← Product Code [DQX](/productcode/DQX) · K203529

# VASSALLO GT Hybrid (K203529)

_Filmecc Co., Ltd. · DQX · May 4, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203529

## Device Facts

- **Applicant:** Filmecc Co., Ltd.
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** May 4, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.

## Device Story

VASSALLO GT Hybrid is a peripheral guide wire used to facilitate placement/exchange of diagnostic and therapeutic devices during intravascular procedures. Device consists of a hybrid Nitinol and stainless-steel core wire with a distal radiopaque coil assembly (Platinum-Nickel). Proximal section features PTFE coating; distal section features hydrophilic and polyurethane coatings to increase lubricity and reduce friction. Available in 0.014-inch outer diameter and 190cm/300cm lengths. Includes a torque device and optional detachable extension wire. Operated by clinicians under fluoroscopic guidance to navigate peripheral vasculature. Radiopaque coil allows visualization of distal tip position. Bench testing confirms performance, kink resistance, and torqueability comparable to predicate devices.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical in vitro bench testing and biocompatibility assessments. Bench tests included dimensional verification, visual inspection, simulated use, tensile strength, torque strength/ability, coating integrity/adhesion, catheter compatibility, corrosion resistance, kink resistance, tip flexibility, and radiopacity. Biocompatibility testing included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, PTT, thromboresistance, and complement activation.

## Technological Characteristics

Hybrid Nitinol and stainless-steel core wire; Platinum-Nickel distal coil; PTFE proximal coating; hydrophilic and polyurethane distal coating. 0.014-inch (0.36mm) outer diameter. Sterilized via Ethylene Oxide. Manual operation; no electronic components or software.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- ASAHI Peripheral Guide Wire ([K150445](/device/K150445.md))
- ASAHI Peripheral Guide Wire ([K163426](/device/K163426.md))

## Reference Devices

- MINAMO ([K190176](/device/K190176.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 4, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

Filmecc Co., Ltd. % Candace Cederman Principal Consultant CardioMed Device Consultants, LLC 1783 Forest Drive, Suite 254 Annapolis, Maryland 21401

Re: K203529

Trade/Device Name: VASSALLO GT Hybrid Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: April 2, 2021 Received: April 5, 2021

## Dear Candace Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203529

Device Name VASSALLO® GT Hybrid Guide Wire

Indications for Use (Describe)

Intended to facilitate the placement and exchange of diagnostic devices during intravascular procedures. This device is intended for peripheral vascular use only.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

(as required by 21 CFR 807.92)

Image /page/3/Picture/2 description: The image shows the logo for Filmecc Peripheral Guide Wires. The logo consists of a red abstract shape on the left, followed by the word "Filmecc" in black with a red square above the "i". Below the word "Filmecc" is the phrase "PERIPHERAL GUIDE WIRES" in smaller, black letters.

### PERIPHERAL GUIDE WIRE VASSALLO® GT Hvbrid

### 510(k) K203529

| Date Prepared:         | 2 April 2021                                                                                                                                                                                                               |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant:             | FILMECC CO., LTD.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya-shi, Aichi 489-0071<br>Japan<br>TEL : +81-52-768-1212, FAX : +81-52-768-1222                                                                                   |
| Contact:               | Takahiro Kuroiwa<br>Regulatory Affairs<br>FILMECC CO., LTD.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya-shi, Aichi 489-0071<br>Japan<br>TEL : +81-52-768-1212, FAX : +81-52-768-1222<br>e-mail: takahiro.kuroiwa@filmecc.com |
| Trade Name:            | VASSALLO® GT Hybrid                                                                                                                                                                                                        |
| Device Classification: | Class 2 per 21 CFR §870.1330                                                                                                                                                                                               |
| Classification Name:   | Catheter, Guide, Wire                                                                                                                                                                                                      |
| Product Code:          | DQX - Catheter Guide Wire                                                                                                                                                                                                  |
| Predicate Devices:     | ASAHI Peripheral Guide Wire, K150445 and<br>K163426                                                                                                                                                                        |
| Reference Devices:     | MINAMO, K190176                                                                                                                                                                                                            |

### INTENDED USE/INDICATIONS FOR USE:

#### VASSALLO® GT Hybrid

This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.

### DEVICE DESCRIPTION:

The VASSALLO GT Hybrid consists of a hybrid NiTi and stainless-steel core wire and a coil assembly on the distal end of the device. The coil assembly is soldered to the NiTi portion of the core. The coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy. Coatings are applied on the surface of the VASSALLO GT Hybrid to increase lubricity and reduce friction. The proximal portion is coated with PTFE. The distal section is

{4}------------------------------------------------

coated with a hydrophilic coating and polyurethane coating. The VASSALLO GT Hybrid has an outer diameter of 0.014 inches (0.36mm) and is available in 190cm and 300cm lengths.

About 2cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the guide wire with a length of less than 300 cm. A Torque device is included in the same package.

## COMPARISON WITH PREDICATE DEVICES:

Comparisons of the VASSALLO GT Hybrid and predicate / reference devices show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life and operating principle are similar to the currently marketed predicate and reference devices. The intended use of the subject device and its predicates are the same.

| Name of<br>Devices              | FILMECC Peripheral<br>Guidewire<br>VASSALLO GT®<br>Hybrid<br>Subject                                                                                                                                                 | ASAHI Peripheral Guide<br>Wires<br>ASAHI Gladius, ASAHI<br>Halberd, ASAHI Gaia PV<br>Predicate                                                                                                                       | MINAMO<br>Reference                                                                                                                                                                                                                                                                        |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                          | TBD                                                                                                                                                                                                                  | K150445 and K163426                                                                                                                                                                                                  | K190176                                                                                                                                                                                                                                                                                    |
| Intended Use<br>and Indications | This product is intended to<br>facilitate the placement and<br>exchange of diagnostic and<br>therapeutic devices during<br>intravascular procedures.<br>This device is intended for<br>peripheral vascular use only. | This product is intended to<br>facilitate the placement and<br>exchange of diagnostic and<br>therapeutic devices during<br>intravascular procedures.<br>This device is intended for<br>peripheral vascular use only. | PCI Guide Wires are intended<br>to facilitate the placement of<br>balloon dilatation catheters<br>during percutaneous coronary<br>intervention (PCI) and<br>percutaneous transluminal<br>angioplasty (PTA).<br>The PCI Guide Wires are not<br>intended for use in the<br>neurovasculature. |
| Nominal OD                      | 0.36mm<br>(0.014in)                                                                                                                                                                                                  | 0.36mm and 0.45mm<br>(0.014in and 0.018in)                                                                                                                                                                           | 0.36mm<br>(0.014in)                                                                                                                                                                                                                                                                        |
| Overall Length                  | 190cm<br>300cm                                                                                                                                                                                                       | 200cm to<br>300cm                                                                                                                                                                                                    | 190cm<br>300cm                                                                                                                                                                                                                                                                             |
| Outer Coil                      | Platinum-Nickel                                                                                                                                                                                                      | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                               | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                                                                                     |
| Tapered Core<br>Wire            | Hybrid<br>Nitinol and Stainless Steel                                                                                                                                                                                | Stainless Steel                                                                                                                                                                                                      | Hybrid<br>Nitinol and Stainless Steel                                                                                                                                                                                                                                                      |
| Inner Structure                 | Stainless Steel Coil                                                                                                                                                                                                 | Stainless Steel Coil                                                                                                                                                                                                 | Stainless Steel Coil                                                                                                                                                                                                                                                                       |
| Tip Shape                       | Straight (shapeable)                                                                                                                                                                                                 | Straight (shapeable)<br>Preshape                                                                                                                                                                                     | Straight (shapeable)<br>Preshape<br>J-tip                                                                                                                                                                                                                                                  |
| Coating                         | Hydrophilic<br>Hydrophobic                                                                                                                                                                                           | Hydrophilic<br>Hydrophobic                                                                                                                                                                                           | Hydrophilic<br>Hydrophobic                                                                                                                                                                                                                                                                 |
| Sterilization                   | Ethylene Oxide                                                                                                                                                                                                       | Ethylene Oxide                                                                                                                                                                                                       | Ethylene Oxide                                                                                                                                                                                                                                                                             |

## NON-CLINICAL TESTING/PERFORMANCE DATA:

Non-clinical laboratory testing was performed on the VASSALLO® GT Hybrid to determine substantial equivalence. The following testing/assessments were performed:

{5}------------------------------------------------

- Dimensional Verification ●
- . Visual Inspection
- Simulated Use ●
- Tensile Strength / Tip Pull ●
- Torque Strength
- Torqueability
- Coating Integrity
- Coating Adhesion ●
- Catheter Compatibility / . Lubricitv
- Corrosion Resistance
- Kink Resistance ●
- Tip Flexibility ●
- Radiopacity

The in vitro bench tests demonstrated that the VASSALLO GT Hybrid met all acceptance criteria and performed similarly to the predicate and reference devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate devices.

## BIOCOMPATIBILITY:

Testing was performed to assess biocompatibility of the VASSALLO GT Hybrid. The following tests were performed:

- Cytotoxicity ●
- Sensitization
- Intracutaneous Irritation ●
- Systemic Toxicity ●
- Pyrogen, Material Mediated
- Hemolysis ●
- Partial Thromboplastin Time
- In Vivo Thromboresistance ●
- SC5b-9 Complement ● Activation

The results from the testing performed showed the VASSALLO GT Hybrid to be biocompatible.

## CONCLUSION:

The VASSALLO GT Hybrid has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference devices. Performance data demonstrates that the device functions as intended. The conclusions drawn from the nonclinical and clinical tests demonstrate that the VASSALLO GT Hybrid is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

Therefore, the VASSALLO GT Hybrid is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K203529](https://fda-staging.innolitics.com/device/K203529)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
