K203586 · Corneagen, Inc. · OTZ · Feb 2, 2022 · Ophthalmic
Device Facts
Record ID
K203586
Device Name
EndoSerter-PL
Applicant
Corneagen, Inc.
Product Code
OTZ · Ophthalmic
Decision Date
Feb 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The EndoSerter®-PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4.0 mm incision during endothelial keratoplasty procedures and for loading and storage of donor transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
Device Story
EndoSerter®-PL is a sterile, single-use, manual ophthalmic surgical instrument. It facilitates loading, storage, and transport of donor corneal endothelial allograft tissue by eye bank technicians; subsequently used by ophthalmic surgeons to deliver the tissue into the anterior chamber during endothelial keratoplasty. The device holds the graft in a rolled configuration. By enabling pre-loading and transport, it streamlines surgical workflows. The device is purely mechanical; no electronic or software components are involved. It benefits patients by providing a standardized, sterile method for graft delivery, reducing potential for tissue damage during insertion.
Clinical Evidence
Bench testing only. Studies included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity), tissue handling/stability (endothelial cell loss assessment), sterilization validation (SAL 10^-6), shelf life/transportation (leak testing), aseptic handling (USP <71>), and dimensional/mechanical functional testing. No clinical data presented.
Technological Characteristics
Handheld, manual, single-use ophthalmic instrument. Materials: polycarbonate, isoplast, polyimide, stainless steel, silicone, ABS, and colorants. Biocompatibility per ISO 10993-1. Sterilization via radiation. Dimensions support ≤8.0 mm graft and ≥4.0 mm incision. No software or energy source.
Indications for Use
Indicated for patients requiring endothelial keratoplasty (EK) for corneal endothelial disease (e.g., Fuchs' dystrophy, bullous keratopathy, transplant failure). Used for insertion of corneal endothelial allograft tissue (≤8.0 mm diameter, ≤100 microns thickness) through ≥4.0 mm incision.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 2, 2022
CorneaGen, Inc % Omid Kodai President Khodai Consulting, Inc. 23 Bellflower Lake Forest, California 92630
Re: K203586
Trade/Device Name: EndoSerter®-PL Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I, reserved Product Code: OTZ Dated: December 22, 2021 Received: December 27, 2021
Dear Omid Kodai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203586
Device Name EndoSerter®-PL
### Indications for Use (Describe)
The EndoSerter® PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4.0 mm incision during endothelial keratoplasty procedures and for loading and storage of donor transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### l. SUBMITTER
CorneaGen, Inc. 101 N Chestnut St Suite 303 Winston Salem NC 27101 Phone: 336-516-9600 Contact Person: Thomas Miller Date Prepared: February 1, 2022
#### DEVICE II.
| Name of Device: | EndoSerter®-PL |
|----------------------|------------------------------------------------------------------------------|
| Common Name: | Graft Insertion Device |
| Classification Name: | Graft Insertion Instrument for Endothelial Keratoplasty (21<br>CFR 886.4300) |
| Regulatory Class: | 1, Reserved |
| Product Code: | OTZ |
#### lll. PREDICATE DEVICE
## Predicate Device:
| 510(k)<br>Number | Trade or Proprietary or<br>Model Name | Manufacturer | Clearance Date |
|------------------|---------------------------------------|-------------------|-----------------|
| K121579 | EK Delivery Device | TDAK Medical Inc. | October 3, 2012 |
This predicate has not been subject to a design-related recall.
## Reference Device:
| 510(k)<br>Number | Trade or Proprietary or<br>Model Name | Manufacturer | Clearance Date |
|------------------|---------------------------------------|--------------|------------------|
| K090626 | EndoSerter® | CorneaGen | January 21, 2011 |
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Endothelial keratoplasty (EK), also referred to as posterior lamellar keratoplasty, is a form of corneal transplantation during which the diseased inner layer of the cornea, the endothelium, is replaced with healthy donor tissue. Diseases that affect the endothelial layer include Fuchs' endothelial dystrophy, aphakic and pseudophakic bullous keratopathy (corneal edema following cataract extraction), and failure or rejection of a previous corneal transplant.
The EndoSerter®-PL is a sterile, single use, handheld, manual ophthalmic surgical instrument. It is used to preload a processed donor corneal endothelial allograft for storage and transportation to the ophthalmic surgeon and to deliver the allograft into the anterior chamber of the eye during corneal surgery. It is designed to deliver a corneal endothelial allograft into the eye during corneal endothelial keratoplasty. The loading and storage of donor tissue for transport to the surgeon is performed by trained technician at the eye bank. Users of this graft insertion instrument for EK are trained ophthalmic surqeons in a healthcare facility or hospital.
#### V. INDICATIONS FOR USE
The indications for use statement for the EndoSerter®-PL by CorneaGen is:
The EndoSerter®-PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4 mm incision during endothelial keratoplasty procedures and for loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
The Indications for Use statement for the EndoSerter®-PL is not identical to the predicate or reference devices; however, the differences do not alter the intended therapeutic use of the device, nor do they affect the safety and/or effectiveness of the device relative to the predicate or reference devices. The subject, predicate, and reference devices have the same intended use for the insertion of corneal endothelial allograft tissue during endothelial keratoplasty procedures.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE AND REFERENCE DEVICES
Graft Insertion for Endothelial Keratoplasty (EK) is the technological principle for both the subject and predicate as well as the reference devices. It is based on the use of insertion instrumentation for corneal tissue graft EK procedures.
At a high level, the subject and predicate devices and when appropriate the reference device are based on the following same technological elements:
- . Same mode of operation in the insertion of corneal endothelial graft for patients requiring Endothelial Keratoplasty (EK)
- Used for loading and storage of donor tissue during transport .
- Delivers a circular endothelial tissue button in a rolled configuration .
- . Provided as a sterile, single use, disposable device
- Loaded by trained technicians at the eye bank .
- . Donor posterior lamellar endothelial graft of central corneal thickness of 100 microns
The following technological differences exist between the subject and predicate devices:
- . Minimum 4 mm incision for the subjective device and 5.1 mm or larger incision for the predicate device. However, the subject device and the reference device are the same in terms of incision size of minimum 4 mm incision.
- Insertion of corneal endothelial allograft tissue measuring less than or equal to . 8.0 mm and the predicate and reference device requiring the insertion of donor tissue diameter of less than or equal to 8.5 mm. This difference of 0.50 mm graft size is negligible and does not introduce any potential impact to safety and/or effectiveness.
- Subject device is manufactured using polycarbonate, isoplast, polyimide, ● stainless steel, silicone, acrylonitrile butadiene styrene (ABS), white, blue, and black colorant. Manufacturing materials for the predicate device is not commercially available or available in public records. The reference device is manufactured using polycarbonate, isoplast, polyimide, stainless steel, white, blue, and black colorant. The addition of silicone and ABS does not introduce any
{6}------------------------------------------------
impact to safety and/or effectiveness based upon biocompatibility testing performed in accordance with ISO 10993-1.
- . Subjective device has tissue storage capability up to 48 hours and the predicate device has tissue storage capability up to 72 hours. This difference does not introduce any potential impact to the safety and/or effectiveness of the subject device in delivering the allograft, and relevant testing has confirmed the stability of the subjective device during tissue handling and stability testing.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for the EndoSerter®-PL device was conducted in accordance with the FDA Guidance issued on September 4, 2020 "Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process"." The battery of testing included the following tests which demonstrated biocompatibility of the device with direct contact to the human tissue:
- Cytotoxicity ●
- Sensitization
- Irritation ●
- Acute Systemic toxicity ●
- . Pyrogenicity
The EndoSerter®-PL device is considered tissue contacting for a duration of less than 24 hours, while the allograft tissue is considered a permanent implant.
## Tissue Handling and Stability Testing
The purpose of this study was to quantitatively determine the endothelial cell damage (endothelial cell loss) when using the EndoSerter®-PL to load, store, transport, and deliver donor allograft tissue. The test results demonstrated that the device did not have any leaking, damage, or shifting of the tissue grafts during the transportation and storage.
{7}------------------------------------------------
# Sterilization Validation
The purpose of this study was to adopt, by equivalency, the EndoSerter®-PL into the existing radiation sterilization process used for the EndoSerter®. The radiation sterilization dose range provided a Sterility Assurance Level (SAL) of 10° in accordance with FDA Guidance issued on January 21, 2016 "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
# Shelf Life and Transportation Testing
The purpose of this study was to perform leak test to demonstrate that the EndoSerter® PL does not leak corneal storage media during storage and transportation over a 48hour period prior to and after accelerated aging. The results of this leak test validated that storage media did not leak from the EndoSerter®-PL during storage and transportation over a 48-hour period prior to and after accelerated aging.
# Aseptic Handling Testing
This testing, performed in accordance with the USP<71>. Method Suitability Test procedures, was conducted to demonstrate the ability to maintain aseptic handling of tissue without increased risk of contamination on transfer to and storage in the EndoSerter®-PL. The testing results demonstrated that the aseptic handling utilized by CorneaGen to load and transport cornea tissue in the EndoSerter®-PL did not increase the risk of contamination of the tissue.
# Dimensional, Functional, and Mechanical Testing
The purpose of this study was to perform dimensional, and mechanical testing utilizing measurement of critical dimensions including additional dimensions relating to mating components, visual inspection, and fit of those components. The study confirmed that the EndoSerter®-PL components perform mechanically and functionally as designed. Therefore, supporting the acceptable use of the device components and materials required in the design and assembly of the EndoSerter®-PL.
#### VIII. CONCLUSION
The EndoSerter®-PL, the subject device, is substantially equivalent to both the predicate and the reference devices' intended use which includes:
- same mode of operation in the insertion of corneal endothelial graft for patients ● requiring Endothelial Keratoplasty (EK)
{8}------------------------------------------------
- . used for loading and storage of donor tissue during transport
- delivering a circular endothelial tissue button in a rolled configuration .
- provided as a sterile, single use, disposable device ●
- . loaded by trained technicians at the eye bank
- . posterior lamellar endothelial graft of central corneal thickness of 100 microns corneal thickness, as well as the duration of time that tissue can be stored within the device.
The minimal technological differences of EndoSerter®-PL, the subject device, to the predicate and the reference devices, do not pose any potential impact to safety and/or effectiveness as demonstrated by the performance testing conducted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.