The Invacare® Platinum® 5NXG Oxygen Concentrator is intended for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.
Device Story
Oxygen concentrator extracts nitrogen from ambient air to provide supplemental oxygen; output concentration 87-96% at 1-5 LPM flow rates. Used in home, long-term care, or non-acute facilities; operated by patients or caregivers. Device connects to 120VAC/60Hz power; delivers oxygen via nasal cannula. Includes flowmeter for flow adjustment and elapsed time meter for usage tracking. Provides oxygen purity fault alarms if concentration drops below 82%. Healthcare providers use device to manage respiratory disorders; output supports patient lifestyle in non-acute settings. Not life-sustaining.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via electrical safety (AAMI ES60601-1, IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6, IEC 62366), alarm systems (IEC 60601-1-8), and biocompatibility (ISO 18562-1/2/3). Oxygen output and purity verified against USP monograph and ISO 80601-2-69 standards.
Technological Characteristics
Oxygen concentrator; nitrogen extraction via molecular sieve. 120VAC/60Hz power. Dimensions: 24x16x12 inches; weight 33 lbs. Output: 1-5 LPM, 87-96% purity. Connectivity: None. Standards: ISO 80601-2-69, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO 14971, ISO 18562-1/2/3. Software: Moderate level of concern.
Indications for Use
Indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates < 1 L/min, use with Invacare Pediatric Flowmeter accessory. Not intended to sustain or support life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the words "U.S. Food & Drug Administration."
April 12, 2021
Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035
Re: K203210
Trade/Device Name: Invacare Platinum 5NXG Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: March 10, 2021 Received: March 11, 2021
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801 ): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-ind-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd D. Courtney -S
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### Device Name
Invacare® Platinum® 5NXG Oxygen Concentrator
Indications for Use (Describe)
The Invacare® Platinum® 5NXG Oxygen Concentrator is indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
{3}------------------------------------------------
#### 510(k) Summary
## SUBMITTER per 21 CFR 807.92(a)(1):
Invacare Corporation One Invacare Way Elyria, OH 44035 Phone: (440) 329-6840 Email: ewreh@invacare.com
#### CONTACT PERSON:
Elijah Wreh Regulatory Affairs Manager
## MANUFACTURER:
Invacare Corporation 1200 Taylor Street Elyria, OH 44035
#### Date Prepared per 21 CFR 807.92(a)(1):
March 8, 2021
#### DEVICE INFORMATION per 21 CFR 807.92(a)(2)
| Name of Device: | Invacare® Platinum® 5NXG Oxygen Concentrator | |
|-----------------------|----------------------------------------------------------------------------------------------------------------|--|
| Common or Usual Name: | Portable Oxygen Generator | |
| Classification Name: | Generator, Oxygen, Portable 21 CFR 868.5440 | |
| Regulatory Class: | 2 | |
| Product Code: | CAW | |
| PREDICATE DEVICE: | Invacare® Perfecto2™V Oxygen Concentrator (K200890) | |
| Patient Population: | Patients with respiratory disorders requiring supplemental<br>oxygen at flow rates of 1 to 5 liters per minute | |
{4}------------------------------------------------
# DEVICE DESCRIPTION per 21 CFR 808.92(a)(4)
The subject Invacare® Platinum® 5NXG Oxygen Concentrator device is used by patients who require supplemental oxygen. The device is not intended to sustain or support life. The oxygen concentration level of the output gas ranges from 87% to 96% at an output flow rate range of 1 to 5 liters per minute (LPM). The oxygen is delivered to the user through the use of a nasal cannula.
The concentrator provides supplemental oxygen to patients with respiratory disorders, in the home, long term care facility or other non-acute patient care facilities. It is not intended to sustain or support life. It is to be used indoors and is intended to support the patient's lifestyle within the home or care facility. The device operates at a nominal 120VAC/60 Hertz supply.
## INTENDED USE per 21 CFR 807.92(A)(5)
The Invacare® Platinum® 5NXG Oxygen Concentrator is intended for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.
## INDICATIONS FOR USE per FORM FDA 3881
The Invacare® Platinum® 5NXG Oxygen Concentrator is indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.
# INDICATIONS FOR USE (IFU) COMPARISON
The Indications for Use statement for the subject device is the same as the previously cleared predicate device.
{5}------------------------------------------------
## COMPARISON of TECHNOLOGICAL CHARACTERISTICS with the PREDICATE DEVICE per 21 CFR 807.92(a)(6)
The device comparison shows that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device. The subject and predicate devices use the same technology to extract nitrogen from air. The subject device components met the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise different questions of safety and effectiveness. The different technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device.
## BASIS of SUBSTANTIAL EQUIVALENCE per 21 CFR 807.100(b)(2)(ii)(A)
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)/' and the technological characteristics which include materials, design, and other device related features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).
The differences between the new and predicate devices are as a result of updating the pneumatic system, complying to the latest version of ISO 80601-2-69, updating the general structure, and upgrading the operating system, hardware and firmware. The sum of the changes can be assessed using the same consensus standards and acceptance criteria that were used for the predicate device.
The design verification testing and device comparison below demonstrates that the subject device is substantially equivalent to the predicate device. The data generated from the subject device design verification test reports support a finding of substantial equivalence regarding device function and performance and comparison of device specifications and characteristics supports they are substantially equivalent.
{6}------------------------------------------------
# Design Characteristics Comparison
| Design and Technological Characteristics | Subject Device | Predicate Device |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Device Name | Invacare® Platinum® 5NXG Oxygen Concentrator | Invacare® Perfecto2™V Oxygen Concentrator (K200890) |
| Manufacturer | Invacare Corporation | Same |
| Intended Use | The Invacare® Platinum® 5NXG oxygen concentrator is<br>intended for patients with respiratory disorders requiring<br>supplemental oxygen at flow rates of 1 to 5 liters per minute. For<br>flow rates below 1 liter per minute use with Invacare Pediatric<br>Flowmeter accessory. It is not intended to sustain or support life. | Same |
| Indications for Use | The Invacare® Platinum® 5NXG oxygen concentrator is<br>indicated for patients with respiratory disorders requiring<br>supplemental oxygen at flow rates of 1 to 5 liters per minute. For<br>flow rates below 1 liter per minute use with Invacare Pediatric<br>Flowmeter accessory. It is not intended to sustain or support life. | Same |
| Type of Use | Prescription (Rx Only) | Same |
| Patient Population | Patients with respiratory disorders requiring supplemental<br>oxygen at flow rates of 1 to 5 liters per minute | Same |
| Oxygen Outlet | Port supplying produced concentrated oxygen | Same |
| Flowmeter | Adjusts oxygen flow to prescribed setting | Same |
| Maximum Flow Rate | 5.0 L/min | Same |
| Adjustable Flow Range | 1 to 5 L/min in 0.5 L/min increments | Same |
| Elapsed Time Meter | Displays the cumulative number of hours the unit has operated | Same |
| Oxygen Purity Fault | If oxygen purity falls below 82% but above 50% | If oxygen purity falls below 85% (±2%) but above 73% (±3%) |
| Power Indicator | Indicates Device Power Status On/Off | Same |
| Rated Current Input | 3.0 A | 3.2 A |
| Electrical Requirements | 120 VAC ±10%, 60 Hz | Same |
| Oxygen Output Concentration Levels | 87% to 96% at all flow rates | 95.6% to 87% at all flow rates |
| Weight | 33 ± 2 lbs. | 40 lbs. |
| Height | 24 ± 3/8" | 23 ± 3/8" |
| Width | 16 ± 3/8" | 15 ± 3/8" |
| Depth | 12 ± 3/8" | Same |
{7}------------------------------------------------
## DESIGN VERIFICATION TESTING DATA
Design verification testing was performed using the same consensus standards and acceptance criteria that were used for the predicate device to demonstrate a finding of substantial equivalence. The subject device met the performance requirements and is substantially equivalent to the predicate device and do not raise any new questions of safety and effectiveness.
## Risk Management
Risk Management has been conducted in accordance with ISO 14971:2012 - Medical Devices -Application of Risk Management to Medical Devices for the subject Invacare® Platinum® 5NXG Oxygen Concentrator. The risk-based assessment concluded that the subject device's risk profile remained unchanged and there is no significant impact on the safety or effectiveness of the subject device. The Risk Management File was updated to our current format.
# Software Verification Testing
Software Verification Testing, including testing per ISO 80601-2-69 on the finished product, was performed to evaluate the functionality of the subject device, including alarms and indicators. Software verification testing was conducted on the subject device as recommended by the FDA's
guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and IEC 62304:2006 – Medical device software – Software Life Cycle Processes.
Level of Concern: The Level of Concern for the subject device is Moderate.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety and Electromagnetic Compatibility (EMC) testing was performed to confirm post-change functionality of the subject device. This includes the following:
- . AAMI ES60601-1: Medical Electrical Equipment - Part 1 : General Requirements for Basic Safety and Essential Performance
- IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements for Basic . Safety & Essential Performance
- ISO 80601-2-69 Medical electrical equipment Part 2: Particular Requirements for the . Basic Safety and Essential Performance of Oxygen Concentrator Equipment
{8}------------------------------------------------
- IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General Requirements for . Basic Safety and Essential performance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- . IEC 60601-1-6: Medical Electrical Equipment Part 1-6 General Requirements for Safety - Collateral Standard: Usabilitv
- . IEC 62366: Medical Devices – Application of Usability Engineering to Medical Devices
- . IEC 60601-1-11: Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential performance – Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- . IEC 60601-1-8: Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
In summary, the sum of the changes can be assessed using the same consensus standards and acceptance criteria that were used for the predicate device and returned substantially the same results as the predicate device demonstrating that the subject device is equivalent in electrical and electromagnetic safety performance to the predicate device.
# Biocompatibility Testing
Biocompatibility testing was performed on the subject device components in accordance with FDA guidance document entitled "Use of International Standard ISO 18562-1, "Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare Applications- Part 1 : Evaluation and Testing within a Risk Management Process." The battery of testing was performed to the following FDA recognized consensus standards:
- ISO 18562-1:2017, Biocompatibility Evaluation of Breathing Gas Pathways in .
- Healthcare Applications Part 1: Evaluation and Testing within a Risk Management .
- . Process
- . ISO 18562-2:2017, Biocompatibility Evaluation of Breathing Gas Pathways in
- Healthcare Applications Part 2: Tests for Emissions of Particulate Matter .
- ISO 18562-3:2017. Biocompatibility Evaluation of Breathing Gas Pathways in .
- Healthcare Applications Part 3 : Tests for Emissions of Volatile Organic Compounds .
Testing to detect inorganic gases including Carbon Monoxide (CO) and Carbon Dioxide (CO2), Oxygen (O2) and odour was performed on the subject device. The objective of the test was to assess CO. CO2. O2. and odour and compare their PPM (parts per million) for eas biocompatibility according to the requirements of the "Oxygen 93 %" USP monograph directly to patients.
{9}------------------------------------------------
## Performance Testing – Bench
The Invacare® Platinum® 5NXG Oxygen Concentrator is tested to recognized consensus standards for basic safety and essential performance, which can be found in Section "Electromagnetic Compatibility and Electrical Safety." Additional respiratory testing has been performed on the subject device to demonstrate a finding of substantial equivalence to the predicate device. The battery of testing includes:
- . Power and oxygen output performance testing
- . ISTA 2A
- Weight .
- Sound Testing .
ISO 80601-2-69 Medical electrical equipment Part 2: Particular Requirements for the Basic Safety and Essential Performance of Oxygen Concentrator Equipment In summary, the testing demonstrates that the subject device meets performance specifications and requirements and is substantially equivalent to the predicate device identified throughout this 510(k) submission. The data generated from the performance testing demonstrated that the
subject device meets all the performance specifications and requirements and is substantially equivalent to the predicate device.
# Animal Study
Animal testing is not required for this submission.
# Clinical Testing
Clinical testing is not required for this submission.
# CONCLUSIONS per 21 CFR 807.92(b)(3)
The subject device has the same intended use and technological characteristics as the predicate device. The design verification results support a finding of substantial equivalence of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.