Marrow-Pack (white)- 11gauge/10cm or 13gauge/10cm, Marrow-Tray (red)- 11gauge/10cm or 13gauge/10cm, Marrow-Set (violet)- 11gauge/10cm or 13gauge/10cm, Marrow-kit (grey)- 11gauge/10cm or 13gauge/10cm, Marrow-Stem (green)- 11gauge/10cm or 13gauge/10cm
Marrow-Pack (white)- 11gauge/10cm or 13gauge/10cm, Marrow-Tray (red)- 11gauge/10cm or 13gauge/10cm, Marrow-Set (violet)- 11gauge/10cm or 13gauge/10cm, Marrow-kit (grey)- 11gauge/10cm or 13gauge/10cm, Marrow-Stem (green)- 11gauge/10cm or 13gauge/10cm
Applicant
Biopsybell S.R.L.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jan 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The BONE MARROW MSC ASPIRATION KIT is intended for use for aspiration / explant of bone marrow through a piston syringe.
Device Story
Single-use disposable bone marrow aspiration needle; includes cannula, stylet, threaded insert, and rotating spacer. Operation: manual insertion into bone; stylet removal; syringe connection; manual rotation of spacer retracts cannula to collect samples from multiple heights/sites via single puncture. Used in clinical settings by healthcare professionals. Benefits: enables multi-site marrow collection through one puncture, reducing patient trauma.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included biocompatibility (cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity), washing process, shelf-life, and mechanical performance (needle retraction, perforation capacity, Luer-lock integrity, sample collection quantity).
Technological Characteristics
Cannula: AISI 304 stainless steel. Syringe: Polycarbonate/ABS/Silicone. Single-use, sterile (Ethylene Oxide). Dimensions: 10 cm length, 11 or 13 gauge. Mechanical operation via threaded insert and rotating spacer for cannula retraction. Biocompatibility compliant with ISO 10993-1.
Indications for Use
Indicated for aspiration or explant of bone marrow in patients requiring bone marrow collection.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Marrow Cellution Bone Marrow Aspiration Needle (K150563)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
January 11, 2021
Biopsybell s.r.l. % Maurizio Pantaleoni Senior Consultant Maytal Doo Kneza Milosa, 79 Belgrade, Serbia 11000 Serbia
Re: K203397
Trade/Device Name: BONE MARROW MSC ASPIRATION KIT Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: November 6, 2020 Received: November 18, 2020
Dear Maurizio Pantaleoni:
We have reviewed your Section 5100k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203397
Device Name BONE MARROW MSC ASPIRATION KIT
Indications for Use (Describe)
The BONE MARROW MSC ASPIRATION KIT is intended for use for aspiration / explant of bone marrow through a piston syringe.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# BIOPSYBELL SRL.
510(K) PREMARKET NOTIFICATION - K203397
- 1. General Information
Submitter :
Biopsybell srl is located at: Via A. Manuzio n.24 41037, Mirandola (MO) ITALY
## Establishment Registration Number: 9617616
| Contact: | Maurizio Pantaleoni<br>Maytal Doo<br>Kneza Miloza 79<br>11000 Beograd (Serbia)<br>Tel. +381 11 3340700<br>Mob. +39 3484435155<br>Email: regulatory@maytal.eu |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Carlo Ricca Prandi Bellini<br>Biopsybell srl<br>Via A. Manuzio n.24,<br>41037, Mirandola (MO), Italy<br>Telephone: +39 0535 27850<br>e-mail: qara@biopsybell.it |
| Summary Preparation Date: | January 06th, 2021 |
#### 2. Name & Classification
| Device Name: | BONE MARROW MSC ASPIRATION KIT |
|--------------------|--------------------------------------------|
| Common Name: | Bone Marrow Biopsy Needle |
| Regulation Name: | Gastroenterology-urology biopsy instrument |
| Regulation Number: | 876.1075 |
| Product Code: | KNW |
| CLASS: | II |
#### 3. Predicate Devices
The BONE MARROW MSC ASPIRATION KIT is substantially equivalent to the following devices:
| Applicant | Device name | 510(k) Number |
|------------|------------------------------------------------------------------------------------------------|---------------|
| RANFAC | Marrow Cellution Bone Marrow<br>Aspiration Needle | K150563 |
| BIOPSYBELL | Manual-bone-marrow-biopsy-needles, Semi-<br>automatic-biopsy-needles, Automatic-biopsy-needles | K130616 |
{4}------------------------------------------------
# BIOPSYBELL SRL.
510(K) PREMARKET NOTIFICATION - K203397
#### 4. Indications for Use
The BONE MARROW MSC ASPIRATION KIT is intended for use for aspiration / explant of bone marrow through a piston syringe.
#### 5. Device Description
The BONE MARROW MSC ASPIRATION KIT is a single use disposable needle intended for the aspiration / explant of bone marrow.
The structure of BONE MARROW MSC ASPIRATION KIT device includes a cannula with a stylet inside, with a threaded insert and a rotating spacer that allow the precise retraction of the cannula to a determined length.
Once the cannula has been inserted and positioned, the stylet can be removed and a syringe is connected to execute the aspiration.
By manually rotating the rotating spacer, the cannula retracts, raising inside the bone marrow, in order to be able to perform aspiration / explant from fresh sites at different heights.
| | Subject device | Predicate device<br>(K150563) | Reference device<br>(K130616) | Reference device<br>(K130616) |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------------|
| | | | Orion Explant | OSTEOBELL T |
| Device | BONE MARROW MSC<br>ASPIRATION KIT | Marrow Cellution<br>Bone Marrow<br>Aspiration Needle<br>(K150563) | ORION EXPLANT –<br>Needle for bone-<br>marrow explant<br>(K130616) | OSTEOBELL T - Needle<br>for bone-marrow biopsy<br>(K130616) |
| 510(K) number | - | | | |
| Applicant | BIOSPYBELL S.R.L. | RANFAC | BIOPSYBELL S.R.L. | BIOPSYBELL S.R.L. |
| Classification | | | | |
| Reg. Number | 876.1075 | 876.1075 | 876.1075 | 876.1075 |
| Product Code | KNW | KNW | KNW | KNW |
| Regulatory Class | II | II | II | II |
| Intended use | | | | |
| Intended use | The BONE MARROW<br>MSC ASPIRATION KIT is<br>intended for use for<br>aspiration / explant of<br>bone marrow through a<br>piston syringe. | The Marrow Cellution<br>Bone Marrow<br>Aspiration Needle is<br>intended for use for<br>aspiration<br>of bone marrow or<br>autologous blood<br>using a standard<br>piston syringe. | needle for bone-marrow<br>explant | needle for bone-marrow<br>biopsy |
#### 6. Comparison with the predicate devices
{5}------------------------------------------------
#### BIOPSYBELL SRL. 510(K) PREMARKET NOTIFICATION - K203397
| | Subject device | Predicate device<br>(K150563) | Reference device<br>(K130616)<br>Orion Explant | Reference device<br>(K130616)<br>OSTEOBELL T |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of Action / Mode of Action | The mechanism of action is bone marrow aspiration / explant.<br>The needle is placed manually, then with a single puncture, is possible to collect bone marrow samples at different heights through the rotation of the rotating spacer. | The mechanism of action is bone marrow aspiration / explant.<br>The needle is placed manually, then with a single puncture, is possible to collect bone marrow samples at different heights through the rotation of the ring nut. | The mechanism of action is bone marrow aspiration / explant.<br>The needle is placed manually, then with a single puncture, is possible to collect bone marrow samples at different heights through the rotation of the ring nut. | The mechanism of action is bone marrow biopsy.<br>The needle is placed manually, then with a single puncture, is possible to collect bone marrow samples. |
| Design Features | | | | |
| Single Use Device | YES | YES | YES | YES |
| Dimensions of cannula (length / gauge) | 10 cm / 11 gauge<br>10 cm / 13 gauge | 9 cm / 11 gauge<br>11,4 cm / 11 gauge | 11-13-14-15 - gauge<br><br>Every gauge diameter is available in the following dimensions: 7, 9, 11, 13 cm | 7-8-9-11-13 gauge<br><br>Every gauge diameter is available in the following dimensions: 10, 15 cm |
| Materials | | | | |
| Cannula | AISI 304 stainless steel | Stainless steel | AISI 304 stainless steel | AISI 304 stainless steel |
| Syringe | Polycarbonate / ABS /<br>Silicone | N/A | N/A | N/A |
| Biocompatibility | | | | |
| Standard | Compliant to ISO 10993-1: | Compliant to ISO 10993-1: | Cytotoxicity<br>Intracutaneous<br>reactivity | Cytotoxicity<br>Intracutaneous<br>reactivity |
| Sterilization | | | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
#### 7. Performance Data
A program of design verification and validation testing was performed that includes the following:
- . Biocompatibility; the following biocompatibility tests have been performed:
- o Cytotoxicity
- o Sensitization
- Irritation or intracutaneous reactivity o
- o Material mediated pyrogenicity
- o Acute systemic toxicity
- . Verification of Washing Process
- . Packaging shelf life accelerated aging tests
- Performance/Functionality/Safety; the following tests have been performed:
- o Needle retraction efficacy
- Excursion of the Needle threaded insert O
- O Maximum useful length of the needle cannula
- o Needle aspiration surface
- o Needle handle Grip during bone insertion procedures
- Needle perforation capacity in the bone o
- o Adequacy of the Needle Luer-Lock Connection
{6}------------------------------------------------
- o Quantity of bone marrow sample collected by the needle
- o Integrity of the aspiration syringe
- EO Sterilization Validation
Results of the evaluations demonstrate that the subject device met the safety and performance requirements as per its indication for use.
#### 8. Clinical data
N/A
#### 9. Conclusions
In light of evidences summarized above and based on classification, intended use, technological characteristics and performance data, the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.