K203487 · Accurate Medical Therapeutics, Ltd. · DQO · Dec 11, 2020 · Cardiovascular
Device Facts
Record ID
K203487
Device Name
Drakon and Sequre Microcatheters
Applicant
Accurate Medical Therapeutics, Ltd.
Product Code
DQO · Cardiovascular
Decision Date
Dec 11, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Drakon™ and Sequre® microcatheters are intended for the infusion of contrast media into all peripheral vessels. The Drakon™ and Sequre® microcatheters are also intended for drug infusion in intraarterial therapy and infusion of embolic materials. The Drakon™ and Sequre® microcatheters should not be used in cerebral vessels.
Device Story
Single-use, single-lumen microcatheters; used for infusion of contrast media, drugs, and embolic materials into peripheral vasculature. Device components include luer lock hub, strain relief, central shaft, and distal tip with radiopaque markers for fluoroscopic visualization. Sequre® model features side holes; Drakon™ model has no side holes. Inner lumen composed of PTFE for smooth passage of fluids/devices; distal shaft features hydrophilic coating for lubricity. Operated by clinicians in interventional procedures. Provides pathway for delivery of therapeutic/diagnostic agents; enables targeted peripheral vessel therapy. Benefits include precise delivery of embolic materials and contrast agents.
Clinical Evidence
Bench testing only. No clinical data provided. Bench tests included bead compatibility, vessel flow dynamics, tensile strength, burst pressure, torque strength, guidewire/catheter compatibility, bend radius, power injection, preconditioning, embolization coil compatibility, particulate matter evaluation, and trackability. All tests met predefined acceptance criteria.
Technological Characteristics
Single-lumen microcatheter; PTFE inner lumen; hydrophilic polymer coating on distal shaft. Radiopaque markers for fluoroscopic visualization. Diameters: 1.7 Fr. to 3.0 Fr.; lengths: 105 cm to 155 cm. Sterile (ISO 11135-1, SAL 10^-6). Mechanical device; no software or electronic components.
Indications for Use
Indicated for infusion of contrast media, drugs, and embolic materials into peripheral vessels. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
December 11, 2020
Accurate Medical Therapeutics Ltd Osnat Harbater R&D RA Manager 19 Eli Hurvitz Street Rehovot, 7608802 Israel
Re: K203487
Trade/Device Name: Drakon and Sequre Microcatheters Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DQO Dated: November 22, 2020 Received: November 27, 2020
Dear Osnat Harbater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203487
Device Name Drakon™ and Sequre® microcatheters
Indications for Use (Describe)
The Drakon™ and Sequre® microcatheters are intended for the infusion of contrast media into all peripheral vessels. The Drakon™ and Sequre® microcatheters are also intended for drug infusion in intraatterial therapy and infusion of embolic materials.
The Drakon™ and Sequre® microcatheters should not be used in cerebral vessels.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary Drakon™ and Sequre® microcatheters 510(k) Number K203487
Date Prepared: November 22nd, 2020
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
#### I. SUBMITTER
#### Company
Accurate Medical Therapeutics Ltd. 19 Eli Hurvitz St Rehovot, Israel 7608802 +972-54-3386871
### Contact Person
Osnat Harbater 19 Eli Hurvitz St Rehovot, Israel 7608802 +972-54-3386871 osnat(@accurmed.com
## II. DEVICE
Name of Device: Common or Usual Name: Classification Name: Regulation: Regulatory Product Code:
Drakon™ and Sequre® microcatheters Drakon™ and Sequre® microcatheters Catheter, Intravascular, Diagnostic 21 CFR 870.1200 Class: II DQO
## III. PREDICATE DEVICE
Accurate Medical Therapeutics Ltd. claims substantial equivalence to the Drakon™ and Sequre® microcatheters, cleared under K173430 (primary predicate) and K202797.
### IV. DEVICE DESCRIPTION
The Drakon™ and Sequre® microcatheters are single use microcatheters primarily comprised of a luer lock hub, a strain relief cover and tube, central shaft, and a distal tip with radiopaque markers for visualization. The two models differ only in the design of the distal tip. The Sequre®'s distal end has side holes and two radiopaque markers while the Drakon™'s distal end has no side holes and one radiopaque marker. These markers allow for the fluoroscopic visualization of the distal tip of the microcatheters.
{4}------------------------------------------------
The inner lumen is made of PTFE (polytetrafluoroethylene), which allows for the smooth passage of fluids, embolic agents and devices such as guide wires. The distal section of the shaft in both models is coated in a hydrophilic polymer laver. which ensures high lubricity when wet with saline or blood.
The Drakon™ and Sequre® microcatheters are sterile single lumen devices and are available in several different diameters (1.7 Fr., 1.9 Fr., 2.4 Fr, 2.7 Fr., 2.8Fr, and 3.0 Fr.) and lengths (105 cm, 130 cm, or 150 or 155 cm).
### V. INDICATIONS FOR USE
The Drakon™ and Sequre® microcatheters are intended for the infusion of contrast media into all peripheral vessels.
The Drakon™ and Sequre® microcatheters are also intended for drug infusion in intraarterial therapy and infusion of embolic materials.
The Drakon™ and Sequre® microcatheters should not be used in cerebral vessels.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device is substantially equivalent to the predicate device with respect to indications for use, principle of operation, fundamental design principles, performance, sterilization, and packaging. The primary reason for submitting this special 510(k) is the addition of the Sequre® 2.7 Fr. XSB.
The Sequre® 2.7 Fr. XSB model differs from the cleared Sequre® 2.7 Fr. primarily in its microsphere bead compatibility range and includes minor material changes.
These differences do not impact product performance or modify the intended use but are mainly designed to enable the Sequre® microcatheter family to be compatible with the same range of microsphere sizes that are compatible with the cleared Drakon™ microcatheters. The Microcatheter's instructions for use identically indicate the size requirements for devices compatible with the Sequre® 2.7 Fr. XSB. In summary, these differences do not raise any new issues of safety and effectiveness.
Although different materials were used in Sequre® 2.7 Fr. XSB, the materials are all either the same as in the cleared models or of the same type. Thus, these changes also do not raise new questions of safety and effectiveness in comparison to predicate device.
## VII. PERFORMANCE DATA
Risk assessment pursuant to ISO 14971 was performed to assess the impact of the changes. The following bench tests were performed to evaluate the design elements and performance characteristics of the modified Drakon™ and Sequre® microcatheters and to demonstrate substantial equivalence to the predicate device. The modified Drakon™ and Sequre® microcatheters met the predetermined acceptance criteria.
{5}------------------------------------------------
## Biocompatibility testing
Biocompatibility was leveraged from previous testing and demonstrated that the biocompatibility evaluation for the modified Drakon™ and Sequre® microcatheters is in compliance with the FDA Guidance Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1, "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process.
## Sterilization and Shelf Life Testing
The Drakon™ and Sequre® microcatheters are provided sterile in compliance with ISO 11135-1 for a SAL 10-6. Shelf life testing to support the labeled shelf life was performed.
## Bench Testing
The following Bench testing was completed successfully by meeting the predefined acceptance criteria:
- Bead Compatibility Bench Test ●
- Vessel Flow Dynamic Indication (Beads Reflux) Bench Test ●
- Tensile Bench Test ●
- Burst Pressure Bench Test
- Torque Strength Bench Test
- Guidewire & Guide Catheter Compatibility; Dimensional and Visual Inspection
- Bend Radius Bench Test ●
- . Power Injection Bench Test
- Preconditioning and Injected Substances Compatibility Bench Test ●
- . Torque Transmission Bench Test
- Embolization Coil Compatibility
- Acute Particulate Matter Evaluation
- Trackability Bench Test ●
## VII. CONCLUSIONS
The modified Drakon™ and Sequre® microcatheters are substantially equivalent in intended use and indications for use, principles of operation, fundamental design, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness. In conclusion, the modified Drakon™ and Sequre® are substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.