinical trial. The UMCU clinical sample cohort was selected on the basis of clinical data related to the patients’ ICU stay. The diagnostic classification of each patient was made based on the infection likelihood of the patient, retrospectively determined by a panel of UMCU physicians, and linked to a sepsis event within 24 hours prior to or 24 hours after ICU admission. Subjects included 46 sepsis cases (23 culture proven, and 23 probable infections based on infection likelihood) and 52 SIRS cases. Indeterminate cases (infection likelihood “possible”) were not included. Thresholds for SeptiCyte RAPID were determined and locked in a two-part study using pilot GMP-grade cartridges. Part 1 was a correlation/method comparison analysis to determine a direct relationship between the scores, bands, and thresholds for SeptiCyte LAB (predicate device) and SeptiCyte RAPID (new device). Part 2 was a resampling-based threshold analysis using training set to determine final locked thresholds. The three SeptiScore interpretation bands that were resolved had the following cut points: - Band 1: [0, 4.95) - Band 2: [4.95, 7.35) - Band 3: [7.35, 15] {19} However, subsequent analysis of the prospectively archived/retrospective study showed a large percentage of subjects in Band 2, which also substantially overlapped the reference range obtained from studies of healthy volunteers. Accordingly, to improve clinical interpretation, banding was subsequently revised to a 4 Band interpretation consistent with the predicate device by evenly splitting the "Intermediate" band of into two bands, as shown below in Table 12. Table 12: Concordance between 3-Band and 4-Band schemes | 3-Band Reporting Scheme | | 4-Band Reporting Scheme | | | --- | --- | --- | --- | | Interpretation Band | SeptiScore Range | Interpretation Band | SeptiScore Range | | 1 | 0 to 4.9 | 1 | 0 to 4.9 | | 2 | 5.0 to 7.3 | 2 | 5.0 to 6.1 | | | | 3 | 6.2 to 7.3 | | 3 | 7.4 to 15 | 4 | 7.4 to 15 | ## E Expected Values/Reference Range: The sponsor studied references ranges in a cohort of 119 'presumably healthy individuals' &gt; 20 years of age balanced by sex (male/female) and race/ethnicity (White, African American, Asian, and Hispanic). Samples were obtained at a single site, Discovery Life Sciences; Huntsville, AL. Results are summarized as follows: Table 13: SeptiCyte RAPID Reference Range by Race/Ethnicity | SeptiCyte RAPID Reference Ranges | | | | | | | --- | --- | --- | --- | --- | --- | | Race/Ethnicity | SeptiScore Interpretation Band | | | | | | | | 1 (0-4.9) | 2 (5.0-6.1) | 3 (6.2-7.3) | 4 (7.4-15) | | Asian | N | 4 | 21 | 5 | 0 | | | % | 13.3 | 70.0 | 16.7 | 0 | | Black | N | 3 | 16 | 8 | 3 | | | % | 10.0 | 53.3 | 26.7 | 10.0 | | Hispanic | N | 13 | 14 | 2 | 0 | | | % | 44.8 | 48.3 | 6.9 | 0 | | White | N | 18 | 11 | 1 | 0 | | | % | 60.0 | 36.7 | 3.3 | 0 | | Total | N | 38 | 62 | 16 | 3 | | | % | 31.9 | 52.1 | 13.4 | 2.5 | This information is essential for proper use of this device and has been included as a section in the device labeling, as well as incorporated into limitations in the labeling for the device. Test results for the device include the following interpretation note: Increased scores were observed in African Americans relative to other racial/ethnic groups, in the clinical study of patients suspected of sepsis. For African-American subjects, only Band 4 showed a higher likelihood of Sepsis versus SIRS. ## VIII Proposed Labeling: K203748 - Page 20 of 21 {20} The labeling supports a finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K203748 - Page 21 of 21
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...