Spark™ Clear Aligner System: The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion. Spark™ Software System: The Spark™ Software System is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays, retainers or bracket bonding templatesor based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the Spark™ Clear Aligner System and Spark™ Software System requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software.
Device Story
System consists of custom-manufactured, thin, clear plastic removable aligners; designed to move teeth incrementally. Clinician uses Spark Software (Web, Design, Anatomy, Approver, Fabrication modules) to manage patient profiles, plan treatments, and submit digital scans. Technicians create 3D models of patient dentition and final treated state; clinician reviews and approves models via Approver software. Fabrication module produces resin molds for thermoforming aligners. Clinician provides sequential aligners to patient; patient wears trays to apply continuous gentle force. Optional features include attachments, hooks, bite ramps, pontics, and posterior bite turbos. Used in clinical settings by orthodontists/dental professionals to correct malocclusion. Benefits include aesthetic, removable tooth movement.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and comparison to legally marketed predicate devices. Bench testing included biocompatibility (ISO 10993 series), software life-cycle (IEC 62304), usability (IEC 62366-1), and mechanical properties (ASTM D790).
Technological Characteristics
Materials: Thermoplastic polyurethane-polyester composite resin. Principle: Orthodontic tooth movement via continuous gentle force from custom-molded aligners. Connectivity: Web-based software portal for clinician submission and review. Software: Proprietary modules for 3D modeling and treatment planning. Standards: ISO 10993-1, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 7405, ISO 14971, IEC 62304, IEC 62366-1, ASTM D790-10, ASTM F1980-16.
Indications for Use
Indicated for alignment of teeth during orthodontic treatment of malocclusion in patients requiring orthodontic intervention.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2021
Ormco Corporation Frank Ray Director of Regulatory Affairs 200 S Kraemer Blvd Brea, California 92821
Re: K203737
Trade/Device Name: Spark Clear Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC, PNN Dated: December 21, 2020 Received: December 22, 2020
Dear Frank Ray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203737
Device Name SparkTM Clear Aligner System
Indications for Use (Describe) SparkTM Clear Aligner System: The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
### SparkTM Software System:
The Spark™ Software System is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays, retainers or bracket bonding templatesor based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Spark™ Clear Aligner System and Spark™ Software System requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ormco. The logo is in blue and consists of the word "Ormco" in a bold, sans-serif font. To the right of the word "Ormco" is the trademark symbol. Below the word "Ormco" is the phrase "YOUR PRACTICE. OUR PRIORITY." in a smaller, sans-serif font.
# Section V - 510(k) Summary for Spark™ Clear Aligner System K203737
#### 1. Submitter Information:
Ormco Corporation 200 S. Kraemer Blvd, Brea, CA 92821
| Contact Person: | Frank Ray |
|-------------------|----------------|
| Telephone Number: | (704) 587-7227 |
| Fax Number: | (704) 587-7250 |
Date Prepared: December 21, 2020
#### 2. Device Name:
3.
| Proprietary Name: | Spark™ Clear Aligner System |
|-----------------------------|-----------------------------|
| Manufacturer: | Ormco Corporation |
| Common Name: | Aligner, Sequential |
| Classification Name: | Orthodontic Plastic Bracket |
| CFR Number: | 872.5470 |
| Device Class: | II |
| Product Code: | NXC, PNN |
| Predicate Device (Primary): | |
- Proprietary Name: Invisalign System (K143630) ●
- Manufacturer: Align Technology, Inc. ●
- Aligner, Sequential ● Common Name:
- Orthodontic Plastic Bracket ● Classification Name:
- CFR Number: 872.5470 ● II
- Device Class:
- NXC Product Code: ●
# Reference Device:
| Proprietary Name: | Ormco™ Spark™ Aligner System (K182826) |
|----------------------|----------------------------------------|
| Manufacturer: | Ormco Corporation |
| Common Name: | Aligner, Sequential |
| Classification Name: | Orthodontic Plastic Bracket |
| CFR Number: | 872.5470 |
| Device Class: | II |
- NXC Product Code: ●
{4}------------------------------------------------
## Reference Device #2:
- . Proprietary Name: Arcad SmileStudio and Aligner System (K192244)
- Manufacturer: ●
- Common Name: Aligner, Sequential
ArcadLab
- Classification Name: Orthodontic Plastic Bracket
=
- CFR Number: 872.5470
- o Device Class:
- Product Code: NXC, PNN o
#### Description of Device: 4.
o
●
0
The Ormco Spark™ Clear Aligner System consists of a series of doctor-prescribed, custom manufactured, thin, clear plastic removable orthodontic appliances (aligners) that gently move the participant's teeth in small increments from their original state to a more optimal, treated state.
Treatment planning, aligner design and aliqner manufacture are supported by a proprietary software system. The Spark™ Clear Aligner system consists of multiple interfacing software modules; Web, Design, Anatomy, Approver, and Fabrication. The Spark Aligner Web software module is an online portal, where clinicians can create new patient profiles, plan treatments, manage patients and submit prescriptions, photos, images and digital scans to Ormco. Trained technicians use the Ormco™ Spark™ Aligner Anatomy and Design software to create a 3D model of the patient's teeth from dental scans and to generate a 3D image of a final, treated state, as well as 3D aligner transitional treatment stage models, and submits them to the clinician. The clinician uses the Approver Software to review and approve the models, then Ormco uses the Fabrication software to produce resin molds for thermoforming the aligners.
The clinician receives the aligners and provides them, in sequential "stages", to the patient, confirming fit and monitoring treatment from the placement of the first aligner to the removal of the final aligner. The trays are held in place by pressure and can be removed by the patient at any time.
The aligners are individually identified and dispensed to patients and are to be worn in a specific, prescribed sequence. Several treatment options may be integrated into the Ormco Spark aligner consisting of:
- . Attachments: Attachments or "buttons" may be prescribed by the dental practitioner to facilitate tooth movement and aligner anchorage. The dental practitioner may choose a standard dental composite and adhesive to bond the attachments to the dentition.
- . Hooks: Hooks may be designed into the aligner, then connected by an elastic to a tooth-bonded button on the opposite arch, to apply additional forces.
- Bite ramps: Bite ramps are step features built into the lingual surfaces of the upper aligner arch that used by the clinician to prevent movement of the teeth during overbite correction treatment.
- Pontics: Pontics are cavity spaces (that may or may not be "tooth-shaped") built into the aligner per clinician's request to fill voids of missing teeth or other gaps that the clinician may wish to retain during treatment.
- Posterior Bite Turbos: Posterior Bite Turbos (PBT) are designed to prevent complete closure of jaws.
{5}------------------------------------------------
# Principle of Operation / Mechanism of Action:
Orthodontic tooth movement occurs through forces applied by the appliance to the dentition as each tooth follows the programmed displacement based on a dental clinician's prescription.
#### 5. Indications for Use:
Spark™ Clear Aligner System:
The Spark™ Clear Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Spark™ Software System:
The Spark™ Software System is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays, retainers or bracket bonding templates based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/ desired treatment objectives.
The use of the Spark™ Clear Aligner System and Spark™ Software System requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software.
#### 6. Description of Substantial Equivalence:
The Spark™ Clear Aligner System design or technical features have not been significantly modified since its clearance in 2018 (Reference Device #1, K182826). This section provides the substantial equivalence rationale for the Spark™ Clear Aligner System and the respective predicate devices with regards to Indications for Use, Technology, and Performance Testing. The differences between the subject device and the predicates do not raise different questions of substantial equivalence.
The similarities between the Spark™ Clear Aligner System (Subject Device) and the Primary Predicate Device Invisalign System (K143630), Reference Device #1 Ormco™ Spark™ Aligner System (K182826), and Reference Device #2 Arcad SmileStudio and Aligner System (K192244) as listed in Table 12.2 below are the Intended Use of the device, Indications for Use, Mode of Action, 3D software description, and treatment process. Also, the Attachments, Hook/Cutouts, Bite Ramps, and Pontics are available options. Furthermore, Biocompatibility testing, Software testing, and Material testing processes are similar.
The difference between the Spark™ Clear Aligner System (Subject Device) and Reference Device #1 Ormco™ Spark™ Aliqner System (K182826) is the material. The Thermoplastic polyurethane-polyester composite resin is the same however the composition slightly varies.
Table 5.2 below compares the Spark™ Clear Aligner System to the Primary predicate device (Invisalign System (K143630)), reference device #1 (Ormco™ Spark™ Aligner
{6}------------------------------------------------
System (K182826)), and reference device #2 (Arcad SmileStudio and Aligner System (K192244)) with respect to intended use, technological characteristics and performance testing.
### Device Comparison Table:
### Table 5.2
| Descriptive<br>Information | Subject Device<br>Spark™ Clear<br>Aligner System | Predicate Device<br>(Primary)<br>Invisalign System<br>(K143630) | Reference<br>Device #1<br>Ormco™ Spark™<br>Aligner System<br>(K182826) | Reference<br>Device #2<br>Arcad<br>SmileStudio<br>and Aligner<br>System<br>(K192244) | Comparison |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pictorial<br>Representation | Image: Spark Clear Aligner System | Image: Invisalign System | Image: Ormco Spark Aligner System | Image: Arcad SmileStudio and Aligner System | N/A |
| Regulatory Classification | | | | | |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR<br>872.5470 | Same as<br>Predicates |
| Regulation Title | Orthodontic<br>Plastic Bracket | Orthodontic Plastic<br>Bracket | Orthodontic Plastic<br>Bracket | Orthodontic<br>Plastic Bracket | Same as<br>Predicates |
| Regulation Class | II | II | II | II | Same as<br>Predicates |
| Product Code | NXC, PNN | NXC | NXC | NXC, PNN | Same as<br>Predicates |
| Indications for Use/Intended Use | | | | | |
| Indications for<br>Use | Spark™ Clear<br>Aligner System:<br>The Spark™ Clear<br>Aligner System is<br>indicated for the<br>alignment of<br>teeth during<br>orthodontic<br>treatment of<br>malocclusion.<br>Spark™ Software<br>System:<br>The Spark™<br>Software System<br>is intended for<br>use as a medical<br>front-end device<br>providing tools<br>for management<br>of orthodontic<br>models,<br>systematic<br>inspection,<br>detailed analysis,<br>treatment<br>simulation and<br>virtual design of<br>a series of dental<br>casts, which may<br>be used for<br>sequential aligner<br>trays, retainers<br>or bracket<br>bonding<br>templates based | The Invisalign System<br>is indicated for the<br>alignment of teeth<br>during orthodontic<br>treatment of<br>malocclusion. | The Ormco™<br>Spark™ Aligner<br>System is indicated<br>for the alignment of<br>teeth during<br>orthodontic<br>treatment of<br>malocclusion in<br>patients with<br>permanent<br>dentition (i.e., all<br>second molars).<br>The Ormco™<br>Spark™ Aligner<br>System positions<br>teeth by way of<br>continuous gentle<br>force. | The Arcad<br>SmileStudio is<br>intended for<br>use as a<br>medical front-<br>end device<br>providing tools<br>for<br>management of<br>orthodontic<br>models,<br>systematic<br>inspection,<br>detailed<br>analysis,<br>treatment<br>simulation and<br>virtual design<br>of a series of<br>dental casts,<br>which<br>may be used<br>for sequential<br>aligner trays or<br>retainers,<br>based on 3D<br>models of the<br>patient's<br>dentition before<br>the start of an<br>orthodontic<br>treatment. It<br>can also be<br>applied during<br>the treatment<br>to inspect and | Same as<br>Predicates<br>The Indications<br>for Use for the<br>bracket is<br>substantial<br>equivalent to<br>the Primary<br>Predicate<br>device.<br>The Indications<br>for Use for the<br>software is<br>substantial<br>equivalent to<br>the Reference<br>device #2. |
{7}------------------------------------------------
| Intended Use | on 3D models of<br>the patient's<br>dentition before<br>the start of an<br>orthodontic<br>treatment. It can<br>also be applied<br>during the<br>treatment to<br>inspect and<br>analyze the<br>progress of the<br>treatment. It can<br>be used at the<br>end of the<br>treatment to<br>evaluate if the<br>outcome is<br>consistent with<br>the<br>planned/desired<br>treatment<br>objectives.<br>The use of the<br>Spark™ Clear<br>Aligner System<br>and Spark™<br>Software System<br>requires the user<br>to have the<br>necessary<br>training and<br>domain<br>knowledge in the<br>practice of<br>orthodontics, as<br>well to have<br>received a<br>dedicated training<br>in the use of the<br>software. | analyze the<br>progress of the<br>treatment. It<br>can be used at<br>the end of the<br>treatment to<br>evaluate if the<br>outcome is<br>consistent with<br>the<br>planned/desired<br>treatment<br>objectives.<br>The Arcad<br>Aligner System<br>is indicated for<br>the treatment<br>of tooth<br>malocclusion in<br>patients with<br>permanent<br>dentition (i.e.<br>all second<br>molars). The<br>Arcad Aligner<br>System<br>positions teeth<br>by way of<br>continuous<br>gentle force.<br>The use of the<br>Arcad Aligner<br>System and<br>SmileStudio<br>requires the<br>user to have<br>the necessary<br>training and<br>domain<br>knowledge in the<br>practice of<br>orthodontics, as<br>well to have<br>received a<br>dedicated<br>training in the<br>use of the<br>software. | N/A | Same as Predicates | | | | | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-----------------------|----------------------------|----------|--|-----|-----------------------|
| | Description of<br>Appliance<br>Application | Orthodontic tooth movement | movement | Orthodontic tooth movement | movement | Orthodontic tooth movement | movement | | N/A | Same as<br>Predicates |
| Intended User | Dental<br>Professional | Dental Professional | Dental Professional | N/A | Same as<br>Predicates | | | | | |
| <b>Technological Characteristics</b> | | | | | | | | | | |
| Mode of Action | Orthodontic tooth<br>movement occurs<br>through forces<br>applied by the<br>appliance to the<br>dentition as each<br>tooth follows the<br>programmed<br>displacement<br>based on a | Orthodontic tooth<br>movement occurs<br>through forces applied<br>by the appliance to<br>the dentition as each<br>tooth follows the<br>programmed<br>displacement based<br>on a doctor's<br>prescription. | Orthodontic tooth<br>movement occurs<br>through forces<br>applied by the<br>appliance to the<br>dentition as each<br>tooth follows the<br>programmed<br>displacement based<br>on a doctor's<br>prescription. | N/A | Same as<br>Predicates | | | | | |
{8}------------------------------------------------
| | doctor's<br>prescription. | | | | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-------------------------|
| | | | | | |
| 3-D Software<br>Description | The Spark™ Clear<br>Aligner System 3-<br>D software uses<br>scanned teeth<br>data, landmarks<br>and the clinician's<br>prescription to<br>design a<br>corrected case<br>setup for the<br>clinician's review.<br>The output files<br>from this<br>software are sent<br>to the clinician,<br>who may suggest<br>improvements or<br>approve as-is for<br>manufacture. The<br>software is used<br>to produce pre-<br>molds and molds<br>needed for the<br>manufacturing of<br>series of custom-<br>made aligners. | The Align 3-D<br>Software uses a scan<br>of a PVS impression<br>or a digital scan<br>(which represents an<br>untreated state) to<br>generate the image of<br>a final, treated state<br>and then interprets a<br>series of images that<br>represent<br>intermediate teeth<br>states. The dental<br>practitioner then<br>reviews these images<br>and has the option to<br>reject or request<br>modifications to the<br>set-up prior to<br>approving it for<br>aligner fabrication.<br>Once the dental<br>practitioner approves<br>the treatment plan,<br>the software converts<br>the files to produce<br>the series of custom-<br>made aligners. | The Ormco™<br>Spark™ Aligner<br>System 3-D<br>software uses<br>scanned teeth data,<br>landmarks and the<br>clinician's<br>prescription to<br>design a corrected<br>case setup for the<br>clinician's review.<br>The output files<br>from this software<br>are sent to the<br>clinician, who may<br>suggest<br>improvements or<br>approve as-is for<br>manufacture. The<br>software is used to<br>produce pre-molds<br>and molds needed<br>for the<br>manufacturing of<br>series of custom-<br>made aligners. | N/A | Same as<br>Predicates |
| Attachments | Available | Available | Available | N/A | Same as<br>Predicates |
| Hook/Cutouts | Available | Available | Available | N/A | Same as<br>Predicates |
| Bite ramps | Available | Available | Available | N/A | Same as<br>Predicates |
| Pontics | Available | Available | Available | N/A | Same as<br>Predicates |
| Single Use | Yes | Yes | Yes | N/A | Same as<br>Predicates |
| Non-Sterile<br>Packaging | Yes | Yes | Yes | N/A | Same as<br>Predicates |
| Treatment<br>Process | The Spark™ Clear<br>Aligner System 3-<br>D software uses<br>scanned teeth<br>data, landmarks<br>and the clinician's<br>prescription to<br>design a<br>corrected case<br>setup for the<br>clinician's review.<br>The output files<br>from this<br>software are sent<br>to the clinician,<br>who may suggest<br>improvements or<br>approve as-is for<br>manufacture. The<br>software is used<br>to produce pre-<br>molds and molds | The Align 3-D<br>Software uses a scan<br>of a PVS impression<br>or a digital scan<br>(which represents an<br>untreated state) to<br>generate the image of<br>a final, treated state<br>and then interprets a<br>series of images that<br>represent<br>intermediate teeth<br>states. The dental<br>practitioner then<br>reviews these images<br>and has the option to<br>reject or request<br>modifications to the<br>set-up prior to<br>approving it for<br>aligner fabrication.<br>Once the dental | The Ormco™<br>Spark™ Aligner<br>System 3-D<br>software uses<br>scanned teeth data,<br>landmarks and the<br>clinician's<br>prescription to<br>design a corrected<br>case setup for the<br>clinician's review.<br>The output files<br>from this software<br>are sent to the<br>clinician, who may<br>suggest<br>improvements or<br>approve as-is for<br>manufacture. The<br>software is used to<br>produce pre-molds<br>and molds needed | N/A | Same as<br>Predicates |
| | needed for the<br>manufacturing of<br>series of custom-made aligners. | practitioner approves<br>the treatment plan,<br>the software converts<br>the files to produce<br>the series of custom-made aligners. | for the<br>manufacturing of<br>series of custom-made aligners. | | |
| Material | Thermoplastic<br>polyurethane-<br>polyester composite<br>resin | Thermoplastic<br>polyurethane-<br>polyester composite<br>resin | Thermoplastic<br>polyurethane-<br>polyester composite<br>resin | N/A | Same as<br>Predicates |
| | Demonstrates<br>sufficient flexural<br>strength, stress<br>retention, stain<br>resistance, and<br>transparency for<br>use as a clear<br>aligner. | Demonstrates<br>sufficient flexural<br>strength, stress<br>retention, stain<br>resistance, and<br>transparency for use<br>as a clear aligner. | Demonstrates<br>sufficient flexural<br>strength, stress<br>retention, stain<br>resistance, and<br>transparency for<br>use as a clear<br>aligner. | N/A | Same as<br>Predicates |
| | Performance testing | | | | |
| Biocompatibility<br>testing | Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within<br>a risk<br>management<br>system). | Biocompatible<br>according to ISO<br>10993-1<br>(Biological evaluation<br>of medical devices -<br>Part 1: Evaluation and<br>testing within a risk<br>management<br>system). | Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within a<br>risk management<br>system). | N/A | Same as<br>Predicates |
| | IEC 62304:<br>2015-06<br>CONSOLIDATED<br>VERSION<br>Edition 1.1 -<br>Medical device<br>software -<br>Software life-<br>cycle processes. | N/A | IEC 62304:<br>2015-06<br>CONSOLIDATED<br>VERSION<br>Edition 1.1 - Medical<br>device software -<br>Software life-cycle<br>processes. | N/A | Same as<br>Predicate #2 |
| | ASTM D790-10:<br>2010-04-01<br>Standard Test<br>Methods for<br>Flexural<br>Properties of<br>Unreinforced and<br>Reinforced<br>Plastics and<br>Electrical<br>Insulating<br>Materials. | N/A | ASTM D790-10:<br>2010-04-01<br>Standard Test<br>Methods for Flexural<br>Properties of<br>Unreinforced and<br>Reinforced Plastics<br>and Electrical<br>Insulating<br>Materials. | N/A | Same as<br>Predicate #2 |
{9}------------------------------------------------
## Non-Clinical Test Data:
Performance bench testing according to international standards for Aligners, Sequential has been conducted to determine conformance in regards to:
- o Biocompatibility has been completed for the applicable components.
- Comparative performance testing of the functions of the Proposed device compared to the . cleared stand-a-lone device.
Furthermore, the performance of the Spark™ Clear Aligner System has been verified utilizing the following standards:
{10}------------------------------------------------
- ISO 10993-1:2018: Biological evaluation of medical devices Part 1: Evaluation and . testing within a risk management process.
- o ISO 10993-5:2009: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
- . ISO 10993-6:2016: Biological evaluation of medical devices - Part 6: Tests for local effects after implantation.
- ISO 10993-10:2010: Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization.
- ISO 10993-11:2017: Biological evaluation of medical devices—Part 11: Tests for systemic ● toxicity.
- ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry.
- ISO 14971:2007: Medical devices Application of risk management to medical devices. ●
- . IEC 62304:2015-06: Medical device software - Software life-cycle processes.
- IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering to medical ● devices.
- o ASTM D790-10: 2010: Standard Test Methods for Flexural Properties of Unreinforced and Reinforced Plastics and Electrical Insulating Materials.
- . ASTM F1980-16: 2016: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
# Clinical Performance Data:
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on wellestablished scientific and engineering principles. Clinical testing has not been conducted on this product.
## Conclusion as to Substantial Equivalence:
Based on a comparison of intended use, indications, material composition, technological characteristics, principle of operation, features and performance data, the Spark™ Clear Aligner System is deemed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.