K203579 · Draeger Medical Systems, Inc. · MSX · Jun 7, 2022 · Cardiovascular
Device Facts
Record ID
K203579
Device Name
Infinity Gateway Suite
Applicant
Draeger Medical Systems, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Jun 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
Device Story
Infinity Gateway Suite is a client-server software application facilitating data exchange between Draeger Infinity Network and hospital information systems. Server software runs on Windows Server, connecting to hospital LAN and Infinity Network. It enables remote viewing of patient data, alarm history, and 12-lead ECG export. Features include HL7/ASTM interfaces, developer tools (WinAccess API), and Service-Oriented Device Compatibility (SDC) with encryption. Users (physicians, nurses, technicians) access information via hospital workstations. The system supports clinical workflows by integrating patient data, lab reports, and admission information, promoting continuity of care. It includes a redundant server option for failover and uses USB license dongles for feature activation. The device does not connect to patients; it serves as a communication and visualization bridge for clinical data.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed the software performs according to its intended use, with all risk mitigations functioning as expected. No clinical data was required or provided.
Technological Characteristics
Client-server software architecture; runs on Windows Server 2016; supports Microsoft SQL 2017. Connectivity via TCP/IP, HL7 2.x, ASTM, and IEEE 11073 standards. Features SDC interface with encryption. Licensed via USB dongle. Compliant with ANSI AAMI ISO 14971:2019, ANSI AAMI IEC 62304:2006/A1:2016, and ANSI AAMI IEC 62366-1:2015.
Indications for Use
Indicated for use by healthcare providers (physicians, nurses, technicians) in a healthcare environment to view patient data remotely and exchange clinical/administrative information between the Infinity Network and hospital network. Not intended for direct patient connection.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, featuring a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text next to it.
Draeger Medical Systems, Inc. Eileen Boyle Manager, Regulatory Affairs 6 Tech Drive Andover, Massachusetts 01810 May 17, 2024
Re: K203579
Trade/Device Name: Infinity Gateway Suite Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MSX
Dear Eileen Boyle:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 7, 2022. Specifically, FDA is updating this SE letter as an administrative correction. A second product code, LNX, was inadvertently included.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact LCDR Stephen Browning, OHT2: Office of Cardiovascular Devices, 240-402-5241, Stephen.Browning@fda.hhs.gov.
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the words "U.S. Food & Drug Administration" in blue text.
June 7, 2022
Draeger Medical Systems, Inc. Eileen Boyle Manager, Regulatory Affairs 6 Tech Drive Andover, Massachusetts 01810
Re: K203579
Trade/Device Name: Infinity Gateway Suite Regulation Number: 21 CFR 870.2300 Regulation Name: System, Network and Communication, Physiological Monitors Regulatory Class: Class II Product Code: MSX, LNX Dated: June 1. 2022 Received: June 2, 2022
Dear Eileen Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203579
Device Name Infinity Gateway Suite
### Indications for Use (Describe)
# Indications for Use:
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
### Intended Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
## Intended Operator
The Infinity Gateway software application is intended for use by Healthcare Providers, ie. Physicians, Nurses, and Technicians.
Intended Patient Population
The Infinity Gateway software application is not intended to be connected to patients.
Intended Use Environment
The infinity Gateway software application is intended for use in a healthcare environment.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is in blue, and the two dots above the "a" are also blue. The word is slightly italicized, and the letters are close together.
# Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810-2434
# 510(k) SUMMARY
I. Submitter: Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810 USA
> Contact Person: Eileen M. Boyle Manager, Regulatory Affairs Draeger Medical Systems, Inc. Phone: 781-439-8120 E-mail:eileen.boyle@draeger.com November 19, 2021 - Date Prepared
#### II. Device
## Names / Common Names / Classification Names:
| Common Name: | Computers and Software, Medical |
|----------------------|---------------------------------------------------------|
| Trade Name: | Infinity® Gateway Suite |
| Classification Name: | System Network and Communication Physiological Monitors |
| Product Code: | MSX |
| Subsequent Code: | LNX |
| Regulatory Class: | II |
| Regulation Number: | §870.1025 |
#### III. Predicate Device
The Infinity® Gateway Suite, VF4 K043549 cleared on January 21, 2005.
| 510(k)<br>No. | Trade Name | Manufacturer | Product Code/Common Name | Regulation<br>Number |
|----------------------|------------------------------------|----------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------|
| K043549<br>1/21/2005 | Infinity<br>Gateway Suite<br>(VF4) | Draeger Medical<br>Systems, Inc. | MSX - System, Network and<br>Communication, Physiological<br>Monitors<br>LNX - Computers and Software,<br>Medical | 21CFR<br>870.1025 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue.
#### Device Description IV.
The Infinity Gateway Suite is a suite of software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
Infinity Gateway Suite is designed to support the unique needs of hospitals, and provide a complete range of options for information connectivity, including:
- Server Software Health Level seven (HL7) Interface Software Options -
- -American society for testing and materials (ASTM) Stat Lab interface
- -Developer's Tools
- -Remote View Applications
- । Pager Interface
- -Alarm history database
- -Time master functions
- -12-lead electrocardiogram (ECG) export
- Service-Oriented Device Compatibility (SDC) interface with encryption -
Infinity Gateway is designed as a client-server application that provides a way for users to view patient information on a hospital workstation that is connected via the hospital network. The server portion runs on a Windows server that is connected directly to the hospital Local Area Network (LAN) and to the Infinity Network. Data access methods are available as options that customers can purchase and enable independently once they install the basic Infinity Gateway application.
The Infinity Gateway Server facilitates the exchange of important clinical information between the Infinity/SDC protocol and existing hospital and patient care systems. The Infinity Gateway Server is designed to provide flexibility by using common healthcare protocols and data format standards for managing communications between multiple disparate systems. The user may select one or more Infinity Gateway Developer's Tools and Interface Options to create a seamless flow of information tailored to support clinical work flow.
Infinity Gateway Developer's Tools and Interface Options are licensed or "unlocked" by using option passwords that are associated with a USB license dongle. An option password is the electronic proof of purchase for the Infinity Gateway software. During the at-hospital setup procedure, the user will be asked for an option or license password which is a unique number associated with the licensing dongle issued. The licenses are always associated with the physical dongle. The Infinity Gateway developer's tools (such as WinAccess API) enable the development of custom applications to support customers' homegrown applications to support clinical research projects or downstream information systems. Finally, Infinity Gateway provides flexible deployment opportunities by leveraging the virtual machine
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color, and the background is white. The two dots above the "a" in "Dräger" are also dark blue.
technology, which facilitates cost-effective software deployments and reduces the total cost of ownership.
Infinity Gateway also promotes patient safety with efficient workflows for timely decisionmaking by integrating patient data and providing continuity-of-care support. Furthermore, it makes comprehensive clinical data available at the point-of-care by facilitating the exchange of lab reports and admission information between the Draeger Infinity Network and other hospital systems.
#### V. Indications for Use / Intended Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
# Intended Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
## Intended Operator
The Infinity Gateway software application is intended for use by Healthcare Providers, ie. Physicians, Nurses, and Technicians.
# Intended Patient Population
The Infinity Gate4way software application is not intended to be connected to patients.
## Intended Use Environment
The infinity Gateway software application is intended for use in a healthcare environment.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the background is white. The two dots above the "a" are also blue.
#### VI. Comnarison of Technological Characteristics with Predicate Device
The intended use, indications for use, performance and technological characteristics are substantially equivalent to the predicate device.
A summary of the main changes compared to the predicate device are listed in the comparison table below:
| Attribute | Predicate Device<br>Infinity Gateway Suite<br>K043549- VF4 | Subject Device<br>Infinity Gateway Suite<br>510(k)#K203579 – VF9.0 | Explanation of Similarities<br>and Differences |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Draeger Medical<br>Systems, Inc. | Draeger Medical Systems,<br>Inc. | Same |
| 510(k) Number | K043549 | K203579 | - |
| Model Number | MS16076 | MS40046 | - |
| Intended Use | The Infinity Gateway<br>software applications are<br>intended to provide<br>clinicians with the<br>capability of viewing<br>patient data remotely via<br>the Infinity Network and<br>for the data exchange of<br>select clinical and<br>administrative<br>information between the<br>Infinity Network and the<br>hospital network. | The Infinity Gateway<br>software applications are<br>intended to provide clinicians<br>with the capability of<br>viewing patient data<br>remotely via the Infinity<br>Network and for the data<br>exchange of select clinical<br>and administrative<br>information between the<br>Infinity Network and the<br>hospital network. | Same, no change since VF4 |
| Update the supported<br>version Microsoft<br>SQL to 2017 | 2008 | 2017 | 2008 will no longer be a<br>supported version |
| Incorporate support<br>for Microsoft SQL<br>2017 and Microsoft<br>Server Operating<br>System 2016 | Windows NT 4.0,<br>Windows 2000 or<br>Windows XP | New feature in VF9.0<br>Support Microsoft SQL 2017<br>Express and Standard.<br>Support Microsoft Server<br>Operating System 2016 for<br>the server application. | Windows NT 4.0, Windows<br>2000, Windows XP will no<br>longer be supported<br>versions. |
| Make the Service-<br>Oriented Device<br>Compatibility (SDC)<br>a locked option with<br>encryption | Not part of VF4<br>software. | SDC defines a<br>communication architecture<br>to establish distributed<br>systems of medical devices<br>in clinical environments. | This new feature is for use<br>with Draeger SDC devices.<br>This license option allows<br>for interoperability and<br>connectivity with the<br>integrated Clinical<br>Assistance Package which |
| Attribute | Predicate Device<br>Infinity Gateway Suite<br>K043549 - VF4 | Subject Device<br>Infinity Gateway Suite<br>510(k)#K203579 - VF9.0 | Explanation of Similarities<br>and Differences |
| | | | consists of an anesthesia<br>workstation, a patient<br>monitoring system and the<br>Hospital Information<br>System:<br>Infinity Gateway software<br>VF9.0<br>Infinity Acute Care System<br>(IACS) patient monitor with<br>software version VG7.1<br>Connectivity Converter<br>CC300 with software<br>version 1.1<br>Perseus A500 Anesthesia<br>Machine<br>Babylog VN500<br>Evita Infinity V500<br>Ventilator |
| Redundant Server<br>Option | Not part of VF4<br>software. | An administrator performs<br>the switchover manually by<br>powering up or enabling the<br>Gateway service on the<br>redundant server. | Gateway supports the use of<br>an optional redundant<br>server. This is a licensed<br>option. It allows a separate<br>server to be configured in<br>advance which can<br>temporarily replace a failed<br>Gateway server. |
| IHE/HL7 Alarm<br>Interface Locked<br>Option (For SDC<br>Only) | Not part of VF4<br>software. | New feature in VF9.0<br>The IHE/HL7 Alarm<br>Communication interface<br>exports Alarm data from<br>devices on the SDC network<br>to a CIS for long term<br>storage and record keeping.<br>The interface support IEEE<br>11073 output and can be<br>configured to be IHE PCD<br>ACM Compliant. | The IHE/HL7 alarm<br>communication<br>management interface is<br>only available for use with<br>SDC devices. |
| Attribute | Predicate Device<br>Infinity Gateway Suite<br>K043549- VF4 | Subject Device<br>Infinity Gateway Suite<br>510(k)#K203579 – VF9.0 | Explanation of Similarities<br>and Differences |
| Intended Population | The communication<br>network and central<br>monitoring device are<br>not connected to patients. | The communication network<br>and central monitoring<br>device are not connected to<br>patients. | Same |
| Intended Environment | A healthcare facility<br>where healthcare<br>professionals provide<br>patient care. | A healthcare facility where<br>healthcare professionals<br>provide patient care. | Same |
| Gateway Pager<br>Access application | Yes | Yes | Same |
| Global Session<br>Manager | Yes | No | Global Session Manager is<br>no longer supported. No<br>change to intended use. |
| Display either the<br>patient name or the<br>patient identifier on<br>WinView/Webviewer<br>Gateway | Yes | Yes, WinView/WebViewer<br>is renamed to PatientWatch | Same.<br>WinView/WebViewer is<br>renamed to PatientWatch |
| Network Technology | • TCP/IP<br>• Health Level 7 (HL7<br>2.3)<br>• ASTM<br>• HTML, Active X<br>• Windows Dynamic<br>Link<br>• Library (DLL)<br>• TAP | • TCP/IP<br>• Health Level 7 (HL7 2.x)<br>• ASTM<br>• HTML, Active X<br>• Windows Dynamic Link<br>• Library (DLL)<br>• TAP | VF9.0 added additional<br>support for protocol<br>interfaces and supports<br>different versions of the<br>HL7 2.x protocols for<br>various interfaces |
| Alarm (audible) | Yes | Yes | Audible alarm information<br>is available only if a paging<br>service provides the<br>capability.<br>No change since VF4, no<br>impact to the intended use. |
| Alarm Display String | Advisory (ADV)<br>Serious (SER)<br>Life Threatening (LT) | Advisory (ADV)<br>Serious (SER)<br>Life Threatening (LT) | Available for informational<br>purposes only. The primary<br>alarm notification is at the<br>patient bedside or the<br>Infinity Central Station.<br>No change since VF4, no<br>impact to the intended use. |
| Alarm (Visual) | Yes | Yes | Same |
| Attribute | Predicate Device<br>Infinity Gateway Suite<br>K043549 - VF4 | Subject Device<br>Infinity Gateway Suite<br>510(k)#K203579 - VF9.0 | Explanation of Similarities<br>and Differences |
| Alarm Grade | Life-Threatening<br>Serious<br>Advisory via<br>WinView/WebViewer | Yes, WinView/WebViewer<br>is renamed to PatientWatch | PatientWatch alarms are<br>available for informational<br>purposes and are not<br>available for SDC<br>The alarm grades have not<br>changed between VF4 and<br>VF9. |
| Alarm State | Active<br>Silenced<br>None | Yes, via PatientWatch | PatientWatch alarms are<br>available for informational<br>purposes and are not<br>available for SDC<br>The alarm states have not<br>changed between VF4 and<br>VF9. |
## Comparison Table Between the Predicate Device K043549, VF4 and the Subject 510(k) K203579, VF9.0:
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is written in blue. The two dots above the "a" are also blue.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Dräger, a German company that manufactures medical and safety technology products. The logo is in blue and features the company's name in a stylized font. The two dots above the 'a' are a distinctive feature of the logo. The logo is simple and recognizable.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the two dots above the "a" are also blue. The letters are large and spaced closely together. The background is white.
#### VII. Performance Data
# Verification and Validation Testing:
The verification and validation that was conducted confirmed the Infinity Gateway product performs according to intended use with no adverse effects upon other medical devices in the Infinity system to which it is connected. Testing confirmed that identified Product Risk mitigations functioned with the new code, with all test cases passing without exception.
Additionally, all possibly known use related hazards, inclusive those beyond the primary operating functions, that are related to usability have been reviewed. As a result of this review, all risk control measures identified in the risk management process are deemed adequate to further mitigate and control the risks at an acceptable level under consideration of the intended use. The results of the Verification and Validation testing confirm the modified software is substantially equivalent to the predicate device.
Biocompatibility: The proposed enhancement to the Infinity Gateway Suite involves software only. No modifications affecting biocompatibility of the device is being proposed at this time and does not apply.
Sterilization: Sterilization and shelf-life do not apply as the Infinity Gateway Suite is software only.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the background is white. The two dots above the "a" are also blue.
# Standards/Compliance Testing:
The Infinity Gateway Suite software modifications have been tested and developed in compliance with the following standards:
- ANSI AAMI ISO 14971:2019 Medical devices Applications of risk management to ﻬ medical devices
- IEEE Std 11073 -10101-2019 Health informatics Point-of-care medical device ﻬ communication. Part 10101: Nomenclature
- IEEE Std 11073-10201-2018 Health informatics Point-of-care medical device r communication Part 10201: Domain Information Model
- ANSI AAMI IEC 62304:2006/a1:2016 ﮩ
- Medical device software Software life cycle processes [Including Amendment 1 r (2016)]
- ANSI AAMI IEC 62366-1:2015 Medical devices Part 1: Application of usability ﻬ engineering to medical devices
#### VIII. Conclusion
The Infinity Gateway Suite significant change modifications are substantially equivalent to the predicate device cleared under K043549 on January 21, 2005. The intended use of the Infinity Gateway Suite as described in the product labeling has not changed as a result of the proposed modifications. Verification and Validation results for the proposed modifications support substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.