Veuron-Brain-mN1 is intended for use in the post-acquisition image enhancement of 3T MR images of the brain acquired through a 3D gradient-echo sequence. When used in combination with other clinical information, the Veuron-Brain-mN1 application may aid the qualified radiologist with diagnosis by providing enhanced visualization of tissue structures with magnetic susceptibility contrasts in brain 3T MR images.
Device Story
Veuron-Brain-mN1 is post-processing software for 3T brain MR images. It imports DICOM data from 3D gradient multi-echo sequences; performs automatic reconstruction of susceptibility map-weighted images (SMWI) using magnitude and Quantitative Susceptibility Mapping (QSM) data; and outputs 3D SMWI images for visualization. Used in hospital settings by radiologists to assist in visual evaluation, manipulation, and assessment of brain pathologies. The software provides enhanced contrast of tissue structures, aiding clinical decision-making by improving diagnostic clarity. It is a non-ML, rule-based image processing tool.
Clinical Evidence
Bench testing only. Performance evaluated using phantom testing across a range of susceptibility values and clinical images. Metrics included Contrast-to-Noise Ratio (CNR) and Signal-to-Noise Ratio (SNR). Testing conducted on 3T MRI scanners (Siemens and Philips). All unit and integration tests met predefined acceptance criteria.
Technological Characteristics
Post-processing software for 3D gradient multi-echo 3T MRI. Inputs: DICOM data. Outputs: 3D SMWI images. Connectivity: USB export. Standards: ISO 14971, ISO 62304, IEC 62366. Algorithm: Rule-based (non-ML).
Indications for Use
Indicated for use in qualified radiologists to aid in the diagnosis of brain pathologies by enhancing visualization of tissue structures with magnetic susceptibility contrasts in 3T MR images acquired via 3D gradient-echo sequences.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 12, 2022
Heuron Co., Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth St. NW WASHINGTON DC 20004
Re: K203279
Trade/Device Name: Veuron-Brain-mN1 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LLZ Dated: June 10, 2022 Received: June 10, 2022
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k203279
Device Name Veuron-Brain-mN1
#### Indications for Use (Describe)
Veuron-Brain-mN1 is intended for use in the post-acquisition image enhancement of 3T MR images of the brain acquired through a 3D gradient-echo sequence. When used in combination with other clinical information, the Veuron-Brain-mN1 application may aid the qualified radiologist with diagnosis by providing enhanced visualization of tissue structures with magnetic susceptibility contrasts in brain 3T MR images.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Neuron logo, which is a combination of a stylized brain-like symbol and the word "neuron" in a combination of blue and gray. Below the logo, the text "510(k) summary" is present. To the right of the logo and text, the number "K203279" is displayed.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Submitter information Submitter name: Heuron Co., Ltd. Address: 10th Floor, 7, Mirae-ro, Namdong-gu, Incheon, 21558 Republic of Korea Phone/Fax: +82-32-429-8508 / +82-32-429-8507 Correspondent Contact: Hogan Lovells US LLP (John J. Smith M.D., J.D., john.smith(@hoganlovells.com), +1-202-637-3638, 555 Thirteenth Street, NW, 20004 USA on behalf of Heuron Co., Ltd. Date of preparation: May 05, 2022
Device information Proprietary name(s): Veuron-Brain-mN1 Common name: Image Processing Software Classification name: System, Nuclear Magnetic Resonance Imaging per 21 CFR 892.1000 Product code: LNH, LLZ Classification panel: Radiology Device class: II
## Device description
Veuron-Brain mN1 is a post-processing software intended to provide visualization, manipulation and reconstruction capabilities, including susceptibility map-weighted images, of 3D gradient multi echo brain 3T MR images. The Veuron-Brain-mN1 aids in the clinical analysis of brain structures from MR images.
- 3D gradient multi echo images are acquired at a specific orientation through the anatomy of interest.
## SMWI Algorithm
- Susceptibility map-weighted imaging (SMWI) are reconstructed from the magnitude and Quantitative Susceptibility Mapping (QSM) images of each echo.
- Algorithm output is a 3D SMWI image. ·
- . The software algorithms are not based on machine learning.
## Predicate device
SPIN-SWI application software (SpinTech, Inc. K173224, Feb 23, 2018)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "Heuron". The logo consists of a stylized graphic to the left of the word "heuron". The graphic is made up of a series of interconnected circles, with one circle colored in blue and the rest in gray. The word "heuron" is written in a rounded, sans-serif font, with the "h" in blue and the rest of the letters in gray.
## Indications for Use
Veuron-Brain-mN1 is intended for use in the post-acquisition image enhancement of 3T MR images of the brain acquired through a 3D gradient-echo sequence. When used in combination with other clinical information, the Veuron-Brain-mN1 application may aid the qualified radiologist with diagnosis by providing enhanced visualization of tissue structures with magnetic susceptibility contrasts in brain 3T MR images.
### Comparison of indications for use statement
The following table compares indications for use statements between the Veuron-Brain-mN1 and the predicate device (i.e. SPIN-SWI application software). Similar to the predicate device, the Veuron-Brain-mN1 is used for managing patient and case base data, collection, analysis and display of the medical images. The device assists the clinician with the visual evaluation, manipulation, and assessment of pathologies derived from brain scans.
| Veuron-Brain-mN1 | SPIN-SWI |
|----------------------------------------------|------------------------------------------------|
| Subject device (K203279) | Predicate device (K173224) |
| Veuron-Brain-mN1 is intended for use in the | The SpinTech, Inc. SPIN- SWI application is |
| post-acquisition image enhancement of 3T MR | intended for use in the post-acquisition image |
| images of the brain acquired through a 3D | enhancement of MRI acquired 3D gradient- |
| gradient-echo sequence. When used in | echo images of the brain. When used in |
| combination with other clinical information, | combination with other clinical information, |
| the Veuron-Brain-mN1 application may aid the | the SPIN-SWI application may aid the |
| qualified radiologist with diagnosis by | qualified radiologist with diagnosis by |
| providing enhanced visualization of tissue | providing enhanced visualization of structures |
| structures with magnetic susceptibility | containing venous blood such as cerebral |
| contrasts in brain 3T MR images. | venous vasculature. |
# Summary of the technological characteristics compared to the predicate device and new device.
The software is similar in uses and application to the predicate device. As stated in the comparison table provided below, both the device and predicates are used to assist the clinician with the visual evaluation, manipulation, and assessment of pathologies derived from brain scans.
| Comparison item | Veuorn-Brian-mN1<br>Subject device (K203279) | SpinTech, Inc.<br>SPIN-SWI application software<br>Predicate device (K173224) |
|-----------------|----------------------------------------------|-------------------------------------------------------------------------------|
| Target user | Qualified radiologist | Qualified radiologist and<br>technologist |
| Anatomical site | Brain | Brain |
| Where used | Hospital | Hospital |
{5}------------------------------------------------
Heuron Co., Ltd. 10th Floor, 7, Mirae-ro, Namdong-gu Incheon, 21558 Republic of Korea Tel. +82-32-429-8508 / Fax. +82-32-429-8507
Image /page/5/Picture/1 description: The image shows the Neuron logo. The logo consists of a stylized neuron symbol on the left and the word "neuron" on the right. The neuron symbol is made up of interconnected circles, with one circle colored in blue. The word "neuron" is written in a rounded sans-serif font, with the "h" in blue and the rest of the letters in gray.
| Design features | • Import DICOM data | • Import DICOM data | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| | • Perform automatic post-processing | • Perform automatic post-processing | |
| | • Provide the user confirmation | • Provide the user confirmation | |
| | • Export the resulting data only<br>through USB | • Export the resulting data through<br>data network | |
| Non-clinical<br>performance | • Bench testing performed to the<br>test functionality of the system. | • Bench testing performed to the test<br>functionality of the system. | |
| Standards | • ISO 14971 | • ISO 14971 | |
| | • ISO 62304 | • ISO 62304 | |
| | • IEC 62366 | | |
| SW verification and<br>validation | Tested in accordance with<br>verification and validation process<br>and planning. The testing results<br>support that all the system<br>requirements have met their<br>acceptance criteria and are<br>adequate for its intended use. | Tested in accordance with<br>verification processes and<br>planning, The testing results<br>support that all the system<br>requirements have met their<br>acceptance criteria and are<br>adequate for its intended use. | |
| | Compatible input data<br>format and modality | • DICOM | • DICOM |
| | | • MRI (Siemens Healthcare, Philips,<br>3.0T) | • MRI |
| | | Input patient data | Manual through keyboard / mouse |
| | Output patient data<br>format | DICOM | • Unknown |
| | Study list functionality | Search, Importing and Exporting | Search, Importing and Exporting |
### Non-clinical study performance
Bench testing is done to show that the system is suitable for its intended use. In order to evaluate the safety of the product, unit tests and integration tests were performed, and all results met the acceptance criteria. The device performance was evaluated using phantom testing representing the range of susceptibility values in the brain tissue. Additionally, the device performance was also evaluated on clinical images using CNR/SNR metrics. The scanner models used for the testing were Siemens Healthcare's and Philips's and field strength was 3T. The predefined acceptance criteria were met to demonstrate substantial equivalence to the predicate.
#### Conclusion
By virtue of its intended use, design features, and technological characteristics, Veuron-BrainmN1 is substantially equivalent to a device that has been approved for marketing in the United States. The non-clinical performance data shows that Veuron-Brain-mN1 is as safe and effective as the predicate device without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.