K203393 · Smith & Nephew, Inc. · GAT · Feb 1, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203393
Device Name
FAST-FIX FLEX
Applicant
Smith & Nephew, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Feb 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FAST-FIX FLEX is indicated for use in meniscal repairs, allograft transplant procedures, and anchoring the allograft to the meniscal rim during allograft transplant procedures.
Device Story
FAST-FIX FLEX is a meniscal repair system used in arthroscopic, all-inside soft tissue procedures. The device consists of an implant-suture construct featuring #2-0 UHMWPE suture and two PEEK polymer implants, preloaded into a needle delivery inserter. Surgeons use the system to reduce meniscal tears. The device includes a sterile, single-use bend tool allowing the surgeon to adjust the distal needle curvature, and a slotted cannula. The system facilitates suture retention to secure tissue. By enabling precise needle placement and implant deployment, the device assists in stabilizing meniscal tissue, potentially improving healing outcomes in orthopedic procedures.
Clinical Evidence
Bench testing only. Performance data included biocompatibility (ISO 10993-1:2018), sterilization validation, implant repair testing, implant bridge strength, tensile strength through cyclic loading, and implant failure testing. All acceptance criteria were met.
Technological Characteristics
Materials: UHMWPE suture, PEEK polymer implants. Delivery: Needle delivery inserter with adjustable distal curvature via bend tool. Accessories: Slotted cannula. Technique: All-inside arthroscopic. Sterilization: Terminal sterilization.
Indications for Use
Indicated for meniscal repairs and allograft transplant procedures, including anchoring allografts to the meniscal rim.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Smith & Nephew FAST-FIX 360 Meniscal Repair System (K121861)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 1, 2021
Smith & Nephew, Inc. Crystal Morales Regulatory Affairs Specialist II 7135 Goodlett Farms Parkway Cordova, Tennessee 38016
Re: K203393
Trade/Device Name: Fast-fix Flex Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: December 4, 2020 Received: December 7, 2020
Dear Ms. Morales:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203393
Device Name FAST-FIX FLEX
Indications for Use (Describe)
The FAST-FIX FLEX Meniscal Repair System is indicated for use in meniscal repairs, allograft transplant procedures, and anchoring the allograft to the meniscal rim during allograft transplant procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Smith & Nephew, Inc. 150 Minuteman Road Andover, MA 01810 Massachusetts, USA
T:+ 1 978 749 1000 T:+ 1 800 343 8386 (USA toll free) www.smith-nephew.com
# 510(k) Summary
Prepared: 26 January 2021
| Submitter Information | Contact Information |
|-----------------------|----------------------------------|
| Smith & Nephew, Inc. | Crystal Morales |
| Endoscopy Division | Regulatory Affairs Specialist II |
| 150 Minuteman Road | Phone: (267)-266-5383 |
| Andover, MA 01810 | Fax: (978) 749-1443 |
| Device Name & Classification | |
|------------------------------|------------------------------------------------|
| Proprietary Name | FAST-FIX ® FLEX |
| Common Name | Suture Retention Device |
| Classification Name | Suture, nonabsorbable, synthetic, polyethylene |
| Classification Regulation | 21 CFR §878.5000 |
| Class | II |
| Product Code(s) | GAT |
| Panel | General and Plastic Surgery |
## Legally Marketed Predicate Devices
The Smith & Nephew FAST-FIX® FLEX Meniscal Repair System is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed devices in commercial distribution:
| Description | Submission Number | Clearance Date |
|-------------------------------------------------------|-------------------|----------------|
| Smith & Nephew FAST-FIX 360<br>Meniscal Repair System | K121861 | 18 OCT 2012 |
## Device Description
FAST-FIX® FLEX is comprised of an implant-suture construct incorporating #2-0 nonabsorbable Ultra-high Molecular Weight Polyethylene (UHMWPE) suture having a pre-tied one-way sliding knot between two non-absorbable polyetheretheketone (PEEK) polymer implants. The suture-implant construct is preloaded into a needle delivery inserter. The needle delivery inserter is offered in two device configurations of curved and reverse curved, whose needle can be modified by the end user with a supplied Bend Tool. The Bend Tool is a sterile, single use optional non-measuring accessory intended to control the bend radius and limit the bend angle to ensure reliable deployment of implants after modification. A sterile, single use slotted cannula is also supplied with the system.
{4}------------------------------------------------
#### Intended Use
The FAST-FIX FLEX is indicated for use in meniscal repairs, allograft transplant procedures, and anchoring the allograft to the meniscal rim during allograft transplant procedures.
The Indications for Use statement for FAST-FIX FLEX is different from that of the predicate device FAST-FIX 360 Meniscal Repair System (K121861). The differences are for clarification of content and does not impact the intended use of the device nor the safety and effectiveness of the device in relation to the predicate.
As with the predicate device, the Smith & Nephew FAST-FIX FLEX Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures.
#### Technological Characteristics
The Smith & Nephew FAST-FIX® FLEX and predicate device FAST-FIX 360 Meniscal Repair System (K121861) uses an all-inside technique through an arthroscopic portal to allow surgeons to reduce tears. The devices use identical suture-implant construct materials that is preloaded into a needle delivery inserter.
Differences in technological characteristics of FAST-FIX FLEX in comparison to the predicate device include, distal needle curvature adjustment with a supplied bend tool, smaller suture bar implants for less disruption to soft tissue and design feature modification to improve implant retention.
The Smith & Nephew FAST-FIX FLEX is substantially equivalent in intended use and fundamental scientific technology as the legally marketed predicate device and raise no new questions of safety and efficacy.
## Summary Performance Data
The nonclinical performance data provided in the premarket notification in support of substantial equivalence included biocompatibility testing, sterilization evaluation, implant repair testing, implant bridge strength, tensile strength through cyclic loading and implant failure testing.
The biocompatibility of the FAST-FIX FLEX was evaluated against the requirements per ISO 10993-1:2018. Evaluation included assessment of existing data, chemical characterization, and cytotoxicity, which provided framework for requiring additional endpoint testing. All acceptance criteria were met and FAST-FIX FLEX was deemed biologically safe.
Sterilization validation methods of FAST-FIX FLEX was compared to the predicate device FAST-FIX 360 Meniscal Repair System (K121861) to demonstrate suitability of the subject device for terminal sterilization under the same conditions.
The performance data demonstrates that FAST-FIX® FLEX met performance specifications for implant repair and implant bridge strength and met performance specifications for tensile strength through cyclic loading and implant failure testing based on established acceptance
{5}------------------------------------------------
criteria and in comparison to the predicate device FAST-FIX 360 Meniscal Repair System (K121861).
Therefore, the FAST-FIX FLEX is considered substantially equivalent to the currently marketed predicate.
# Substantial Equivalence Information
The substantial equivalence of the FAST-FIX® FLEX is based on similarities in indications for use, general design features, operational principles, material biocompatibility and composition, and performance to the predicate, FAST-FIX 360 Meniscal Repair System (K121861). Based on the similarities to the predicate, the FAST-FIX FLEX is equivalent to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.