K203415 · Yanbian Pacific Textile Co., Ltd. · FYC · May 20, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203415
Device Name
Surgical Isolation Gown
Applicant
Yanbian Pacific Textile Co., Ltd.
Product Code
FYC · General, Plastic Surgery
Decision Date
May 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Surgical Isolation Gown is intended to protect patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. In addition, the Surgical Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The Surgical Isolation Gown is a single use, disposable medical device provided non-sterile.
Device Story
Surgical Isolation Gown; single-use, disposable, non-sterile protective apparel. Constructed from blue polypropylene SMS non-woven material. Provides AAMI Level 3 barrier protection against microorganisms, body fluids, and particulates. Used in healthcare settings by personnel to protect themselves and patients; not for operating room use. Available in seven sizes (S-4XL). Device functions as a physical barrier; no electronic or software components.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against AAMI PB70:2012 (Level 3), ASTM D3776 (mass per unit area), ASTM D5034 (breaking strength), ASTM D5587 (tearing strength), 16 CFR 1610 (flammability), AATCC 42/127 (water resistance), and ISO 9073-10 (linting). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Polypropylene SMS non-woven fabric; blue color. AAMI Level 3 barrier protection. Dimensions range from 105cmx125cm to 135cmx155cm. Single-use, non-sterile. Tested per ASTM D3776, D5034, D5587; AATCC 42, 127; ISO 9073-10.
Indications for Use
Indicated for healthcare patients and personnel to protect against transfer of microorganisms, body fluids, and particulate material. Not for use in operating rooms. Single-use, non-sterile, disposable.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Yanbian Pacific Textile Co., LTD % Rafi Wong Manager Pacific Fortune Management Inc. 2350 Mission College BLVD, STE 475 Santa Clara, California 95054
Re: K203415
Trade/Device Name: Surgical Isolation Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC, Dated: April 13, 2021 Received: April 19, 2021
Dear Rafi Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Ryan Ortega Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K203415
Device Name
Surgical Isolation Gown
Indications for Use (Describe)
The Surgical Isolation Gown is intended to protect patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. In addition, the Surgical Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSU AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The Surgical Isolation Gown is a single use, disposable medical device provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| ☐ Research Use (Part 21 CFR 601 Subpart D) |
|---------------------------------------------------|
| ☑ Source Tissue Controlled (21 CFR 601 Subpart C) |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 4. Indications for Use Statement
The Surgical Isolation Gown is intended to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. In addition, the Surgical Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The Surgical Isolation Gown is a single use, disposable medical device provided non-sterile.
The Surgical Isolation Gown is considered under the product code, FYC, under the classification of a surgical isolation gown per the new Guidance for Industry and FDA Staff - Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings issued by the FDA on December 9, 2015 because this isolation gown has a moderate barrier protection.
The Surgical Isolation Gowns are non-sterile and provided in ONE product model in seven sizes to meet the needs of healthcare patients and health care personnel, Refer to Table 1 below. The medical device will be available in one model in S(160), M(165), L(170), XL(175), XXL(180), 3XL(185), 4XL(190) sizes.
The proposed Surgical Isolation Gown is constructed of Polypropylene SMS non-woven material with the color in blue. The Surgical Isolation Gowns consist of a one critical zone throughout the entire gown including seams but excluding cuffs, hems, and bindings. The product has been tested for barrier performance per ANSI/AAMI PB70:2012. Testing was performed according to the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and ANSI/AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities. All results of testing met AATCC-42/AATCC-127, and meets AAMI PB70:2012 Level 3 requirements.
| Product<br>Name | Model<br>Number | Sterility | Color | Dimensions | Size |
|-------------------------------|-----------------|-----------------|-------|-------------|----------|
| Surgical<br>Isolation<br>Gown | | Non-<br>sterile | Blue | 105cm×125cm | S(160) |
| | | | | 110cm×130cm | M(165) |
| | 12265 | | | 115cm×135cm | L(170) |
| | | | | 120cm×140cm | XL(175) |
| | | | | 125cm×145cm | XXL(180) |
| | | | | 130cm×150cm | 3XL(185) |
| | | | | 135cm×155cm | 4XL(190) |
Table 1 Specifications and dimensions of proposed device
{4}------------------------------------------------
## 510K SUMMARY
# K203415
### Date of Summary Prepared:
May 20, 2021
#### Submitter Information 1.
| Submitter Contact: | YANBIAN PACIFIC TEXTILE CO., LTD<br>Longjing industry concentration area, Longjing city<br>Jilin, China, 133400 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Submitter Contact Person: | |
| Name: | Juan Zhuang |
| Phone Number: | (+86) 138-6397-7758 |
| Email: | zhuangjuan777@126.com |
| Designated Submission Correspondent: | |
| Name: | Rafi Wong |
| Phone Number: | +1 (408) 646-6537 |
| Email: | rafi.wong@pfmfinance.com |
#### 2. Device Name: Surgical Isolation Gown
#### 3. Regulatory Information
| Classification Name: | Surgical Isolation Gown |
|----------------------|-------------------------|
| Common Name: | Surgical Apparel |
| Classification: | Class II |
| Product Code: | FYC |
| Regulation Number: | 21 CFR 878.4040 |
#### 4. Predicate Device
| 510K Number: | K190306 - AMD Medicom Inc. |
|---------------|----------------------------------|
| Device Name: | AMD Ritmed AssureWear™ VersaGown |
| Cleared date: | January 17, 2020 |
#### 5. Intended Use/Indications for Use
The Surgical Isolation Gown is intended to protect patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. In addition, the Surgical Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). The Surgical Isolation Gown is a single use, disposable medical device provided non-sterile.
{5}------------------------------------------------
#### 6. Device Description
The proposed Surgical Isolation Gown is constructed of Polypropylene SMS non-woven material with the color in blue. The melt-blown nonwovens are made of polypropylene. The Surgical isolation gown is different in size. The size of surgical isolation gown is divided into seven groups: S(160), M(165), L(170), XL(175), XXL(180), 3XL(185), 4XL(190).
The Surgical Isolation Gown is a single use, disposable medical device provided non-sterile.
The device description of the Surgical Isolation Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and ANSI/AAMI PB70:2012 Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities.
#### 7. Summary of Comparison and Technological Characteristics
| Items | Proposed Device<br>K203415 | Predicate Device<br>K190360 | Result | | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------------|------|
| Product<br>Common Name | Surgical Isolation Gown | AMD Ritmed AssureWear™<br>VersaGown | Difference | | |
| Manufacturer | YANBIAN PACIFIC TEXTILE CO.,LTD | AMD Medicom Inc. | - | | |
| 510K Number | K203415 | K190306 | - | | |
| Product Code | FYC | FYC | Same | | |
| Classification | Class II (21 CFR 878.4040) | Class II (21 CFR 878.4040) | Same | | |
| Intended<br>Use/Indication<br>for Use | The Surgical Isolation Gown is intended to<br>protect patients and health care personnel<br>from the transfer of microorganisms, body<br>fluids and particulate material. Not intended<br>for use in the operating room. In addition,<br>the Surgical Isolation Gown meets the<br>requirements of an AAMI Level 3 barrier<br>protection for an isolation gown per<br>ANSI/AAMI PB70:2012 Liquid Barrier<br>Performance and Classification of Protective<br>Apparel Drapes Intended for Use in Health<br>Care Facilities (ANSI/AAMI PB70). The<br>Surgical Isolation Gown is a single use,<br>disposable medical device provided non-<br>sterile. | AMD Ritmed AssureWear™ VersaGown is<br>intended to be worn by healthcare personnel<br>to protect health care patients and health<br>care personnel from the transfer of<br>microorganisms, body fluids and particulate<br>material. AssureWear™ VersaGown meets the<br>requirements of an AAMI Level 3 barrier<br>protection for an isolation gown per<br>ANSI/AAMI PB70:2012 Liquid Barrier<br>Performance Classification of Protective<br>Apparel Drapes Intended for Use in Health<br>Care Facilities (ANSI/AAMI PB70). AMD<br>Ritmed AssureWear™ VersaGown is a single<br>use, non-sterile disposable medical device<br>and not intended for use in operating rooms. | Similar | | |
| Materials | SMS PP + PE non-woven fabric material | PP SMS non-woven + PE | Same | | |
| Color | Blue | Blue | Same | | |
| Design | Tape Neck Closure Belt Tie<br>Elastic Cuffs | Thumb loop Elastic cuffs Extended cuff<br>(Thumb loop) Flexneck™ Tie (neck) Straight<br>sleeve Inclined sleeve Blue belt tie<br>Reinforced seams | Similar for all<br>designs<br>except for<br>the Flexneck<br>patented<br>design | | |
| OTC Use | Yes | Yes | Same | | |
| Single Use | Yes | Yes | Same | | |
| Sterile | No | No | Same | | |
| Level of barrier<br>protection<br>AAMI PB70 | Level 3 | Level 3 | Same | | |
| Physical and mechanical performance | | | | | |
| Basic weight<br>ASTM D3776 | 60.7 g/m2 (1.79 oz/yd2) Seam between<br>areas C&D: <0.1g | 39.97 ± 1.61 g/m2<br>(1.17 oz/yd2 ± 0.05) | Similar | | |
| Flammability<br>16 CFR Part<br>1610.7 | Class 1 | Class 1 | Same | | |
| Breaking<br>strength (MD)<br>ASTM D5034 | Mean: 175.5 N | 18.17 ± 0.31 lbf | Similar | | |
| Breaking<br>strength (CD)<br>ASTM D5034 | Mean: 118.0 N | 11.78 ± 0.33 lbf | Similar | | |
| Tearing<br>strength (MD)<br>ASTM D5733-<br>1999 | Mean: 63.5 N | 11.01 ± 0.64 lbf (ASTM D5587) | Similar | | |
| Tearing<br>strength (CD)<br>ASTM D5733-<br>1999 | Mean: 34.5 N | 5.30 ± 0.35 lbf (ASTM D5587) | Similar | | |
| Linting (ISO<br>9073-10) | Particulate size range(μm): 3 to 25<br>A: Face: Measured value Coefficient of<br>lingting log10 Min:2.2, Max:2.8,<br>Mean: 2.5; | Particulate size range(μm): 1 to 25<br>Outside: Total linting >0.3:2.07;>0.5:1.97<br>Index for Particulate Matter (IPM):1.50 | Similar | | |
| | B: Face: Measured value Coefficient of<br>lingting log10 Min:2.5, Max:2.9,<br>Mean:2.74 | Inside: Total<br>linting >0.3:2.16;>0.5:2.00<br>Index for Particulate Matter (IPM):1.35 | | | |
| | Biocompatibility | | | | |
| | Cytotoxicity<br>ISO 10993-5:<br>2009 | Non-Cytotoxic | | Non-Cytotoxic | Same |
| Irritation ISO<br>10993-<br>10: 2010 | Non-Sensitizing | Non-Sensitizing | Same | | |
| Sensitization<br>ISO 10993-<br>10: 2010 | Non-Irritating | Non-Irritating | Same | | |
## Table I - General Comparison
{6}------------------------------------------------
{7}------------------------------------------------
*The difference in the materials and colors does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials and color additives.
#### 8. Summary of Non-clinical Tests Performed on the Proposed Device
The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes on August, 1993 STANDARDS:
AAMI PB70:2012, Liquid Barrier Performance Classification of Protective Apparel and 0 Drapes Intended for Use in Health Care Facilities.
o ASTM D3776/D3776M-09 (2013), Test Methods for Mass Per Unit Area (Weight) of Woven Fabric.
ASTM D5034-09(2013), Standard Test Method for Breaking Strength and Elongation of ● Textile Fabrics (Grab Test).
ASTM D5587-2015, Standard Test Method for Tearing Strength of Fabrics by Trapezoid ● Procedure.
- o (CPSC), Part 1610, Standard For The Flammability Of Clothing Textiles.
- AATCC 42-2017, Water Resistance: Impact Penetration Test.
- AATCC 127-2018, Water Resistance: Hydrostatic Pressure Test. ●
- o ISO 10993- 1: 2009/(R)2013, Biological Evaluation of Medical Devices- Part 1: Evaluation and testing within a risk management process.
o ISO 10993-5: 2009 Biological Evaluation Of Medical Devices – Part 5: Tests For In Vitro Cytotoxicity.
ISO 10993-10: 2010 Biological Evaluation Of Medical Devices – Part 10: Tests For o Irritation And Skin Sensitization.
o ISO 9073-10:2004, Textiles-Test methods for nonwovens-Part 10: Lint and other particles generation in the dry state.
{8}------------------------------------------------
#### Clinical Test 9.
There is no clinical study included in this submission.
#### 10. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the proposed device Surgical Isolation Gown is as safe, as effective, and performs as well as or better than the predicate device, the AMD Ritmed AssureWear™ VersaGown (K190306) manufactured by AMD Medicom Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.