← Product Code [NBW](/productcode/NBW) · K203711

# IWL2020 Blood Glucose Monitoring System (K203711)

_Roche Diabetes Care, Inc. · NBW · May 6, 2022 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203711

## Device Facts

- **Applicant:** Roche Diabetes Care, Inc.
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** May 6, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The IWL2020 Blood Glucose Monitoring System is comprised of the WL2020 test strips. The IWL2020 Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip as an aid in monitoring the effectiveness of glucose control. The IWL2020 Blood Glucose Monitoring System is intended for in vitro diagnostic single patient use by people with diabetes at home. The IWL2020 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is not for use in diagnosis or screening of diabetes mellitus, nor for neonatal use.

## Device Story

Handheld blood glucose monitoring system; includes meter, test strips, and control solutions. Uses amperometric assay with FAD-dependent glucose dehydrogenase (GDH) enzyme on test strips. User applies capillary blood from fingertip to strip via capillary action; meter detects electrons generated by glucose-enzyme reaction; magnitude of current proportional to glucose concentration. Results displayed on screen, stored in memory, and transmissible via Bluetooth Low Energy (BLE). Intended for home use by patients with diabetes. Provides quantitative glucose readings to aid in monitoring glucose control. Factory calibrated; no user calibration required. Meter supports cleaning and disinfection for five-year expected lifetime.

## Clinical Evidence

Clinical user evaluation study (n=350) with lay users performing self-testing compared to Roche/Hitachi cobas c 501 PCA-HK reference method. Results: 99% of results within ±20% of laboratory values; 96% within ±15%; 88% within ±10%; 63% within ±5%. Study confirmed lay users can obtain accurate readings and are satisfied with ease of operation.

## Technological Characteristics

Amperometric glucose test system; FAD-GDH enzyme. Handheld meter powered by two 3V CR2032 batteries. Connectivity via Bluetooth Low Energy (BLE). Operating range: 6-44°C, 10-90% RH. Factory calibrated. Compliant with ISO 14971, IEC 62304, IEC 61010-1, and CLSI EP06-A.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Accu-Chek Guide Blood Glucose Monitoring System ([K160944](/device/K160944.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number

K203711

B Applicant

Roche Diabetes Care Inc.

C Proprietary and Established Names

IWL2020 Blood Glucose Monitoring System

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NBW | Class II | 21 CFR 862.1345
Glucose test system | CH- Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

New device

B Measurand:

Glucose in capillary whole blood from fingertip

C Type of Test:

Quantitative amperometric assay, glucose dehydrogenase (FAD-GDH)

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The IWL2020 Blood Glucose Monitoring System is comprised of the IWL2020 meter and IWL2020 test strips.

The IWL2020 Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip as an aid in monitoring the effectiveness of glucose control.

The IWL2020 Blood Glucose Monitoring System is intended for in vitro diagnostic single patient use by people with diabetes at home.

The IWL2020 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

This system is not for use in diagnosis or screening of diabetes mellitus, nor for neonatal use.

C Special Conditions for Use Statement(s):

OTC - Over The Counter

Do not use the meter at high hematocrit levels above 65% or low hematocrit levels below 10%.

Not for use in diagnosis or screening of diabetes mellitus.

Not for neonatal use.

The IWL2020 Blood Glucose Monitoring System is not indicated for Alternative Site Testing (AST).

Abnormally high concentrations (greater than 5 mg/dL) of ascorbic acid (vitamin C) may cause inaccurate results. High-dose vitamin C therapy that would result in abnormally high concentrations is typically prescribed by your healthcare professional. If you are not sure if this applies to you, please check with your healthcare professional.

Do not use the meter system to measure blood glucose in people who are experiencing cardiovascular collapse (severe shock) or decreased peripheral blood flow.

K203711 - Page 2 of 12

{2}

Do not use during or soon after xylose absorption testing since xylose may cause inaccurate results. Xylose absorption testing is performed under the supervision of a doctor. Ask your doctor how long to wait after xylose testing before performing a blood glucose test.

Not for use on critically ill patients, patients in shock, dehydrated patients, or hyperosmolar patients with or without ketosis.

This system has not been tested at altitudes higher than 10,150 feet.

For single patient use only

Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.

## D Special Instrument Requirements:

IWL2020 Meter

## IV Device/System Characteristics:

### A Device Description:

The IWL2020 Blood Glucose Monitoring System consists of the following:

- IWL2020 Meter (with two 3-volt lithium CR2032 batteries)
- IWL2020 Test Strips (may be sold separately)
- IWL2020 Control Solutions (Level 1 and Level 2, may be sold separately)
- Quick Start Guide

The IWL2020 Blood Glucose Monitoring System is a handheld device that incorporates features to aid in self-monitoring of blood glucose. The blood glucose results are displayed on the screen and stored in the meter's memory, and may also be transmitted via Bluetooth Low Energy (BLE) wireless communication.

### B Principle of Operation:

The IWL2020 Blood Glucose Monitoring System quantitatively measures glucose amperometrically, and includes the enzyme on the test strip, FAD-dependent glucose dehydrogenase (GDH), which converts the glucose in the blood to gluconolactone. When a drop of whole blood is applied to the reaction cell on the test strip, glucose in the blood sample reacts in the presence of GDH-FAD and a mediator. The electrons generated during this reaction are detected by the meter. The magnitude of the resultant current is proportional to the concentration of glucose in the blood and is converted to a glucose concentration. The glucose concentration is displayed on the meter display for the user.

### C Instrument Description Information:

1. Instrument Name:

IWL2020 Meter

K203711 - Page 3 of 12

{3}

K203711 - Page 4 of 12

2. Specimen Identification:

Capillary whole blood from the user's fingertip

3. Specimen Sampling and Handling:

The system is intended to be used with capillary whole blood from the fingertip. The whole blood sample is applied directly to the test strip by capillary action. Samples are to be tested immediately upon collection.

4. Calibration:

The IWL2020 meter is factory calibrated. No user calibration is required.

5. Quality Control:

IWL2020 control solutions (Level 1 and Level 2) are for use with the IWL2020 Blood Glucose Monitoring System to check that the meters and test strips are working together properly and that the test is performing correctly. The ranges for Level 1 and Level 2 controls are printed on the test strip container label. The meter automatically recognizes the difference between the control solution and blood. Instructions on when to perform a control test, details about the control solution, performing a control test, understanding out-of-range control results are provided in the meter manual. Control solution ranges are printed on the control solution package insert.

V Substantial Equivalence Information:

A Predicate Device Name(s):

ACCU-CHEK Guide Blood Glucose Monitoring System

B Predicate 510(k) Number(s):

K160944

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K203711 | K160944  |
| --- | --- | --- |
|  Device Trade Name | IWL2020 Blood Glucose Monitoring System | Accu-Chek Guide Blood Glucose Monitoring System  |

{4}

|  General Device Characteristic Similarities |  |   |
| --- | --- | --- |
|  Intended Use/Indications For Use | Same | Intended to quantitatively measure glucose (sugar) in fresh capillary whole blood samples as an aid in monitoring the effectiveness of glucose control.  |
|  Test principle | Same | Amperometric Detection  |
|  Sample volume | Same | 600 nanoliters  |
|  Measurement Range | Same | 20-600 mg/dL  |
|  General Device Characteristic Differences |  |   |
|  Site Testing | Fingertip | Palm, Forearm, and Upper Arm  |

VI Standards/Guidance Documents Referenced:

- ISO14971-Third edition 2019-12-“Medical Devices-Application of Risk Management to Medical Devices”
- IEC-62304 Edition 1.1 2015-06-“Medical device software-Software life cycle processes”
- CLSI-EP06-A-“Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach”
- IEC-61010-1 edition 3.0 2010-06-“Safety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements”

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Repeatability (within-run precision) studies were performed with contrived venous whole blood samples at 5 glucose concentration ranges using three test strip lots and 10 IWL2020 meters. Ten runs (vials) were performed on each sample with 10 replicates per strip lot resulting in a total of 100 replicates collected for each test strip lot and each glucose level tested. Results are summarized below:

K203711 - Page 5 of 12

{5}

K203711 - Page 6 of 12

|  Glucose Level | Strip Lot | Target Glucose (mg/dL) | N | Mean Glucose (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Level 1 (30-50) | 1 | 40 | 100 | 38.2 | 1.1 | 2.9  |
|   |  2 | 40 | 100 | 38.0 | 1.1 | 2.8  |
|   |  3 | 40 | 100 | 38.7 | 1.2 | 3.1  |
|   |  Combined | 40 | 300 | 38.3 | 1.2 | 3.0  |
|  Level 2 (51-110) | 1 | 80 | 100 | 82.6 | 1.5 | 1.8  |
|   |  2 | 80 | 100 | 83.2 | 1.7 | 2.0  |
|   |  3 | 80 | 100 | 81.8 | 1.9 | 2.3  |
|   |  Combined | 80 | 300 | 82.5 | 1.8 | 2.2  |
|  Level 3 (111-150) | 1 | 130 | 100 | 138.6 | 3.1 | 2.3  |
|   |  2 | 130 | 100 | 140.5 | 2.4 | 1.7  |
|   |  3 | 130 | 99 | 137.0 | 2.3 | 1.6  |
|   |  Combined | 130 | 299 | 138.7 | 3.0 | 2.1  |
|  Level 4 (151-250) | 1 | 200 | 100 | 218.1 | 4.0 | 1.9  |
|   |  2 | 200 | 100 | 219.3 | 3.7 | 1.7  |
|   |  3 | 200 | 100 | 212.8 | 4.4 | 2.1  |
|   |  Combined | 200 | 300 | 216.7 | 4.9 | 2.3  |
|  Level 5 (251-400) | 1 | 325 | 100 | 369.7 | 8.1 | 2.2  |
|   |  2 | 325 | 100 | 370.9 | 8.4 | 2.3  |
|   |  3 | 325 | 100 | 360.3 | 9.1 | 2.5  |
|   |  Combined | 325 | 300 | 367.0 | 9.8 | 2.7  |

Intermediate precision was evaluated using three levels of glucose control solutions. Each sample was measured in replicates of ten using ten IWL2020 meters, three lots of test strips over ten days, resulting in a total of 300 tests for each glucose level. Results are summarized below:

|  Glucose Level | Strip Lot | N | Mean Glucose (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- | --- |
|  Level 1 (51-110) | 1 | 100 | 44.5 | 1.3 | 2.8  |
|   |  2 | 100 | 44.8 | 1.3 | 2.8  |
|   |  3 | 100 | 44.0 | 1.4 | 3.2  |
|   |  Combined | 300 | 44.4 | 1.3 | 3.0  |
|  Level 2 (111-150) | 1 | 100 | 113.3 | 3.0 | 2.7  |
|   |  2 | 100 | 114.9 | 2.8 | 2.5  |
|   |  3 | 100 | 112.5 | 2.5 | 2.3  |
|   |  Combined | 300 | 113.6 | 3.0 | 2.6  |
|  Level 3 (251-400) | 1 | 100 | 291.7 | 6.6 | 2.3  |
|   |  2 | 100 | 294.8 | 5.9 | 2.0  |
|   |  3 | 100 | 289.4 | 6.5 | 2.2  |
|   |  Combined | 300 | 291.9 | 6.7 | 2.3  |

{6}

K203711 - Page 7 of 12

2. Linearity:

Linearity was evaluated using three test strip lots and 36 IWL2020 meters by testing 11 venous whole blood samples with glucose concentrations ranging from 5.6 to 627.8 mg/dL (5.6, 14.8, 37.7, 58.4, 94.2, 127.1, 153.1, 211.0, 282.0, 471.6, 627.8 mg/dL). One run (across all 36 meters) was performed with each sample type using each of the strip lots. The values from the IWL2020 meter were compared with those obtained from the comparator method (hexokinase reagent on the Roche Cobas c 501 analyzer). The results from regression analysis are summarized below:

Test strip lot #1: y=1.080 x+2.3817; R² = 0.9985
Test strip lot #2: y=1.06 x+1.0307; R² = 0.9987
Test strip lot #3: y=1.06 x+0.6098; R² = 0.9986

The results of the study support the sponsor's claimed glucose measurement range of 20 to 600 mg/dL. The meter will display "LO" when the result is less than 20 mg/dL and "HI" when result is greater than 600 mg/dL. The sponsor validated the "LO" and "HI" functions and demonstrated that they functioned as intended.

3. Analytical Specificity/Interference:

To assess potential interference, the sponsor used venous whole blood samples adjusted to three different glucose levels of 60, 120, 250 mg/dL, split into control sample and test samples. Various endogenous and exogenous substances were then added to the test sample only. The % difference between the test and control sample was calculated and the concentration tested at which no significant interference (defined as ± 10 mg/dL for glucose concentrations &lt; 100 mg/dL and ± 10% for glucose concentrations ≥ 100 mg/dL) was observed is presented in the table below:

|  Potential Interfering Substance | Highest concentration at which no significant interference is observed (mg/dL)  |
| --- | --- |
|  Acetaminophen | 20  |
|  Ascorbic acid | 5  |
|  Bilirubin (unconjugated) | 60  |
|  Cholesterol | 500  |
|  Creatinine | 30  |
|  Dopamine | 0.09  |
|  Galactose | 300  |
|  Gentisic Acid | 1.8  |
|  Glutathione | 6.14  |
|  Hemoglobin | 1000  |
|  Heparin | 8000 U/dL  |
|  Ibuprofen | 50  |
|  Lactitol | 100  |

{7}

|  Potential Interfering Substance | Highest concentration at which no significant interference is observed (mg/dL)  |
| --- | --- |
|  Isomalt | 0.09  |
|  Potential Interfering Substance | Highest concentration at which no significant interference is observed (mg/dL)  |
| --- | --- |
|  Sodium | 165 mmol/L  |
|  Xylitol | 200  |
|  L-Dopa | 2  |
|  Maltitol | 20.2  |
|  Maltose | 685  |
|  Mannitol | 1800  |
|  Methyl Dopa | 1.5  |
|  Pyridine aldoxime Methiodide (PAM) | 25  |
|  Salicylic Acid | 60  |
|  Sorbitol | 70  |
|  Tolazamide | 200  |
|  Tolbutamide | 100  |
|  Triglycerides | 1800  |
|  Uric Acid | 23.5  |
|  Xylose | 10  |
|  EDTA | 360  |
|  Sorbitol | 70  |

The sponsor has the following limitations in their labeling:

- Abnormally high concentrations (greater than 5 mg/dL) of ascorbic acid (vitamin C) may cause inaccurate results. High-dose vitamin C therapy that would result in abnormally high concentrations is typically prescribed by your healthcare professional. If you are not sure if this applies to you, please check with your healthcare professional.
- Do not use during or soon after xylose absorption testing since xylose may cause inaccurate results. Xylose absorption testing is performed under the supervision of a doctor. Ask your doctor how long to wait after xylose testing before performing a blood glucose test.

4. Assay Reportable Range:

The results of the linearity study support the sponsor's claimed glucose measurement range of 20 to 600 mg/dL.

K203711 - Page 8 of 12

{8}

K203711 - Page 9 of 12

5. Traceability, Stability, Expected Values:

Traceability: According to the sponsor, the IWL2020 Blood Glucose Monitoring System is traceable to the NIST Standard Reference Material (SRM) 917 glucose reference material. A method comparison was performed using the IWL2020 Blood Glucose Monitoring System and hexokinase reagent on the Roche Cobas c 501 analyzer as the comparator method (see section B.1).

Test Strip Stability: IWL2020 Test Strip stability assessed using real time stability studies. The protocols and acceptance criteria were reviewed and found to be acceptable to support the labeling claims that the test strips are stable both closed (shelf life) and open-vial for within 18 months when stored at the recommended storage temperatures of 39-86°F (4-30°C) and 10-90% relative humidity (RH). The labeling instructs the users not to freeze the test strips.

6. Detection Limit:

The reportable range for the IWL2020 Blood Glucose Monitoring System is 20 to 600 mg/dL.

7. Assay Cut-Off:

Not applicable.

8. Accuracy (Instrument):

See method comparison study (section VII.C3).

9. Carry-Over:

Not applicable.

B Comparison Studies:

1. Method Comparison with Predicate Device:

See section VII.C3.

2. Matrix Comparison:

Not applicable. Only for use with capillary whole blood sample collected from the fingertip.

C Clinical Studies:

1. Clinical Sensitivity:

Not applicable

2. Clinical Specificity:

{9}

Not applicable

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

To assess the performance of the IWL2020 Blood Glucose Monitoring System in the hands of the intended users, the sponsor performed a study with 350 lay user participants. Users obtained their own fingertip capillary sample and performed a glucose test using only the labeling instructions provided in English. Three test strip lots were used in the study. Results were analyzed by comparing blood glucose results from finger stick samples using the IWL2020 meter, obtained by the lay users, against results obtained on the comparator method (hexokinase reagent on the Roche Cobas c 501 analyzer) with samples obtained by a trained technician. The samples ranged from 49 to 395 mg/dL as measured by the comparator method. All samples used were unaltered. The results are summarized in the tables below:

Results from glucose concentrations less than 75 mg/dL

|  Within ± 5% | Within ± 10% | Within ± 15%  |
| --- | --- | --- |
|  3/6 (50 %) | 5/6 (83 %) | 6/6 (100%)  |

Results from glucose concentrations greater than 75 mg/dL

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20 %  |
| --- | --- | --- | --- |
|  217/344 (63%) | 304/344 (88%) | 331/344 (96%) | 340/344 (99%)  |

Regression results:

|  Slope | Y-intercept | R²  |
| --- | --- | --- |
|  0.974 | 4.8 | 0.975  |

Readability:

A Flesch-Kincaid readability assessment was conducted on the IWL2020 Meter Manual, IWL2020 Test Strip Insert, and Quick Start Guide and the results demonstrated that the labeling was written at lower than 8th grade level.

Accuracy at Extreme Glucose Values:

An additional study was conducted to assess the accuracy of the IWL2020 Blood Glucose Monitoring System at the extreme low and high ends of the claimed measurement range using 46 capillary whole blood samples with glucose concentrations below 80 mg/dL, and 56 samples greater than 250 mg/dL. Samples were altered by spiking or glycolysis in order to obtain the appropriate glucose concentrations. Samples were tested with the IWL2020 Blood Glucose Monitoring System with test strips from 3 lots and compared to the results obtained using the comparator method (hexokinase reagent on the Roche Cobas c 501 analyzer). The glucose concentrations in the samples ranged from 22 – 590 mg/dL, as measured by the comparator method. Results are summarized below:

K203711 - Page 10 of 12

{10}

Accuracy results for Glucose concentrations less than 80 mg/dL:

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  33/51 (65%) | 48/51 (94%) | 51/51 (100%) | 51/51 (100%)  |

Accuracy results for Glucose concentrations greater than 250 mg/dL:

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  39/50 (78%) | 49/50 (98%) | 50/50 (100%) | 50/50 (100%)  |

## D Clinical Cut-Off:

Not applicable

## E Expected Values/Reference Range:

The following is included in the labeling:

|  Time of day | People without diabetes  |
| --- | --- |
|  Fasting and before meals | <100 mg/dL  |
|  2 hours after meals | <140 mg/dL  |

American Diabetes Association. 2. Classification and diagnosis of diabetes; Standards of Medical Care in Diabetes- 2019. Diabetes Care 2019; 42 (Suppl. 1): S13-S28.

## F Other Supportive Instrument Performance Characteristics Data:

1) Hematocrit Study: The effect of different hematocrit levels on the performance of the IWL2020 Blood Glucose Monitoring System was evaluated using contrived venous whole blood samples with hematocrit levels of 10 – 65% (10, 15, 20, 25, 30, 35, 42 (nominal), 45, 50, 55, 60, 65, 70%) spiked with glucose to achieve target concentrations between 30-50, 51-110, 111-150, 151-250, 251-400mg/dL. A total of 90 replicates per sample were performed on IWL2020 Blood Glucose Monitor and 3 test strip lots and the results on the candidate system were compared to comparator method (hexokinase reagent on the Roche Cobas c 501 analyzer). The results demonstrated that the IWL2020 Blood Glucose Monitoring System performs as intended with samples containing hematocrit concentrations spanning the claimed hematocrit range of 10 – 65%.

2) Altitude Study: To evaluate the effects of altitude on the IWL2020 Blood Glucose Monitoring System results, contrived venous whole blood samples were altered to five glucose concentrations (60, 110, 160, 350 and 500 mg/dL) and tested at 10,000 ft feet above sea level. Each venous whole blood glucose concentration was tested in duplicate using two strip lots and five glucose meters per strip lot. The meter results (20 glucose measurements for each glucose concentration) were compared to those obtained with the comparator method (hexokinase reagent on the Roche Cobas c 501 analyzer, tested at 719 ft). The results support the claims that the system functions as intended at altitudes up to the claimed

K203711 - Page 11 of 12

{11}

altitude of 10,000 feet.

3) System Operating Conditions Testing: To evaluate device performance over varying operating temperature and humidity conditions, contrived venous whole blood samples at glucose concentrations (60-80, 105-135, and 225-275 mg/dL) were tested at temperatures ranging from 4-45°C (39.2-113°F) and relative humidity from 10-90%. Conditions tested included 4°C (39.2°F)/30% (relative humidity, r.h.), 4°C (39.2°F)/90% r.h., 10°C (50°F)/10% r.h., 28°C (82.4°F)/90% r.h., 45°C (113°F)/10% r.h., and 45°C (113°F)/36% r.h, and meter results compared to the comparator method. The results support the claimed range of operating conditions: 6-44°C (42.8-111.2°F) and 10-90% relative humidity.

4) Flex Studies: Short sample detection, error codes for samples outside the measuring range, sample perturbation, testing with used test strips, intermittent sampling was completed. The testing performed demonstrated that the device is robust to short sample detection, sample perturbation, testing with used test strips, intermittent sampling, and that an error message is returned to the user if a used test strip is inserted into the meter.

5) Infection Control Studies: The device is intended for single patient use. A cleaning study was performed by cleaning the meter with mild dishwashing liquid detergent and water. Three test meters were subjected to 260 cleaning cycles, representing one cleaning per week for five years. Three meters were used for control/reference purposes that were not subjected to cleaning. The study demonstrated that cleaning the meter did not impact device durability or function. Disinfection efficacy studies were performed on the materials compromising the meter with the chosen disinfectant, Super Sani-Cloth (EPA registration #9480-4). Robustness studies were also performed by the demonstrating that there was no change in performance or external materials of the meter after 260 cleaning and disinfection cycles (representing weekly disinfection for five years, the expected lifetime of the meter).

6) The sponsor has provided appropriate documentation certifying that acceptable electrical safety and electromagnetic compatibility (EMC) testing had been performed with the candidate device, and that the system was found to be compliant.

7) Test strip lot release criteria: The test strip lot release protocols and criteria were reviewed and found to be acceptable.

VIII Proposed Labeling:

The labeling does support the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and does support a substantial equivalence decision.

K203711 - Page 12 of 12

---

**Source:** [https://fda-staging.innolitics.com/device/K203711](https://fda-staging.innolitics.com/device/K203711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
