K203542 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · NHA · Aug 2, 2021 · Dental
Device Facts
Record ID
K203542
Device Name
Neodent Implant System - Mini Abutment 60°
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S.A.
Product Code
NHA · Dental
Decision Date
Aug 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mini Conical Abutments are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. It may be used with single-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Device Story
Neodent Implant System - Mini Abutment 60° is a dental abutment used in conjunction with Zygomatic Implants. Manufactured from titanium alloy (Ti6Al4V-ELI), the device features an anti-rotational implant-to-abutment interface compatible with GM Zygomatic Implants. It is screw-retained to the implant and supports multi-unit restorations. Used in clinical dental settings by trained professionals, the abutment serves as a connector between the zygomatic implant and the prosthetic restoration. By providing a stable interface, it enables the restoration of aesthetics and chewing function in patients with severe jaw resorption. The device is provided sterile for single use.
Clinical Evidence
Bench testing only. Evidence includes dynamic fatigue testing per ISO 14801, static torsional loading, sterilization validation (ISO 11135), ethylene oxide residual assessment (ISO 10993-7), biocompatibility (ISO 10993-5, ISO 10993-18), and shelf-life validation (ASTM F1980). No clinical data provided.
Technological Characteristics
Material: Titanium alloy (Ti6Al4V-ELI) per ASTM F136. Design: 60° angulation, various gingival heights, screw-retained, anti-rotational GM interface. Sterilization: Ethylene oxide (SAL 1x10^-6). Single-use. Compatible with Neodent GM Line impression copings, provisional cylinders, and protective cylinders.
Indications for Use
Indicated for patients with severe jaw resorption requiring Zygomatic Implants to restore aesthetics and chewing function. Suitable for single-stage procedures and multiple unit restorations; allows immediate loading given good primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Neodent Implant System - Zygomatic Implants and Abutments (K190718)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
JJGC Industria e Comercio de Materiais Dentarios S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K203542
Trade/Device Name: Neodent Implant System - Mini Abutment 60° Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 2, 2021 Received: July 8, 2021
### Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203542
Device Name Neodent Implant System - Mini Abutment 60°
#### Indications for Use (Describe)
The Mini Conical Abutments are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. It may be used with single-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <b> <span style="font-size: 10pt"> <span style="text-decoration: overline">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </b> | <span style="font-size: 10pt"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K203542
ADMINISTRATIVE INFORMATION
| Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent) |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| | Av. Juscelino Kubitschek de Oliveira, 3291 |
| | Curitiba, Parana, Brazil 81270-200 |
| | Registration No.: 3008261720 |
| | Owner/Operator No.: 10031702 |
| Contact Person | Jennifer M. Jackson, MS |
| | Director of Regulatory Affairs, |
| | Straumann USA |
| | E-mail: jennifer.jackson@straumann.com |
| | Telephone (978) 747-2509 |
| Date Prepared | 29/Jul/2021 |
| Preparer / Alternate Contact | Mariana Soares Hartmann |
| | Regulatory Affairs Analyst |
| | JJGC Indústria e Comércio de Materiais Dentários SA |
| | E-mail: mariana.hartmann@neodent.com |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/ Proprietary Name | Neodent Implant System - Mini Abutment 60° |
| Common Name | Endosseous dental implant |
| Classification Name(s) | Endosseous dental implant Abutment |
| Classification Regulation(s) | 21 CFR 872.3630, Class II |
| Product Code(s) | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
| PREDICATE DEVICE INFORMATION | |
| Primary Predicate Device | K190718 - Neodent Implant System - Zygomatic Implants and<br>Abutments, JJGC Indústria e Comércio de Materiais Dentários SA |
JJGC Indústria e Comércio de Materiais Dentários S.A.
{4}------------------------------------------------
K161598 - NobelZygoma 0°, Nobel Biocare AB Reference Predicate Devices K182620 - MRI Safety Information Labeling Change, JJGC Indústria e Comércio de Materiais Dentários SA
#### INDICATIONS FOR USE
The Mini Conical Abutments are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. It may be used with single-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
#### SUBJECT DEVICE DESCRIPTION
#### Mini Abutments 60°
- Intended for single use;
- Provided sterile via ethylene oxide gas;
- Manufactured of titanium alloy (Ti6Al4V-ELI) per ASTM F136;
- Available in different gingival heights;
- Screw-retained to the implant;
- Provided with an anti-rotational implant-to-abutment interface compatible with GM Zygomatic Implants;
- Provided with coronal geometries in rotational (non-indexed) versions to support multi-unit restorations
- . The subject devices are compatible with Impression copings, Provisional cylinders, Protective cylinders and Copings of the Neodent GM Line previously cleared per market.
### TECHNOLOGICAL CHARACTERISTIC COMPARISON
The Substantial Equivalence Comparison table is provided on the pages that follow below.
{5}------------------------------------------------
| | SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATE | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| COMPARISON | Subject<br>Mini Abutments 60°<br>JJGC Indústria e Comércio de Materiais Dentários<br>S.A. | K190718<br>Neodent Implant System - Zygomatic Implants and<br>Abutments<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | K161598<br>NobelZygoma 0°<br>Nobel Biocare AB | EQUIVALENCE DISCUSSION |
| Indications for Use | The Mini Conical Abutments are indicated for<br>use with Zygomatic Implants, in cases of severe<br>jaw resorption, in order to restore patient<br>aesthetics and chewing function. It may be<br>used with single-stage or two-stage<br>procedures, for multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | Zygomatic Implants are indicated for surgical<br>installation in the zygoma region, in cases of severe<br>jaw resorption, in order to restore patient esthetics<br>and chewing function. Zygomatic Implants are<br>recommended for the posterior (pre-molar/molar)<br>region, one implant on each side, with at least two<br>standard dental implants in the anterior region to<br>support a fixed restoration. Zygomatic Implants may<br>be loaded immediately when good primary stability<br>is achieved and with appropriate occlusal loading. | NobelZygoma implants are end<br>osseous dental implants intended<br>to be surgically placed in the bone<br>of the upper jaw arch to provide<br>support for prosthetic devices,<br>such as artificial teeth, in order to<br>restore patient esthetics and<br>chewing function. The<br>NobelZygoma Implants are<br>appropriate for immediate loading<br>when good primary stability is<br>achieved and with appropriate<br>occlusal loading. | Equivalent<br>Subject and primary<br>predicate devices are<br>indicated to be used<br>together. The wording of the<br>subject devices is a slightly<br>different from the one of the<br>primary predicate because<br>the current submission is<br>regarding abutments only<br>and the predicate devices<br>contained also implants in<br>the submission. |
| Implant-to-Abutment<br>Interface | GM | GM | GM | Identical<br>The implant-to abutment<br>interface is equivalent to that<br>of the primary predicate<br>device. |
| Abutment Design | Gingival Height: 1.5; 2.5 mm<br>Angulation: 60° | Gingival Height: 1.5; 2.5 mm<br>Angulation: 45° | Gingival Height: 6; 8 mm<br>Angulation: 45° and 60° | Equivalent<br>The subject device gingival<br>heights are equivalent to the<br>primary predicate devices<br>and the angulation is<br>equivalent to the reference<br>devices. Dynamic fatigue<br>testing of the subject devices<br>indicates performance<br>suitable for the intended use. |
| Material | Titanium alloy (Ti6Al4V-ELI) per ASTM F136 | Titanium alloy (Ti6Al4V-ELI) per ASTM F136 | Titanium alloy (Ti6Al4V-ELI) per ASTM<br>F136 | Identical<br>Subject and predicate devices<br>have the same material of<br>construction. |
JJGC Indústria e Comércio de Materiais Dentários S.A.
{6}------------------------------------------------
| | SUBJECT DEVICES | PRIMARY PREDICATE | REFERENCE PREDICATE | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| COMPARISON | Subject<br>Mini Abutments 60°<br>JJGC Indústria e Comércio de Materiais Dentários<br>S.A. | K190718<br>Neodent Implant System - Zygomatic Implants and<br>Abutments<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | K161598<br>NobelZygoma 0°<br>Nobel Biocare AB | EQUIVALENCE DISCUSSION |
| Single Use | Yes | Yes | Yes | Identical<br>Subject and predicate devices<br>are not reusable. |
| Sterilization Method | Ethylene Oxide to a SAL of 1x10-6 | Ethylene Oxide to a SAL of 1x10-6 | Gamma radiation | Identical<br>Subject and primary predicate<br>devices utilize the same<br>sterilization method and<br>minimum SAL. |
| Compatible devices | GM Line Impression copings, Provisional cylinders,<br>Protective cylinders and Copings already cleared<br>per market. | GM Line Impression copings, Provisional cylinders,<br>Protective cylinders and Copings already cleared per<br>market. | | Identical<br>Subject and primary predicate<br>devices are compatible with the<br>same devices. |
The subject devices have been assessed to deternine whether the previously conducted studies of the MR compatibility of the devices of the Neodent Implant System also apply to the subject devices. It has been determined that the subject devices do not result in new worst-asse constructs for the purpose of assessing MR compatibility. The subject devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per K182620.
{7}------------------------------------------------
## SUMMARY OF NON-CLINICAL PERFORMANCE DATA
Comparative dynamic fatigue test per ISO 14801 and FDA guidance was performed following the Zygomatic surgical protocol to determine the fatigue strength using the subject devices as compared to similar constructs of the reference devices.
Torsion testing was performed to evaluate the subject device's screws under static torsional loading. The results met the acceptance criteria.
Sterilization of the subject abutments via ethylene oxide gas was validated per ISO 11135. A minimum Sterility Assurance Level (SAL) of 1 x 10-6 has been validated.
Ethylene oxide residuals have been assessed per ISO 10993-7. Residuals are within accepted limits.
Biological Safety Assessment guided by ISO 10993-1 and FDA guidance. Reference to previous biocompatibility testing is supplied as follows:
- -Cytotoxicity testing was performed per ISO 10993-5.
- -Chemical characterization was performed per ISO 10993-18.
Product and package stability has been validated per ASTM F1980. The subject device's shelflife was determined to be 5 years.
The MR Conditional Labeling was leveraged from K182620.
### CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the subject devices are substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.