The EyeCTester Model iOS is a mobile software as a medical device (SaMD) app for adults ages 22 and above. It is intended to serve as an aid to eyecare providers to monitor and detect central / paracentral visual distortions (maculopathies) as well as central 10-degree visual field scotomas. EyeCTester is not intended to screen or diagnose, but it is intended to alert a healthcare provider of any changes in a patient's central visual status. The device is intended for both at-home and clinical use. The EyeCTester is indicated to be used only with compatible mobile devices.
Device Story
Mobile SaMD app for remote monitoring of central vision; inputs include patient-reported visual distortions/scotomas via interactive Amsler Grid test on smartphone; patient fixates on pulsating point while manually outlining missing or distorted areas on grid; app aggregates responses into report for physician review; used at home or in clinic; physician interprets data to monitor changes in visual status over time; aids clinical decision-making between office visits; benefits patient through continuous monitoring of maculopathies and visual field defects.
Clinical Evidence
Prospective, randomized, cross-over, 2-site clinical studies compared EyeCTester to paper Amsler Grid in healthy volunteers and patients with neuro-ophthalmic/retinal diseases. Studies evaluated repeatability and agreement between digital and paper formats. Results confirmed EyeCTester is consistent and repeatable with paper analogs. Human error rate in app testing was <1%. Human factors validation confirmed safe use and acceptable risk profile.
Technological Characteristics
Mobile software application (SaMD) for iOS; utilizes Amsler Grid testing; cloud-based data storage with HIPAA/HiTech compliance; physician dashboard for data review; software developed per IEC 62304.
Indications for Use
Indicated for adults ages 22+ to aid eyecare providers in monitoring and detecting central/paracentral visual distortions (maculopathies) and central 10-degree visual field scotomas. Not for screening or diagnosis. Used in home or clinical settings.
Regulatory Classification
Identification
An Amsler grid is a device that is a series of charts with grids of different sizes that are held at 30 centimeters distance from the patient and intended to rapidly detect central and paracentral irregularities in the visual field.
{0}------------------------------------------------
September 26, 2022
Image /page/0/Picture/13 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on a separate line below "U.S. Food & Drug". The logos are placed side-by-side.
Neuro-Eye Diagnostic Systems LLC % Isabella Schmitt Director of Regulatory Affairs Proxima Clinical Research, Inc. 2450 Holcombe Blvd. Houston, Texas 77071
Re: K203594
Trade/Device Name: EyeCTester Regulation Number: 21 CFR 886.1330 Regulation Name: Amsler Grid Regulatory Class: Class I Product Code: OTW
Dear Isabella Schmitt:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 7, 2022. Specifically, FDA is updating this SE Letter as an administrative correction for a typographical error in your Trade/Device Name.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Elvin Ng, OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 240-402-4662, Elvin.Ng@fda.hhs.gov.
Sincerely,
Elvin Y. Ng-S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
September 7, 2022
Image /page/1/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features an abstract symbol, while the FDA part includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Neuro-Eye Diagnostic Systems LLC % Isabella Schmitt Director of Regulatory Affairs Proxima Clinical Research, Inc. 2450 Holcombe Blvd. Houston, Texas 77071
Re: K203594
Trade/Device Name: EyeQTester Regulation Number: 21 CFR 886.1330 Regulation Name: Amsler grid Regulatory Class: Class I Product Code: QTW Dated: June 4, 2021 Received: June 14, 2021
Dear Isabella Schmitt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203594
Device Name EyeCTester Monitoring App
Indications for Use (Describe)
The EyeCTester Model iOS is a mobile software as a medical device (SaMD) app for adults ages 22 and above. It is intended to serve as an aid to eyecare providers to monitor and detect central visual distortions (maculopathies) as well as central 10-degree visual field scotomas.
EyeCTester is not intended to screen or diagnose, but it is intended to alert a healthcare provider of any changes in a patient's central visual status. The device is intended for both at-home and clinical use. The EyeCTester is indicated to be used only with compatible mobile devices.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Owner: | Neuro-Eye Diagnostic Systems, LLC |
|------------------------------------|-------------------------------------|
| | 2020 Quenby Street |
| | Houston, TX 77005 |
| | |
| Official Contact: | Jade Schiffman, MD |
| | Telephone: +1 (281) 701-0577 |
| | E-mail: jschiffman@neuroeye.com |
| | |
| Representative Consultant Contact: | Isabella Schmitt, RAC |
| | Proxima Clinical Research, Inc. |
| | 2450 Holcombe Blvd, Suite J |
| | Houston, TX 77021 |
| | Telephone: +1 (404) 205-4653 |
| | E-mail: Isabella@ProximaCRO.com |
| Date Summary Prepared: | 06 September 2022 |
| Trade Name: | EyeCTester - Monitoring application |
| Common Name: | Digital Amsler Grid |
| Classification: | Class I |
| Classification Number(s): | 21 CFR 886.1330 |
| Product Code(s): | QTW |
| Classification Advisory Committee: | Ophthalmology |
| Predicate Device(s): | Alleye (K180895) |
{5}------------------------------------------------
#### Device Description:
The EyeCTester - Monitoring Application comprises a survey and a series of tests to provide remote monitoring of the patient's visual parameters from the patient's home. The EyeCTester -Monitoring application is available to patients with a prescription from the healthcare provider managing their vision. Prescribed, routine patient testing via the app allows providers to monitor vision health in the interim period between visits to the physician's clinical practice. The tool is not intended to replace the need for an in-person eye exam with a professional eve care provider. EyeCTester is not intended to diagnose the patient; diagnosis and management is the responsibility of the qualified team prescribing and interpreting the app measurements.
Prior to regular, at-home use, patients will receive training in the clinic and will then undergo training within the app and verify their understanding of how to complete testing. Pop-ups will appear throughout the app reminding the patient that the app's purpose is to monitor changes in vision, not to provide a diagnosis.
The app uses an Amsler Grid test to detect changes in visual distortions / scotomas / field cuts within the central 10 degrees of vision. The test prompts patients to outline and define distorted and/or missing areas on a series of grids while staring at a pulsating fixation point. At the end of the test, the grids are combined to display the patient's responses. The results of the assessments are summarized in a report that is provided to the prescribing healthcare provider to be interpreted and evaluated for any changes over time. A comparison of the EyeCTester application with its predicate is shown in the table below.
This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, for functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review.
| Specification | EyeCTester (Subject Device) | Alleye |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K203594 | K180895 |
| Product Name | EyeCTester App | Alleye |
| Sponsor | Neuro-Eye Diagnostic Systems, LLC | Oculocare Medical AG |
| Product Code(s) | QTW | HOQ |
| Regulation Number(s) | 21 CFR 886.1330 | 21 CFR 886.1330 |
| Classification | Class I | Class I |
| Rx / OTC | Rx Only | Rx Only |
| Indications for Use | The EyeCTester Model iOS is a mobile<br>software as a medical device (SaMD)<br>app for adults ages 22 and above. It is<br>intended to serve as an aid to eyecare<br>providers to monitor and detect central /<br>paracentral visual distortions<br>(maculopathies) as well as central 10-<br>degree visual field scotomas. | The Alleye is a mobile medical<br>software application indicated for the<br>detection and characterization<br>of metamorphopsia, a visual distortion,<br>in patients with age-related macular<br>degeneration (AMD) and as an aid in<br>the monitoring of the progression of<br>this condition in respect of |
Table 1: Comparison of EyeCTester - Monitoring Application to Identified Predicate
{6}------------------------------------------------
| Specification | EyeCTester (Subject Device) | Alleye |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | EyeCTester is not intended to screen or<br>diagnose, but it is intended to alert a<br>healthcare provider of any changes in a<br>patient's central visual status. The<br>device is intended for both at-home and<br>clinical use. The EyeCTester is<br>indicated to be used only with<br>compatible mobile devices. | metamorphopsia. It is intended to be<br>used by patients who have the<br>capability to regularly perform a simple<br>self-test at home. |
| Testing Setting | At home and in clinic | At home |
| Hardware Platform | Software application to be used at<br>home on a patient's personal<br>smartphone or in clinic on an office<br>supplied smartphone | Software application to be used at home<br>on a patient's personal smartphone |
| Amsler Grid / Visual<br>Field Algorithm | Measures 10-degree visual field for<br>distortions, including crooked and<br>doubling, and missingness defects, via<br>presentation of a series of grids on<br>which a user fixes their vision and<br>manually selects and identifies areas of<br>visual defects. | Measures visual field for distortions,<br>including crooked defects, through an<br>alignment hyperacuity task |
| Data Upload and<br>Storage | Vision test data are uploaded from the<br>patient's smartphone to a HIPAA and<br>HiTech-compliant server. The data are<br>presented on an online physician<br>dashboard and within the provider's<br>app for review and comparison by the<br>prescribing healthcare professional. | Vision test data are uploaded from the<br>patient's phone and are stored and<br>transmitted in encrypted form. The<br>Alleye platform is HIPAA-compliant,<br>and Alleye data is stored on physical<br>servers. The data are presented on an<br>online physician dashboard for review<br>and comparison by the prescribing<br>healthcare professional. |
## Intended Use / Indications for Use:
The EyeCTester Model iOS is a mobile software as a medical device (SaMD) app for adults ages 22 and above. It is intended to serve as an aid to eyecare providers to monitor and detect central / paracentral visual distortions (maculopathies) as well as central 10-degree visual field scotomas.
EyeCTester is not intended to screen or diagnose, but it is intended to alert a healthcare provider of any changes in a patient's central visual status. The device is intended for both at-home and clinical use. The EyeCTester is indicated to be used only with compatible mobile devices.
## Technological Characteristics:
The EyeCTester application provides remote visual monitoring for patients through Amsler Grid testing. Patients complete prescribed testing protocols within the application for one or both eyes
{7}------------------------------------------------
as assigned by their eyecare provider. Providers are then able to access and interpret the results of these tests through the app or an online server and share them with the patient after interpretation, as appropriate. Providers may also be alerted when changes in visual parameters are reported by patients through the testing, based on prespecified rules set by the provider or clinic.
#### Software Testing:
The software development and testing was executed in compliance with IEC 62304 Medical Device Software Life Cycle Processes. The results of these tests indicate that EyeCTester Amsler Grid/ Visual Field functionality is substantially equivalent to the predicate device.
#### Human Factors / Usability Engineering Testing:
NEDS conducted a Human Factors Validation study in which participants without visual defects were enrolled to assess the user performance and test results of the EyeCTester compared to the standard, in-office examination tools. Each participant underwent the tests twice to ensure the validity of the findings. The Human Factors Validation study indicated minimal human error risks, many of which are shared with the non-digital test analogs. Additionally, the digital, at-home, mobile (smartphone) format allow considerable improvements to the test environment and test preparation (e.g., education in a quiet and self-paced environment), compared to non-digital test analogs. Overall, all participants were observed to safely use EyeCTester's features. Based on the results from this study, the risk profile of the application is confirmed to be acceptable. The Human Factors Validation study has shown that the EyeCTester application is safe and effective for the intended users, uses, and use environments.
#### Clinical Testing:
Clinical testing was performed to validate the visual parameter assessments of the EyeCTester compared to the standard of care modality (paper Amsler Grid) in both healthy volunteers and patients with neuro-ophthalmic, retinal, and other diseases. Both studies were approved by wcgIRB.
These clinical studies were prospective, randomized, cross-over, 2-site studies that involved a test and re-test to evaluate repeatability and agreement between the subject device and the paper format. These studies demonstrated that a patient may use the EyeCTester application to gather psychophysiological measurements of central and paracentral vision parameters related to normative functioning of the human visual system, and these measurements are useful for remote monitoring in between clinic visits.
The data from this clinical study provide affirmation that the EyeCTester app-based test is repeatable and consistent with the results of the paper analogue. The Amsler Grid testing was consistent and repeatable across the two testing methods. An additional metric assessed during the study was rate of human error in completing the test. It was confirmed that the human error rate was below 1% in the EyeCTester app test. Overall, the clinical evaluations support the use of the EyeCTester for testing patients using the Amsler Grid test.
{8}------------------------------------------------
#### Conclusion:
In conclusion, the NEDS EyeCTester - Monitoring Application is demonstrated to be as safe and effective and perform as well as the identified predicate device. The Amsler Grid test has the same intended use, indication, technological characteristics, and principles of operation as the predicate. Any differences between the EyeCTester application's functionalities and the predicate device's do not alter the intended use of the device and do not raise new or different questions regarding its safety and effectiveness when use as labelled. Clinical and Human Factors testing demonstrate that the device performs as intended. Thus, the EyeCTester application is substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.