← Product Code [LLZ](/productcode/LLZ) · K203696

# RBknee (K203696)

_Radiobotics Aps · LLZ · Aug 27, 2021 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203696

## Device Facts

- **Applicant:** Radiobotics Aps
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 27, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K203696 · Aug 27, 2021 | RBknee | Radiobotics Aps | Osteoarthritis Initiative (OAI) database | The OAI database was used to perform a standalone clinical performance validation of the RBknee software, comparing device outputs against adjudicated ground truth grades for Kellgren-Lawrence, OARSI criteria, and joint space width measurements. | Osteoarthritis Initiative; OAI; Observational study; Clinical validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Osteoarthritis Initiative (OAI); Prospective multicenter observational study; Follow-up/Duration: 12 years | Patients with or at risk for knee osteoarthritis; Sample Size: Up to 4279 images / 421 unique subjects (varies by endpoint); Number of Sites: Multicenter | Not applicable for this study | Kellgren-Lawrence grade, OARSI grades (osteophyte, sclerosis, joint space narrowing), and minimum joint space width |

## Indications for Use

RBknee is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the measurement of minimum joint space width; the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence of radiographic knee OA based on Kellgren-Lawrence grading of standing. fixed-flexion radiographs of the knee. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, radiologists, orthopedics, physicians and medical technicians.

## Device Story

RBknee is a standalone, fully automated radiological image processing software; inputs are digital (CR/DX) AP/PA knee radiographs. Device utilizes computer vision and machine learning algorithms to detect and measure OA-related features; outputs include a visual report with ROI overlays marking minimum joint space width (JSW) and a table/text summary of OARSI-based findings (sclerosis, narrowing, osteophytes) and Kellgren-Lawrence (KL) scores. Used in clinical settings by radiologists, orthopedists, physicians, or technicians; integrates with PACS/DICOM viewers. Provides objective measurements to assist clinicians in diagnosis; does not replace full patient evaluation. Benefits include standardized, automated quantification of radiographic knee OA markers.

## Clinical Evidence

Standalone clinical performance validation using the Osteoarthritis Initiative (OAI) database (4,279 images/421 subjects). Ground truth established by two physicians with third-party adjudication. Results: KL-grade ≥ 2 (Sens 0.88, Spec 0.87); Joint Space Narrowing (Sens 0.82, Spec 0.87); Osteophytosis (Sens 0.89, Spec 0.78); Subchondral Sclerosis (Sens 0.84, Spec 0.87). Minimum JSW measurements validated via orthogonal linear regression against OAI data, showing slopes of 1.00 for both medial and lateral measurements.

## Technological Characteristics

Standalone software; operates in Linux environment; compatible with Docker. Utilizes computer vision and machine learning algorithms. Inputs: digital (CR/DX) radiographs. Outputs: visual reports with overlays and textual summaries of OARSI/KL parameters. No direct patient interaction; no life-sustaining function.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- IB Lab GmbH, KOALA ([K192109](/device/K192109.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 20, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.

Radiobotics ApS % John Smith Partner Hogan Lovells US LLP 555 13th Street. NW Washington, District of Columbia 20004

Re: K203696

Trade/Device Name: RBknee Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, JAK

Dear John Smith:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 27th, 2021. Specifically, FDA is updating this SE Letter due to the SE package containing the wrong version of the 510(k) Summary and Indications for Use statement.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Thalia Mills, OHT7: Office of In Vitro Diagnostics and Radiological Health by email (Thalia.Mills(@fda.hhs.gov) or phone (301-796-6641).

Sincerely,

Image /page/0/Figure/9 description: The image shows a digital signature. The signature is for Laurel M. Burk -S. The date of the signature is 2021.09.20, and the time is 09:15:35 -04'00'.

Laurel Burk Chief Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

August 27, 2021

Image /page/1/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Radiobotics ApS % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 13th Street, NW WASHINGTON DC 20004

Re: K203696

Trade/Device Name: RBknee Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, JAK Dated: June 1, 2021 Received: June 1, 2021

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{2}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara
For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

510(k) Number (if known) K203696

Device Name RBknee

Indications for Use (Describe)

RBknee is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the measurement of minimum joint space width; the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence of radiographic knee OA based on Kellgren-Lawrence grading of standing. fixed-flexion radiographs of the knee.

It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.

The system is to be used by trained professionals including, but not limited to, radiologists, orthopedics, physicians and medical technicians.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Health Department of and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY Radiobotics Aps's RBknee

### Submitter

Radiobotics ApS

Esplanaden 8C, 1-tv, 1263 Copenhagen, Denmark

Phone: +45 26 14 61 84

Contact Person: Martin Christian Axelsen, PhD; Co-founder & CSO

Date Prepared: June 1, 2021

Name of Device: RBknee

Trade/proprietary name of device: RBknee

Common or Usual Name: System, Image Processing, Radiological

Classification Name: Medical Image Management and Processing System (892.2050)

Regulatory Class: Class II

Product Codes: LLZ, JAK

Classification Advisory Committee: Radiology

Predicate Device: IB Lab GmbH, KOALA (K192109)

#### Intended Use / Indications for Use

RBknee is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the measurement of minimum joint space width; the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren-Lawrence grading of standing, fixed-flexion radiographs of the knee.

It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.

The system is to be used by trained professionals including, but not limited to, radiologists, orthopedics, physicians and medical technicians.

#### Technological Characteristics

In general, the technological characteristics of the RBknee is directly comparable to the predicate device, KOALA.

{5}------------------------------------------------

Similar to KOALA, RBknee as a fully-automated image processing stand-alone software is a software device designed to assist clinicians in analyzing and measuring radiographic abnormalities during review of posterior-anterior (PA) and anterior-posterior (AP) radiographs.

The predicate and subject software utilize computer vision and machine learning algorithms trained on medical images. The machine-learning algorithms allow for high accuracy in the detection and measurement of OA related symptoms visible on knee radiographs.

In brief, RBknee takes digital radiographs (AP/PA) as input and as output provides:

- 1. A visual report with
	- a. A copy of regions of interest from the original radiographs with overlays that mark the minimum Joint Space Width (JSW)
	- A table with the measurement of the minimum JSW in millimeters (mm), information ﻒ on the presence or absence of joint space narrowing, osteophytes, and sclerosis based on OARSI gradings, and information on the presence or absence of radiographic knee osteoarthritis (OA) based on the Kellgren-Lawrence (KL) score.
- 2. A text report with
	- a. A textual summary of the presence or absence of joint space narrowing, osteophytes, and sclerosis based on OARSI gradings (called findings), and a textual summary of the presence or absence of radiographic knee osteoarthritis (OA) based on the Kellgren-Lawrence score (called impression).

RBknee is not interpreting any results but makes an objective measurement and providing an output based on a well established scale (OARSI/KL).

RBknee can be integrated to a PACS and the outputs of RBknee can be reviewed in a DICOM viewer.

RBknee operates in a Linux environment and can be compatible with any operating system supporting the third-party software Docker. The integration environment has to support RBknee data input and output requirements. The device does not interact with the patient directly, nor does it control any life-sustaining devices.

## Performance Data

Software verification and validation testing was completed for the subject device. The software functioned as intended and all results observed were as expected.

A standalone clinical performance validation of RBknee was performed using the open-access database from the Osteoarthritis Initiative (OAI), a large U.S. prospective multicenter observational study conducted over a period of 12 years. Ground truth grading for Kellgren Lawrence grades, as well as osteophyte, sclerosis and joint space narrowing grades according to the OARSI (Osteoarthritis Research Society International) guidelines, and measurements of the minimum joint space width was established by two physicians following adjudication procedures with a third reviewer for discrepancies.

The performance of RBknee for the status indicators of the radiographic findings are listed in the table below.

{6}------------------------------------------------

|                                     | RBknee                                    |              |              |
|-------------------------------------|-------------------------------------------|--------------|--------------|
|                                     | Data Support<br>(images/ unique subjects) | Sensitivity  | Specificity  |
| Kellgren-Lawrence status*           | 4279/421                                  | 0.88         | 0.87         |
| 95% CI                              |                                           | 0.85    0.91 | 0.84    0.89 |
| Joint Space Narrowing Status#       | 4279 / 421                                | 0.82         | 0.87         |
| 95% CI                              |                                           | 0.79    0.86 | 0.84    0.90 |
| Osteophytosis status#               | 2427/289                                  | 0.89         | 0.78         |
| 95% CI                              |                                           | 0.87    0.91 | 0.72    0.84 |
| Subchondral Sclerosis status#       | 2329/272                                  | 0.84         | 0.87         |
| 95% CI                              |                                           | 0.80    0.87 | 0.84    0.90 |
| * KL-grade ≥ 2<br># OARSI-grade > 0 |                                           |              |              |

The accuracy of the measurements of the minimum Joint Space Width by RBknee was quantified using orthogonal linear regression by comparing the measurements by RBknee with equivalent measurements from the OAI. The slope and intercept can be seen in the table below.

{7}------------------------------------------------

|                | RBknee    |            |
|----------------|-----------|------------|
|                | Slope     | Intercept  |
| Medial         | 1.00      | -0.07      |
| Medial, 95%CI  | 0.98 1.02 | -0.17 0.03 |
| Lateral        | 1.00      | -0.10      |
| Lateral, 95%CI | 0.97 1.03 | -0.27 0.06 |

## Summary

- Software verification and validation testing was completed for the subject device. The -RBknee functioned in all instances as intended and all results observed were as expected.
- RBknee is an effective image processing device that provides reliable measurements and accurate indicators for presence/absence of radiographic features relevant for the analysis of radiographic knee osteoarthritis. Thus, the device performs as intended and is substantially equivalent to the predicate device.

### Substantial Equivalence

The RBknee is as safe and effective as the KOALA. The RBknee has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. Performance data demonstrates that the RBknee is as safe and effective as KOALA. Thus, the RBknee is substantially equivalent.

## Conclusions

In conclusion, RBknee is as safe, as effective, and performs as well as the predicate device, KOALA. The RBknee has a similar intended use, indication, technological characteristics, and principles of operation as KOALA. The differences between RBknee and KOALA do not alter the intended use of the device and do not raise new or different questions regarding its safety and effectiveness when used as labelled. Performance data demonstrates that the device performs as intended. Thus, RBknee is substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K203696](https://fda-staging.innolitics.com/device/K203696)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
