Philips Avent Double electric breast pump Advanced SCF394/61, Philips Avent Double electric breast pump Advanced SCF394/62
K203383 · Philips Consumer Lifestyle - Innovation Site Eindhoven · HGX · Dec 15, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K203383
Device Name
Philips Avent Double electric breast pump Advanced SCF394/61, Philips Avent Double electric breast pump Advanced SCF394/62
Applicant
Philips Consumer Lifestyle - Innovation Site Eindhoven
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Dec 15, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Philips Avent double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the breast. The device is intended for a single user.
Device Story
Electrically powered double breast pump; extracts milk via suction applied to nipple; silicone diaphragm prevents backflow into vacuum system. Motor unit features press-button interface, 2-digit display, rechargeable battery or mains adapter. Expression kits include cushion or flange types. Includes breast pump belt for hands-free carrying. Used by lactating women in home environments. Provides milk collection in containers for storage. Benefits include efficient milk expression and engorgement relief.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, including Cytotoxicity (ISO 10993-5), Skin Sensitization (ISO 10993-10), and Skin Irritation (ISO 10993-10).
Technological Characteristics
Materials: ABS, mABS, HTV Silicone, Polypropylene, Polybutylene Terephthalate. Compliant with 21 CFR 177.1520/177.2600. Vacuum range: -45 to -270 mmHg. Cycle speed: 53 to 120 cycles/min. Power: 3.6V 2600mAh Li-ion battery or 100-240V AC adapter. Connectivity: Micro-USB (power only). Interface: 4 buttons, 2-digit display. Non-sterile.
Indications for Use
Indicated for lactating women to express and collect breast milk and alleviate breast engorgement. Intended for single-user, home healthcare environment use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Philips Avent Double electric breast pump Advanced (K201381)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 15, 2020
Philips Consumer Lifestyle - Innovation Site Eindhoven Aurore Millet Senior Regulatory Affairs Manager High Tech Campus 37 Eindhoven, 5656 AE NETHERLANDS
Re: K203383
> Trade/Device Name: Philips Avent Double electric breast pump Advanced SCF394/61 Philips Avent Double electric breast pump Advanced SCF394/62 Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Regulatory Class: Class II Product Code: HGX Dated: November 13, 2020 Received: November 17, 2020
Dear Aurore Millet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Monica Garcia Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203383
Device Name
Philips Avent Double electric breast pump Advanced SCF39461, Philips Avent Double electric breast pump Advanced SCF394/62
Indications for Use (Describe)
The Philips Avent double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the breast. The device is intended for a single user.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K203383
### 1. Submitter Information
| Applicant: | Philips Consumer Lifestyle - Innovation Site<br>Eindhoven |
|------------|-------------------------------------------------------------|
| Address: | High Tech Campus 37<br>Eindhoven 5656 AE<br>The Netherlands |
### 2. Correspondent Information
| Contact: | Aurore Millet |
|----------|---------------------------|
| | Senior Regulatory Manager |
| Phone: | +31 6 181 91 888 |
| Email: | aurore.millet@philips.com |
#### 3. Date prepared: December 11, 2020
#### 4. Device Information
| Device Name: | Philips Avent Double electric breast pump Advanced |
|--------------------|----------------------------------------------------|
| Common Name: | Powered Breast Pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
#### 5. Predicate Device Information
| Device Name: | Philips Avent Double electric breast pump Advanced |
|----------------|----------------------------------------------------|
| 510(k) Number: | K201381 |
| Manufacturer: | Philips Consumer Lifestyle BV |
| Product Code: | HGX (Pump, Breast, Powered) |
The predicate device has not been subject to a design-related recall.
### 6. Device Description
The Philips Avent Double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman. The device is intended for a single user.
The subject devices are for repeated use by a single user and are for use in home and professional healthcare environments. The devices are not sterile.
The subject devices are electrically powered double (Double Advanced) breast pumps consisting of the following key components: a motor unit with a press-button user interface, two expression kits with two milk collection containers connected with the motor unit via silicone tubes. The motor unit operates on embedded software. Software updates by end-users are not supported.
The breast pump extracts milk from the breasts by creating a seal around the nipple and applying and releasing suction to the nipple. The milk is collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, the expression kit includes a silicone diaphragm that
{4}------------------------------------------------
physically separates the milk-contacting pathway from the vacuum system.
The motor unit operates on a rechargeable battery or on an external mains adapter that is provided with the system via a micro-USB type connection. The rechargeable battery can be charged from the external mains adapter if the motor unit is not in operation.
The breast pump is provided with the same accessories as the predicate device (K201381), with the addition of a breast pump belt. The breast pump belt intended to carry the electrical breast pump while the device is used to express and collect milk. The belt is intended to be worn over clothes.
The expression kit is provided in two types: cushion-type (SCF394/61) and flange-type (SCF394/62).
The cushion-type expression kit has a silicone cushion which is mounted on the expression kit pump body. The cushion comes into contact with the breast for the duration of the expression session.
In the flange-type expression kit, the silicone cushion is integrated in a rigid frame. This integrated part (flange) can be attached to the pump body with a screw connection. The silicone part comes into contact with the breast for the duration of the expression. The rigid frame is not in contact with the breast.
All other components of the subject devices are not in contact with the breast.
Table 1: Subject Device variants
| Trade/Device<br>Name | Number of<br>pumps | User<br>Interface | Battery | Expression<br>kit type | Model<br>Identification |
|----------------------------------------------------------------|--------------------|----------------------------------|--------------|------------------------|-------------------------|
| Philips Avent<br>Double<br>electric breast<br>pump<br>Advanced | 2 | 4 buttons and<br>2-digit display | With battery | Cushion | SCF394/61 |
| | | | | Flange | SCF394/62 |
The subject devices consist of the following materials:
- · Motor unit: ABS, mABS, HTV Silicone
- · Cushion, tube, milk valve, diaphragm: Silicone
- · Pump body, milk collection container: Polypropylene
- · Flange: Polypropylene, Polybutylene Terephthalate, Silicone
All milk contacting components are compliant with 21 CFR 177.1520 and 21 CFR 177.2600.
# 7. Indications for Use
The Philips Avent double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the device is intended for a single user.
# 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
{5}------------------------------------------------
| | Philips Avent Double<br>electric breast pump<br>Advanced<br>K203383 | Philips Avent Double<br>electric breast pump<br>Advanced<br>K201381 | Comparison |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device | |
| Product Name | Double electric breast pump<br>Advanced (SCF394/61 and<br>SCF394/62) | Double electric breast pump<br>Advanced (SCF394/61 and<br>SCF394/62) | Same |
| Manufacturer | Philips Consumer Lifestyle BV<br>Tussendiepen 4<br>9206 AD, Drachten<br>The Netherlands | Philips Consumer Lifestyle BV<br>Tussendiepen 4<br>9206 AD, Drachten<br>The Netherlands | Same |
| Product Code | HGX | HGX | Same |
| Regulation No. | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Class | Class II | Class II | Same |
| Patient Population | Lactating Women | Lactating Women | Same |
| Environment of use | Home Healthcare Environment | Home Healthcare Environment | Same |
| Indications for Use | The Philips Avent double electric<br>breast pump Advanced is intended<br>to express and collect milk from<br>the breast of a lactating woman,<br>and to alleviate engorgement of the<br>breast. The device is intended for a<br>single user. | The Philips Avent single/double<br>electric breast pump Advanced is<br>intended to express and collect<br>milk from the breast of a<br>lactating woman, and to alleviate<br>engorgement of the breast. The<br>device is intended for a single<br>user. | Same |
| Pump Options | Double | Single or Double | Similar: The affected<br>models of the predicate<br>are exclusively double<br>pumping |
| Backflow Protection | Yes<br>(Silicone diaphragm on the<br>expression kit, which separates<br>the vacuum system of the motor<br>unit from milk-contacting parts) | Yes<br>(Silicone diaphragm on the<br>expression kit, which separates<br>the vacuum system of the motor<br>unit from milk-contacting parts) | Same |
| Suction Modes | Stimulation Mode and Expression<br>Mode | Stimulation Mode and<br>Expression Mode | Same |
| Transition between<br>suction modes | Automatic or Manual | Automatic or Manual | Same |
| Adjustable suction levels | Yes | Yes | Same |
| Vacuum range:<br>Stimulation | -45 to -150 mmHg | -45 to -150 mmHg | Same |
| Vacuum range:<br>Expression | -45 to -270 mmHg | -45 to -270 mmHg | Same |
| Cycle Speed:<br>Stimulation | 105 to 120 cycles/min | 105 to 120 cycles/min | Same |
| Cycle Speed:<br>Expression | 53 to 85 cycles/min | 53 to 85 cycles/min | Same |
| Controls | 4 Buttons (On-Off-Pause-<br>Resume, Stimulation or<br>Expression mode. Decrease | 4 Buttons (On-Off-Pause-<br>Resume, Stimulation or<br>Expression mode. Decrease | Same |
| | vacuum level, Increase vacuum level) | vacuum level, Increase vacuum level) | |
| Indicators | LED indicators:<br>8-dot display (versions without battery)<br>2-digit 7-segment display<br>(versions with battery) | LED indicators:<br>8-dot display (versions without battery)<br>2-digit 7-segment display<br>(versions with battery) | Same |
| Power connection | Micro-USB (power only) | Micro-USB (power only) | Same |
| Power Supply | AC-Adapter provided (100-240 V primary and 5 V secondary) or Lithium- ion battery 3.6 V 2600 mAh (for battery-operated devices) | AC-Adapter provided (100-240 V primary and 5 V secondary) or Lithium- ion battery 3.6 V 2600 mAh (for battery-operated devices) | Same |
| Accessories | Pump body<br>Cushion<br>Flange<br>Slave cap<br>Membrane<br>Tube<br>Milk valve<br>4oz bottle<br>Breast shield cover<br>Screw ring<br>Sealing disc<br>Teat (nipple)<br>Dormal cap<br>Pouch<br>Travel bag<br>Disposable breast pads<br>Breast pump belt | Pump body<br>Cushion<br>Flange<br>Slave cap<br>Membrane<br>Tube<br>Milk valve<br>4oz bottle<br>Breast shield cover<br>Screw ring<br>Sealing disc<br>Teat (nipple)<br>Dormal cap<br>Pouch<br>Travel bag<br>Disposable breast pads | Different: The subject device is packaged with a breast pump belt accessory. Through performance testing validation it was determined that the accessory does not raise additional questions of substantial equivalence. |
# Table 2: Comparator Table for Subject and Predicate Devices
{6}------------------------------------------------
The indications for use of the subject and predicate device are identical.
The subject and predicate devices have similar technological features, including design, user interface, vacuum pressure range, cycle speeds, and power source. The different technological characteristics of the subject device do not raise different types of safety and effectiveness questions.
# 9. Summary of Non-Clinical Performance Testing
# Biocompatibility
Biocompatibility studies, including Skin Irritation Testing, Cytotoxicity, and Skin Sensitization testing were performed in accordance with the 2016 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows:
- Cytotoxicity (ISO 10993-5:2009)
- Skin Sensitization (ISO 10993-10:2010) ●
- Skin Irritation (ISO 10993-10:2010) ●
The testing supports the biocompatibility of the device.
# 10. Conclusion
The performance testing described above demonstrate that the subject devices are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.