K203250 · Our United Corporation · IWB · Mar 12, 2021 · Radiology
Device Facts
Record ID
K203250
Device Name
TaiChiC
Applicant
Our United Corporation
Product Code
IWB · Radiology
Decision Date
Mar 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The system is a teletherapy device intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions in the brain and head where radiation treatment is indicated.
Device Story
TaiChiC is a stereotactic radiotherapy system using Cobalt-60 rotary focusing technology; delivers radiation to brain/head lesions based on a prescribed plan. System components: gantry with slip ring, focusing gamma ray treatment head, image guidance subsystem (CBCT, kV-kV), patient couch, power supply, and control console. Operated by clinicians in a clinical setting. Integrates gamma beam therapy with image-guided radiotherapy (IGRT) to enhance precision. Healthcare providers use interface software to monitor treatment and control radiation delivery. Benefits include precise targeting of intracranial lesions while sparing surrounding healthy tissue. Device uses 18 Cobalt-60 sources with higher activity to achieve dose rates comparable to predicate systems with more sources.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed conformance to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-11, IEC 60601-1-3, IEC 60601-2-68, IEC 61217, IEC 62274, IEC 62366-1, IEC 60601-1-6, ISO 15223-1, and ISO 10993-1. Hardware and software verification/validation performed per 21 CFR 820 and ISO 13485.
Technological Characteristics
Radionuclide radiation therapy system using Cobalt-60 sources. Features gantry with slip ring, 7 sets of collimators (6mm-35mm), and CBCT/kV-kV imaging. Dimensions/form factor include treatment bore and patient couch. Connectivity includes control console interface. Software level of concern: major. Sterilization: not applicable (non-invasive). Standards: IEC 60601-1, IEC 60601-2-11, IEC 60601-2-68, ISO 10993-1.
Indications for Use
Indicated for patients with lesions, tumors, and conditions in the brain and head requiring stereotactic radiosurgery or precision radiotherapy.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
March 12, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Our United Corporation % Mr. Qi Liu Regulatory Affairs Engineer Room 10301, 3rd Floor, Unit 1, Block 28, ShouChuang International Business Center, No.66 Xi'an, Shaanxi 710018 CHINA
Re: K203250
Trade/Device Name: TaiChiC Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: Class II Product Code: IWB Dated: February 3, 2021 Received: February 10, 2021
Dear Mr. Qi Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203250
Device Name TaiChiC
Indications for Use (Describe)
The system is a teletherapy device intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions in the brain and head where radiation treatment is indicated.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification [510(k)] Summary OUR UNITED CORPORATION TaiChiC Rotating Gamma System
The following information is provided following the format of 21 CFR 807.92.
# I. GENERAL INFORMATION
K203250
| Submitter's Name: | OUR UNITED CORPORATION |
|-------------------|----------------------------------------------------------|
| | Room 10301, 3rd Floor, Unit 1, Block 28, Shou Chuang |
| | International Business Center, No.66 Fengcheng 12 Road, |
| | Xi'an Economic & Technological Development Zone, 710018, |
| | Shaanxi, China. |
| | |
| Contact Person: | Qi Liu |
|-----------------|-----------------------------|
| | Phone: +86 15389012257 |
| | Email: qi.liu@ourunited.com |
Date Prepared: Oct. 30th 2020
## II. DEVICE INFORMATION
| Proprietary Name: | TaiChiC |
|----------------------|---------------------------------------|
| Common/Usual Name: | Rotating Gamma System |
| Classification Name: | Radionuclide radiation therapy system |
| Device Regulation: | 21 CFR 892.5750 |
| Regulatory Class: | II |
| Product Code: | IWB |
## III. PREDICATE DEVICE
| Primary Device | Akesis Galaxy RTx (K200050) |
|----------------------|---------------------------------------|
| Classification Name: | Radionuclide radiation therapy system |
| Device Regulation: | 21 CFR 892.5750 |
| Regulatory Class: | II |
{4}------------------------------------------------
| Product Code: | IWB |
|----------------------|---------------------------------------------------|
| Reference Device | TaiChiA Medical Linear Accelerator (K193207) |
| Classification Name: | Medical charged-particle radiation therapy system |
| Device Regulation: | 21 CFR 892.5050 |
| Regulatory Class: | II |
| Product Code: | IYE |
### IV. Device Description:
TaiChiC Rotating Gamma System has been designed to deliver radiation treatments in accordance with a prescribed plan. This system is a stereotactic radiotherapy equipment developed by applying the principle and mature technology of Cobalt-60 rotary focusing. It is suitable for brain and head lesions and tumors requiring radiotherapy, and provides image guided stereotactic radiosurgery and precision radiotherapy.
TaiChiC consists of a gantry with a slip ring, a focusing gamma ray treatment head, image guidance subsystem, couch, power supply and electrical cabinet, main cover and treatment bore etc. TaiChiC combines gamma beam stereotactic radiation therapy technology with contemporary advanced image guidance technology (IGRT) to provide image guided stereotactic radiotherapy and precision radiotherapy.
## v. Intended Use:
The system is a teletherapy device intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions in the brain and head where radiation treatment is indicated.
### vi. Indications for Use:
The system is a teletherapy device intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions in the brain and head where radiation treatment is indicated.
{5}------------------------------------------------
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
TaiChiC and predicate Akesis Galaxy RTx are both Radionuclide radiation therapy system. falling within 21 CFR 892.5750, product code IWB. The proposed TaiChiC is comparable to the predicate device with respect to the intended use, indications for use, technological characteristics and similar performance specifications.
The TaiChiC is also appropriately comparable to another device, TaiChiA Medical Linear Accelerator, cleared under 510(k) K193207. TaiChiA Medical Linear Accelerator is classified as a Medical charged-particle radiation therapy system, falling within 21 CFR 892.5050, product code IYE. This reference device is included as it is similar to TaiChiC with respect to indications for use, imaging techniques, gantry rotation, patient immobilization, couch and other features.
With regards to technological characteristics, the TaiChiC Rotating Gamma System, the predicate device and the reference device all have similar features and components. TaiChiC and the predicate device have same intended use, and both fall within the broader scope of the reference device. TaiChiC and Akesis Galaxy RTx both utilize Cobalt-60 radiation source to generate the treatment beam. All three systems have patient supporting system (couch) to support and position the patient during the treatment. All three systems have control consoles and interface software to control and monitor the systems. All three systems protect the operator and the general public from the radiation from the machine.
The main differences between TaiChiC and the predicate device include the number of sources used, the image guidance and the number of the collimator sets.
Comparing with the predicate device with 30 radioactive sources, TaiChiC has only 18 sources but with higher activities to provide a comparable dose rate at the focal spot, which is appropriate for the intended use.
The CBCT and kV-kV imaging techniques used in TaiChiC are similar as those in the reference device, while the predicate device does not provide this technique.
TaiChiC provides 7 sets of collimators with beam diameter at the focal spot of 6mm, 9mm, 12mm, 16mm, 20mm, 25mm and 35mm. While the predicate device provides 4 sets of collimators ¢ 4 mm, Φ 8 mm, Φ 14 mm and Φ 18 mm. The first four sets of TaiChiC collimators ( d 6mm, o 12mm, φ16mm) are comparable to those of the predicate device and all the collimators of TaiChiC fall within the broad beam collimating sizes of the reference device.
The above differences among the proposed device, the predicate and the reference device do not raise any safety and effectiveness issues.
The comparisons between the proposed TaiChiC Rotating Gamma System and the predicate device Akesis Galaxy RTx, the reference device TaiChiA Medical Linear Accelerator are provided in the table below.
{6}------------------------------------------------
| N<br>O. | Feature | New Device | Predicate<br>Device<br>(K200050) | Reference Device<br>(K193207) | Analysis of<br>Differences |
|---------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Product<br>Name | OUR TaiChiC Rotating<br>Gamma System. | Akesis Galaxy RTx | OUR TaiChiA Medical<br>Linear<br>Accelerator | |
| 2 | Intended<br>Use | TaiChiC is a<br>teletherapy device<br>intended to provide<br>stereotactic<br>radiosurgery and<br>precision radiotherapy<br>for lesions, tumors, and<br>conditions in the brain<br>and head where<br>radiation treatment is<br>indicated. | The Akesis Galaxy<br>RTx system is a<br>teletherapy device<br>intended for the<br>stereotactic<br>irradiation of head<br>structures. | TaiChiA is a Medical<br>Linear<br>Accelerator, intended to<br>provide stereotactic<br>radiosurgery and<br>precision radiotherapy for<br>lesions,<br>tumors, and conditions<br>anywhere in<br>the body where radiation<br>treatment<br>is indicated. | The new device has<br>same intended use<br>as the Predicate<br>device,<br>and has same<br>intended use as the<br>Reference device<br>for treatment of<br>lesions in the brain<br>and head |
| 3 | Indications<br>for Use | ThiChiC is a<br>teletherapy device<br>intended to provide<br>stereotactic<br>radiosurgery and<br>precision radiotherapy<br>for lesions, tumors, and<br>conditions in the brain,<br>and head, where<br>radiation treatment is<br>indicated. | The Akesis Galaxy<br>RTx system is a<br>teletherapy device<br>intended for the<br>stereotactic<br>irradiation of head<br>structures. | TaiChiA, is a Medical<br>Linear Accelerator,<br>intended to provide<br>stereotactic radiosurgery<br>and precision radiotherapy<br>for lesions, tumors, and<br>conditions anywhere in<br>the body where radiation<br>treatment is indicated. | The new device has<br>same intended use<br>as the Predicate<br>device,<br>and has same<br>intended use as the<br>Reference device<br>for treatment of<br>lesions in the brain<br>and head |
| 4 | Radiation<br>source | Cobalt 60 | Cobalt 60 | 6MV X-ray | The new device<br>radiation source is<br>the same with the<br>Predicate Device |
| 5 | Initial Total<br>activity | 22860 Ci ±5% | 6000 Ci ±10% | Treatment beam 6 MV<br>nominal photon beam<br>energy. | The new device<br>total cobalt-60<br>activity at loading is<br>higher than the<br>Predicate Device,<br>but the dose rate at<br>the focal spot is at<br>the same level. |
| 6 | Number of<br>radiation<br>sources | 18 | 30 | Treatment beam 6 MV<br>nominal photon beam<br>energy. | The radioactive<br>sources of the new<br>device have higher<br>activities, and so<br>less radiation<br>sources are needed<br>compared with<br>Predicate device |
| | | Initial Dose<br>rate at the<br>focal spot | $\textgreater 3.0Gy / min$ | Up to 14 Gy / min | Radiation dose rate<br>at focal spot at<br>loading for the new<br>device is very<br>comparable to the<br>Predicate device<br>which is<br>appropriate for the<br>intended use. |
| 8 | Number of<br>collimator<br>sizes<br>(and<br>nominal<br>aperture<br>size) | 7:<br>$φ$ 6mm, $φ$ 9mm,<br>$φ$ 12mm, $φ$ 16mm,<br>$φ$ 20mm, $φ$ 25mm<br>and $φ$ 35mm | 4:<br>$φ$ 4 mm, $φ$ 8 mm,<br>$φ$ 14 mm, $φ$ 18 mm | NA | The new device<br>utilizes 7 sets of<br>collimator cones to<br>collimate the<br>Gamma rays, the<br>collimator sizes of<br>the new device<br>collimator sizes<br>( $φ$ 6mm, $φ$ 9mm,<br>$φ$ 12mm, $φ$ 16mm)<br>are similar with the<br>Predicate device. |
| 9 | Radiologic<br>accuracy | $\textless 0.5mm$ | $\textless 0.5mm$ | $\textless 1.0mm$ | The new device<br>radiologic accuracy<br>is the same as the<br>Predicate device |
| 10 | Patient<br>Couch | Yes | Yes | Yes | All systems have a<br>patient couch to<br>support patient<br>during treatment. |
| 11 | Imaging<br>techniques | CBCT、kV-kV | NA | CBCT、kV-kV | The new device<br>imaging techniques<br>is the same with the<br>Reference device. |
| 12 | Patient<br>immobilization<br>method | The Encompass™ SRS | Stereotactic Head<br>Frame and<br>Positioning System | The Encompass™ SRS<br>Immobilization System | The new device<br>lesion localization<br>falls within the<br>range of the<br>predicate, and same<br>as the reference<br>device. |
{7}------------------------------------------------
{8}------------------------------------------------
#### V. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence:
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and electromagnetic compatibility (EMC) testing were conducted on this medical device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
#### Bench Test
Successful testing was performed in accordance with following standards:
IEC 60601-2-11: 2013, particular requirements for the basic safety and essential performance of gamma beam therapy equipment
IEC60601-1-3:2013, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic Xray equipment
IEC60601-2-68:2014, Medical electrical equipment - Part 2-68: particular requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment
IEC61217:2011, Radiotherapy equipment - Coordinates, movements and scales
IEC62274 :2005, Medical electrical equipment - Safety of radiotherapy record and verify systems
IEC62366-1:2015, Medical devices - Part 1: Application of usability engineering to medical devices
IEC60601-1-6:2013, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
RT2:2017 Radiation therapy readiness check
ISO 15223-1: 2016. Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements
ISO10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
### Hardware and Software Verification and Validation Testing
Hardware and software verification and validation process were conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards.
{9}------------------------------------------------
Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
#### Animal or clinical test
Not applicable for TaiChiC
#### Summary
The verification and validation testing demonstrated that TaiChiC has met its specifications, demonstrated substantially equivalent performance to the predicate and met its intended use.
#### VI. CONCLUSIONS
The TaiChiC has the same intended use, indications for use, fundamental scientific technology as the predicate device.
The difference among TaiChiC and predicate does not raise any new safety or effectiveness issues. The differences are substantial equivalent to the reference device
The Verification and Validation demonstrates that the device is as safe and effective as the predicate device.
Based on the comparison and analysis above, OUR therefore believes that TaiChiC is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.