The RibFix Advantage System is indicated for the fixation, stabilization, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Device Story
The RibFix Advantage System is a thoracic fixation device consisting of titanium bridges with threaded locking posts, locking caps, and optional washers. It is used for the thoracoscopic fixation and stabilization of rib fractures and osteotomies. The device is implanted by a surgeon; the bridge is placed on the underside of the rib, with locking posts extending through pre-drilled holes to be secured by locking caps on the anterior side. The system provides mechanical stabilization to facilitate bone fusion. It is supplied non-sterile and requires steam sterilization by the user. The device is intended to benefit patients by providing rigid fixation for rib fractures, potentially improving healing outcomes.
Clinical Evidence
No clinical data was required or provided for this submission. Substantial equivalence was established through non-clinical performance testing, including MR safety evaluations (displacement force, torque, RF heating, and image artifacts) conducted in accordance with ASTM F2052-15, ASTM F2213-17, ASTM F2182-11a, and ASTM F2119-07.
Technological Characteristics
Implants manufactured from commercially pure titanium (ASTM F67) and titanium alloys (Ti-6Al-7Nb per ASTM F1295; Ti-6Al-4V per ASTM F136/F1472). System includes bridges with threaded locking posts, locking caps, and washers. Mechanical fixation principle. Provided non-sterile; steam sterilization required. No software or electronic components.
Indications for Use
Indicated for fixation, stabilization, and fusion of rib fractures and osteotomies in patients with normal or osteoporotic bone. Contraindicated in patients with active/latent infection, sepsis, device material sensitivity, or those unable to follow postoperative instructions. Not for use on clavicle, spine, or highly comminuted fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Biomet Microfixation RibFix Blu Thoracic Fixation System (K162974)
Biomet Microfixation SternaLock 360 Sternal Closure Fixation System (K163007)
Submission Summary (Full Text)
{0}------------------------------------------------
January 28, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biomet Micofixation Lauren Jasper Regulatory Affairs Manager 1520 Tradeport Drive Jacksonville, Florida 32218
Re: K203474
Trade/Device Name: Biomet Microfixation RibFix Advantage System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC. HTN Dated: November 24, 2020 Received: November 25, 2020
Dear Lauren Jasper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203474
Device Name Biomet MicrofixationR ibFix Advantage System
Indications for Use (Describe)
TheR ibFix Advantage System is indicated for the fixation, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for BIOMET Microfixation. The word "BIOMET" is in large, bold, dark gray letters on the top line. Below that, in smaller, lighter gray letters, is the word "MICROFIXATION". A thin, light green line is below the word "MICROFIXATION". The words "One Surgeon. One Patient" are below the green line in a smaller font.
## 510(k) Summary
Prepared January 28, 2021
| Submitter: | Biomet Microfixation<br>1520 Tradeport Drive<br>Jacksonville, FL 32218 |
|------------|----------------------------------------------------------------------------------------------------------|
| Contact: | Lauren Jasper, Regulatory Affairs Manager<br>lauren.jasper@zimmerbiomet.com<br>Telephone: (904) 741-9259 |
Device Name: Biomet Microfixation RibFix Advantage System
| | Device Classification: |
|--|------------------------|
|--|------------------------|
| Product Code | Classification Name | Device<br>Classification | Regulation<br>Number | Regulation Description |
|---------------|-----------------------------|--------------------------|----------------------|------------------------------------------------------------------------------------|
| HRS (Primary) | Plate,<br>Fixation,<br>Bone | 2 | 888.3030 | Single/multiple component<br>metallic bone fixation<br>appliances and accessories |
| HWC | Screw,<br>Fixation,<br>Bone | 2 | 888.3040 | Smooth or threaded metallic<br>bone fixation fastener |
| HTN | Washer, Bolt<br>Nut | 2 | 888.3030 | Single/multiple component<br>metallic bone fixation<br>appliances and accessories. |
Indications for Use: The RibFix Advantage System is indicated for the fixation, stabilization, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Contraindications: The RibFix Advantage system is contraindicated for:
- Use in patients with latent or active infection, sepsis, and/or device material sensitivity. 1.
- 2. Use in patients who are unwilling or incapable of following postoperative care instructions.
- 3. This device is not intended for locking post attachment or fixation to the clavicle or spine.
- 4. This device is not intended for highly comminuted fractures.
Device Description: The RibFix Advantage System consists of bridges (with locking posts) and locking caps for the thoracoscopic fixation and stabilization of ribs. These implants are manufactured from commercially pure titanium (per ASTM F67) and titanium alloys (Ti-6A1-7Nb
{4}------------------------------------------------
per ASTM F1295 and Ti-6Al-4V per ASTM F136 or ASTM F1472). The system is comprised of the following implants:
- . Titanium alloy RibFix Advantage Bridge with threaded locking posts
- . Commercially pure titanium locking caps
- . Commercially pure titanium washers (optional)
When fully assembled, the bridge is placed on the underside of the rib, the threaded locking posts extends through pre-drilled holes in the rib, and the locking caps are fixed to the locking post on the anterior side of the rib. Washers may be placed on the posts prior to assembling with locking caps. The combined threaded locking post/locking cap provides for fixation of the bridge. The devices are sold non-sterile and intended to be sterilized by the user prior to implantation. The system also includes reusable instrumentation to aid in implantation of the devices.
The design, materials, sterilization method, and indications for use are identical between the subject device and the predicate device.
## Substantial Equivalence:
Primary Predicate Device: K183317, AdvantageRib System
The similarities of the subject devices to the predicate devices are as follows:
- . The indications for use are identical to that of the predicate device.
- The design of the subject devices is equivalent to the predicate devices. ●
- The sterilization method (steam sterilization) of the subject devices is identical to the predicate devices.
- The materials of the subject devices are identical to the predicate devices. ●
The differences of the subject devices as compared to the predicate devices are as follows:
- . Update package insert to include MR conditional parameters and symbol
- Update outer package label to include MR conditional symbol
- Update cleaning instructions for the processing of reusable instruments and . new/uncompromised implants.
- . Add additional sterilization parameters applicable for other geographic locations.
- Revise the contraindications statement.
# Additional Reference Devices:
K162974. Biomet Microfixation RibFix Blu Thoracic Fixation System K163007, Biomet Microfixation SternaLock 360 Sternal Closure Fixation System
The similarities of the subject devices to these predicate devices are as follows:
- . The cleaning and sterilization instructions provided in the package insert are identical.
{5}------------------------------------------------
Non-Clinical Performance Data: The following nonclinical tests were submitted and relied on in this premarket notification submission for a determination of substantial equivalence. Testing identified in this summary has all passed acceptance criteria established by the predicate device where applicable.
MR Safety Evaluation following standards listed below:
- ASTM F2052-15, "Standard Test Method for Measurement of Magnetically Induced ● Displacement Force on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2213-17, "Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2182-11a, "Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging"
- ASTM F2119-07 (Reapproved 2013), "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
Clinical Performance Data: Clinical testing was not necessary for the determination of substantial equivalence.
Sterilization Information: There have been no changes to the sterilization method. The implants and instruments are provided non-sterile to be sterilized by steam at the end user facility.
Conclusion: The proposed device has identical indications for use as the predicate devices. The submission demonstrates that (1) any differences in technological characteristics of the predicates do not raise any new questions of safety and efficacy and (2) the proposed device is at least as safe and effective as the predicates. It is concluded that the information included in this summary supports substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.