PlusCare Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes. These devices are indicated for use by qualified medical personnel only.
Device Story
Reusable temperature probe; connects to compatible patient monitors (e.g., Philips, Marquette, Mindray) via YSI 400 series interface. Input: patient skin or rectal thermal energy. Operation: NTC thermistor resistance changes linearly with temperature; monitor converts resistance to temperature reading. Output: continuous temperature data displayed on host monitor. Used in clinical settings by medical personnel. Benefits: enables continuous core or skin temperature monitoring for clinical decision-making.
Clinical Evidence
Bench testing only. Device performance validated against recognized standards: IEC 60601-1 (safety), ISO 80601-2-56 (clinical thermometer performance), and ISO 10993 series (biocompatibility: cytotoxicity, sensitization, and skin irritation).
Indicated for adult patient temperature monitoring by qualified medical personnel. Skin-type probes for contact <30 days; rectum-type probes for contact <24h.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
August 31, 2021
JKH USA, LLC Bill Quanqin Dai Manager 14271 Jeffrey Rd. #246 Irvine, California 92620
Re: K203636
Trade/Device Name: PlusCare Temperature Probe, Skin Types: PT2252-AS, PTHP-AS, PTMQ-AS, PTSW-AS, PTSM-15AS, PTSL-AS, PTMR-AS; General Purpose Types: PT2252-AG, PTHP-AG, PTMQ-AG, PTSW-AG, PTSM-15AG, PTSL-AG, PTMR-AG Regulation Number: 21 CFR 880,2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: July 20, 2021 Received: July 28, 2021
Dear Bill Quanqin Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203636
#### Device Name
PlusCare Temperature Probe Skin Types: PT2252-AS, PTHP-AS, PTMQ-AS, PTSW-AS, PTSM-15AS, PTSL-AS, PTMR-AS: Rectum Types: PT2252-AG, PTHP-AG, PTMQ-AG, PTSW-AG, PTSM-15AG, PTSL-AG, PTMR-AG
#### Indications for Use (Describe)
PlusCare Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes.
These devices are indicated for use by qualified medical personnel only.
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203636 - 510(k) Summary
| Submitter: | Name: JKH USA, LLC | | |
|---------------------|--------------------------------------------------------------|--|--|
| | Mailing Address: 14271 Jeffrey Rd. #246, Irvine, CA 92620 | | |
| Contact Person: | Name: Bill Quanqin Dai | | |
| | Phone Number: 909-929-9896 | | |
| | Email Address: Bill@jkhUSA.com | | |
| Date Prepared: | 08/31/2021 | | |
| Device Trade Name: | PlusCare Temperature Probe | | |
| Device Common Name: | Clinical electronic thermometer | | |
| Model: | Skin Types: PT2252-AS, PTHP-AS, PTMQ-AS, PTSW-AS, PTSM-15AS, | | |
| | PTSL-AS, PTMR-AS; | | |
| | Rectum Types:<br>PT2252-AG, PTHP-AG,<br>PTMQ-AG, PTSW-AG, | | |
| | PTSM-15AG, PTSL-AG, PTMR-AG | | |
| Regulation Name: | Clinical electronic thermometer | | |
| Regulation Number: | 21 CFR 880.2910 | | |
| Product Code: | FLL | | |
| Device Class: | Class II | | |
| Predicate Device: | | | |
| 510(k) Number: | K121427 | | |
| Device Name: | Unimed Temperature Probe | | |
| Manufacturer: | UNIMED MEDICAL SUPPLIES INC | | |
### Description of Devices:
The Skin Temperature Probe and Rectum Temperature Probe are used during patient temperature measurement. These probes consist of a connector on the monitor end and a thermistor on the patient end. These probes are to be used with compatible temperature measurement systems only.
Temperature probes measure temperature through a resistor that is sensitive to temperature changes.
The probe is connected to the monitor either directly by using the connector or by an extension cable. These probes have a skin or core contact with a patient.
| Model | Application site | length |
|-----------|------------------|--------|
| PT2252-AS | Skin | 3.0m |
| PTHP-AS | Skin | 3.0m |
| PTMQ-AS | Skin | 3.0m |
Table 1. Basic information of devices
{4}------------------------------------------------
| PTSW-AS | Skin | 3.0m |
|---------------|--------|------|
| PTSM-15AS | Skin | 1.5m |
| PTSL-AS | Skin | 3.0m |
| PTMR-AS | Skin | 3.0m |
| PT2252-AG | Rectum | 3.0m |
| PTHP-AG | Rectum | 3.0m |
| PTMQ-AG | Rectum | 3.0m |
| PTSW-AG | Rectum | 3.0m |
| PTSM-<br>15AG | Rectum | 1.5m |
| PTSL-AG | Rectum | 3.0m |
| PTMR-AG | Rectum | 3.0m |
## Operating Principle:
The operating principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. It is to measure differences in resistance and equates that to changes intemperature.
### Indications for Use:
PlusCare Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes.
These devices are indicated for use by qualified medical personnel only.
## Substantial Equivalence Discussion:
The subject and predicate devices are exactly the same. They have the same intended use, the same design principle, the same production process, and the same material composition. See the detailed comparison in the table below.
### Table 2. Comparison of Technological Characteristics
| Description | Subject Device (K203636) | Predicate Device (K121427) | Comparison | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade name | PlusCare Temperature Probe | Unimed Temperature Probe | N/A | |
| Population | Adult | Adult | Same | |
| Indications<br>for Use | PlusCare Temperature Probes<br>are intended to be used for<br>monitoring temperature. The<br>temperature probes are<br>reusable and designed for use<br>with monitors of Philips,<br>Marquette, Mindray,<br>Spacelabs, Siemens,<br>Artema/S&W and other<br>monitors compatible with YSI<br>400 series temperature probes.<br>These devices are indicated for<br>use by qualified medical<br>personnel only. | Unimed Temperature Probes<br>are intended to be used for<br>monitoring temperature. The<br>temperature probes are<br>reusable and designed for use<br>with monitors of Philips,<br>Marquette, Mindray,<br>Spacelabs, Siemens,<br>Artema/S&W and other<br>monitors compatible with YSI<br>400 series temperature probes.<br>These devices are indicated for<br>used by qualified medical<br>personnel only. | Same | |
| Contact<br>duration<br>period of use | Skin-Type Probe: less than 30 days<br>Rectum-Type Probe: less than 24h | Skin-Type Probe: less than 30 days<br>Rectum-Type Probe: less than 24h | Same | |
| Prescription/<br>OTC use | Prescription | Prescription | Same | |
| Energy source | Powered by compatible<br>devices | Powered by compatible<br>devices | Same | |
| Operating<br>Principle | Thermistor resistance based on<br>the metal conductor increase<br>with temperature decrease, and<br>the linear changes to the<br>characteristics of the<br>temperature measurement. | Thermistor resistance based on<br>the metal conductor increase<br>with temperature decrease, and<br>the linear changes to the<br>characteristics of the<br>temperature measurement. | Same | |
| Technology | Temperature sensitive resistor | Temperature sensitive resistor | Same | |
| Application<br>site | Skin, rectum | Skin, rectum | Same | |
| Material | ABS, copper, PA6, PVC,<br>Gilded copper needle, TPU,<br>NTC thermistor, Epoxy resin,<br>Stainless steel plate | ABS, copper, PA6, PVC,<br>Gilded copper needle, TPU,<br>NTC thermistor, Epoxy resin,<br>Stainless steel plate | Same | |
| Measurement<br>range | 25-45°C/77-113°F | 25-45°C | Same | |
| Thermistor<br>resistance | NTC resistance<br>$2.25 K\Omega @25°C$ | NTC resistance<br>$2.25 K\Omega @25°C$ | Same | |
| Accuracy | $\pm 0.1°C/\pm 0.18°F$ | $\pm 0.1°C$ | Same | |
| Component | plug, cable and<br>temperature sensing probe | plug, cable and<br>temperature sensing probe | Same | |
| Usage | Reusable | Reusable | Same | |
| | | | | |
| Operation<br>Environment | Temperature:<br>0°C~40°C (32°F ~104°F);<br>Relative humidity range: 15%<br>to 85% | Temperature:<br>0°C~40°C (32°F ~104°F);<br>Relative humidity range: 15%<br>to 85% | Same | |
| | Storage<br>environment | Temperature:<br>-25°C~55°C (-13°F~131°F)<br>Relative humidity<br>range: ≤85% | Temperature:<br>-25°C~55°C (-13°F~131°F)<br>Relative humidity<br>range: ≤85% | Same |
| Biocompatibility | | All the patient-<br>contacting materials are<br>evaluated by the<br>biocompatibility<br>standard ISO10993-5,<br>ISO 10993-10. | All the patient-<br>contacting materials are<br>evaluated by the<br>biocompatibility<br>standard ISO10993-5,<br>ISO 10993-10. | Same |
| | | Sterile | Non-sterile | Non-sterile |
| | Standards met | IEC 60601-1 | IEC 60601-1 | Same |
| | | ISO 80601-2-56 | ISO 80601-2-56 | |
| | | ISO 10993-1 | ISO 10993-1 | |
| | | ISO 10993-5 | ISO 10993-5 | |
| ISO 10993-10 | | ISO 10993-10 | | |
{5}------------------------------------------------
{6}------------------------------------------------
#### Non-Clinical Test Data:
The subject device meets the following recognized standards:
- IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for Safety
- ISO 80601-2-56 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- ISO 10993-1 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process
- ISO 10993-5, Biological evaluation of medical devices-Part 5: Tests for In Vitrocytotoxicity
- ISO 10993-10, Biological evaluation of medical devices-Part 10: Tests for irritation and delayed-type hypersensitivity
The subject device is classified as a skin contacting device with a contact duration of less than 30 days. Biocompatibility tests have been conducted on the subject device, including cytotoxicity, sensitization, and skin irritation. The test results show that the subject device met the requirements of the standards.
### Substantial Equivalence:
The subject and predicate devices are exactly the same. Therefore, the PlusCare Temperature Probe is substantially equivalent to the Unimed Temperature Probe cleared under K121427 with respect to the intended use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.