← Product Code [KDN](/productcode/KDN) · K203262

# Cannula for Organ Perfusion (K203262)

_Bridge TO Life, Ltd. · KDN · Jul 1, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203262

## Device Facts

- **Applicant:** Bridge TO Life, Ltd.
- **Product Code:** [KDN](/productcode/KDN.md)
- **Decision Date:** Jul 1, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5880
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Cannula for Organ Perfusion is a single-use device indicated for hypothermic flushing and replacement of residual blood in donor organs with a legally marketed organ preservation solution at the time of organ removal from the donor during the preparation of these organs for transplantation. This device is indicated for use with gravity flow; no testing has been provided for use with pump systems, such as mechanical perfusion devices.

## Device Story

Cannula for Organ Perfusion is a single-use, sterile, non-toxic, nonpyrogenic device used during organ transplantation procedures. It functions as a conduit for flushing and replacing residual blood in donor organs with organ preservation solution. The device features a tapered proximal connector and an atraumatic distal tip; it is available in multiple sizes (8F to 34F) to accommodate different organs. Surgeons select the appropriate size and perform the procedure using gravity flow. The device is intended for use during the organ removal and preparation process. It facilitates organ preservation by ensuring effective flushing, which is critical for maintaining organ viability for transplantation. No mechanical pump systems are used with this device.

## Clinical Evidence

No clinical data. Bench testing only. Testing included tensile strength, leakage, and tightness in accordance with ISO 10555-1, and flow rate verification. Biocompatibility testing was performed per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, blood compatibility, and rabbit pyrogenicity.

## Technological Characteristics

Single-use, sterile cannula. Materials comply with EN ISO 10993-1. Sterilization via Ethylene Oxide (ETO) to SAL 10^-6. Design includes tapered proximal connector and atraumatic distal tip. Sizes range from 8F to 34F. Operates via gravity flow. No electronic components or software.

## Regulatory Identification

An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.

## Predicate Devices

- Cannula Accessory of the LifePort Kidney Perfusion Transporter ([K021362](/device/K021362.md))
- Fresenius Multiorgan Perfusion Cannula ([K902394](/device/K902394.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

July 1, 2021

Bridge to Life Ltd. % H. Carl Jenkins Regulatory Affairs Counsel Wood Burditt Group 10 E. Scranton Ave., Ste. 201 Lake Bluff, IL 60044

Re: K203262

> Trade/Device Name: Cannula for Organ Perfusion Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated Kidney Perfusion and Transport System and Accessories Regulatory Class: II Product Code: KDN, MSB Dated: May 27, 2021 Received: May 28, 2021

Dear H. Carl Jenkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203262

Device Name

Cannula for Organ Perfusion

#### Indications for Use (Describe)

Cannula for Organ Perfusion is a single-use device indicated for hypothermic flushing and replacement of residual blood in donor organs with a legally marketed organ preservation solution at the time of organ removal from the donor during the preparation of these organs for transplantation. This device is indicated for use with gravity flow; no testing has been provided for use with pump systems, such as mechanical perfusion devices.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| XI Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "The Wood Burditt Group". The words "the", "WOOD", "BURDITT", and "group" are arranged in a stacked format. A teal-colored curved line is present on the right side of the text, adding a design element to the logo.

The Wood Burditt Group LLC 10 E. Scranton Ave., Suite 201 Lake Bluff, IL 60044 847. 234. 7500 (tel.) 847. 574. 0728 (e-fax) www.woodburditt.com

# K203262 -- 510(k) Summary

Date prepared: June 16, 2021

| Submitter / Contact Person | H. Carl Jenkins<br>The Wood Burditt Group<br>10 E. Scranton Ave, Suite 201<br>Lake Bluff, IL 60044<br><br>(ph) 847-234-7500 x 205<br>(fax) 847-578-0728<br>(email) hcjenkins@woodburditt.com |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Applicant & 510(k) Owner | Bridge to Life Ltd.<br>128 Suber Rd., Suite A<br>Columbia, SC 29210<br>(ph) 847-796-3070<br><br>FDA Establishment Registration Number: 3009110022<br>FEI Number: 3009110022<br>Owner/Operator Number: 10038889 |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

#### Device Information

| Trade Name           | Cannula For Organ Perfusion                                    |
|----------------------|----------------------------------------------------------------|
| Proprietary Name     | Cannula For Organ Perfusion                                    |
| Common Name          | Organ Perfusion Cannula                                        |
| Classification Name  | Isolated Kidney Perfusion and Transport System and Accessories |
| Classification Panel | Gastroenterology/Urology                                       |
| Regulation           | 21 CFR 876.5880                                                |
| FDA Product Code     | KDN; MSB                                                       |
| FDA Classification   | II                                                             |

#### Reason for 510(k) Submission

The applicant device is a new device.

{4}------------------------------------------------

### Device Description

The Cannula for Organ Perfusion is a device designed and intended as a component to an organ perfusion and transport system / organ preservation solution, used in an organ transplantation procedure. The device is a cannula (available in several different sizes) intended for flushing and replacement of residual blood in donor organs with organ preservation solution during the preparation processes of these organs for transplantation. The cannulas are single use, sterile, non-toxic and nonpyrogenic devices. As a component to an organ perfusion and transport system / organ preservation solution, the Cannula for Organ Perfusion is also a Class II medical device, subject to Product Code KDN and MSB, 21 CFR 876.5880: "This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set." (21 CFR 876.5880(a)).

Image /page/4/Figure/3 description: The image shows a medical instrument with several labeled parts. The instrument is a long, thin tube with a blue flexible connector on one end. The other end has an athraumatic tip, and there are also fixed and slidable rings along the tube's length, labeled d1, d2, and d3 respectively.

The Cannulas come in multiple sizes, according to the French catheter scale, including 8F, 10F, 12F, 16F (adjustable to 18F with a slidable ring), 20F (adjustable to 28F with a slidable ring), 25F (adjustable to 34F with a slidable ring). In addition, there is a configuration of the cannula that is intended for "back table" procedures, which is available in a 16F size.

| Cannula size | d1 mm (F)  | d2 mm (F)   | d3 mm (F)   |
|--------------|------------|-------------|-------------|
| 8 F          | 2.7 (8 F)  | 4.7 (14 F)  | -           |
| 10 F         | 3.3 (10 F) | 5.3 (16 F)  | -           |
| 12 F         | 4.0 (12 F) | 6.3 (19 F)  | -           |
| 16 F         | 5.3 (16 F) | 7.3 (22 F)  | 8.7 (26 F)  |
| 20 F         | 6.7 (20 F) | 9.3 (28 F)  | 10.7 (32 F) |
| 25 F         | 8.3 (25 F) | 11.3 (34 F) | 13.3 (40 F) |

{5}------------------------------------------------

The surgeon performing the organ perfusion procedure will select the appropriate cannula for the particular organ to be perfused.

## Indications for Use:

Cannula for Organ Perfusion is a single-use device indicated for hypothermic flushing and replacement of residual blood in donor organs with a legally marketed organ preservation solution at the time of organ removal from the donor during the preparation of these organs for transplantation. This device is indicated for use with gravity flow; no testing has been provided for use with pump systems, such as mechanical perfusion devices.

#### Predicate / Reference Device Summary Table

Based on the comparison of the device features, materials, intended use and performance the Cannula for Organ Perfusion was shown to be substantially equivalent to the lawfully marketed predicate devices indicated in the table below.

| Device                                                         | Applicant                    | 510(k) # | Featured Element(s) of Substantial Equivalence         |
|----------------------------------------------------------------|------------------------------|----------|--------------------------------------------------------|
| Cannula Accessory of the LifePort Kidney Perfusion Transporter | Organ Recovery Systems, Inc. | K021362  | Intended Use / Indications for Use                     |
| Fresenius Multiorgan Perfusion Cannula                         | Fresenius                    | K902394  | Intended Use / Indications for Use; Technology; Design |

#### Technological Characteristics and Comparison:

The Cannula for Organ Perfusion is substantially equivalent to the cannula accessory of the LifePort Kidney Perfusion Transporter-specifically, the perfusion cannula component of the LifePort Kidney Perfusion Transporter-in terms of intended use, indications for use, FDA classification and FDA Product Code. Like the subject device, the cannula accessory of the LifePort Kidney Perfusion Transporter is used for flushing and replacement of residual blood in donor organs with organ preservation solution during the preparation processes of these organs for transplantation.

In addition, a comparison of the subject device to the Fresenius Multiorgan Perfusion Cannula demonstrates the safety and efficacy of the Cannula for Organ Perfusion in terms of substantially equivalent design, technology, use and suitability.

{6}------------------------------------------------

| Feature                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate / Reference Devices                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use / Indications | Cannula for Organ Perfusion<br>is a single-use device<br>indicated for hypothermic<br>flushing and replacement of<br>residual blood in donor organs<br>with a legally marketed organ<br>preservation solution at the<br>time of organ removal from<br>the donor during the<br>preparation of these organs for<br>transplantation. This device is<br>indicated for use with gravity<br>flow; no testing has been<br>provided for use with pump<br>systems, such as mechanical<br>perfusion devices. | LifePort Kidney Perfusion<br>Transporter (K021362):<br>System component cannula for organ<br>perfusion, intended to be used in an<br>organ transplantation procedure.<br>Fresenius Multiorgan Perfusion<br>Cannula (K902394):<br>"For placement in the [organ] of a<br>heart beating cadaver to allow in situ<br>flushing of the cadaver organs with<br>organ preservation solution." |
| Single Use / Disposable    | Single Use / Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Fresenius Multiorgan Perfusion<br>Cannula (K902394):<br>Single Use / Disposable                                                                                                                                                                                                                                                                                                       |
| Device Design              | The Cannula for Organ Perfusion<br>has a tapered connector at<br>proximal end of the cannula, and<br>an atraumatic distal end; cone<br>shaped.                                                                                                                                                                                                                                                                                                                                                     | Fresenius Multiorgan Perfusion<br>Cannula (K902394):<br>Perfusion cannula has a proximal end<br>which consists of a funnel-shaped<br>connector (i.e., a cone shape)<br>Perfusion cannula has a distal end<br>which consists of an atraumatic tip<br>and cone-shaped cuff.                                                                                                             |
| Sterilization              | Ethylene Oxide<br>SAL 1 x 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Fresenius Multiorgan Perfusion<br>Cannula (K902394):<br>Ethylene Oxide<br>SAL 1 x 10-6                                                                                                                                                                                                                                                                                                |
| Regulation                 | 21 CFR 876.5880                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LifePort Kidney Perfusion<br>Transporter (K021362):<br>21 CFR 876.5880                                                                                                                                                                                                                                                                                                                |
| FDA Product Code           | KDN; MSB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | LifePort Kidney Perfusion<br>Transporter (K021362): KDN                                                                                                                                                                                                                                                                                                                               |
| FDA Classification         | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | LifePort Kidney Perfusion<br>Transporter (K021362): II<br>Fresenius Multiorgan Perfusion<br>Cannula (K902394): II                                                                                                                                                                                                                                                                     |

{7}------------------------------------------------

#### Clinical Testing:

No clinical tests were conducted in support of this 510(k) submission.

## Non-Clinical Tests Performed:

Bench testing demonstrates that the Cannula for Organ Perfusion is safe, effective and meets device specifications:

- Testing related to tensile strength vielded Pass results in accordance with ISO 10555-1. -
- Testing related to leakage and tightness yielded Pass results in accordance with ISO -10555-1.
- Testing related to flow rate demonstrated adequate flow of liquid in accordance with ISO -10555-1.

#### Sterilization & Shelf Life:

Product is sterilized using ETO, with a Sterility Assurance Level (SAL) of 10-6.

The shelf life of the product has been established in accordance with real time aging testing results.

#### Biocompatibility:

All component materials of this medical device carry adequate quality certificates confirming their biological, physical-chemical and mechanical properties, and they are all intended for ETO sterilization.

In particular, the materials used in the manufacturing of the cannula comply with requirements of EN ISO 10993-1.

{8}------------------------------------------------

In accordance with ISO 10993-1, the Cannula for Organ Perfusion is classified as:

- surgical invasive medical device
- having external contact with the body through the circulating blood,
- remaining in contact with the body for up to 24 hours.

The following biological tests were selected for the Cannula for Organ Perfusion:

- cytotoxicity
- sensitizing effect
- intracutaneous reactivity
- systemic (acute) toxicity
- blood compatibility
- rabbit pyrogenicity

In accordance with ISO 10993-1:2009, all tests were performed using the final product after the sterilization process of the product, using a validated sterilization process.

Additionally, the manufacturer has conducted a separate biological evaluation of the subject medical device. This study included the defined requirements, product classification, description of the adopted test selection criteria, specifications and approvals of materials used to manufacture the evaluated product, a list and summary of biological tests collected or performed and analysis of the data obtained from the evaluation.

#### Conclusion:

The applicant device is substantially equivalent in its intended use, technology / principal of operation, materials, and performance to the predicate and reference devices identified in this 510(k) submission. There is no significant difference that raises any issues of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K203262](https://fda-staging.innolitics.com/device/K203262)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
