Mobile ImagingRing System, IRm, Loop-X Mobile Imaging Robot, Loop-X
Applicant
Medphoton GmbH
Product Code
OWB · Radiology
Decision Date
Dec 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IRm is a mobile x-ray system to be used for 2D planar and fluoroscopic and 3D imaging for adult and pediatric patients. It is intended to be used where 2D and 3D information of anatomic structures such as bony anatomy and soft tissue and objects with high X-ray attenuation such as (metallic) implants is required. The IRm provides an interface that can be used by system integrators for integration of the IRm with image guidance systems such as surgical navigation systems.
Device Story
Mobile ImagingRing System (IRm/Loop-X) is a robotic, digital X-ray system for 2D fluoroscopy and 3D Cone-Beam Computed Tomography (CBCT). Input: X-ray projections acquired via a ring-gantry with independently movable arms (X-ray source/collimator and flat-panel detector). Output: 2D/3D images for diagnostic/interventional procedures. Used in sterile/non-sterile areas (OR, clinic) by physicians (surgeons, radiologists). Integrates with surgical navigation/therapy systems (linear accelerators, surgical robots) via DICOM or proprietary network interfaces. Internal/external cameras track surgical instruments/pointers in the imaging coordinate system. Output assists clinicians in intraoperative guidance, patient positioning, and therapy control, potentially improving clinical outcomes through precise visualization of anatomy and implants.
Clinical Evidence
No clinical data. Bench testing only. System-level validation confirmed requirements met. Imaging performance evaluated using phantoms and cadavers compared against the predicate device, demonstrating equivalent resolution and superior dynamic range.
Indicated for 2D planar, fluoroscopic, and 3D imaging in adult and pediatric patients requiring visualization of bony anatomy, soft tissue, and high-attenuation objects (e.g., metallic implants).
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
December 22, 2020
medPhoton GmbH % Daniel Schaffarzick Ouality and Regulatory Affairs Manager Karolingerstraße 16 Salzburg, Salzburg 5020 AUSTRIA
Re: K203281
Trade/Device Name: Mobile ImagingRing System, IRm, Loop-X Mobile Imaging Robot, Loop-X Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB Dated: October 30, 2020 Received: November 6, 2020
Dear Daniel Schaffarzick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K203281
Device Name Mobile ImagingRing System, IRm Loop-X Mobile Imaging Robot, Loop-X
#### Indications for Use (Describe)
The IRm is a mobile X-ray system to be used for 2D planar and fluoroscopic and 3D imaging for adult and pediatric patients. It is intended to be used where 2D and 3D information of anatomic structures such as bony and soft tissue and objects with high X-ray attenuation such as (metallic) implants is required.
The IRm provides an interface that can be used by system integration of the IRm with image guidance systems such as surgical navigation systems.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 3.375L5.75 10.875L2.75 7.875" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K203281
In accordance with the requirements of the Safe Medical Device Act, medPhoton GmbH herewith submits a 510(k) Summary.
This 510(k) summary for the Mobile ImagingRing System meets the requirements of 21 CFR 807.92.
| Date Prepared: | December 22, 2020 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Information: | medPhoton GmbH<br>Karolingerstraße 16<br>5020 Salzburg<br>Austria<br>Phone: 0043 699 161 053 86<br>Registration Number: 3012631246<br>Owner/Operator Number: 10054933 |
| Official Correspondent: | Dr. Daniel Schaffarzick<br>medPhoton GmbH<br>Karolingerstraße 16<br>5020 Salzburg<br>Austria<br>Phone: 0043 699 161 054 06 |
| US Agent (Contact): | Michael Friedl<br>Österr. AußenwirtschaftsCenter New York<br>120 West 45th Street, 9th Floor<br>New York, NY 10036<br>Phone: T +1 212 421 52 50<br>E-mail: newyork@wko.at |
| Device(s) Identification: | |
| Device Trade Name(s): | Mobile ImagingRing System, IRm,<br>Loop-X Mobile Imaging Robot, Loop-X |
| Device common name: | Interventional Fluoroscopic X-Ray System |
| Device Classification Name: | Interventional Fluoroscopic X-Ray System |
| Regulation Description: | Image-intensified fluoroscopic x-ray system |
| Product Code: | OWB |
| Reference: | per 21 CFR 892.1650 |
| Review Panel: | Radiology |
| Device Class: | 2 |
| medPhoton GmbH | Traditional 510(k) Submission | Page 1 of 9 |
|----------------|-------------------------------|-------------|
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for medPhoton. The logo consists of the letters 'm' and 'Φ' in blue, with the word 'medPhoton' in gray underneath. The 'm' is stylized with a curved line extending from the top of the letter.
### Device Description:
The Mobile ImagingRing System (IRm, "Loop-X" or "Loop-X mobile Imaging Robot") is a robotic, digital X-ray system which supports both 2D planar and fluoroscopic X-ray and 3D Cone-Beam Computed Tomography (CBCT) image acquisition and reconstruction of a patient's anatomical regions of interest (ROI).
The design of the Mobile ImagingRing is based on a ring-gantry ("ring") with two arms mounted: the first arm on gantry carries a monoblock X-ray source and a collimator, the second holds a flat-panel detector. The two arms are independently moveable to accommodate individual patient setups and off-center imaging of regions of interest.
The ImagingRing m / Loop-X Imaging Robot is intended for mobile use on patients in general and interventional radiology, in sterile and non-sterile areas. The system can be used for diagnostic and interventional purposes by volumetric or planar X-ray imaging and enables image-quided maneuvers and operations in medical fields such as general surgery. traumatology, orthopedics, neurology or radiotherapy. The system is designed for flexible examinations and treatments with patients lying, sitting or standing, with and without contrast agent. It can also be combined with therapy devices such as linear accelerators, particle beam delivery systems or other irradiation devices with active sources and patient positioning systems in Image Guided Radiation Therapy, Intraoperative Radiation Therapy, Brachytherapy or surqical navigation and assistance systems as well as in combination with surgical robots in lmage Guided Surgery, in which the image data provided by the ImagingRing IRm / Loop-X m can be used directly with patient in situ to control the respective therapy maneuvers. Using additional equipment, such as external tracking cameras or the built-in optical cameras, the position of objects such as tracked surgical instruments or pointer tools, for example, can also be determined in the imaging coordinate system. This position data, in combination with the patient's pre- and intra-interventional imaging data, can be used in guided or navigated workflows. The internal cameras of the IRm can also be used to capture pictures and film sequences of the medical procedures.
#### Intended Use / Indications for Use:
The intended use / indications for use statement has been specified as follows:
"The IRm is a mobile x-ray system to be used for 2D planar and fluoroscopic and 3D imaging for adult and pediatric patients. It is intended to be used where 2D and 3D information of anatomic structures such as bony anatomy and soft tissue and objects with high X-ray attenuation such as (metallic) implants is required. The IRm provides an interface that can be used by system integrators for integration of the IRm with image guidance systems such as surgical navigation systems."
The interface between the subject device and other systems is twofold:
- A standard DICOM interface for communication with picture archiving and . communication systems (PACS), radiological information systems (RIS) or hospital
| medPhoton GmbH | Traditional 510(k) Submission | Page 2 of 9 |
|----------------|-------------------------------|-------------|
|----------------|-------------------------------|-------------|
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for medPhoton. The logo consists of the letters 'm' and 'Φ' (phi) in blue, with the word 'medPhoton' written in gray below. The 'm' is stylized with a curved shape, and the 'Φ' is a Greek letter.
information systems (HIS) facilitates import and export of DICOM files from the subject device to an external DICOM server as well as the import of patient data (using DICOM Modality Worklist) from an external DICOM server to the subject device.
- In parallel, a proprietary navigation network-interface for communication of the subject . device with external systems (such as surgery planning and support systems or radiation therapy control systems in brachytherapy) is provided. External optical tracking systems can be used to track the position of the IRm's gantry based on passive reflectors (markers) that are attached to it. The IRm's imaging coordinate system can then be registered with an external reference coordinate system and based on this the external system can, for example, visualize tracked surgical instruments on the images provided by the IRm.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for medPhoton. The logo consists of a stylized lowercase "m" connected to a Greek letter Phi (Φ), both in blue. Below the symbols, the word "medPhoton" is written in gray, with the "m" aligned under the stylized "m" above.
### Predicate device:
Device Trade Name: Applicant: 510(k) No .: Device common name: Device Classification Name: Regulation Description: Classification Product Code: Reference: Review Panel: Device Class:
Ziehm Vision RFD 3D Ziehm Imaging GmbH K142740 Interventional Fluoroscopic X-Ray System Interventional Fluoroscopic X-Ray System Image-intensified fluoroscopic x-ray system OWB 21CFR892.1650 Radiology 2
## Reference device1:
Device Trade Name: ImagingRing System on Rails Applicant: medPhoton GmbH 510(k) No .: K191267 Device common name: ImagingRing System on Rails Device Classification Name: Accelerator, Linear, Medical Regulation Description: Medical charged-particle radiation therapy system Classification Product Code: IYE Reference: 21CFR892.5050 Review Panel: Radiology Device Class: 2
The Mobile ImagingRing System is considered substantial equivalent to the Ziehm Vision RFD 3D (K142740).
- 1. Intended use, medical application and treatment method as well as the basic parameter settings are equivalent for the Mobile ImagingRing System and the predicate device.
- 2. There is no significant difference in intended use or technology.
<sup>1</sup> As proposed in the FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" (December 27, 2011), a reference device is used to support the scientific methods used to evaluate the different technological characteristics' effect on safety and effectiveness.
| medPhoton GmbH | Traditional 510(k) Submission | Page 4 of 9 |
|----------------|-------------------------------|-------------|
|----------------|-------------------------------|-------------|
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for medPhoton. The logo features a stylized lowercase "m" connected to the Greek letter Phi (Φ) in blue. Below the symbol, the word "medPhoton" is written in gray, with "med" and "Photon" appearing on the same line.
## Summary of Technological Characteristics:
| Characteristics | Ziehm Vision RFD 3D<br>(predicate device) | Mobile ImagingRing System<br>(IRm or "Loop-X mobile<br>imaging robot")<br>(subject device) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | |
| 510(k) | K142740 | (K-number not yet assigned) |
| Manufacturer | Ziehm Imaging GmbH | medPhoton GmbH |
| Product code | OWB | OWB |
| Intended Use / Indications for Use | The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, interventional electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurology, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for Computer-Assisted Surgery procedures. The device is also intended to provide 3D medical imaging of | The IRm is a mobile x-ray system to be used for 2D planar and fluoroscopic and 3D imaging for adult and pediatric patients. It is intended to be used where 2D and 3D information of anatomic structures such as bony anatomy and soft tissue and objects with high X-ray attenuation such as (metallic) implants is required. The IRm provides an interface that can be used by system integrators for integration of the IRm with image guidance systems such as surgical navigation systems. |
| medPhoton GmbH | Traditional 510(k) Submission | Page 5 of 9 |
|----------------|-------------------------------|-------------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for medPhoton. The logo consists of the letters 'm' and 'p' connected in a stylized, cursive font, with the Greek letter 'Phi' (Φ) placed to the right of the 'p'. The text 'medPhoton' is written in a smaller, gray font below the symbol.
| | neurological,<br>intra-operative | |
|------------------------------|------------------------------------------|-----------------------------|
| | surgical procedures and where the | |
| | clinician benefits from 3D | |
| | visualization<br>complex | |
| | anatomical structures, such as but | |
| | not limited to those of high contrast | |
| | objects, and<br>bones, bones,<br>joints, | |
| | maxillofacial, cervical, thoracic, | |
| | and lumbar regions of the spine, | |
| | pelvis, acetabulum and joint | |
| | fractures of the upper and lower | |
| | extremities, and where digital | |
| | image and C-arm positioning data | |
| | is required for Computer-Assisted | |
| | Surgery procedures. | |
| | The visualization of of<br>such | |
| | anatomical structures assists the | |
| | clinician in the clinical outcome. At | |
| | the discretion of a physician, the | |
| | device may be used for other | |
| | imaging applications. This device | |
| | does not support d direct | |
| | radiographic film exposures and is | |
| | not intended for use in performing | |
| | mammography. The system is not | |
| | intended<br>for use near<br>MRI | |
| | systems. | |
| System Design | C-arm | Gantry |
| Imaging Modes | | |
| Isocentric imaging | > | く |
| Non-isocentric imaging | V | V |
| 2D planar imaging mode | V | V |
| 2D Fluoroscopic imaging mode | √ (pulsed and continuous) | V (pulsed and continuous) |
| | | |
| 3D imaging mode | V | V |
| СВСТ | イ | イ |
| X-ray System | | |
| Power Rating [kW@100kVp] | max. 25 | max. 15 |
| Voltage Range [kV] | 40-120 | 40-120 |
| Current Range [mA] | 0.2-250 | 5-80 |
| Pulse length [ms] | 4-40 | 2-35 |
| Tube Type | Rotating anode | Rotating anode |
| Anode Material | Tungsten rhenium molybdenum | Tungsten rhenium molybdenum |
| Fixed Filtration | ≥ 4.3 mm Al eq. | 4.4 mm Al eq. |
| Image Detector | | |
| medPhoton GmbH | Traditional 510(k) Submission | Page 6 of 9 |
|----------------|-------------------------------|-------------|
|----------------|-------------------------------|-------------|
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for medPhoton. The logo consists of a stylized lowercase 'm' in blue, followed by a Greek letter Phi, also in blue. Below these symbols, the word "medPhoton" is written in gray, with the 'm' aligned under the stylized 'm' above.
| Detector Technology | Amorphous silicon flat panel detector (aSi) | Amorphous silicon flat panel detector (aSi) |
|-----------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Panel Size [cm²] | 30 x 30 | 43.2 x 43.2 |
| Max. detector resolution [pixels] | 1536 x 1536 | 2880 x 2880 |
| User Interface | | |
| Remote Control Panels (incl. radiation buttons, radiation warning lamps and emergency stop buttons) | ✓ | ✓ |
| Wired footswitch | ✓ | ✓ |
| Other Components | | |
| Anti-Scatter Grids | ✓ | ✓ |
| Beam Limiter / Collimator | ✓ | ✓ |
| Lasers | ✓ (class 2M) | ✓ (class 1) |
| Monitor(s) | ✓ | -- |
| Collision protection | ✓ | ✓ |
| Electrical requirements | | |
| Input specification | 100-240 VAC (± 10%), 50/60 Hz | 120-230 VAC (± 10%), 50/60 Hz |
| Environmental Conditions | | |
| Operation | | |
| Temperature [°C] | 10 - 35 | 15 - 32 |
| Humidity [%] (non-condensing) | 20 - 70 | 30 - 60 |
| Atmospheric Pressure [mbar] | 790 - 1060 | 800 - 1100 |
| Storage | | |
| Temperature [°C] | -5 - 55 | 0 - 40 |
| Humidity [%] (non-condensing) | 20 - 70 | 10 - 75 |
| Atmospheric Pressure [mbar] | 790 - 1060 | 750 - 1100 |
| Transport | | |
| Temperature [°C] | -5 - 55 | -10 - 50 |
| Humidity [%] (non-condensing) | 20 - 70 | 5 - 75 |
| Atmospheric Pressure [mbar] | 790 - 1060 | 750 - 1100 |
| Other system specifications | | |
| Opposing source detector distance (SDD) [cm] | 104.5 | 126 |
| Max. rotational speed [°/s] | 15 | 16 (7 for certain IGRT applications) |
Table 1: Comparison of technological characteristics of IRS and the predicate XVI R5.0
| medPhoton GmbH |
|----------------|
| |
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image contains a logo for a company called medPhoton. The logo features the letters 'm' and 'p' in a stylized, cursive font, with the 'p' resembling the Greek letter phi (Φ). The letters are in blue, and the word "medPhoton" is written in gray below the symbol.
Like the listed predicate device, the medPhoton Mobile ImagingRing System (IRm) is a mobile x-ray system intended to be used for 2D fluoroscopic and 3D imaging (adult and pediatric patients). Both devices are intended to be used where 2D and 3D information of anatomic structures such as bony anatomy and objects with high X-ray attenuation such as (metallic) implants is required. Also, both devices are prepared for (equipped with respective interfaces) integration with image guidance systems, such as surgical navigation systems.
The technological characteristics (materials, energy sources, other features) of the subject device are not significantly different - i.e. can be considered the same - as those of the predicate device.
The device design of the subject device and the predicate device (gantry vs. C-arm) is significantly different, however, this difference raises no new questions of safety and effectiveness. The existing questions of safety and effectiveness have been addressed and answered in the submission of the reference device (K191267) which has the same design as the subject device. The Mobile ImagingRing System does not introduce new potential hazards or safety risks.
Since the subject device can therefore be considered to be as safe and effective as the predicate device and since they also have the same intended use, medPhoton is certain that the devices are substantially equivalent.
## Summary of performance testing:
#### Bench Testing:
Non-clinical bench performance testing has been conducted to further support substantial equivalence of the subject device and the predicate device. The main aspects in determining substantial equivalence are the intended use of the devices, their technological characteristics as well as safety and effectiveness of the devices.
A system level validation has been conducted, which provides evidence that the Mobile ImagingRing System (IRm, Loop-X Mobile Imaging Robot, Loop-X) meets all customer and system requirements and is able to perform according to its intended use.
Clinical use of the device in general and its clinical imaging performance in particular have been tested using phantoms and cadavers. Those clinical imaging tests were conducted with the subject device as well as the predicate device. Results suggest that the subject device performs at least as good as the predicate device (in terms of resolution capabilities) or better (in terms of a larger dynamic range) and therefore support the demonstration of substantial equivalence.
Although safety and effectiveness of the subject device's design have been demonstrated for a
| medPhoton GmbH | Traditional 510(k) Submission | Page 8 of 9 |
|----------------|-------------------------------|-------------|
|----------------|-------------------------------|-------------|
{11}------------------------------------------------
Image /page/11/Picture/2 description: The image shows the logo for medPhoton. The logo features the letters 'm' and 'p' in a stylized, connected design. The 'm' is rendered in a flowing, cursive style, while the 'p' is represented by the Greek letter Phi, which is in blue. Below the symbol, the word 'medPhoton' is written in a sans-serif font, with 'med' in gray and 'Photon' in a slightly darker shade of gray.
reference device (with the same design), additional bench testing pertaining to a variety of different mechanical characteristics of the device design were conducted. Results demonstrate the safety and effectiveness of the subject device in terms of its design.
Voluntary consensus standards applied during design and development include but are not limited to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54, IEC 62304, IEC 62366 and ISO 14971. Standard compliance has been demonstrated in a variety of tests conducted internally (by medPhoton) and by an external accredited testing laboratory.
In summary, the results of all non-clinical bench performance tests support substantial equivalence of the subject device and the predicate device and demonstrate safety and effectiveness of the subject device.
#### Animal Testing:
medPhoton GmbH did not perform any animal testing for the Mobile ImagingRing System.
#### Clinical Testing:
medPhoton GmbH did not perform any clinical testing for the Mobile ImagingRing System.
Conclusion: medPhoton GmbH believes that the Mobile ImagingRing System is substantially equivalent to the currently legally marketed devices. The Mobile ImagingRing System does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.
| medPhoton GmbH | Traditional 510(k) Submission | Page 9 of 9 |
|----------------|-------------------------------|-------------|
|----------------|-------------------------------|-------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.