The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions. The Multiple Brain Mets SRS application as one RT Element provides optimized planning and display for cranial multimetastases radiation treatment planning. The Cranial SRS application as one RT Element provides optimized planning and display for cranial radiation treatment planning. The Spine SRS application as one RT Element provides optimized planning and display for single spine metastases. The Cranial SRS w/ Cones application as one RT Element provides planning and display for functional diseases (e.g. trigeminal neuralgia) or cranial lesion radiation treatment. The RT QA application as one RT Element contains features for patient specific quality assurance. Use RT QA to recalculate patient treatment plans on a phantom to verify that the patient treatment plan fulfills the planning requirements. The Dose Review application as one RT Element contains features for review of isodose lines, review of DVHs, dose comparison and dose summation.
Device Story
RT Elements is a software-only radiation treatment planning system used in clinical settings by medical physicists, radiation oncologists, and dosimetrists. It processes patient imaging data to generate optimized radiation treatment plans for Linac-based stereotactic radiosurgery (SRS). The device utilizes the Circular Cone dose calculation algorithm to compute dose distributions based on tissue phantom ratios, relative output factors, and single beam profiles. It provides tools for plan calculation, review of isodose lines and dose-volume histograms (DVHs), and patient-specific quality assurance (QA) via phantom recalculation. The output assists clinicians in verifying treatment plan requirements and optimizing dose delivery to lesions. By enabling precise, computer-planned radiation therapy, the device aims to improve the accuracy and safety of SRS procedures for cranial and extracranial lesions.
Clinical Evidence
No clinical tests were included in this submission. Evidence consists of software verification and validation, including usability testing and dose measurements, confirming design specifications are met. Bench testing demonstrated dose calculation accuracy within 1%/1mm.
Technological Characteristics
Software-only radiation treatment planning system. Uses Circular Cone dose calculation algorithm based on tissue phantom ratios, relative output factors, and single beam profiles. Accuracy: 1%/1mm. Supports conical collimators for photon-based Linac radiation therapy. No patient contact. Software level of concern: Major.
Indications for Use
Indicated for radiation treatment planning in stereotactic, conformal, Linac-based radiation therapy for cranial, head and neck, and extracranial lesions, including functional diseases like trigeminal neuralgia and specific metastases (brain, spine).
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
April 9, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Brainlab AG % Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, Bavaria 81829 GERMANY
Re: K203681
Trade/Device Name: RT Elements. Cranial SRS w/ Cones. Multiple Brain Mets SRS. Cranial SRS. Spine SRS, RT OA, Dose Review Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: March 15, 2021 Received: March 18, 2021
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203681
#### Device Name
RT Elements , Cranial SRS w/ Cones, Multiple Brain Mets SRS, Cranial SRS, Spine SRS, RT QA, Dose Review
Indications for Use (Describe)
The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, Linac based radiation treatment of cranial, head and neck and extracranial lesions.
The Multiple Brain Mets SRS application as one RT Element provides optimized planning and display for cranial multimetastases radiation treatment planning.
The Cranial SRS application as one RT Element provides optimized planning and display for cranial radiation treatment planning.
The Spine SRS application as one RT Element provides optimized planning and display for single spine metastases.
The Cranial SRS w/ Cones application as one RT Element provides planning and display for functional diseases (e.g. trigeminal neuralgia) or cranial lesion radiation treatment.
The RT QA application as one RT Element contains features for patient specific quality assurance. Use RT QA to recalculate patient treatment plans on a phantom to verify that the patient treatment plan fulfills the planning requirements.
The Dose Review application as one RT Element contains features for review of isodose lines, review of DVHs, dose comparison and dose summation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is an abstract design featuring a pattern of interconnected triangles. The triangles vary in size and shade, transitioning from a darker blue on the left to a lighter blue and almost white on the right. The white lines between the triangles create a sense of depth and dimension, giving the design a modern and geometric appearance. The overall effect is visually appealing and dynamic.
Image /page/3/Picture/1 description: The image shows the word "BRAINLAB" in bold, black letters. To the left of the word is a small, stylized logo that appears to be a symbol or abstract design. The logo is also in black and is positioned to the left of the company name.
K203681
## 510(k) Summary
March 15, 2021
| General Information | |
|----------------------------|-----------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme Str.9, 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Device Name | System, planning, radiation therapy treatment |
| Trade Name | RT Elements, Cranial SRS w/ Cones, Multiple Brain Mets SRS,<br>Cranial SRS, Spine SRS, RT QA, Dose Review |
| Classification Name | Medical charged-particle radiation therapy system |
| Product Code | MUJ |
| Regulation Number | 892.5050 |
| Regulatory Class | II |
| Panel | Radiology |
| Predicate Device(s) | iPlan RT K103246 |
| Contact Information | |
|-----------------------------------|----------------------------------------|
| Primary Contact | Alternate Contact |
| Chiara Cunico | Regulatory Affairs Brainlab |
| Manager RA | Phone: +49 89 99 15 68 0 |
| Phone: +49 89 99 15 68 0 | Fax: +49 89 99 15 68 5033 |
| Email: chiara.cunico@brainlab.com | Email: regulatory.affairs@brainlab.com |
#### 1. Intended Use
The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.
The Multiple Brain Mets SRS application as one RT Element provides optimized planning and display for cranial multi-metastases radiation treatment planning.
The Cranial SRS application as one RT Element provides optimized planning and display for cranial radiation treatment planning.
The Spine SRS application as one RT Element provides optimized planning and display for single spine metastases.
{4}------------------------------------------------
The Cranial SRS w/ Cones application as one RT Element provides planning and display for functional diseases (e.g. trigeminal neuralgia) or cranial lesion radiation treatment.
The RT QA application as one RT Element contains features for patient specific quality assurance. Use RT QA to recalculate patient treatment plans on a phantom to verify that the patient treatment plan fulfills the planning requirements.
The Dose Review application as one RT Element contains features for review of isodose lines, review of DVHs, dose comparison and dose summation.
In addition to the more generic indications for use, as also present in the predicate device, the RT Elements also include specific indications for each element. Cranial SRS w /Cones is the newest addition to the RT Elements group and the main focus within this submission. Since the functionality of Cranial SRS w /Cones is based on the same Circular Cone Algorithm as in the predicate device and now there is more information provided in label, the change in wording does not affect the safety and effectiveness of the device.
### 2. Device Description
The RT Elements group consist of different applications, referred to as Elements, intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions.
The newest addition to the RT Elements group is the Cranial SRS w/ Cones element, which is a software application used for automatic radiation treatment planning for either functional areas (e.g. trigeminal neuralgia) or cranial lesions using conical collimators.
Cranial SRS w/Cones 3.0 UDI-DI: 04056481000530 Cranial SRS w/ Cones has no variants.
The device can be used in a hospital environment, in a clinical planning office by medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.).
Cranial SRS w/ Cones offers features for plan calculation (selection of machine profile and clinical protocols, dose calculation, etc.), plan review (isodose lines, Conformity index, min and max doses review. etc.), as well as, saving and exporting plan options. It is a software only device, with no patient contact.
For the dose calculation optimization, the Circular Cone dose algorithm is used on different publications:
- · Rice, R. K., Hansen, J. L., Svensson, G. K., Siddon, R. L .; Measurements of dose distributions in small 6MV x-rays. Phys. Med. Biol. 32; 1987; 1087-1099
- · Luxton G, Jozsef G, Astrahan M A; Algorithm for dosimetry of multiarc linear-accelerator stereotactic radiosurgery. (1991) Med. Phys. 18 pp 1211 - 1221.
- · ]Hartmann G. H. (Editor) Quality Assurance Program on Stereotactic Radiosurgery: Report from a Quality Assurance Task Group. (1995) Published by Springer-Verlag, Berlin Heidelberg.
{5}------------------------------------------------
In qeneral, the main assumption of stereotactic dose calculation models, like the circular cone dose calculation algorithm, is that secondary scatter can be assumed to be of limited significance. Several authors have proposed and investigated this hypothesis (Rice et al 1997, Luxton et al 1991, Hartmann et al 1995) and dose calculations have tended to be a function of only three basic beam parameters: tissue phantom ratios, relative output factors and single beam profiles, Scatter (in particular phantom scatter) is considered to be implicit to these measurements and does not vary significantly with depth in a medium.
The accuracy (reproducibility of measured beam data) of the circular cone dose algorithm is within 1%/1mm.
The circular cone dose calculation algorithm has limited accuracy for dose calculations near inhomogeneous areas such as lunq or bone tissue or close to the tissue border (both within a range of a few centimeters). When using the circular cone dose calculation algorithm in dose calculations for depths different from the depth at which the off-axis ratios have been measured, the results may be inaccurate. The calculated penumbra width may be different from the penumbra in the real dose delivered.
#### 3. Substantial Equivalence
Conical Collimator Planning is the technological principle for both the subject and predicate devices. It is based on the circular cone dose calculation algorithm which is basically identical in both devices.
At a high level, the subject and predicate devices are based on the following same technological elements:
- -Radiation treatment planning software
- י Circular Cone dose calculation algorithm
- Radiation modality (=photons) =
- e Support of conical collimators
- = Similar GUI layouts (e.g. arc plane view layout)
The following technological differences exist between the subject and predicate devices:
- -New GUI Design
However, these differences do not raise any questions of safety or efficacy since these functionalities have been successfully tested for their efficacy.
There is also a change in wording on the indications for use, which now provides more information, with no impact on the use of the device.
#### 4. Performance Data
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of
{6}------------------------------------------------
Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
The verification of the subject device has been carried out thoroughly both at the top-level including compatibility testing and on the underlying modules. The verification was done according to verification plan to demonstrate that the design specifications are met. The validation was done according to clinical evaluation plan and included usability testing as well as dose measurements. All test reports were finally rated as successful according to their acceptance criteria. The nonclinical validation has been performed with software and units that are considered equivalent to the final version of the product, as warranted by 21 CFR 820.30 (g) and which have the UI as planned for the release.
No clinical tests have been included in this pre-market submission.
### 5. Conclusion
The comparison of the Cranial SRS w/ Cones (subject device) with the predicate device(s) iPlan RT shows that the subject device has similar functionality, intended use, technological characteristics, and typical users as the predicate device(s). Verification and validation activities ensured that the design specifications are met and that the subject device does not introduce new issues concerning safety and effectiveness. Hence, Brainlab believes that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.