K203319 · Medos International SARL · NKG · Feb 5, 2021 · Orthopedic
Device Facts
Record ID
K203319
Device Name
SYMPHONY™ OCT System
Applicant
Medos International SARL
Product Code
NKG · Orthopedic
Decision Date
Feb 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3075
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3): - · Traumatic spinal fractures and/or traumatic dislocations; - Instability or deformity; - · Failed previous fusions (e.g. pseudarthrosis); - · Tumors involving the cervical/thoracic spine; · Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and · Degenerative disease of the facets with instability. The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods. The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine. The SYMPHONY OCT System may be connected to the EXPEDIUM Spine System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.
Device Story
Posterior spinal fixation system; provides immobilization/stabilization of craniocervical, cervical, and upper thoracic spine segments. System comprises bone anchors (screws), longitudinal components (offset rods), interconnection mechanisms (set screws), and optional transverse connectors. Surgeon-assembled to fit patient anatomy. Used in surgical settings to treat spinal fractures, deformities, tumors, and degenerative conditions. Provides mechanical stability to facilitate fusion or restore spinal integrity in palliative tumor cases. Compatible with various DePuy Synthes spinal systems (SYNAPSE, MOUNTAINEER, EXPEDIUM, VIPER, USS, MATRIX) and SONGER wire/cable systems.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by static torsion, static compression, and dynamic compression testing per ASTM F1717. MR compatibility verified via ASTM F2052, F2119, F2182, and F2213.
Technological Characteristics
Posterior spinal fixation system. Components: Titanium alloy (ASTM F136). Mechanical fixation via screws, rods, and connectors. MR Conditional. No software or electronic components.
Indications for Use
Indicated for patients with acute/chronic instabilities of craniocervical junction, cervical spine (C1-C7), and upper thoracic spine (T1-T3) requiring immobilization as an adjunct to fusion. Conditions include traumatic fractures/dislocations, instability/deformity, failed fusions, tumors, and degenerative disease (radiculopathy, neck/arm pain, facet instability). Also indicated for spinal column integrity restoration in patients with advanced cervical tumors where fusion is not achievable due to limited life expectancy.
Regulatory Classification
Identification
Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
{0}------------------------------------------------
February 5, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Medos International Sarl % Eric Zhu Sr. Regulatory Affairs Specialist DePuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767
Re: K203319
Trade/Device Name: SYMPHONY™ OCT System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP Dated: November 9, 2020 Received: November 12, 2020
Dear Eric Zhu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203319
Device Name SYMPHONY™ OCT System
## Indications for Use (Describe)
The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3):
- · Traumatic spinal fractures and/or traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudarthrosis);
- · Tumors involving the cervical/thoracic spine;
· Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
· Degenerative disease of the facets with instability.
The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.
The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.
The SYMPHONY OCT System may be connected to the EXPEDIUM Spine System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| A. | Submitter Information | |
|----|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturer: | Medos International SARL<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland |
| | Submitter: | DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, MA 02767 |
| | Contact Person: | Eric Zhu<br>325 Paramount Drive<br>Raynham, MA 02767 |
| | Telephone: | (508) 828-3730 |
| | Fax: | (508) 828-3797 |
| | Email: | Ezhu@its.jnj.com |
| B. | Date Prepared | November 9, 2020 |
| C. | Device Name | |
| | Trade/Proprietary Name: | SYMPHONY™ OCT System |
| | Common/Usual Name: | Posterior Cervical Screw System |
| | Regulatory Class: | II |
| | Review Panel: | Orthopedic |
| | Product Codes: | NKG – Class II – 21 CFR §888.3075<br>Posterior Cervical Screw System |
| | | KWP – Class II – 21 CFR §888.3050<br>Appliance, Fixation, Spinal Interlaminal |
| D. | Predicate Device Names | |
| | Primary Predicate: | SYMPHONY OCT System (K181949) |
| | Additional Predicates: | SYMPHONY OCT System (K190895, K192646)<br>SYNAPSE OCT System (K142838)<br>MOUNTAINEER OCT Spinal System (K151885)<br>CONNECTOR System by Orthofix Inc. (K190751) |
{4}------------------------------------------------
#### E. Submission Purpose
Obtain clearance for additional offset rod components of the SYMPHONY OCT System and update in MR Conditional product labeling to include additional scanning parameters.
### F. Device Description
The SYMPHONY OCT System is a posterior spinal fixation system intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3). The system is composed of multiple components to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The system consists of bone anchors (such as screws) for connection by longitudinal components (such as offset rods) via an interconnection mechanism (e.g., set screws) with optional transverse connectors (e.g., cross connectors) to link the longitudinal components for additional stability.
### G. Indications for Use
The SYMPHONY OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3):
- · Traumatic spinal fractures and/or traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudarthrosis);
- · Tumors involving the cervical/thoracic spine;
- · Degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and
- · Degenerative disease of the facets with instability.
The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.
{5}------------------------------------------------
The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.
The SYMPHONY OCT System may be connected to the EXPEDIUM® Spine System and VIPER® System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.
## H. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use
The intended use, technological characteristics, and performance of the SYMPHONY OCT System subject devices are consistent with those of the predicate devices.
## I. Materials
The additional SYMPHONY OCT System components in scope are comprised of Titanium alloy conforming to ASTM F136.
### J. Performance Data
## Offset Rods:
Non-clinical testing was conducted in alignment with the following standards:
- . ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
The following tests were completed to support substantial equivalence:
- Static Torsion
- Static Compression ●
- Dynamic Compression .
## MR Compatibility:
Non-clinical testing was conducted in alignment with the following standards:
- ASTM F2213 Standard Test Method for Measurement of Magnetically Induced . Torque on Medical Devices in the Magnetic Resonance Environment
- ASTM F2052 Standard Test Method for Measurement of Magnetically Induced . Displacement Force on Medical Devices in the Magnetic Resonance Environment
- ASTM F2119 Standard Test Method for Evaluation of MR Image Artifacts from . Passive Implants
{6}------------------------------------------------
- ASTM F2182 Standard Test Method for Measurement of Radio Frequency . Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
Results demonstrated compatibility conditions of the subject devices in the MR environment.
#### K. Conclusion
Evaluation of the subject device intended use, technological characteristics, and performance data demonstrates substantial equivalence with the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.