K203740 · Riverpoint Medical · MBI · Feb 12, 2021 · Orthopedic
Device Facts
Record ID
K203740
Device Name
JuggerKnot Soft Anchor
Applicant
Riverpoint Medical
Product Code
MBI · Orthopedic
Decision Date
Feb 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
JuggerKnot® Soft Anchors are intended for use in soft tissue to bone fixation for the following indications: Knee MPFL Knee Patellar tendon repair Knee MCL Knee Quadriceps tendon repair Hip Acetabular labral repair Hip Proximal hamstring repair Hip Hip labral reconstruction Foot and Ankle Achilles tendon repair Foot and Ankle Medial/lateral repair and reconstruction Foot and Ankle Plantar plate repair Foot and Ankle Mid- and forefoot repair Foot and Ankle Metatarsal ligament/tendon repair or reconstruction Shoulder Rotator Cuff Shoulder Shoulder Instability Shoulder Biceps Tenodesis Elbow Lateral epicondylitis repair Elbow Biceps tendon reattachment
Device Story
JuggerKnot® Soft Anchor is a suture-based fixation device for soft tissue-to-bone attachment. Device consists of a non-absorbable braided polyester suture anchor sleeve with spliced non-absorbable braided UHMWPE working sutures. Inserter features a metallic shaft with overmolded handle. Used in hospital/clinic/surgical settings by surgeons. Device is implanted into bone; anchor sleeve deploys to secure sutures, which are then used to reattach soft tissue (tendons/ligaments) to bone. Provides mechanical fixation to facilitate healing. Benefits include secure tissue reattachment in orthopedic procedures. Single-use, sterile, ethylene oxide sterilized.
Clinical Evidence
Bench testing only. Included biocompatibility (ISO 10993-1), sterilization validation, packaging stability (ISO 11607-1), endotoxin/pyrogenicity testing (ANSI/AAMI ST72, USP <161>, <151>, <85>), and usability engineering (EN 62366). Mechanical performance verified via insertion, cyclic, and pullout testing compared to predicate, demonstrating comparable fixation strength.
Technological Characteristics
Materials: UHMWPE and polyester sutures, metallic inserter shaft. Dimensions: USP standard sizes. Energy: None (mechanical). Connectivity: None. Sterilization: Ethylene oxide. Form factor: Suture sleeve anchor with inserter. No software/algorithm.
Indications for Use
Indicated for soft tissue to bone fixation in patients requiring repair or reconstruction of tendons and ligaments in the knee, hip, foot, ankle, shoulder, and elbow.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text. The overall design is clean and professional, reflecting the FDA's role in regulating food and drugs in the United States.
February 12, 2021
Riverpoint Medical Amanda Cole Regulatory Affairs Associate 825 NE 25th Ave. Portland, Oregon 97232
Re: K203740
Trade/Device Name: JuggerKnot® Soft Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: December 21, 2020 Received: December 22, 2020
Dear Amanda Cole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K203740
Device Name JuggerKnot® Soft Anchor
Indications for Use (Describe) Indicated for use in soft tissue to bone fixation for the following indications: Knee MPFL Knee Patellar tendon repair Knee MCL Knee Quadriceps tendon repair Hip Acetabular labral repair Hip Proximal hamstring repair. Hip Hip labral reconstruction. Foot and Ankle Achilles tendon repair Foot and Ankle Medial/lateral repair and reconstruction Foot and Ankle Plantar plate repair Foot and Ankle Mid- and forefoot repair Foot and Ankle Metatarsal ligament/tendon repair or reconstruction Shoulder Rotator Cuff Shoulder Shoulder Instability Shoulder Biceps Tenodesis Elbow Lateral epicondylitis repair Elbow Biceps tendon reattachment
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## JuggerKnot® Soft Anchor
## Submitter Information
| Submitter's Name: | Riverpoint Medical |
|----------------------|----------------------------------------|
| Address: | 825 NE 25th Ave.<br>Portland, OR 97232 |
| Phone Number: | (503) 517-8001 or 866 445-4923 |
| Fax Number: | (503) 517-8002 |
| Registration Number: | 3006981798 |
| Contact Person: | Amanda Cole<br>(503) 517-8001 |
| Date of Preparation: | December 21, 2020 |
## Device Name
| Trade Name: | JuggerKnot® Soft Anchor |
|------------------------|----------------------------------------------------|
| Common or Usual Names: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Name: | Smooth Or Threaded Metallic Bone Fixation Fastener |
## Device Classification
| FDA Class: | II |
|-------------------------|----------------------------------------------------------------------------------------------------------------|
| Product Classification: | 888.3040: Smooth Or Threaded Metallic Bone Fixation<br>Fastener |
| Classification Code: | MBI |
| Review Panel | Orthopedic |
| Premarket Review | Office of Device Evaluation<br>Division of Orthopedic Devices, Restorative and Repair<br>Devices Branch (RRDB) |
# Predicate Device
K150768 – Zimmer Biomet JuggerKnot® Soft Anchor
No reference devices were used in this submission.
{4}------------------------------------------------
## Device Description
The JuggerKnot® Soft Anchor is comprised of a suture sleeve structure and working suture. Non-absorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a non-absorbable braided polyester suture anchor sleeve. Up to three nonabsorbable round and flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white), blue, black or with or without a stainless steel needle attached. Available Suture sizes are standard according to USP requirements (dependent on suture type).
Suture supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of metallic shaft with overmolded handle. The device is sterilized by ethylene oxide gas, and is provided sterile for single use. JuggerKnot® Soft Anchors are available in common sizes and lengths with or without pre-attached 302 stainless steel needles and will be sold sterile for single use with no components or accessories. The device is intended for use in a hospital/clinic/surgical setting.
The classification for the JuggerKnot® Soft Anchor is FDA Class II device with product classification 21 CFR §888.3040: Smooth or threaded metallic bone fixation fastener, Product Code MBI.
## Intended Use and Indications for Use
JuggerKnot® Soft Anchors are intended for use in soft tissue to bone fixation for the following indications:
| Knee | MPFL |
|----------------|-----------------------------------------------------|
| Knee | Patellar tendon repair |
| Knee | MCL |
| Knee | Quadriceps tendon repair |
| Hip | Acetabular labral repair |
| Hip | Proximal hamstring repair |
| Hip | Hip Labral reconstruction |
| Foot and Ankle | Achilles tendon repair |
| Foot and Ankle | Medial/lateral repair and reconstruction |
| Foot and Ankle | Plantar plate repair |
| Foot and Ankle | Mid- and forefoot repair |
| Foot and Ankle | Metatarsal ligament/tendon repair or reconstruction |
| Shoulder | Rotator Cuff |
| Shoulder | Shoulder Instability |
| Shoulder | Biceps Tenodesis |
| Elbow | Lateral epicondylitis repair |
| Elbow | Biceps tendon reattachment |
{5}------------------------------------------------
#### Performance Data
The sutures used to construct the JuggerKnot® Soft Anchors meet requirements established by the United States Pharmacopeia (USP), except for diameter. The UHMWPE sutures are tested per USP performance requirements for needle attachment and tensile strength. FDA Guidance "Bone Anchors - Premarket Notification (510(k)) Submissions Guidance for Industry and Food and Drug Administration Staff' and FDA Guidance "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" were followed during the preparation of this submission. Non-clinical performance testing for the JuggerKnot® Soft Anchor included a sterilization adoption validation, biocompatibility testing per ISO10993-1:2018 - Biological Evaluation of Medical Devices, stability testing on the product packaging per ISO 11607-1:2006 - Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems, usability engineering validation with simulated use in a cadaveric models performed per EN62366: 2015- Medical devices -Application of usability engineering to medical devices. Endotoxin/pyrogenicity testing was performed per ANSI/AAMI ST72:2019, USP <161>, USP <151> and USP <85>. Non-clinical mechanical testing was performed to verify the fixation strength of the JuggerKnot® Soft Anchor using insertion, cyclic and pullout testing as compared to the predicate device. Results of performance testing for the JuggerKnot® Soft Anchor device concluded that the device performed comparably to the predicate device and to other currently marketed soft anchor devices in insertion, cyclic and pullout testing and the validations performed demonstrated that the JuggerKnot® Soft Anchor met all requirements for its intended use.
#### Substantial Equivalence and Comparison of Technical Characteristics
The JuggerKnot® Soft Anchor is substantially equivalent to the previously cleared JuggerKnot® Soft Anchor cleared per K150768 "predicate device." The JuggerKnot® Soft Anchor has the same intended use, the same principles of operation, and similar technological characteristics as the predicate device. Both the JuggerKnot® Soft Anchor and the predicate device are comprised of the same materials, packaged using the same packaging materials and sterilized using the same processes. The JuggerKnot® Soft Anchor subject device contains slight technological differences from the JuggerKnot® predicate device in the following ways: i) the flat braid configuration of UHMWPE sutures ii) triple loaded anchor configuration (up to three sutures spliced into the anchor) and iii) fixed (non-sliding) positioning of the suture anchor. The JuggerKnot® Soft Anchor predicate device anchor is a sliding anchor with up to two round braid UHMWPE working sutures spliced in the anchor. However, these technical characteristics is within the range of currently marketed devices. Therefore, the JuggerKnot® Soft Anchor "subject device" is substantially equivalent to the predicate devices in both technological characteristics and intended use and does not raise any issues of safety or effectiveness.
#### Conclusion
The information provided in this Traditional 510(k) demonstrates that the JuggerKnot® Soft Anchor substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.