Reprocessed Webster CS Bi-Directional Diagnostic EP Catheter, Reprocessed Webster CS Bi-Directional Diagnostic EP Catheter (with Auto ID)

K203655 · Innovative Health, LLC · NLH · May 4, 2021 · Cardiovascular

Device Facts

Record IDK203655
Device NameReprocessed Webster CS Bi-Directional Diagnostic EP Catheter, Reprocessed Webster CS Bi-Directional Diagnostic EP Catheter (with Auto ID)
ApplicantInnovative Health, LLC
Product CodeNLH · Cardiovascular
Decision DateMay 4, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

The Reprocessed Webster CS Bi-directional Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. The catheter is designed for use in the coronary sinus.

Device Story

Reprocessed Webster CS Bi-directional Diagnostic EP Catheter is a steerable, multi-electrode catheter used for cardiac electrophysiological mapping. Device features high-torque shaft, bi-directional deflectable tip, and platinum electrode array. Operated by physicians in clinical settings to facilitate stimulation and recording within the coronary sinus. Tip deflection controlled via rocker lever; shaft rotation allows precise positioning. Certain models include Auto ID technology using EEPROM to store identification data for recognition by CARTO EP Navigation Systems. Device is reprocessed by Innovative Health, LLC, involving cleaning, decontamination, inspection, and functional testing. Reprocessing limited to maximum of three cycles. Output consists of electrical signals for mapping cardiac structures, aiding clinical diagnosis of cardiac arrhythmias.

Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning and sterilization validation, functional testing (visual inspection, dimensional verification, electrical continuity/resistance, simulated use, mechanical characteristics, Auto ID verification), electrical safety (dielectric and current leakage), and packaging validation.

Technological Characteristics

7FR, 115cm usable length, steerable catheter with bi-directional deflectable tip. Materials include platinum electrodes. Features rocker lever for deflection and high-torque shaft. Some models include EEPROM for Auto ID functionality. Reprocessed via cleaning and decontamination; sterilized. No changes to materials or design compared to predicate.

Indications for Use

Indicated for electrophysiological mapping of cardiac structures (stimulation and recording) in the coronary sinus.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 4, 2021 Innovative Health, LLC Christina Fleming VP, Compliance and Regulatory Affairs 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257 Re: K203655 Trade/Device Name: Reprocessed Webster CS Bi-Directional Diagnostic EP Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: March 5, 2021 Received: March 8, 2021 Dear Christina Fleming: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Aneesh Deoras Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Details of Reprocessed Devices: The item numbers included in the scope of this submission are as follows: | Item Number | Description | Sheath<br>Usable<br>Length<br>(cm) | French<br>Size | Curve | Spacing<br>(mm) | |---------------|---------------------------------------------------------------------------------------|------------------------------------|----------------|-------|-----------------| | BD710DF282RTS | Reprocessed<br>Webster CS<br>Bi-Directional<br>Diagnostic EP<br>Catheter | 115 | 7F | D-F | 2-8-2 | | BD710FJ282RTS | Reprocessed<br>Webster CS<br>Bi-Directional<br>Diagnostic EP<br>Catheter | 115 | 7F | F-J | 2-8-2 | | BD710DF282CT | Reprocessed<br>Webster CS<br>Bi-Directional<br>Diagnostic EP<br>Catheter with Auto ID | 115 | 7F | D-F | 2-8-2 | | BD710FJ282CT | Reprocessed<br>Webster CS<br>Bi-Directional<br>Diagnostic EP<br>Catheter with Auto ID | 115 | 7F | F-J | 2-8-2 | Table 10.1: Device Scope (The remainder of this page has been intentionally left blank) {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203655 #### Device Name Reprocessed Webster CS Bi-directional Diagnostic Electrophysiology (EP) Catheters Indications for Use (Describe) The Reprocessed Webster CS Bi-directional Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. The catheter is designed for use in the coronary sinus. **Type of Use (Select *one or both*, as applicable)** > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # SECTION 5: 510(k) SUMMARY As required by 21 CFR 807.92(c) ### Submitter's Name and Address: Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257 ### Contact Name and Information: Christina Fleming VP. Compliance and Requlatory Affairs Innovative Health, LLC. (480) 525-5972 (office) (888) 965-7706 (fax) tfleming@innovative-health.com ### Date prepared: December 14, 2020 ### Device Information: Trade/Proprietary Name: Common Name: Classification Name: Classification Number: Product Code: Reprocessed Webster CS Bi-directional Diagnostic Electrophysiology Catheters Diagnostic Electrophysiology Catheter Catheter, Recording, Electrode, Reprocessed Class II, 21 CFR 870.1220 NLH ### Predicate Device: | 510(k) Number | 510(k) Title | Manufacturer | |---------------|--------------------------------------------------------------------------------------------------------------|-------------------------| | K170922 | Reprocessed Webster CS Bi-Directional<br>Diagnostic Electrophysiology Catheter | Innovative Health, LLC. | | K101345 | Webster CS Catheter with EZ Steer<br>Technology, Webster CS Catheter with<br>EZ Steer Technology and Auto ID | Biosense Webster | ### Device Description: The Reprocessed Webster Coronary Sinus (CS) Diagnostic Electrophysiology (EP) Catheter is a steerable, multi-electrode catheter with a deflectable tip designed to facilitate electrophysiological mapping of the heart. The device is a 7FR catheter with a usable length of 115cm. The catheter has a high-torque shaft with a bi-directional deflectable tip section containing an array of platinum electrodes that can be used for stimulation and recording. Standard features of this catheter include a braided bi-directional deflectable tip section with an array of platinum electrodes that includes a 2mm tip dome. Additionally, two asymmetric curve types are available providing two 180º opposed, single plane curves. The curve types include DF and FJ. A Rocker Lever is used to deflect the tip. The hightorque shaft allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site. {5}------------------------------------------------ # For Devices with Auto ID Technology: The catheter is equipped with Electronically Erasable Programmable Read Only Memory (EEPROM) which is used to store unique catheter identification information. CARTO EP Navigation Systems equipped with Auto ID Technology can access the stored information and automatically recognize he catheter information. The catheter interfaces with CARTO EP Navigation Systems equipped with Auto ID Technology via interface cables with the appropriate connectors. Note: Only the catheter is the subject of this submission. Any other related equipment is not included in the scope of this submission. | Item Number | Description | Sheath<br>Usable<br>Length<br>(cm) | French<br>Size | Curve | Spacing<br>(mm) | |---------------|---------------------------------------------------------------------------------|------------------------------------|----------------|-------|-----------------| | BD710DF282RTS | Reprocessed Webster CS<br>Bi-Directional Diagnostic<br>EP Catheter | 115 | 7F | D-F | 2-8-2 | | BD710FJ282RTS | Reprocessed Webster CS<br>Bi-Directional Diagnostic<br>EP Catheter | 115 | 7F | F-J | 2-8-2 | | BD710DF282CT | Reprocessed Webster CS<br>Bi-Directional Diagnostic<br>EP Catheter with Auto ID | 115 | 7F | D-F | 2-8-2 | | BD710FJ282CT | Reprocessed Webster CS<br>Bi-Directional Diagnostic<br>EP Catheter with Auto ID | 115 | 7F | F-J | 2-8-2 | The item numbers included in the scope of this submission are as follows: Table 5.1: Item Numbers This 510(k) adds two (2) reprocessing cycles to the same device cleared under K170922 and minor changes to the reprocessing and sterilization release method. # Indications for Use: The Reprocessed Webster CS Bi-directional Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. The catheter is designed for use in the coronary sinus. ### Technoloqical Characteristics: The purpose, design, materials, function, and intended use of the Webster CS Bidirectional Diagnostic EP Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling. # Functional and Safety Testing: Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Webster CS Bi-directional Diagnostic EP Catheters. {6}------------------------------------------------ This included the following: - Biocompatibility - Cleaning Validation . - Sterilization Validation . - Functional Testing 0 - o Visual Inspection - . Dimensional Verification - Electrical Continuity and Resistance ● - Simulated Use ● - Mechanical Characteristics ● - Auto ID Verification . - Electrical Safety Testing ● - Dielectric and Current Leakage ● - Packaging Validation ● The Reprocessed Webster Coronary Sinus (CS) Bi-directional Diagnostic Electrophysiology (EP) Catheters are reprocessed no more than three (3) times. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors. # Predicate Comparison: A comparison of the device and reprocessing methods with the predicates are provided in the table below: | | This 510(k) | K170922 | K101345 | |----------------------|-----------------------------------------------------------------------------------------------------------|------------------------------|-----------| | Device: | Identical | Identical | Identical | | Reprocessing Cycles: | 3 | 1 | 0 | | Reprocessing Method: | Change to processing<br>parameters. Packaging<br>configuration and label<br>change. Shelf life<br>change. | Cleared/validated<br>process | N/A | | Sterilization: | Change to release<br>method. No change to<br>the sterilization method<br>or SAL. | Cleared/validated<br>process | N/A | | Routine Monitoring: | Change to frequency<br>and adjustment of limits. | Cleared process | N/A | Table 5.2: Predicate Comparison # Conclusion: Innovative Health concludes that the Reprocessed Webster Coronary Sinus (CS) Bidirectional Diagnostic Electrophysiology (EP) Catheters are substantially equivalent to the predicate devices described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...