K203340 · Siemens Medical Solutions USA, Inc. · KPR · Jan 6, 2021 · Radiology
Device Facts
Record ID
K203340
Device Name
MULTIX Impact C
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPR · Radiology
Decision Date
Jan 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
MULTIX Impact C is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact C enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact C is not intended for mammography. MULTIX Impact C uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact C is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
Device Story
Radiographic X-ray system for hospitals/clinics; operated by healthcare professionals. Inputs: X-ray radiation; outputs: diagnostic images via digital detectors, film/screen, or CR cassettes. System components: ceiling-mounted tube stand (manual/motorized), bucky wall stand, bucky table, generator, and wireless/fixed detectors. Features include motorized tube tilting, automatic collimator, and touch user interface. Ortho function options (Smart Ortho, Smart Virtual Ortho, Auto Full-Spine & Long-Leg Collimation) assist in planning and positioning. Wireless remote-control console enables motion control and positioning. System monitors performance; blocks functions and displays error messages upon failure. Benefits: enables diagnostic imaging across diverse patient populations and clinical settings.
Clinical Evidence
Bench testing only. Verification and validation testing performed to support hardware and software modifications. Conformance to standards including ES60601-1, IEC 60601-1-3, IEC 60601-1-2, IEC 62366-1, ISO 14971, IEC 62304, and ISO 10993-1. Risk analysis and integration/functional testing confirmed specifications met acceptance criteria.
Technological Characteristics
Stationary X-ray system. Components: ceiling-mounted tube stand, bucky wall stand, bucky table, X-ray generator, digital detectors. Materials conform to ISO 10993-1. Energy source: electrical. Connectivity: wireless/fixed detectors. Software: VA20 version. Standards: ES60601-1, IEC 60601-1-3, IEC 60601-1-2, IEC 62366-1, ISO 14971, IEC 62304, IEC 60601-2-28, IEC 60601-2-54, NEMA PS 3.1-3.20.
Indications for Use
Indicated for radiographic imaging of the whole body (skull, chest, abdomen, extremities) in pediatric, adult, and bariatric patients. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
January 6, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Siemens Medical Solutions USA, Inc. % Mr. Martin Rajchel Senior Regulatory Affairs Specialist 40 Liberty Blvd. Mail code: 65-1A MALVERN PA 19355
### Re: K203340
Trade/Device Name: MULTIX Impact C Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MQB Dated: November 11, 2020 Received: November 12, 2020
Dear Mr. Rajchel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203340
Device Name MULTIX Impact C
#### Indications for Use (Describe)
MULTIX Impact C is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact C enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, or in the prone position. MULTIX Impact C is not intended for mammography.
MULTIX Impact C uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact C is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: MULTIX Impact C
## K203340
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
Date Prepared: November 11, 2020
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
## 1. General Information
Importer/ Distributor Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
### Location of Manufacturing Sites
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road Shanghai, 201318, China Establishment Registration Number: 3003202425
Siemens Healthcare GmbH Siemensstrasse 1 Forchheim, Germany 91301 Establishment Registration Number: 3004977335
## 2. Contact Person
Martin Rajchel Sr. Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 martin.rajchel(@siemens-healthineers.com
{4}------------------------------------------------
## 3. Subject Device Name and Classification
| Trade Name: | MULTIX Impact C |
|----------------------------|-------------------------|
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
| Secondary Product Code: | MQB |
# 4. Legally Marketed Predicate Device
| Trade Name: | MULTIX Impact |
|----------------------------|-------------------------|
| 510(k) #: | K193089 |
| Clearance Date: | March 20, 2020 |
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
| Reference Devices: | |
| Trade Name: | YSIO X.Pree |
| 510(k) #: | K201670 |
| Clearance Date: | October 21, 2020 |
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
| Trade Name: | Multix Fusion Max |
| 510(k) #: | K191418 |
| Clearance Date: | June 19, 2019 |
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
{5}------------------------------------------------
### 5. Device Description
The MULTIX Impact C Radiography X-ray system is a modular system of x-ray components (ceiling suspension with x-ray tube, bucky wall stand, bucky table, x-ray generator, and portable wireless and fixed detectors) based on the predicate device, the MULTIX Impact (K193089). The detectors for the subject device, MULTIX Impact C, are the same as the detectors of the predicate device. The following modifications have been made to the predicate device:
- 1. New ceiling suspension with motorized tube tilting support for ortho function and new ceiling suspension with manual tube tilting
- Modified automatic collimator 2.
- 3. New Bucky Wall Stand
- Upgraded software version from VA11 to VA20 to support hardware modifications 4.
- Modified patient table న్.
- Modified touch user interface (TUI) 6.
- Modified wireless remote-control console (WRCC) with new control design 7.
The new system will be branded as the MULTIX Impact C.
#### 6. Indications for Use
MULTIX Impact C is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact C enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX
Impact C is not intended for mammography.
MULTIX Impact C uses digital detectors for generating diagnostic images by converting Xrays into image signals. MULTIX Impact C is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
#### 7. Substantial Equivalence
MULTIX Impact C is a modification of the predicate device, the MULTIX Impact (K193089). The subject device is within the same classification regulation, has the same indications for use, and the same mechanical design as the predicate device. The MULTIX Impact C is substantially equivalent to the predicate device and documentation is provided to support a claim of substantial equivalence.
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device
MULTIX Impact C is substantially equivalent to the commercially available predicate device, MULTIX Impact (K193089) in terms of the indications for use, design, material, functionality, technology, and energy source. The subject device uses the same or similar components cleared in the MULTIX Impact (e.g. tube, generator, collimator, patient table, detectors and imaging system).
{6}------------------------------------------------
The components of the subject device have many of the same technological characteristics as those in the predicate device. Some technological characteristics that differ slightly are shown in the comparison tables below. Verification and validation testing have been successfully completed and test results show that the subject device, MULTIX Impact C with all its components, is substantially equivalent to the predicate device.
The modifications made to the subject device, MULTIX Impact C, do not affect the intended use of the device nor do they alter its fundamental scientific technology compared to the predicate device, the MULTIX Impact (K193089).
{7}------------------------------------------------
The following tables compare the main performance data of the subject device with the predicate device.
| Attribute | MULTIX Impact C<br>(Subject) | MULTIX Impact<br>K193089<br>(Predicate) | Comparison<br>Results |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | MULTIX Impact C is a<br>radiographic system used in<br>hospitals, clinics, and medical<br>practices. MULTIX Impact C<br>enables radiographic exposures<br>of the whole body including:<br>skull, chest, abdomen, and<br>extremities and may be used on<br>pediatric, adult and bariatric<br>patients. Exposures may be<br>taken with the patient sitting,<br>standing, or in the prone<br>position. MULTIX Impact C is<br>not intended for<br>mammography.<br>MULTIX Impact C uses digital<br>detectors for generating<br>diagnostic images by<br>converting X- rays into image<br>signals.<br>MULTIX Impact C is also<br>designed to be used with<br>conventional film/screen or<br>Computed Radiography (CR)<br>cassettes. | The MULTIX Impact system<br>is a radiographic system used<br>in hospitals, clinics, and<br>medical practices. MULTIX<br>Impact enables radiographic<br>exposures of the whole body<br>including: skull, chest,<br>abdomen, and extremities and<br>may be used on pediatric,<br>adult and bariatric patients.<br>Exposures may be taken with<br>the patient sitting, standing, or<br>in the prone position. The<br>MULTIX Impact system is not<br>meant for mammography.<br>The MULTIX Impact uses<br>digital detectors for generating<br>diagnostic images by<br>converting X-rays into image<br>signals.<br>The MULTIX Impact is also<br>designed to be used with<br>conventional film/screen or<br>Computed Radiography (CR)<br>cassettes. | Same<br>(Minor<br>grammatical<br>changes only) |
| Attribute | MULTIX Impact C<br>(Subject) | MULTIX Impact<br>K193089<br>(Predicate) | Comparison<br>Results |
| Tube Stand<br>(TS) | Ceiling mounted manual TS<br>- Manual tube tilting<br>- Motorized tube lifting<br>Ceiling mounted motorized TS<br>- Motorized tube tilting<br>- Motorized tube lifting | Floor mounted semi motorized<br>TS<br>- Manual tube tilting<br>- Manual longitudinal<br>movement<br>- Motorized tube lifting | Improved to support<br>more operational<br>possibilities. |
| | Hardware: Ortho Support<br>Software: three options as below | N.A. | |
| Ortho<br>function<br>(option) | Smart Ortho<br>-<br>Smart Virtual Ortho<br>-<br>Auto Full-Spine & Long-<br>-<br>Leg Collimation | N.A. | New option to support<br>more operational<br>possibilities. |
| Collimator | Automatic collimator | Automatic collimator | Same |
| | N.A. | Manual collimator | New option |
| | Modified automatic collimator | N.A. | |
| Bucky Wall<br>Stand<br>(BWS) | New BWS with motorized height<br>adjustment<br>- 5 field AEC<br>- Option for detector unit with<br>motorized lifting scope of<br>315mm~1750mm<br>- Detector unit with manual<br>tilting | BWS with manual or<br>motorized height adjustment<br>- 3 field AEC<br>- Option for detector unit with<br>manual or motorized lifting<br>scope of 330mm~1800mm | Improved to support<br>more operational<br>possibilities. |
| Imaging<br>System | Detectors<br>- Trixell MAX wi-D= Pixium<br>3543EZh<br>- iRay Core XL=Mars1717VS<br>- iRay Core<br>Static=Venu1717X | Detectors<br>- Trixell MAX wi-D= Pixium<br>3543EZh<br>- iRay Core XL=Mars1717VS<br>- iRay Core<br>Static=Venu1717X | Same |
| | Software version: VA20 | Software version: VA11 | New software version<br>with features like Ortho<br>function,<br>SmartPositioning, and<br>modifications to support<br> |
| Attribute | MULTIX Impact C<br>(Subject) | MULTIX Impact<br>K193089<br>(Predicate) | Comparison<br>Results |
| | User interface:<br>- Operation Tabs are on the<br>top area | User interface:<br>- Operation Tabs are on the<br>right sidebar | Improved appearance of<br>user interface. |
| Other minor modifications | | | |
| Patient table | Elevating patient table with<br>- 5 field AEC<br>- Tray for wireless or fixed<br>detector | Fixed or elevating patient table<br>with<br>- 3 field AEC<br>- Tray for wireless detector | Improved to support<br>more operational<br>possibilities. |
| Touch User<br>Interface<br>(TUI) | Touch user interface<br>- physical button with<br>software indicator<br>- Modified user interface for<br>Ortho function | Touch user interface<br>- physical button with<br>silkscreen | Improved to support<br>more operational<br>possibilities. |
| Wireless<br>Remote<br>Control<br>Console<br>(WRCC) | Wireless Remote-Control<br>Console (WRCC)<br>- Collimation control<br>- BWS lifting control<br>- Enabled tracking<br>- Tube stand motion (lifting)<br>control<br>- SmartPositioning button<br>- Light localizer button | Wireless Remote-Control<br>Console (WRCC)<br>- Collimation control<br>- BWS lifting control<br>- Enabled tracking<br>- Light localizer button | Improved to support<br>more operational<br>possibilities. |
Table 1: Indications for Use Comparison:
{8}------------------------------------------------
# Table 2: Subject Device Compared to Predicate
{9}------------------------------------------------
{10}------------------------------------------------
| Product<br>Function | | MULTIX Impact C<br>(Subject) | YSIO X.pree<br>(K201670)<br>(reference) | Comparison<br>Results |
|-------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Hardware | | Ortho Support(Ortho<br>Stand) | Ortho Support(Ortho<br>Stand) | Same |
| Software<br>(optional<br>license<br>keys) | Smart Ortho | Tilting Ortho: Ortho range<br>set by adjusting collimator<br>and tube tilt manually | Tilting Ortho: Ortho range<br>set by adjusting collimator<br>and tube tilt manually | |
| | Smart Virtual<br>Ortho | Ortho range set by 2D<br>camera in the image<br>system manually | Ortho range set by 3D<br>camera in the image<br>system manually | New option to<br>support more<br>operational<br>possibilities. |
| | Auto Full-Spine<br>& Long-Leg<br>Collimation | Otho range automatically<br>planned for Full-Spine &<br>Long-Leg by 2D camera<br>with manual adjustment | Otho range automatically<br>planned for Thorax by 3D<br>camera with manual<br>adjustment | |
## Table 3: Comparison of the Ortho function to the reference device
## Table 4: Comparison of the Tube Stand to the reference device
| Attribute | MULTIX Impact C<br>(Subject) | Multix Fusion Max<br>K191418<br>(reference) | Comparison<br>Results |
|--------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Tube Stand<br>(TS) | Ceiling mounted manual TS<br>- Manual tube tilting<br>- Motorized tube lifting | Ceiling mounted semi-motorized TS<br>- Manual tube tilting<br>- Motorized tube lifting | New option to<br>support more<br>operational<br>possibilities. |
| | Ceiling mounted motorized TS<br>- Motorized tube tilting<br>- Motorized tube lifting | | |
{11}------------------------------------------------
### 9. Nonclinical Performance Testing
Non-clinical tests were conducted for the MULTIX Impact C during product development. The modifications described in this Premarket Notification are supported with verification and validation testing.
MULTIX Impact C conforms to the following standards: ES60601-1:2005/(R)2012 and A1:2012: IEC 60601-1-3:2008+A1:2013: IEC 60601-1-2:2014: IEC 62366-1:2015: ISO 14971:2007; IEC 60601-1-6:2013; IEC 62304:2015; IEC 60601-2-28:2017; IEC 60601-2-54:2018; NEMA PS 3.1-3.20 (2016) and ISO 10993-1:2009.
All components of the x-ray system MULTIX Impact C were tested and found adequate. All test results are a pass and support our claim of device safety and effectiveness.
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005, is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted on the MULTIX Impact C during product development.
The risk analysis was completed and risk controls were implemented to mitigate identified hazards. The test results support that all the software specifications have met the acceptance criteria. Verification and validation testing were found acceptable to support the claim of substantial equivalence.
#### General Safety and Effectiveness Concerns 10.
Instructions for use are included within the device labeling and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features, including visual and audible warnings, are incorporated into the system design. In addition, the MULTIX Impact C Radiography X-ray system is continually monitored and if an error occurs the system functions will be blocked and an error message will be displayed.
Risk management is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are healthcare professionals familiar with and responsible for the x-ray examinations to be performed.
#### 11. Conclusion as to Substantial Equivalence
The MULTIX Impact C has the same indications for use as the predicate device, MULTIX Impact (K193089). The operating environment and mechanical design are similar.
Verification and validation testing demonstrate that the MULTIX Impact C performs as intended. The non-clinical test data demonstrate that the MULTIX Impact C device
{12}------------------------------------------------
performance is comparable to the predicate device that is currently marketed for the same intended use.
In summary, Siemens concludes that the MULTIX Impact C does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as, the predicate device, MULTIX Impact (K193089).
#### 12. Guidance documents
The following FDA guidance documents were utilized in the documentation of this Premarket Notification:
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: October 2, 2014
- Information to Support a Claim of Electromagnetic Compatibility (EMC) of . Electrically-Powered Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on July 11, 2016.
- Pediatric Information for X-ray Imaging Device Premarket Notifications . Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.
- Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016
- Guidance for Industry and FDA Staff Guidance for the Content of ● Premarket Submissions for Software contained in Medical Devices Document issued on: May 11, 2005
- Appropriate Use of Voluntary Consensus Standards in Premarket . Submissions for Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 14, 2018
- . The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff Document issued on: July 28, 2014
{13}------------------------------------------------
- Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: August 14, 2013
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.