AccelX™ Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations. All digitally designed abutments for use with AccelX™ CAD-CAM Abutments are intended to be sent to an Integrated Dental Systems validated milling center for manufacture. AccelX™ Abutments are compatible with MegaGen AnyRidge Internal Implant System components as listed below.
Device Story
AccelX™ Abutments are dental implant components used to support prosthetic restorations; compatible with MegaGen AnyRidge Internal Implant System. Components include cover screws, healing abutments, temporary abutments, prepable straight/angled abutments, castable base abutments, multi-unit abutments, and CAD-CAM titanium base/blank abutments. Used in dental clinics by dentists/prosthodontists. CAD-CAM abutments require digital design sent to a validated milling center. Titanium base abutments are two-piece (titanium base + zirconia coping) cemented with CemPlant cement. Output is a stable prosthetic foundation for the patient, restoring chewing function.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: static compression and compression fatigue testing per ISO 14801, biocompatibility testing per ISO 10993-5/12, and sterilization validation per ISO 17665.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136), Co-Cr-Mo alloy (ASTM F1537), Zirconia (ISO 13356). Interface: Internal connection. Sterilization: Moist heat (SAL 10^-6). CAD-CAM workflow: Digital design to validated milling center. Mechanical testing: ISO 14801.
Indications for Use
Indicated for partially or fully edentulous patients requiring prosthetic support for dental restorations (crowns, bridges, overdentures) in the maxillary or mandibular arch. Compatible with MegaGen AnyRidge Internal Implant System.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
Integrated Dental Systems LLC % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K203240
Trade/Device Name: AccelXTM Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 29, 2021 Received: March 29, 2021
## Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203240
#### Device Name
AccelXTM Abutments
#### Indications for Use (Describe)
AccelX™ Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.
All digitally designed abutments for use with AccelX™ CAD-CAM Abutments are intended to be sent to an Integrated Dental Systems validated milling center for manufacture.
AccelX™ Abutments are compatible with MegaGen AnyRidge Internal Implant System components as listed below.
| Implant System Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|---------------------------------------------|---------------------------|----------------------|
| MegaGen AnyRidge Internal<br>Implant System | 4.0 | 3.5 |
| | 4.4 | 3.5 |
| | 4.9 | 3.5 |
| | 5.4 | 3.5 |
| | 5.9 | 3.5 |
| | 6.4 | 5.0 |
| | 6.9 | 5.0 |
| | 7.4 | 5.0 |
| | 7.9 | 5.0 |
| 8.4 | 5.0 | |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K203240 AccelX™ Abutments Integrated Dental Systems, LLC
April 27, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Integrated Dental Systems, LLC<br>145 Cedar Lane<br>Englewood, New Jersey 07631 | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| | Telephone | +1 917-693-5595 | |
| Official Contact | David Singh, Chief Operating Officer | | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | |
| | Telephone | +1 858-792-1235 | |
| | Fax | +1 858-792-1236 | |
| | Email | kthomas@paxmed.com<br>flarson@paxmed.com | |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | AccelX™ Abutments |
|------------------------|---------------------------------------------------------------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K191986, DESS Dental Smart Solutions, Terrats Medical SL Additional Predicate Devices K123988, AnyOne™ Internal Implant System, MegaGen Implant Company, Limited K182448, AnyRidge Octa 1 Implant System, MegaGen Implant Co., Ltd. K140091, Xpeed AnyRidge Internal Implant System, MegaGen Implant Co., Ltd. K140728, Dental Implants and Abutments, Ditron Precision Ltd.
{4}------------------------------------------------
K191986 is the primary predicate for support of substantial equivalence in terms of Indications for Use, CAD-CAM abutments, and referenced sterilization and biocompatibility data. The additional predicates K123988 and K182448 are for support of substantial equivalence in terms of component designs. The predicate K140091 is for support of substantial equivalence in terms of the OEM compatibilities. The additional predicate K140728 is for referenced sterilization and biocompatibility data.
## INDICATIONS FOR USE STATEMENT
AccelX™ Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.
All digitally designed abutments for use with AccelX™ CAD-CAM Abutments are intended to be sent to an Integrated Dental Systems validated milling center for manufacture.
AccelX™ Abutments are compatible with MegaGen AnyRidge Internal Implant System components as listed below.
| Implant System Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|---------------------------------------------|---------------------------|----------------------|
| MegaGen AnyRidge Internal<br>Implant System | 4.0 | 3.5 |
| | 4.4 | 3.5 |
| | 4.9 | 3.5 |
| | 5.4 | 3.5 |
| | 5.9 | 3.5 |
| | 6.4 | 5.0 |
| | 6.9 | 5.0 |
| | 7.4 | 5.0 |
| | 7.9 | 5.0 |
| | 8.4 | 5.0 |
## SUBJECT DEVICE DESCRIPTION
AccelX™ Abutments comprises conventional and CAD-CAM abutments, abutment screws, and other associated components compatible with MegaGen AnyRidge Internal Implant System.
AccelX™ conventional abutments and prosthetic components include Cover Screws. Healing Abutments. Temporary Abutments, Prepable Straight Abutments, Angled Abutments, Castable Base Abutments, and Multi-Unit Abutments. All conventional abutments and prosthetic components are manufactured from Ti-Al-4V alloy conforming to ASTM F136 except the Castable Base Abutments that are manufactured from Co-Cr-Mo alloy conforming to ASTM F1537.
AccelX™ CAD-CAM abutments include Titanium Base Abutments for use with a zirconia coping, and Titanium Blank Abutments to be milled to a final one-piece abutment. Titanium Base Abutments are twopiece abutments composed of the titanium base component and a zirconia top-half, which comprises the final abutment once the two components are cemented together using CemPlant Implant Cement. The Titanium Base Abutments and Titanium Blank Abutments are manufactured from Ti-Al-4V alloy conforming to ASTM F136. The zirconia copings for the Titanium Base Abutments will be manufactured
{5}------------------------------------------------
from zirconia conforming to ISO 13356. All digitally designed abutments for use with AccelX™ CAD-CAM abutments are intended to be sent to an Integrated Dental Systems validated milling center for manufacture. The AccelX™ abutments and prosthetic components are summarized in the following table.
| | Coronal Ø, mm | Angle | Material |
|---------------------------------------------------------------------------------------|------------------------------|--------------|--------------------------------------------------|
| Cover Screws | 3.5 | n/a | Titanium alloy, ASTM F136 |
| Healing Abutments | 4, 5, 6, 7, 8, 10 | n/a | Titanium alloy, ASTM F136 |
| | Prosthetic Platform Ø, mm | | |
| Temporary Abutments | 4 | n/a | Titanium alloy, ASTM F136 |
| Prepable Straight Abutments | 4, 5, 6, 7;<br>4.5, 5.5, 6.5 | n/a | Titanium alloy, ASTM F136 |
| Angled Abutments | 4, 5, 6, 7 | 15°, 25° | Titanium alloy, ASTM F136 |
| Castable Base Abutment - Co-Cr-Mo Alloy | 3.8 | 0° | Co-Cr-Mo alloy, ASTM F1537 |
| Multi-Unit Abutments | 4.8 | 0°, 17°, 30° | Titanium alloy, ASTM F136 |
| Titanium Base Abutments | 4.0, 4.4 | 0° (coping) | Titanium alloy, ASTM F136<br>Zirconia, ISO 13356 |
| Titanium Blank Abutments | 6.9 (or per available space) | Up to 30° | Titanium alloy, ASTM F136 |
| Abutment Multi Post Screw<br>Multi-Unit Abutment Screw<br>Multi-Unit Prosthetic Screw | n/a | n/a | Titanium alloy, ASTM F136 |
The subject device abutments and healing components are compatible with MegaGen AnyRidge Internal System components (K140091).
## PERFORMANCE DATA
The subject device was evaluated and tested as recommended in the FDA guidance documents Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments (issued May 12, 2004), and Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (issued September 4, 2020).
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: moist heat (steam) sterilization according to ISO 17665-1, and ISO 17665-2 (included in this submission and referenced from K140728 and K191986); biocompatibility data for the ASTM F136 titanium alloy (included in this submission and referenced from K191986), for the ASTM F1537 Co-Cr-Mo alloy (referenced from K191986), and included in this submission for the ISO 13356 zirconia material used to fabricate copings for Titanium Base Abutments. Testing to demonstrate compatibility to the OEM system included reverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws, and static compression and compression fatigue testing according to ISO 14801.
No clinical data were included in this submission.
{6}------------------------------------------------
### EQUIVALENCE TO MARKETED DEVICES
The subject devices are substantially equivalent in indications and design principles to the primary predicate device and the additional predicate devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and a table comparing the technological characteristics of the subject device, the primary predicate device, and the additional predicate devices.
The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K191986 and to the additional predicate devices. Slight differences in the language of the IFUS do not affect the intended use as dental implant abutments for support of a prosthesis to restore chewing function.
Minor differences between the IFUS for the subject device and the primary predicate device K191986 include differences in tradenames, differences in validated milling centers, differences in the compatible OEM implant systems, and K191986 is limited to abutments for CAD-CAM technology. These minor differences do not raise new questions of safety or effectiveness because both IFUS express equivalent intended use.
Minor differences between the IFUS for the subject device and the additional predicate devices K123988. K182448, and K140091 include: the subject device IFUS includes the term "single-unit or multi-unit prosthetic restorations" and the predicate devices use alternative terminology "Crown, bridges, and overdentures": the subject device IFUS includes references to CAD-CAM technology to fabricate patientspecific abutments, validated milling centers, and compatible OEM implant systems and the additional predicate IFUS does not include these elements; and differences in tradenames. These minor differences do not raise new questions of safety or effectiveness because all IFUS express equivalent intended use.
#### Subject Device Abutments
The following subject device designs are substantially equivalent to the predicate device K123988: Cover Screws, Healing Abutments, Temporary Abutments, Prepable Straight Abutments, and Angled Abutments. These subject device prosthetic components and the corresponding predicate device prosthetic components are all manufactured from titanium allov and are compatible with the MegaGen AnyRidge Internal Implant System. The subject device Cover Screws have a coronal diameter 3.5 mm, and provided in cuff heights ranging from 1 mm to 2.8 mm. The predicate device cover screws have a coronal diameter 3.5 mm, and provided in cuff heights ranging from 0.8 mm to 2.6 mm. Both the subject device and predicate device are threaded to match the internal thread of the implant. The subject device Healing Abutments and the predicate device healing abutments have a coronal diameter ranging from 4 mm to 10 mm, cuff heights ranging from 3 mm to 7 mm, and are threaded to match the internal thread of the implant. The subject device Temporary Abutments and the predicate device temporary abutments have a 2 mm cuff height, a 4 mm prosthetic platform diameter, and a 10 mm prosthetic post. The subject device Prepable Straight Abutments and the predicate device EZ Post Abutments have a cuff height ranging from 2 mm to 5 mm, prosthetic platform diameters ranging from 4 mm to 7 mm, and a prosthetic post of 5.5 mm or 7 mm. The subject device Angled Abutments and predicate device angled abutments have a cuff height ranging from 2 mm to 5 mm, prosthetic platform diameters ranging from 4 mm to 7 mm, a prosthetic post of 7 mm, and an angle of 15° or 25°.
The subject device Multi-Unit Abutments designs are substantially equivalent to the predicate device K 182248. The subject device and predicate device abutments are for multi-unit restorations, have a
{7}------------------------------------------------
prosthetic diameter of 4.8 mm, include angulations of 0°, 17°, and 30°, have an internal implant interface connection, and are manufactured from titanium alloy. The subject device and predicate device abutments with angulation of 0° are provided in cuff heights ranging from 1.3 mm. The subject device abutments with angulation of 17° are provided with cuff heights ranging from 2.5 mm. The subject device abutments with angulation of 30° are provided with cuff heights 3.5 mm and 4.5 mm. The predicate device abutments with angulation of 17° and 30° are provided with cuff heights ranging from 2.3 mm to 4.3 mm. The subject device Multi Unit Abutment Cylinder (coping) is substantially equivalent in material and design to the predicate device cylinder (coping) cleared in K182248.
The subject device Castable Base Abutment – Co Cr Mo Alloy design is substantially equivalent to the predicate device K182248. The subject and predicate device abutments are bases for a final abutment fabricated by the traditional cast-to (lost wax) method with non-precious alloys, without angulation. The subject and predicate devices have a prosthetic diameter of 3.8 mm.
The subject device Titanium Base Abutments designs are substantially equivalent to the predicate K123988. The subject device Titanium Base Abutments have a prosthetic post height of 4.5 mm or 6 mm, with a cut out section in one side of the post to allow for an angled access to the screw channel, and have a cuff height of 0.6 mm or 1.5 mm. The predicate device titanium base abutments have a prosthetic post height of 4.5 mm, 6.0 mm, and 8.0 mm, with a cut out section in one side of the post to allow for an angled access to the screw channel, and have a cuff height of 0.6 mm, 1.5 mm, 3.0 mm, or 4.0 mm. Both the subject device and predicate device titanium base abutments are two-piece abutments composed of a titanium base component and a zirconia top-half, which comprises the final abutment once the two components are cemented together.
The subject device Titanium Blank Abutments are substantially equivalent to the predicate K191986. The overall (milling) diameter of the subject device is 10 mm or 12 mm, and the overall (milling) diameter of the predicate device is 10 mm and 14 mm. Both the subject device and predicate device are manufactured from titanium alloy and are compatible with MegaGen AnyRidge Internal Implant System.
All subject device prosthetic components are provided non-sterile and are to be sterilized to a sterility assurance level (SAL) of 10-6 by the end user using the same sterilization method (moist heat) and parameters for devices previously cleared. Sterilizations for the subject devices were performed according to ISO 17665-1 and ISO TR 17665-2.
The subject device Cover Screws, Healing Abutments, Angled Abutments, Multi Unit Abutments, and Titanium Base Abutments are contract manufactured by Ditron Dental Ltd. for Integrated Dental Systems. The sterilization method and parameters for these subject devices are the same as the predicate device K140728. Therefore, no new sterilization was performed because the subject device components are substantially equivalent to the predicate device K140728.
The subject device Temporary Abutments are contract manufactured by Terrats Medical SL for Integrated Dental Systems. The sterilization method and parameters for the Subject device Temporary Abutments are the same as the predicate device K191986. Therefore, no new sterilization was performed because the subject device components are substantially equivalent to the predicate device K191986.
The subject device Prepable Abutments are contract manufactured by La Precision for Integrated Dental Systems. Sterilization method and parameters were validated according to ISO 17665-1 and ISO TR 17665-2. Similarly, the sterilization method and parameters for subject device Castable Base Abutment –
{8}------------------------------------------------
Co-Cr-Mo Alloy and Titanium Blank Abutments with zirconia superstructure were validated according to ISO 17665-1 and ISO TR 17665-2.
The subject device prosthetic components were evaluated for biocompatibility or tested according to ANSI/AAMI/ISO 10993-5 and ANSI/AAMI/ISO 10993-12.
The subject device Cover Screws, Healing Abutments, Angled Abutments, Multi Unit Abutments, Multi-Unit Abutment Cylinder, and Titanium Base Abutments are contract manufactured by Ditron Dental Ltd. for Integrated Dental Systems. These subject devices are manufactured from titanium alloy, conforming to ASTM F136, in the same facilities using the same manufacturing processes and same materials as used for Ditron Precision components previously cleared in K140728. Confirmatory biocompatibility testing of the titanium alloy was performed according to ISO 10993-5 and ISO 10993-12.
The subject device Temporary Abutments, Titanium Blank abutments, and Castable Base Abutment -Co-Cr-Mo Alloy are contract manufactured by Terrats Medical SL for Integrated Dental Systems. Temporary Abutments and Titanium Blank Abutments are manufactured from titanium alloy conforming to ASTM F136. The Castable Base Abutment – Co-Cr-Mo Alloy are manufactured from cobalt chromium alloy conforming to ASTM F1537. These subject devices are manufactured from the same materials, in the same facilities using the same manufacturing processes as used for Terrats Medical component previously cleared in K191986. Therefore, no new biocompatibility testing was performed, as the subject device Temporary Abutments. Titanium Blank abutments, and Castable Base Abutment - Co-Cr-Mo Alloy are substantially equivalent to the predicate device K191986 with regards to materials and processing.
The subject device Prepable Abutments are contract manufactured by La Precision for Integrated Dental Systems. Prepable Abutments are manufactured from titanium alloy conforming to ASTM F136. Confirmatory biocompatibility testing was tested according to ANSI/AAMI/ISO 10993-5 and ANSI/AAMI/ISO 10993-12 because these subject devices are not manufactured from the same materials, in the same facilities using the same manufacturing processes as a predicate device.
Additional cytotoxicity testing was conducted on the subject device Titanium Base Abutments with zirconia superstructure.
Mechanical performance testing of the subject device was performed in conformance to ISO 14801. The fatigue limit data demonstrated that constructs of the subject device abutments, including abutments fabricated to the limits stated in the proposed labeling, in combination with previously-cleared MegaGen AnyRidge Internal Implant System implants have sufficient strength for their intended use.
Minor differences in the designs, dimensions, sizes, or compatible OEM implant lines among the subject device, the primary predicate device, and the additional predicate devices do not affect substantial equivalence. These minor differences do not impact substantial equivalence because these differences are related to the compatible OEM implant designs or are mitigated by the mechanical performance testing.
### CONCLUSION
The subject device, the primary predicate device, the additional predicate devices, and the reference device have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device, the primary predicate, the additional predicate devices, and the
{9}------------------------------------------------
reference device encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{10}------------------------------------------------
# Table of Substantial Equivalence– Indications for Use Statements
| | | Indications for Use Statement | | |
|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|----------------------|
| Subject Device | | | | |
| K203240<br>AccelX™ Abutments<br>Integrated Dental Systems LLC | | AccelX™ Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support.<br>All digitally designed abutments for use with AccelX™ CAD-CAM Abutments are intended to be sent to an Integrated Dental Systems validated.<br>AccelX™ Abutments are compatible with MegaGen AnyRidge Internal Implant System components as listed below. | | |
| | | Implant System Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
| | MegaGen AnyRidge Internal Implant System | 4.0 | 3.5 | |
| | | 4.4 | 3.5 | |
| | | 4.9 | 3.5 | |
| | | 5.4 | 3.5 | |
| | | 5.9 | 3.5 | |
| | | 6.4 | 5.0 | |
| | | 6.9 | 5.0 | |
| | | 7.4 | 5.0 | |
| | | 7.9 | 5.0 | |
| | | 8.4 | 5.0 | |
| Primary Predicate Device | | | | |
| K191986<br>DESS Dental Smart Solutions<br>Terrats Medical SL | | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular.<br>All digitally designed custom abutments for use with Ti Base abutments or Pre-milled (Blank) abutments are to be sent to a Terrats Medical validated. | | |
| | Compatible Implant System | Implant Body Diameter, mm | Implant Platform | |
| | Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm | |
| | | 3.6 | 2.9 mm | |
| | Astra Tech EV | 4.2 | 3.5 mm | |
| | | 4.8 | 4.1 mm | |
| | | 3.0 | 3.0 mm | |
| | Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 mm | |
| | | 4.5/5.0 | 4.5/5.0 mm | |
| | | 3.25 | 3.45 mm | |
| | Biomet 3i Certain® | 4.0 | 4.1 mm | |
| | | 5.0 | 5.0 mm | |
| | | 3.25 | 3.4 mm | |
| | Biomet 3i OSSEOTITE® | 3.75, 4.0 | 4.1 mm | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.