K203359 · Becton, Dickinson and Company · FMF · Jul 28, 2021 · General Hospital
Device Facts
Record ID
K203359
Device Name
BD Flu+ Syringe
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
Jul 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD™ Flu+ Syringe is intended for aspiration and injection of fluids.
Device Story
BD Flu+ Syringe is a sterile, two-piece, single-use piston syringe with an integrated needle; designed for variable dose delivery up to 1 mL. Features low dead space to minimize medication waste. Used in clinical settings for general-purpose aspiration and injection of fluids from vials or ampoules. Operated by healthcare professionals. Device provides visual dose markings (0.5 mL, 1.0 mL, and 0.25 mL increments) to assist clinicians in accurate dosing. Benefits include reduced medication waste and simplified fluid delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing (leakage, dose accuracy, dead space, activation forces, cannula/shield pull force, scale mark permanency) and biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility) per ISO 10993 and USP standards.
Technological Characteristics
Two-piece piston syringe (plastic barrel/plunger) with integrated stainless steel needle. Low dead space design. Sterilized via Ethylene Oxide (SAL 10^-6). Materials: plastic barrel, plastic/colorant plunger, stainless steel cannula, silicone lubricant, epoxy adhesive. Dimensions: 1.0 mL max capacity. Standards: ISO 7886-1 (performance), ISO 7864 (cannula), ISO 10993 (biocompatibility), ASTM F1980-16 (shelf life).
Indications for Use
Indicated for aspiration and injection of fluids in patients requiring medication delivery via syringe.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the full name "U.S. Food & Drug Administration" in blue text.
July 28, 2021 Becton, Dickinson and Company Nikita Kumar Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
### Re: K203359
Trade/Device Name: BD Flu+ Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: June 26, 2021 Received: June 30, 2021
### Dear Nikita Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203359
Device Name BD™ Flu+ Syringe
| Indications for Use (Describe) |
|---------------------------------------------------------------------------------------|
| The BD <sup>TM</sup> Flu+ Syringe is intended for aspiration and injection of fluids. |
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular symbol on the left and the letters "BD" on the right. The circular symbol is orange and features a stylized image of a person with outstretched arms inside of a circle. The letters "BD" are in blue and are written in a bold, sans-serif font.
## K203359 510(k) Summary
# BD™ Flu+ Syringe
| Submitter<br>Information | Submitter Name: | Becton, Dickinson and Company |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| | Submitter Address: | 1 Becton Drive<br>Franklin Lakes, NJ 07417 |
| | Contact Person: | Nikita Abirami Mahendra Kumar<br>Senior Regulatory Affairs Specialist |
| | Email Address: | Nikita.Abirami.Mahendra.Kumar@bd.com |
| | Phone Number: | (201) 847-5641 |
| | Date of Preparation: | July 28, 2021 |
| Subject Device | Trade Name: | BD™ Flu+ Syringe |
| | Common Name: | Piston Syringe |
| | Regulation Number: | 21 CFG 880.5860 |
| | Regulation Name: | Piston Syringe |
| | Regulatory Class: | Class II device |
| | Product Code: | FMF |
| | Classification Panel: | General Hospital |
| Predicate Device | Trade Name: | BD™ FLU+ Syringe |
| | 510(k) Reference: | K091377 |
| | Common Name: | Piston Syringe |
| | Regulation Number: | 21 CFR 880.580 |
| | Regulation Name: | Piston Syringe |
| | Regulatory Class: | Class II Device |
| | Product Code: | FMF |
| | Classification Panel: | General Hospital |
| Reason for<br>Submission | The basis of this submission is to modify the indications for use<br>of the BD™ Flu+ Syringe. | |
| Device Description | The BD Flu+ Syringe is a sterile two-piece single use, sterile<br>syringe with an integral needle. It allows for a variable dose up<br>to 1 ml to be aspirated and injected. It is intended for general-<br>purpose aspiration and injection of fluids from a vial or ampoule.<br>The BD Flu+ syringe has been designed for low dead space to<br>reduce medication waste. | |
| | The BD Flu+ Syringe is a 1.0mL maximum dosage with 0.5mL<br>and 1.0mL barrel marking and 0.25mL incremental markings.<br>The Flu+ syringe is assembled with a pre attached needle in the<br>following gauges and sizes: | |
| | 23G (0.6mm) x 1 inch (25mm) | |
| | 25G (0.5mm) x 5/8 inch (16mm.) | |
| Indications for Use | The BD™ Flu+ Syringe is intended for aspiration and injection of fluids. | |
| Technological<br>Characteristics | The subject devices are equivalent to the predicate devices in intended use, materials and performance characteristics. | |
{4}------------------------------------------------
# 25G (0.5mm) x 1 inch (25mm)
| Element of<br>Comparison | Subject Device:<br>BD™ Flu+ Syringe | Predicate Device<br>BD™ Flu+ Syringe | Comparison |
|----------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------|
| 510(k) Number | K203359 | K091377 | N/A |
| Intended Use | Intended for<br>aspiration and<br>injection of fluids | Intended for<br>aspiration and<br>injection of the<br>influenza vaccine | Different; Subject<br>device is intended<br>for general use |
| Syringe Type | 2 Piece (barrel and<br>plunger) | 2 Piece (barrel and<br>plunger) | Same |
| Dose Saving<br>Feature | Low Dead<br>Space/Volume | Low Dead<br>Space/Volume | Same |
| Integrated<br>Needle | Yes | Yes | Same |
| Dose<br>Setting/Volumes | Variable dose scale<br>markings with<br>1.0mL max delivery | Variable dose scale<br>markings with<br>1.0mL max delivery | Same |
| Needle Size<br>(Gauge x Length) | 23G x 1"<br>25G x 1"<br>25G x 5/8" | 23G x 1"<br>25G x 1"<br>25G x 5/8" | Same |
| Barrel Material | Plastic | Plastic | Same |
| Plunger Material | Plastic+Colorant<br>(Blue, Orange) | Plastic+Colorant<br>(Blue, Orange) | Same |
| Integrated<br>Cannula | Stainless Steel | Stainless Steel | Same |
| Shield | Plastic | Plastic | Same |
| Adhesive | Epoxy | Epoxy | Same |
| Cannula<br>Lubricant | Silicone | Silicone | Same |
| Sterilization | Ethylene Oxide (SAL<br>10-6) | Ethylene Oxide (SAL<br>10-6) | Same |
| Shelf Life | 5 years | 5 years | Same |
| Functional Testing: | | | |
| Leakage | ISO 7886-1 | ISO 7886-1 | Same |
| Dose Accuracy | ISO 7886-1 | ISO 7886-1 | Same |
| Dead Space | ISO 7886-1 | ISO 7886-1 | Same |
| Activation Forces | ISO 7886-1 | ISO 7886-1 | Same |
| Cannula Pull Force | ISO 7864 | ISO 7864 | Same |
| Shield Pull Force | BD Internal Requirement | BD Internal Requirement | Same |
| Scale Mark Permanency | BD Internal Requirement | BD Internal Requirement | Same |
| <b>Biocompatibility Testing:</b> | | | |
| Testing per ISO 10993:2018 | | | |
| Cytotoxicity | ISO 10993-5 & USP <87>, Non-cytotoxic<br>• LLNA: ISO 10993-10 & ASTM F2148 | ISO 10993-5 & USP <87>, Non-cytotoxic<br>• LLNA: ISO 10993-10 & ASTM F2148 | Same |
| Sensitization | • Maximization: ISO 10993-11<br>Non-sensitizer | • Maximization: ISO 10993-11<br>Non-sensitizer | Same |
| Intracutaneous Reactivity | Per ISO 10993-10 & USP<88>, Non-irritant | Per ISO 10993-10 & USP<88>, Non-irritant | Same |
| Primary Dermal Irritation | ISO 10993-10 | ISO 10993-10 | Same |
| Acute Systemic Toxicity | ISO 10993-11 & USP<88>, Non-toxic | ISO 10993-11 & USP<88>, Non-toxic | Same |
| Pyrogenicity | Per ISO 10993-11:2017 & USP<151>, Non-pyrogenic | Per ISO 10993-11:2017 & USP<151>, Non-pyrogenic | Same |
| Extractables/Leachables | Per ISO 10993-18:2005, Acceptable | Per ISO 10993-18:2005, Acceptable | Same |
| Hemolysis | Per ISO 10993-4, Non-hemolytic | Per ISO 10993-4, Non-hemolytic | Same |
{5}------------------------------------------------
Performance Tests
BD has performed the following bench and biocompatibility testing on BD™ Flu+ Syringe. Modifications of the subject device's intended use does not affect the results of this analysis.
Bench Performance:
- Leakage (ISO 7886-1:2017) .
- Dose Accuracy (ISO 7886-1:2017) .
Confidential & Proprietary Page 3 of 5
{6}------------------------------------------------
- Dead Space (ISO 7886-1:2017) ●
- Activation Forces (ISO 7886-1:2017)
- Cannula Pull Force (ISO 7864:2016) ●
- Shield Pull Force (BD Internal Requirement)
- Scale Mark Permanency (BD Internal Requirement) ●
- Package Integrity (ISO 11607-1:2010). ●
Material Biocompatibility Performance:
In accordance with ISO 10993-1, the Flu+ Syringe is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). The following testing was conducted:
- Cytotoxicity ●
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity
- . Hemocompatibility
Particulate matter testing was conducted in accordance with USP<788> Particulate Matter in Injections and met the USP acceptance criteria.
## Sterility, Shipping and Shelf-Life
| Sterilization<br>Method | Ethylene Oxide (ETO) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Residuals | Maximum EO mg/device limit:<br>4mg/device<br>Maximum ECH mg/device Limit:<br>9mg/device<br>The above limits are acceptable per<br>ISO 10993-7:2008 |
| Validation<br>Method | The sterilization process is validated<br>in accordance with Standard EN ISO<br>11135-1."Sterilization of health care<br>products-Ethylene oxide- Part 1:<br>Requirements for development,<br>validation and routine control of a<br>sterilization process for medical<br>devices".<br>The sterilized product meets the |
{7}------------------------------------------------
| | | requirements of EN 556 "Sterilization<br>of Medical Devices - Requirements for<br>Medical Devices to be Labelled Sterile<br>". |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Pyrogenicity<br>/Endotoxin<br>Testing | Test methods are per USP 40-NF 35<br>monographs 85 (Bacterial Endotoxins<br>Test) and 161 (Medical Devices<br>Bacterial Endotoxin and Pyrogens<br>Test). |
| | Packaging integrity testing, after environmental conditioning are<br>simulated transportation in accordance with ISO 11607-1:2010<br>and 11135:2014, was conducted on the final packaged, and<br>sterile device. All packaging deemed acceptable for protection of<br>product and sterility maintenance. | |
| | Shelf Life of 5 years is validated using the FDA recognized<br>standards ASTM F1980-16 Standard Guide for Accelerated Aging<br>of Sterile Barrier Systems for Medical Devices. | |
| | The subject device continue to meet all the predetermined<br>acceptance criteria for the above-listed performance tests,<br>demonstrating substantial equivalence to the predicate device. | |
| Clinical Testing | | Clinical testing was not required for this submission |
| Summary of<br>Substantial<br>Equivalence | | The BD™ Flu+ Syringe is substantially equivalent to the<br>predicate device in principles of operation, technology, design, |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.