K203560 · Tso3 Inc., Now A Part of Stryker · FRC · Mar 24, 2021 · General Hospital
Device Facts
Record ID
K203560
Device Name
STERIZONE VP4 Test Pack
Applicant
Tso3 Inc., Now A Part of Stryker
Product Code
FRC · General Hospital
Decision Date
Mar 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERIZONE® VP4 Test Pack is intended to be used for routine monitoring of the STERIZONE® VP4 Sterilizer cycle (Cycle 1) and for the performance validation of the STERIZONE® VP4 Sterilizer system.
Device Story
The STERIZONE® VP4 Test Pack is a Process Challenge Device (PCD) used to monitor the STERIZONE® VP4 Sterilizer. It consists of a 10 mL plastic syringe, a plunger, a diffusion restrictor (18-gauge PTFE cannula), and a Terragene Bionova® BT96 self-contained biological indicator (SCBI) inoculated with Geobacillus stearothermophilus spores. A chemical indicator is included to distinguish processed from unprocessed packs. The device is assembled by placing the SCBI inside the syringe, inserting the plunger, and attaching the diffusion restrictor. The pack is placed in the sterilizer load; after the cycle, the SCBI is retrieved and incubated to assess spore kill. The BT96 indicator provides a rapid 30-minute readout compared to the 18-hour readout of the previous version. The device ensures the sterilization process achieves necessary conditions to kill at least 10^6 viable spores, providing a greater challenge to the process than the BI alone.
Clinical Evidence
Bench testing only. Performance evaluated per FDA guidance for Biological Indicator (BI) Premarket Notification. Testing included half-cycle survivability, full-cycle inactivation, and resistance verification. Results confirmed the SCBI in the Test Pack showed growth at half-cycle, 0% survival at full-cycle, and higher resistance compared to a naked SCBI, meeting all predetermined acceptance criteria.
Technological Characteristics
Components: 10 mL Becton-Dickinson plastic syringe with Luer-Lok, 18-gauge (1.02 mm) 2-inch (50.8 mm) PTFE diffusion restrictor, Terragene Bionova® BT96 SCBI (Geobacillus stearothermophilus ATCC 7953), and STERIZONE® CI+ chemical indicator. Single-use, disposable. Operates as a passive process challenge device for vaporized hydrogen peroxide and ozone sterilization. No software or electronic components.
Indications for Use
Indicated for routine monitoring and performance validation of the STERIZONE® VP4 Sterilizer (Cycle 1). Used as a Process Challenge Device (PCD) to simulate sterilization loads and assess process effectiveness.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
March 24, 2021
TSO3 Inc., Now a part of Stryker Karan Modi, MS, RAC Staff Regulatory Affairs Specialist 1941. Stryker Wav Portage, Michigan 49002 USA
Re: K203560
Trade/Device Name: STERIZONE VP4 Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: February 19, 2021 Received: February 23, 2021
Dear Karan Modi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, PhD. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203560
Device Name STERIZONE® VP4 Test Pack
Indications for Use (Describe)
The STERIZONE® VP4 Test Pack is intended to be used for routine monitoring of the STERIZONE® VP4 Sterilizer cycle (Cycle 1) and for the performance validation of the STERIZONE® VP4 Sterilizer system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a chemical structure diagram. The diagram includes the letters 'T' and 'S' on the left side. To the right of the letters, there is a chemical structure with an 'O' in the center and a subscript '3' at the bottom right.
K203560
# 1. Submitter
510(k) Owner:
TSO3 Inc., Now a part of Stryker 2505, avenue Dalton Québec (Québec) G1P 3S5 Canada
FDA Establishment Registration Number:
Contact Person:
3004148947
Karan Modi, MS, RAC Staff Regulatory Affairs Specialist Stryker Instruments 1941 Stryker Way Portage, Michigan 49002 USA P 269 888 0305 karan.modi@stryker.com
### Date Submitted:
March 24, 2021
#### Subject Device Name 2.
| Trade Name: | STERIZONE® VP4 Test Pack |
|------------------------------|---------------------------------------------|
| Common Name: | Biological Indicator (Test Pack) |
| Classification: | Class II |
| Classification Product Code: | FRC |
| Classification Name: | Indicator, Biological Sterilization Process |
| Classification Regulation: | 21CFR 880.2800 |
| Review Panel: | General Hospital |
| Submission Type: | Traditional 510(k) |
{4}------------------------------------------------
$$\mathbb{S}^{\mathbb{S}}$$
#### Legally Marketed Predicate Device 3.
| Predicate Device | |
|--------------------------------|---------|
| STERIZONE® VP4 Test Pack | K141580 |
| Reference Device | |
| Terragene Bionova® SCBI (BT96) | K191021 |
#### 4. Device Description
The STERIZONE® VP4 Test Pack is a Process Challenge Device (PCD) designed to have greater resistance than the worst-case sterilization load of the STERIZONE® VP4 Sterilizer. The STERIZONE® VP4 Test Pack is intended to simulate products to be sterilized and to constitute a defined challenge to the sterilization process and used to assess the effective performance of the process.
The STERIZONE® VP4 Test Pack cleared under K141580 contained a biological indicator (STERIZONE® BI+) which required 18 hours of incubation time. This Traditional 510(k) submission aims to qualify an additional biological indicator -Terragene's Bionova® BT96 (hereinafter referred to as "BT96") within the STERIZONE® VP4 Test Pack (hereinafter referred to as "Test Pack"). BT96 is a rapid biological indicator with 30 mins of incubation time and is cleared under K191021 for monitoring the effectiveness of sterilization processes that use vaporized hydrogen peroxide as the primary sterilant, including the STERIZONE® VP4 Sterilizer. The Test Pack with STERIZONE® BI+ cleared under K141580 will continue to be an option for customers to use.
The STERIZONE® VP4 Test Pack (subject device) is composed of a Self-contained Biological Indicator (the Terragene's Bionova® BT96), a 10 mL syringe and its plunger, and a diffusion restrictor. A STERIZONE® CI+ Chemical Indicator is also added, external to the syringe, to allow differentiating processed from unprocessed test packs. All components of the Test Pack are single-use, disposable items.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a chemical structure diagram. The diagram includes the letters 'T' and 'S' on the left side. To the right of the letters, there is a chemical structure with 'O' and the number '3' at the bottom right.
#### 5. Intended Use/Indications for use
The STERIZONE® VP4 Test Pack is intended to be used for routine monitoring of the STERIZONE® VP4 Sterilizer cycle (Cycle 1) and for the performance validation of the STERIZONE® VP4 Sterilizer system.
### Comparison of Technological Characteristics with the Predicate 6. Device
The following table identifies technological characteristics shared between the Predicate and Subject device:
| Features | STERIZONE® VP4 Test Pack<br>(Predicate Device- K141580) | STERIZONE® VP4 Test Pack (Subject Device) |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product description | A Self-contained BI, and a syringe with diffusion restrictor for holding the BI during the sterilization cycle, along with a process chemical indicator. | Same |
| Picture of the device | Image: STERIZONE VP4 Test Pack (Predicate Device- K141580) | Image: STERIZONE VP4 Test Pack (Subject Device) |
| Intended Use | Routine monitoring of the STERIZONE® VP4 Sterilizer cycle and performance validation of the STERIZONE® VP4 Sterilizer system | Same |
| Indications for Use | The STERIZONE® BI+ Self-contained Biological Indicator (SCBI) is intended for routine monitoring of the STERIZONE® VP4 Sterilizer, which offers a single pre-set sterilization cycle (“Cycle 1”). The SCBI should only be used in a Test Pack configuration to monitor Cycle 1. The SCBI placed within the STERIZONE® VP4 Test Pack monitors exposure to both vaporized hydrogen peroxide (H2O2 or VHP) and ozone (O3) which are both used in the STERIZONE® VP4 Sterilizer. The STERIZONE® VP4 Test Pack is intended to have equivalent to greater resistance than worst case devices and loads in any load configuration. | The STERIZONE® VP4 Test Pack is intended to be used for routine monitoring of the STERIZONE® VP4 Sterilizer cycle (Cycle 1) and for the performance validation of the STERIZONE® VP4 Sterilizer system. |
| Features | STERIZONE® VP4 Test Pack<br>(Predicate Device- K141580) | STERIZONE® VP4 Test<br>Pack (Subject Device) |
| | The STERIZONE® VP4 Test Pack is a device composed of<br>the STERIZONE® BI+ Self-contained Biological Indicator<br>(TSO3 product code 42602, including crusher), a 10 mL<br>syringe and its plunger, and a diffusion restrictor (sold in the<br>form of a kit - TSO3 product code: 44020). An external<br>STERIZONE® CI+ Chemical Indicator (TSO3 product code<br>43810) is also added to allow differentiating processed from<br>unprocessed test packs. All components of the Test Pack are<br>single-use, disposable items.<br><br>The STERIZONE® VP4 Test Pack is constructed by first<br>inserting the STERIZONE® BI+ Self-contained Biological<br>Indicator inside the syringe, with the SCBI cap facing to the<br>Luer-lock of the syringe. The plunger is then inserted to the<br>10 mL mark of the syringe. The diffusion restrictor is<br>screwed to the Luer-lock of the syringe. The chemical<br>indicator is then inserted in the opening between the plunger<br>and the syringe.<br><br>The test pack is then placed within the load on the upper<br>shelf of the STERIZONE® VP4 Sterilizer loading rack.<br><br>After processing, the SCBI is retrieved from the test pack.<br>The SCBI is intended to provide users with a means to<br>assess spore kill by the STERIZONE® VP4 Sterilizer. A<br>"no growth" result from the SCBI after 18 hours of<br>incubation indicates that the process achieved the conditions<br>necessary to kill at least $1 \times 10^6$ viable spores of Geobacillus<br>stearothermophilus (6 logs) on the SCBI inoculated stainless<br>steel carrier. | |
| Contraindication | Do not use in a system other than the STERIZONE® VP4 Sterilizer.Do not reuse any test pack component after it has been processed in the STERIZONE® VP4 Sterilizer. | Same |
| Sterilizer to<br>be<br>monitored<br>by the test<br>pack | STERIZONE® VP4 Sterilizer | Same |
| Biological<br>indicator | STERIZONE® BI+ Self-contained Biological Indicator<br>(K141580) | Terragene Bionova® SCBI<br>(BT96) (K191021) |
| Features | STERIZONE® VP4 Test Pack<br>(Predicate Device- K141580) | STERIZONE® VP4 Test<br>Pack (Subject Device) |
| | Geobacillus stearothermophilus<br>ATCC 7953 spores | Geobacillus<br>stearothermophilus ATCC<br>7953 spores |
| Holder | Syringe and plunger-<br>10 mL Becton-Dickinson (BD) plastic syringe with Luer-<br>Lok™ connector and its plunger | Same |
| Diffusion<br>restrictor | 18 gauge (1.02 mm diameter),<br>2-inch (50.8 mm) long polytetrafluoroethylene (PTFE)<br>cannula with Luer-lock attachment | Same |
| Chemical<br>Indicator | STERIZONE® CI+ Chemical Indicator | Same |
| Resistance<br>characteristics | Demonstrated to have equivalent or greater resistance than<br>the worst case devices and loads in any load configuration.<br>Demonstrated to be more resistant than the half-cycle,<br>including exposure to hydrogen peroxide and ozone. | Same |
### Table 1. Comparison of the subject and the predicate (K141580) STERIZONE® VP4 Test Pack devices
{6}------------------------------------------------
TSO₃
{7}------------------------------------------------
$$\mathbb{S}^{\mathbb{S}}$$
The intended use of the modified device, as described in the labeling, has not changed as a result of the modification.
The STERIZONE® VP4 Test Pack components are identical to its original filing with the exception of the Self-Contained Biological Indicator.
- Bionova® BT96 Super Rapid Readout Biological Indicator from Terragene -(K191021)
- A diffusion restrictor (same as predicate): PTFE canula of 18 gauge and 2 inches -(5.1 cm) in length (with a Luer-lock connector) (K141580)
- A plastic syringe (same as predicate): A 10 mL Becton Dickinson (BD) plastic syringe with Luer-Lok™ tip (K141580)
- A chemical indicator (same as predicate): STERIZONE® CI+ Chemical Indicator -(K141580 and K141698)
The difference in the self-contained biological indicator between the subject and predicate device was considered in relation to the substantial equivalence determination:
The predicate Test Pack (PCD) contains the STERIZONE® BI+ Self-contained Biological Indicator (BI+) which was cleared under K141580. The subject Test Pack
{8}------------------------------------------------
TSO₃
(PCD) contains the Bionova® BT96 Super Rapid Readout Biological Indicator from Terragene. The BT96 biological indicator is cleared for monitoring the effectiveness of STERIZONE® VP4 Sterilizer under K191021 and is therefore used in this submission as a reference device. Similar to the BI+, the BT96 is a self-contained biological indicator inoculated with at least 106 Geobacillus stearothermophilus viable bacterial spores. The qualification of BT96 as a biological challenge device for use within the Test Pack would provide a rapid readout option to the users. The BT96 biological indicator offers rapid readout (30 mins incubation time) compared to 18 hours with the current BI+.
#### 7. Performance Testing - Bench
The performance of the subject STERIZONE® VP4 Test Pack (using BT96) was evaluated in accordance with FDA guidance document for Industry and FDA Staff, Biological Indicator (BI) Premarket Notification [510(k)] Submissions," issued on October 4, 2007. All results from testing meet the predetermined acceptance criteria.
The following performance data were provided in support of the substantial equivalence determination:
| Test Name | Test Purpose | Test Acceptance<br>Criteria | Subject Device Test<br>Result |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Half-cycle<br>Survivability | Determine if the SCBI, when used<br>within the Test Pack, can provide a<br>resistance to the sterilization process<br>that is equal to or greater than the most<br>difficult item routinely processed in the<br>sterilizer, per ISO 14937, ISO 11138-7<br>and the Guidance for Industry and Staff<br>– Biological-Indicator (BI) Premarket<br>Notification [510(k)] Submission. | The SCBI in the Test<br>Pack must show<br>growth at half-cycle<br>in all the limit<br>parameter loads<br>tested. | The SCBI in the Test<br>Pack showed growth at<br>half-cycle in all the limit<br>parameter loads tested. |
| Full cycle<br>inactivation | Determine if the SCBI used within the<br>Test Pack can repeatedly and<br>consistently show an inactivation point<br>in the second half of the sterilization<br>cycle. | The SCBI in the Test<br>Packs must<br>demonstrate a<br>survival ratio of 0%<br>at least at the full<br>cycle using the worst-<br>case challenge load. | The SCBI in the Test<br>Pack shows survival<br>ratio of 0% in the second<br>half of the sterilizer<br>cycle, thus shows total<br>inactivation by the end<br>of a full cycle. |
| Increased<br>resistance | Determine if the use of the Test Pack<br>provides a greater challenge to the<br>process than the BI itself, per Guidance<br>for Industry and Staff - Biological-Indicator (BI) Premarket Notification<br>[510(k)] Submission. | The SCBI in the VP4<br>Test Pack must show<br>a higher survival ratio<br>than<br>naked<br>its<br>counterpart. | When exposed to the<br>same partial cycles, the<br>SCBI in the Test Pack<br>showed growth while the<br>naked SCBI was fully<br>inactivated. |
### Table 2. Summary of Nonclinical Testing:
{9}------------------------------------------------
Traditional 510(k) – K203560 STERIZONE® VP4 Test Pack
(ປີ
## K203560 - 510(k) Summary
#### Conclusion 8.
The performance testing demonstrates that the subject STERIZONE® VP4 Test Pack is substantially equivalent to the identified predicate STERIZONE® VP4 Test Pack (K141580).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.