M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System
K203671 · Medacta International S.A. · OUR · Jan 13, 2021 · Orthopedic
Device Facts
Record ID
K203671
Device Name
M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System
Applicant
Medacta International S.A.
Product Code
OUR · Orthopedic
Decision Date
Jan 13, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The M.U.S.T. Sacral Illac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroiliitis.
Device Story
System consists of Ti6Al4V ELI metallic screws with hydroxyapatite (HA) coating; designed for sacroiliac joint fusion and pelvic fracture fixation. Screws feature hollow-body threaded design, multiple fenestrations, self-tapping tips, and tapered shafts to facilitate bone ingrowth and arthrodesis. Available in standard (with modular washers) and headless configurations in various diameters and lengths. Used by surgeons in clinical settings for orthopedic trauma and joint stabilization. HA coating promotes biological fixation. Device provides mechanical stability to bone segments; aids in fusion process. System is provided sterile.
Clinical Evidence
No clinical data. Bench testing only. Leveraged performance data from predicate device including ASTM F543-13 (bone screws), ASTM F2193-14 (spinal fixation components), ISO 13779-3:2008 (HA characterization), and ASTM F1185-03 (HA composition). Includes sterilization validation, shelf-life testing, and cadaver/sawbones workshop evaluations.
Technological Characteristics
Materials: Ti6Al4V ELI with hydroxyapatite (HA) coating. Standards: ASTM F543-13, ASTM F2193-14, ISO 13779-3:2008, ASTM F1185-03. Design: Hollow-body threaded, fenestrated, self-tapping, tapered tip. Configurations: Standard (with modular washers) and headless. Diameters: 7.5-11 mm. Lengths: 25-80 mm. Sterilization: Provided sterile.
Indications for Use
Indicated for skeletally mature patients requiring fracture fixation of small and long bones of the pelvis, or sacroiliac joint fusion due to sacroiliac joint disruptions or degenerative sacroiliitis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System (K193083)
Reference Devices
M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System (K171595)
Submission Summary (Full Text)
{0}------------------------------------------------
May 4, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Chris Lussier Senior Director, Quality and Regulatory Medacta International SA 3973 Delp Street Memphis, Tennessee 38118
Re: K203671
Trade/Device Name: M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC, HTN
Dear Chris Lussier:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 13, 2021. Specifically, FDA is updating this SE Letter as an administrative correction for an incorrect Indications for Use Statement.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Colin O'Neill, M.B.E., OHT6: Office of Orthopedic Devices, (301) 796-6428, colin.oneill@fda.hhs.gov.
Sincerely.
# Colin O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
January 13, 2021
Image /page/1/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration" with the word "Administration" on the second line.
Chris Lussier Senior Director, Quality and Regulatory Medacta International SA 3973 Delp Street Memphis, Tennessee 38118
### Re: K203671
Trade/Device Name: M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC, HTN Dated: December 15, 2020 Received: December 16, 2020
Dear Mr. Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Colin O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K203671
Device Name
M.U.S.T. Sacro-Iliac Screws and Pelvic Trauma System
Indications for Use (Describe)
The M.U.S.T. Sacral Illac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroiliitis.
Type of Use (Select one or both, as applicable)
| <div style="display:flex; align-items:center;"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 2.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director Quality / Regulatory, Medacta USA Date Prepared: December 15, 2020
#### II. Device
| Device Proprietary Name: | M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System |
|--------------------------|------------------------------------------------------|
| Common or Usual Name: | Pelvic Joint Fixation |
| Classification Name: | Smooth or threaded metallic bone fixation fastener |
| Primary Product Code: | OUR |
| Secondary Product Codes: | HWC, HTN |
| Regulation Number: | 21 CFR 888.3040, 21 CFR 888.3030 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
Primary Predicate Device:
- . M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K193083, Medacta International SA
### Reference Predicate Device
- M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K171595, Medacta ● International SA
{5}------------------------------------------------
#### IV. Device Description
The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is designed for sacroiliac joint fusion in degenerative sacroiliitis, as well as for the fixation of small and long bone fractures in trauma cases.
The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System consists of various sized screws, manufactured from Ti6A14V ELI and coated with rough Hydroxyapatite (HA). The HA coating allows for biological fixation and potentially leads to arthrodesis.
The sacroiliac (SI) joint screws are hollow-body threaded fusion devices with a multiplefenestrated pattern shaft to promote arthrodesis, a self-tapping design to facilitate screw insertion, and a tapered tip to aid in guidance through pilot hole. Radial windowed slots along the shaft facilitate bone ingrowth after implantation and "bone filling" after insertion.
The SI joint screws are provided in standard and headless designs. The standard screw can be coupled to modular washers of different diameters to optimize and stabilize the contact between the screw head and the cortical bone and to improve the compression activity of the screw. The standard screws are provided sterile in three diameters (8, 9, and 10 mm) and multiple lengths (25 – 80 mm).
The headless screw has an anatomical shape which allows for full insertion into the bone. The headless screws are provided sterile in three diameters (7.5, 9, and 11 mm) and multiple lengths (30 - 75 mm). The M.U.S.T. Sacral Iliac Screws Extension introduces new sizes of these screws keeping the same diameters (7.5, 9, and 11 mm) with new lengths from 80 mm (5 mm increment).
#### Indications for Use V.
The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroillitis.
#### VI. Comparison of Technological Characteristics
The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws and the predicate devices share the following characteristics:
- material of construction; ●
- diameters; ●
- coating and coating composition; ●
- biocompatibility; ●
{6}------------------------------------------------
- sterility;
- shelf life; and ●
- packaging. ●
The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws are technologically different from the predicate device with respect to length of the screw and number of the radial windowed slots on the screw shaft.
These differences do not raise different questions of safety or effectiveness and design verification testing supports there are no new risks associated with the subject devices.
#### Performance Data VII.
The introduction of new sizes of M.U.S.T. SI-Joint Headless screws does not create a new worst case; therefore, the following performance testing from the predicate device was leveraged to support this submission:
- ASTM F543-13 Standard Specification and Test Methods for Metallic Medical Bone ● Screws;
- ASTM F2193-14 Standard Specifications and Test Methods for Components Used in ● the Surgical Fixation of the Spinal Skeletal System;
- ISO 13779-3: 2008 Implants for Surgery Hydroxyapatite Part 3: Chemical Analysis ● and Characterization of Crystallinity and Phase Purity;
- ASTM F1185-03 (Reapproved 2014) Standard Specification for Composition of ● Hydroxyapatite for Surgical Implants;
- sterilization validation; and ●
- shelf-life testing ●
- Cadaver/Sawbones Workshop ●
A comparative analysis of the subject devices to the identified predicate and reference devices was performed to determine if this line-extension created a new worst-case product size. It was determined that the subject devices are substantially equivalent to the previously cleared predicate device and that the line-extension can be considered within the scope of the design verification and validation testing previously completed for the initial M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System product range.
### VIII. Conclusion
The information provided within this submission supports that the M.U.S.T. SI-Joint Headless screws are are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.