K203710 · Storz Medical AG · IPF · May 3, 2021 · Physical Medicine
Device Facts
Record ID
K203710
Device Name
Storz Medical MAGNETOLITH Muscle Stimulator
Applicant
Storz Medical AG
Product Code
IPF · Physical Medicine
Decision Date
May 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MAGNETOLITH is indicated for - Relaxation of muscle spasms, - Prevention or retardation of disuse atrophy, - Increasing local blood circulation, - Muscle re-education, - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, - Maintaining or increasing range of motion.
Device Story
MAGNETOLITH delivers high-energy electromagnetic pulses to stimulate muscle tissue via magnetic induction field principle; non-invasive; no direct skin contact required. System comprises control unit, handpiece applicator, applicator holder, mains cable, and trolley. Applicator uses treatment coil to generate electromagnetic induction, creating voltage across electrical conductors in changing magnetic field; water-cooled applicator prevents tissue heating. Used in clinical settings by healthcare professionals. Output consists of electromagnetic pulses; healthcare provider monitors therapy via touch screen interface. Device facilitates muscle contraction to achieve therapeutic goals like spasm relaxation and atrophy prevention.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed device meets design specifications. Software validated as moderate level of concern. Compliance with IEC 60601-1:2012, IEC 60601-1-2:2014, and IEC 60601-1-6:2012 demonstrated. No clinical data provided.
Technological Characteristics
Powered muscle stimulator using electromagnetic induction. Components: control unit, water-cooled handpiece applicator, trolley. Energy source: 100-240 VAC, 50/60 Hz. Magnetic field intensity: 0.4T (surface), 0.08T (coil center). Pulse duration: 125μs ± 20%. Pulse repetition rate: 1-10Hz. Connectivity: touch screen interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, ISO 14971, ISO 13485.
Indications for Use
Indicated for patients requiring muscle spasm relaxation, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical calf muscle stimulation to prevent venous thrombosis, and maintenance/increase of range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo, which features a stylized eagle. To the right of the eagle is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
May 3, 2021
Storz Medical AG % Michael Dayton President & CEO Biomed Research, Inc. 3959 Van Dyke Road Suite 245 Lutz, Florida 33558
Re: K203710
Trade/Device Name: Storz Medical MAGNETOLITH Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, NGX Dated: February 1, 2021 Received: February 3, 2021
Dear Michael Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 53 1 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-ind-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
Patrick Antkowiak, PhD Assistant Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203710
Device Name
MAGNETOLITH Muscle Stimulator
Indications for Use (Describe)
The MAGNETOLITH Muscle Stimulator is indicated for:
-Relaxation of muscle spasms,
-Prevention or retardation of disuse atrophy,
-Increase local blood circulation.
-Muscle re-education,
-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
-Maintaining or increasing range of motion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# Storz Medical AG
### MAGNETOLITH Muscle Stimulator
#### 1. SPONSOR
Storz Medical AG Lohstampfestr. 8 CH-8274 Tägerwilen Switzerland
Contact Person: Pavel Novak, Ph.D. Telephone: 41 (0) 71 677 45 13 FAX: 41 (0) 71 677 45 05
Date Prepared: May 3, 2021
#### DEVICE NAME 2.
Proprietary Name: MAGNETOLITH Regulation Name: Powered Muscle Stimulator
| Regulation Number: | 21 CFR 890.5850 |
|--------------------|-----------------|
| Product Code: | IPF, NGX |
| Regulatory Class: | II |
#### 3. PREDICATE DEVICES
Equivalence is claimed to the following predicate devices: HPM-6000 (K160992, primary), and Bemer Classic & Pro (K151834, secondary).
#### 4. DEVICE DESCRIPTION
The MAGNETOLITH delivers electromagnetic pulses to stimulate muscle tissues by applying the Magnetic Induction Field Principle (high-energy magnetic pulses) directly over (but not in direct contact with) the patient in order to achieve the same powered muscle stimulation without having to place adhesive electrode pads directly onto the patient's body. Key components of the device are the Control Unit, Handpiece Applicator, Applicator holder, a Mains cable, and a Trolley. These device components are provided non-sterile and are intended to be reusable.
The principle of operation of the MAGNETOLITHis by the application of a treatment coil to generate electromagnetic induction. i.e., the production of voltage across an
{4}------------------------------------------------
electrical conductor in a changing magnetic field. The duration of the individual pulses is short and therefore the system does not increase temperature within the tissues. In addition, the MAGNETOLITH Applicator is cooled by circulating water.
#### 5. INTENDED USE
The MAGNETOLITH is indicated for
- · Relaxation of muscle spasms,
- Prevention or retardation of disuse atrophy, .
- · Increasing local blood circulation,
- · Muscle re-education,
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, .
- Maintaining or increasing range of motion. .
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The technological characteristics of the MAGNETOLITH device and the predicate device is substantially equivalent in that each device is designed to non-invasively use electromagnetic induction to initiate action potential of nerves resulting in muscle contraction.
#### 7. PERFORMANCE TESTING
Verification and validation testing was performed and demonstrated that the MAGNETOLITH meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed.
The MAGNETOLITH software was validated and demonstrated to be of a moderate level of concern; while hazard analysis / risk management was performed and demonstrated that all risks are mitigated to an acceptable level. While the MAGNETOLITH has cleaning and sanitation requirements, there are no body contacting components and therefore biocompatibility testing was not performed. The MAGNETOLITH was tested and demonstrated to conform to the general safety requirements of IEC 60601-1:2012; IEC 60601-6:2013; and electromagnetic compatibility requirements of IEC 60601-1-2:2014.
No performance standards applicable to this device have been adopted under Section 514 of the Act. The MAGNETOLITH complies with the applicable requirements of the following international consensus standards:
- ISO 14971:2000/A1:2007: Medical devices: Application of risk management to . medical devices.
{5}------------------------------------------------
- IEC 60601-1:2012, (Ed. 3.1): Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance.
- IEC 60601-1-2: 2014(4th Ed.): Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- . IEC 60601-1-6: 2012 (3.1 Ed.): Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- IEC 62304:2015: Medical Device Software - Software Life CycleProcess.
- ISO 13485:2016: Medical devices Ouality management systems -. Requirements for regulatory purposes.
The performance testing demonstrated that the MAGNETOLITH Muscle Stimulator is substantially equivalent to the predicate devices and that it is as safe, as effective, and performs as well as or better than the predicate devices. The table below compares the characteristics of the MAGNETOLITH to the predicate devices.
| 510(k) number<br>Device name<br>Company name | K160992<br>HPM-6000<br>BTL Industries, Inc. | K151834<br>Bemer Classic & Pro<br>Bemer Int'l AG | K203710<br>MAGNETOLITH<br>Storz Medical AG | Comparison |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code and<br>Regulation | Physical Medicine<br>21 CFR 890.5850<br>IPF - Stimulator,<br>Muscle, Powered | Physical Medicine<br>21 CFR 890.5850<br>NGX - Stimulator, Muscle,<br>Powered | Physical Medicine<br>21 CFR 890.5850<br>IPF - Stimulator,<br>Muscle, Powered<br>NGX - Stimulator, Muscle,<br>Powered | Same |
| Intended Use /<br>Indications for Use | The HPM-6000 device is intended<br>to be used under medical<br>supervision for adjunctive therapy<br>for the treatment of medical<br>diseases and conditions.<br>The HPM-6000 device is indicated<br>for use in stimulating<br>neuromuscular tissue for bulk<br>muscle excitation in the legs or<br>arms for rehabilitative purposes.<br>Indications for Use for Muscle<br>Stimulators:<br>-Relaxation of muscle spasms<br>-Prevention or retardation of disuse<br>atrophy<br>-Increasing local blood circulation<br>-Muscle re-education<br>-Immediate post- surgical<br>stimulation of calf muscles to<br>prevent venous thrombosis<br>-Maintaining or increasing range<br>of motion | Indications for Use:<br>-To temporarily increase local<br>blood circulation in healthy leg<br>muscles<br>-To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance | Indications for Use:<br>-Relaxation of muscle spasms<br>-Prevention or retardation of<br>disuse atrophy<br>-Increasing local blood<br>circulation<br>-Muscle re-education<br>-Immediate post- surgical<br>stimulation of calf muscles to<br>prevent venous thrombosis<br>-Maintaining or increasing range<br>of motion<br>-To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance | Similar |
#### 8. SUBSTANTIAL EQUIVALENCE COMPARISON
{6}------------------------------------------------
| Principle of Action | Non- invasive tissue stimulation<br>via magnetic field induction | Non- invasive tissue stimulation<br>via magnetic field induction | Non- invasive tissue stimulation<br>via magnetic field induction | Same |
|--------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|----------|
| Clinical Use | Prescription use | Over the counter use | Prescription use | Same |
| Electrical<br>Protection | Class II, BF | unknown | Class I, B | Similar¹ |
| User<br>Interface | Touch screen | Touch screen | Touch Screen | Same |
| Firmware<br>Controlled | Yes | Yes | Yes | Same |
| Type of<br>Energy | Magnetic field | Magnetic field | Magnetic field | Same |
| Number<br>of outputs | 1 | 2 | 1 | Same |
| Number of<br>Magnetic Coils in<br>the Applicator | 1 | unknown | 1 | Same |
| Magnetic Field<br>Intensity | BTL 299-1 applicator: 0.5 - 1.8 T<br>BTL 299-2 applicator: 0.7 - 2.5 T | 35-100 μT | 0.4T ±20% at surface<br>0.08T ±20% at center of coil | Similar |
| Pulse Repetition<br>Rate | 1 - 150 Hz | 10-30 Hz | 1-10Hz | Similar |
| Pulse Duration | BTL 299-1 applicator: 280μs ±<br>20%<br>BTL 299-2 applicator: 280μs ±<br>20% | 10-33μs | 125μs ± 20% | Similar |
| Therapy Time | Up to 60 min | 20 minutes | 10-20 minutes | Similar |
| Energy Source | 100 - 240 VAC,<br>50-60 Hz | 100 - 240 VAC,<br>50/60 Hz | 100 - 240 VAC,<br>50/60 Hz | Same |
| System Dimensions<br>(W×H×D) | 500×970×580 mm (20×38×23 in) | 320 x 320 x 70 mm | 454 x 187 x 460 mm | Similar |
| Ambient<br>Temperature | 10° - 30°C | 10° - 30°C | 10° - 30°C | Same |
| Relative Humidity | 30 - 75% | unknown | 5-55% | Similar |
| Environmental<br>Specifications | For indoor use only | For indoor use only | For indoor use only | Same |
'Electrical Protection: The electrical security measures in devices are slightly different, but both comply with theIEC60601-1 standard for basic safety and essential performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.