K203291 · Medtronic Sofamor Danek USA, Inc. · OLO · Dec 8, 2020 · Neurology
Device Facts
Record ID
K203291
Device Name
CD Horizon Spinal System Instruments
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
OLO · Neurology
Decision Date
Dec 8, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Medtronic Navigated Reusable Instruments for use with StealthStation™ and IPC™ POWEREASE™ Systems Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Reusable Instruments are also compatible with the IPC™ POWEREASE™ System. Medtronic Surgical Instruments for use with MAZOR X Stealth™ Edition Medtronic Surgical Instruments are intended to be used during the preparation and placement of Medtronic implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Surgical Instruments are specifically designed for use with the MAZOR X Stealth™ Edition, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Surgical Instruments can be navigated or non-navigated manual instruments that may or may not be guided through the MAZOR X Stealth™ Edition Arm Guide. Medtronic surgical drills shall only be used through the MAZOR X Stealth™ Edition arm guides, Medtronic cannulas, and Medtronic drill guides. Some of the Mectronic Surgical Instruments are also compatible with the IPC™ POWEREASE™ System or AO* style quick connect drilling motors. An instrument may incorporate a measuring function which has uses as described on the label and the instrument.
Device Story
CD Horizon Spinal System Instruments are non-sterile, reusable manual surgical instruments used for screw/implant placement during spinal surgery. Instruments integrate with Medtronic StealthStation and MAZOR X Stealth Edition systems via NavLock trackers and sterile spheres; these allow computer-assisted tracking of instrument position relative to patient anatomy. Used by surgeons in OR settings for open or minimally invasive procedures. Instruments may be used with IPC POWEREASE or manual handles. Navigation provides real-time visual feedback on surgical displays, assisting precise anatomical localization. Benefits include improved surgical accuracy and guidance during complex spinal instrumentation.
Clinical Evidence
Bench testing only. Testing included navigation accuracy verification, CAD model verification, tools package functional verification, and anatomical simulated use testing to validate design outputs against user needs.
Technological Characteristics
Reusable manual surgical instruments; materials meet national/international standards for orthopedic/neurological use. Features include NavLock tracker compatibility for optical/electromagnetic tracking. Compatible with IPC POWEREASE and AO-style drilling motors. Non-sterile delivery.
Indications for Use
Indicated for patients undergoing spinal surgery requiring preparation and placement of Medtronic screws or implants; applicable for open or minimally invasive procedures where stereotactic navigation is appropriate using CT, MR, fluoroscopy, or digitized anatomical landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
December 8, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medtronic Sofamor Danek USA, INC. Erikka Edwardsen Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
#### Re: K203291
Trade/Device Name: CD Horizon Spinal System Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 6, 2020 Received: November 9, 2020
#### Dear Erikka Edwardsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K203291
Device Name CD Horizon™ Spinal System Instruments
#### Indications for Use (Describe)
Indications for Use Medtronic Navigated Reusable Instruments for use with StealthStation™ and IPC™ POWEREASE™ Systems
Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Reusable Instruments are also compatible with the IPC™ POWEREASE™ System.
Medtronic Surgical Instruments for use with MAZOR X Stealth™ Edition
Medtronic Surgical Instruments are intended to be used during the preparation and placement of Medtronic implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Surgical Instruments are specifically designed for use with the MAZOR X Stealth™ Edition, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Surgical Instruments can be navigated or non-navigated manual instruments that may or may not be guided through the MAZOR X Stealth™ Edition Arm Guide. Medtronic surgical drills shall only be used through the MAZOR X Stealth™ Edition arm guides, Medtronic cannulas, and Medtronic drill guides. Some of the Mectronic Surgical Instruments are also compatible with the IPC™ POWEREASE™ System or AO* style quick connect drilling motors. An instrument may incorporate a measuring function which has uses as described on the label and the instrument.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
| Department of Health and Human Services |
|-----------------------------------------|
| Food and Drug Administration |
| Office of Chief Information Officer |
| Paperwork Reduction Act (PRA) Staff |
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# November 6, 2020
| I. Submitter: | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901)396-3133 |
|---------------|-----------------------------------------------------------------------------------------------------------------|
|---------------|-----------------------------------------------------------------------------------------------------------------|
| Contact: | Erikka Edwardsen<br>Sr Regulatory Affairs Specialist<br>Telephone Number: (901) 399-2485<br>Email: rikka.r.edwardsen@medtronic.com |
|----------|------------------------------------------------------------------------------------------------------------------------------------|
|----------|------------------------------------------------------------------------------------------------------------------------------------|
## II. Device:
| Proprietary Trade Name: | CD Horizon™ Spinal System |
|-------------------------|-----------------------------------------------|
| Common Name: | Stereotaxic Instrument, Navigated Instruments |
| Classification Name: | Stereotaxic Instrument |
| Regulation Numbers: | Class II (21 CFR 882.4560) |
| Classification: | Class II |
| Product Code: | OLO |
#### III. Predicates:
Predicate:
- Primary Predicate: CD Horizon™ SOLERA Instruments (K140454, S.E. . 05/22/2014)
- Additional Predicate: CD Horizon™ Spinal System Instruments (K182121, S.E. . 11/02/2018)
The predicate devices have not been subject to a design-related recall.
## IV. Product Description:
The subject CD Horizon™ Spinal System drivers are non-sterile instruments that may be used during placement of various Medtronic screws during spinal surgery. The Medtronic
{4}------------------------------------------------
instruments are made from materials commonly used in orthopedic and neurological procedures which meet available national or international standards specifications. The subject CD Horizon™ Spinal System drivers are compatible with POWEREASE, Medtronic NavLock trackers and Medtronic single-use sterile spheres to allow a Medtronic computer-assisted surgery system to track the instruments in the surgical field.
## V. Indications for Use
Medtronic Navigated Reusable Instruments for use with StealthStation™ and IPC™ POWEREASE™ Systems
Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Reusable Instruments are also compatible with the IPC™ POWEREASE™ System.
### Medtronic Surgical Instruments for use with MAZOR X StealthTM Edition
Medtronic Surgical Instruments are intended to be used during the preparation and placement of Medtronic implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Surgical Instruments are specifically designed for use with the MAZOR X Stealth™ Edition, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Surgical Instruments can be navigated or non-navigated manual instruments that may or may not be guided through the MAZOR X Stealth™ Edition Arm Guide. Medtronic surgical drills
{5}------------------------------------------------
shall only be used through the MAZOR X Stealth™ Edition arm guides, Medtronic cannulas, and Medtronic drill guides. Some of the Medtronic Surgical Instruments are also compatible with the IPC™ POWEREASE™ System or AO* style quick connect drilling motors. An instrument may incorporate a measuring function which has uses as described on the label and the instrument.
## VI. Comparison of Technological Characteristics:
The subject CD Horizon™ instruments are intended to be used during the placement of Medtronic screws during spinal surgery. The subject devices have similar designs, principle of operation, fundamental scientific technology, material, intended use, as the predicate devices and incorporate the same design features to enable navigation and compatibility and use with IPC™ POWEREASE™ System when desired.
The only change to the subject CD Horizon™ Spinal System drivers (K182121, S.E. 11/02/2018) is the expanded indications. Expanding the indications will provide surgeons additional driver options during procedures currently enabled on the StealthStation™ system. The subject drivers will have the same indications for use as the CD Horizon Solera™ drivers (K140454, S.E. 05/22/2014). The devices continue to be compatible with the Mazor X Stealth Edition, NavLock Trackers, the IPC POWEREASE System (K111520, S.E. 10/26/2011) quick connect manual handles and CD Horizon Spinal System bone screws
#### VII. Discussion of the Performance Testing:
Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. The test results show that the subject device and the predicate device are equivalent.
| Test | Description |
|--------------------------------|------------------------------------------------------------------------------------------|
| Navigation Accuracy<br>Testing | Confirmed navigated instrument accuracy |
| CAD Model Verification | Confirm the driver's CAD models are accurately reflected<br>in the application software. |
{6}------------------------------------------------
| Tools Package<br>Functional Verification<br>Testing | Provides confirmation that the Spine tools package has met<br>the required interface needs of the spine application<br>software. |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Navigated Simulated Use | Validate that the design outputs meet the requirements<br>associated with customer needs. |
| Anatomical Simulated<br>Use Testing | User needs are satisfied by performing simulated testing per<br>the instructions outlined in the surgical techniques |
## VIII. Conclusion:
Based on the supporting information provided in this pre-market notification, the subject CD Horizon™ Spinal System Drivers are substantially equivalent to CD HORIZON SOLERA Instruments (K140454, S.E. 05/22/2014)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.