← Product Code [DSA](/productcode/DSA) · K203556

# Medtronic Model 5492A, 5492V, 5492AL, 5492VL Patient Cables (K203556)

_Medtronic, Inc. · DSA · Jul 12, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203556

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DSA](/productcode/DSA.md)
- **Decision Date:** Jul 12, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Medtronic Model 5492A, 5492V, 5492AL, 5492VL patient cables are intended for use as part of a temporary cardiac pacing system. They connect atrial and ventricular pacing lead systems to a temporary external pacemaker. They are intended for use by trained clinicians in a clinical environment. For more information about the intended use of and the indications for temporary cardiac pacing systems, refer to the technical manuals for the cardiovascular stimulating instruments. The Model 5492A, 5492V, 5492AL, 5492VL reusable patient cables are intended to be used with: · Medtronic endocardial or myocardial pacing lead connector pins 0.38 to 2.41 mm (0.015 to 0.095 in.) in diameter and up to 25.4 mm (1.0 in.) long. · Medtronic cardiovascular stimulating instruments.

## Device Story

Reusable patient cables; connect atrial/ventricular pacing leads to temporary external pacemakers. Used by trained clinicians in clinical settings. Facilitates temporary cardiac pacing by providing electrical connection between lead systems and stimulating instruments. Reusable design; supplied non-sterile. Benefits patient by enabling temporary pacing therapy.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included sterilization validation, biocompatibility, EMC/electrical safety, and performance testing.

## Technological Characteristics

Reusable patient cables for temporary cardiac pacing. Connects pacing leads to external pacemakers. Compatible with lead connector pins 0.38-2.41 mm diameter. Supplied non-sterile. Evaluated for biocompatibility, electrical safety, and performance.

## Regulatory Identification

A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.

## Predicate Devices

- Medtronic Model 5487/5487L Disposable Patient Cables ([K960446](/device/K960446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

July 12, 2021

Medtronic, Inc. Alexandra Theisen Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112

#### Re: K203556

Trade/Device Name: Medtronic Model 5492A. 5492V. 5492AL. 5492VL Patient Cables Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: June 14, 2021 Received: June 15, 2021

#### Dear Alexandra Theisen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

K203556

Device Name

Medtronic Model 5492A, 5492V, 5492AL, 5492VL Patient Cables

Indications for Use (Describe)

The Medtronic Model 5492A, 5492V, 5492AL, 5492VL patient cables are intended for use as part of a temporary cardiac pacing system. They connect atrial and ventricular pacing lead systems to a temporary external pacemaker. They are intended for use by trained clinicians in a clinical environment. For more information about the intended use of and the indications for temporary cardiac pacing systems, refer to the technical manuals for the cardiovascular stimulating instruments.

The Model 5492A, 5492V, 5492AL, 5492VL reusable patient cables are intended to be used with: · Medtronic endocardial or myocardial pacing lead connector pins 0.38 to 2.41 mm (0.015 to 0.095 in.) in diameter and up to 25.4 mm (1.0 in.) long.

· Medtronic cardiovascular stimulating instruments.

| Type of Use (Select one or both, as applicable)                                        | <table><tr><td><input checked="checked" type="checkbox"/> Remediation Use (Part 21 CFR 601 Subpart D)     </td></tr><tr><td><input type="checkbox"/> Same Therapeutic Context Use (21 CFR 601 Subpart C)     </td></tr></table> | <input checked="checked" type="checkbox"/> Remediation Use (Part 21 CFR 601 Subpart D) | <input type="checkbox"/> Same Therapeutic Context Use (21 CFR 601 Subpart C) |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <input checked="checked" type="checkbox"/> Remediation Use (Part 21 CFR 601 Subpart D) |                                                                                                                                                                                                                                 |                                                                                        |                                                                              |
| <input type="checkbox"/> Same Therapeutic Context Use (21 CFR 601 Subpart C)           |                                                                                                                                                                                                                                 |                                                                                        |                                                                              |

|X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of

{3}------------------------------------------------

# 510(k) Summary

[As required by 21 CFR 807.92]

Date Prepared: December 04, 2020

| General Information |                                                                                                                                                                      |                                                                                                                                                         |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name          | Medtronic Model 5492A, 5492V, 5492AL, 5492VL Patient Cables                                                                                                          |                                                                                                                                                         |
| Common Name         | Patient cable                                                                                                                                                        |                                                                                                                                                         |
| Product Code        | DSA                                                                                                                                                                  |                                                                                                                                                         |
| Classification      | II                                                                                                                                                                   |                                                                                                                                                         |
| Regulation Number   | 21 CFR 870.2900                                                                                                                                                      |                                                                                                                                                         |
| Panel               | Cardiovascular                                                                                                                                                       |                                                                                                                                                         |
| Predicate Device    | Medtronic Model 5487/5487L Disposable Patient Cables K960446                                                                                                         |                                                                                                                                                         |
| Submitter           | Medtronic, Inc.<br>Cardiac Rhythm Management<br>8200 Coral Sea Street<br>Mounds View, MN 55112                                                                       |                                                                                                                                                         |
| Contact Person      | Primary Contact<br>Alexandra Theisen<br>Senior Regulatory Affairs Specialist<br>alexandra.r.theisen@medtronic.com<br>Tel: 763.526.6758 (direct)<br>Fax: 651.367.0603 | Alternate Contact<br>Wendy Saunders<br>Regulatory Affairs Director<br>wendy.a.saunders@medtronic.com<br>Tel: 763.526.8134 (direct)<br>Fax: 651.367.0603 |

# Brief Device Description

The 5492 A-V Patient Cables are reusable patient cables that are designed to connect a temporary pacemaker to atrial and ventricular pacing lead systems.

# Indications for Use

The Medtronic Model 5492A, 5492V, 5492AL, 5492VL patient cables are intended for use as part of a temporary cardiac pacing system. They connect atrial and ventricular pacing lead systems to a temporary external pacemaker. They are intended for use by trained clinicians in a clinical environment. For more information about the intended use of and the indications for temporary cardiac pacing systems, refer to the technical manuals for the cardiovascular stimulating instruments.

The Model 5492A, 5492V, 5492AL, 5492VL reusable patient cables are intended to be used with:

- . Medtronic endocardial or myocardial pacing lead connector pins 0.38 to 2.41 mm (0.015 to 0.095 in.) in diameter and up to 25.4 mm (1.0 in.) long.
- Medtronic cardiovascular stimulating instruments. ●

{4}------------------------------------------------

# Technological Characteristics

Intended use, design, performance and technological characteristics are substantially equivalent to the predicate devices referenced.

When compared to the predicate device (K960446), the 5492 A-V Patient Cables presented in this submission have similar:

- o Intended use/indications for use
- Operating principle ●
- Design features ●
- Device functionality .

The 5492 A-V Patient Cables and the predicate device differ in the following:

- Usage (i.e. reusable versus disposable) ●
- Sterility (i.e. supplied non-sterile versus supplied sterile) ●

### Substantial Equivalence and Summary of Studies:

Technological differences between the subject and predicate devices have been evaluated through bench tests. The following test data provided in this Traditional 510(k) submission has been completed:

- Sterilization (reference Sterilization)
- Biocompatibility (reference Biocompatibility) ●
- o Electrical Safety (reference EMC and Electrical Safety)
- o Performance Testing (reference Performance Testing - Bench)

# Conclusion

Medtronic has demonstrated that the 5492 A-V Patient Cables described in this submission result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are substantially equivalent from the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K203556](https://fda-staging.innolitics.com/device/K203556)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
