Classic is a device that consist in maxillary and mandibular devices connected with a personalized mechanism to reduce snoring and mild to moderate sleep apnea by holding the mandible forward during sleep.
Device Story
Classic is a mandibular advancement device (MAD) for treating snoring and mild-to-moderate obstructive sleep apnea in adults. The device consists of upper and lower dental splints connected by a screw-based adjustment mechanism. A physician or dentist customizes the device to hold the mandible in a forward position during sleep, increasing muscle tone at the airway level to improve airflow. The device allows for controlled mandibular advancement up to 10 mm, while permitting jaw opening, closing, and lateral movements. It is a prescription-use, non-sterile, removable intraoral appliance. The patient uses the device at home during sleep, adjusting the advancement via an included screwdriver as directed by their clinician. By maintaining airway patency, the device reduces apnea events and snoring, potentially improving patient sleep quality.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including flexural properties (ISO 178:2010), tensile properties (ASTM D638-02), and biocompatibility (ISO 10993-1, -5, -10).
Technological Characteristics
Materials: Polymer PET G, TPU, Polymethylmethacrylate (PMMA), stainless steel, and titanium alloy (ISO 5832-3, ISO 5832-1). Form factor: Two-splint intraoral appliance with screw-driven mandibular advancement mechanism. Max advancement: 10 mm. Non-sterile. No software or electronics.
Indications for Use
Indicated for adults (18+) with mild to moderate obstructive sleep apnea (OSA) and to alleviate or reduce snoring.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2021
Orthoapnea S.L. Jose Llecha Managing Director C / Flauta Magica 22 Malaga, Malaga 29006 SPAIN
Re: K203477
Trade/Device Name: Classic Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: November 15, 2020 Received: November 27, 2020
Dear Jose Llecha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
#### 510(k) Number (if known)
#### K203477
Device Name
Classic
Indications for Use (Describe)
Classic is a Mandibular Advancement Device (MAD) indicated for mild to moderate obstructive sleep apnea (OSA) and to alleviate or reduce snoring in adults.
| Type of Use (Select one or both, as applicable) | <table style="border:none"><tr><td><span style="border: 1px solid black; padding: 3px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><span style="border: 1px solid black; padding: 3px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="border: 1px solid black; padding: 3px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid black; padding: 3px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| <span style="border: 1px solid black; padding: 3px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid black; padding: 3px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
{3}------------------------------------------------
## ) ( OrthoApnea
Classic 510(k) Premarket Notification Submission
### SECTION 5 - 510(k) SUMMARY
#### K203477
| DATE OF SUBMISSION: | 2020-11-15 |
|---------------------|---------------------------------------------|
| SUBMITTER NAME: | Orthoapnea S.L |
| SUBMITTER ADDRESS: | Calle Flauta Mágica 22 29006 Málaga SPAIN |
| CONTACT: | Jose Repollés Llecha |
|------------|--------------------------|
| TELEPHONE: | +34 952 122 174 ext. 154 |
| e-mail: | jrepolles@orthoapnea.com |
DEVICE TRADE NAME: Classic
COMMON NAME: MANDIBULAR ADVANCEMENT DEVICE CLASSIFICATION NAME: INTRAORAL DEVICES FOR SNORING PRODUCT CODE: LRK REGULATION DESCRIPTION: Intraoral devices for snoring and intraoral
device for snoring and obstructive sleep apnea (21 CFR 872.5570)
PREDICATE DEVICE(S): K060388 (Airway Management Inc.) - Primary predicate K062951 (Airway Management Inc.) Reference device
#### A. Description of the Device
Classic device is an intraoral device specially indicated for the treatment of snoring and obstructive sleep apnea syndrome. Its design makes it simple to use and comfortable to wear. It is intended for single-patient use by adults who are 18 years or older.
Classic device consists of two splints united to each other by a screw that allows controlled advancement of the mandible by the increase of the muscular tone at the airway level. Consequently, the flow of air through the upper airway is increased, and reducing snoring and apnea events.
#### patient's
The maximum protrusion of Classic arch form is 10 mm, measuring from plate. This device needs an external controlling component such as screwdriver included in the kit. The device is according arches, maximizes tongue space and allow open, close and laterality movements during wear. These movements can be personalized by the doctor.
{4}------------------------------------------------
## ) ( OrthoApnea
#### Classic 510(k) Premarket Notification Submission
Each Classic device has the following components:
Image /page/4/Picture/3 description: The image shows a blue oral appliance with a metal adjustment mechanism. The appliance consists of an upper tray (1) and a lower tray (2), both made of a transparent blue material. A metal adjustment mechanism (3, 4, 5) connects the two trays, allowing for controlled movement of the lower jaw. The appliance is designed to treat sleep apnea and snoring by advancing the lower jaw and opening the airway.
| Component # | Component Name |
|-------------|------------------|
| 1 | Upper splint |
| 2 | Lower splint |
| 3 | Activation key |
| 4 | Guide (Bar) |
| 5 | OrthoApnea screw |
#### B. Intended Use
Classic is a device that consist in maxillary and mandibular devices connected with a personalized mechanism to reduce snoring and mild to moderate sleep apnea by holding the mandible forward during sleep.
#### C. Summary of Testing and Comparison to the Predicate Device
The devices are designed and manufactured in accordance with the following standards:
- ISO 20795-2 Second edition 2013-03-01 Dentistry Base polymers Part 2: Orthodontic base polymers
- ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- -ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
- -ISO 10993-10:2009 Biological evaluation of medical devices – Part 10: Test for irritation and delayed-type hypersensitivity
- -ISO 5832-3 2016 Implants for surgery - Metallic materials -- Part 3: Wrought titanium 6-aluminium 4-vanadium alloy
- -ISO 5832-1 2016 Implants for surgery - Metallic materials - Part 1: Wrought stainless steel
{5}------------------------------------------------
## ) | ( OrthoApnea
### Classic
510(k) Premarket Notification Submission
And according the following guidance:
- -Intraoral Devices for Snoring and/or Obstructive Sleep Apnea-Guidance for Industry and FDA 2002
#### Comparison of Technological Characteristics
#### Table 1: Device Comparison
| Characteristic | Subject Device | Primary predicate | Reference predicate | SE / Comments |
|-----------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------|
| Model Name | Classic | TAP II | TAP III | - |
| 510K | K203477 | K060388 | K062951 | |
| Intended use | To reduce or alleviate nighttime snoring and mild to moderate obstructive sleep apnea (OSA). | The TAP II is intended to reduce or alleviate night time snoring and mild to moderate obstructive sleep apnea, OSA. | The TAP III is intended to reduce or alleviate nighttime snoring and mild to moderate obstructive sleep apnea, OSA. | Yes |
| Material | Polymer PET G and TPU<br>Polymethylmethacrylate<br>Stainless steel and Titanium alloy | Polycarbonate<br>Stainless steel | Polycarbonate<br>Stainless steel | Yes |
| Fixed /removable | Removable | Removable | Removable | Yes |
| Sterile | Non- sterile | Non-sterile | Non-sterile | Yes |
| Mandibular adjustment | Customized by physician or dentist. | Customized by physician or dentist. | Customized by physician or dentist. | Yes |
| Mandibular adjustment range | Up to 10.0 mm | Up to 7.0 mm | Up to 7.0 mm | Yes |
Classic device is substantially equivalent intended use as the identified predicates. Classic device is similar in fundamental scientific technology to the predicate devices in that they all have been designed, manufactured and tested in compliance with FDA'S Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea.
Classic device is substantially equivalent in materials, indications and intended use, packaging, labeling and performance to the predicate devices currently marketed in the U.S.
The only differences the subject device and the predicate are slight differences in adjustment and mandibular advancement range.
{6}------------------------------------------------
# ) | ( OrthoApnea
#### Classic
510(k) Premarket Notification Submission
#### Non-Clinical performance tests
The proposed devices have been subject to bench testing to determine fulfillment of design and performance requirements. Bench testing followed the recommendations provided in FDA Guidance Document - Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea.
Performance testing on the Classic device included:
- . Determination of flexural properties (ISO 178:2010)
- . Determinations of the Tensile properties (ASTM D638-02)
- Biocompatibility evaluation in accordance (ISO 10993-1:2009) .
but as we use widely used standard components in the market there are considered not necessary to perform additional tests.
Test data from the device was leveraged for the previous performance tests based on the use of identical device and packaging materials as well as identical manufacturing, packaging and cleaning processes.
#### D. Clinical Testing
No clinical testing was performed. Non-clinical testing was used to support the determination of substantial equivalence.
#### E. Conclusion of Substantial Equivalence
Based on the similarities observed and results of non-clinical testing performed, we conclude that the proposed devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.