The Navigated Everest Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants. The Navigated Everest Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The Navigated Everest Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Everest Spinal Fixation System using the Stryker Spine Navigation System. The Navigated Mesa Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants. The Navigated Mesa Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The Navigated Mesa Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Mesa Spinal Fixation System using the Stryker Spine Navigation System. The Navigated Yukon Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants. The Navigated Yukon Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The Navigated Yukon Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Yukon Spinal Fixation System using the Stryker Spine Navigation System. The Navigated Xia 4.5- Xia CT Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants. The Navigated Xia 4.5- Xia CT Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The Navigated Xia 4.5- Xia CT Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Xia 4.5- Xia CT Spinal Fixation System using the Stryker Spine Navigation System. The Navigated Xia 3- Serrato Deformity Screwdriver is intended to be used as an accessory to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants. The Navigated Xia 3-Serrato Deformity Screwdriver may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a regid anatomical structure can be identified. The Navigated Xia 3- Serrato Deformity Screwdriver is intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of the Stryker Spine XIA® 3 and ES2® Spinal Fixation Systems using the Stryker Spine Navigation System.
Device Story
Navigated Spine Instruments (screw inserters/taps) function as manual surgical accessories for the Stryker Spine Navigation System; used to facilitate placement of spinal fixation implants (Mesa, Everest, Yukon, Xia 3, Xia 4.5, ES2). Instruments attach to a Rotational Navigation Adapter and trackers; provide intraoperative guidance by referencing rigid anatomical structures during surgery. Operated by surgeons in clinical settings; output consists of physical instrument positioning relative to patient anatomy as tracked by the navigation system. Benefits include enhanced precision in implant placement during spinal procedures.
Clinical Evidence
Bench testing only; no clinical data. Performance testing confirmed functionality, compatibility with the Stryker Spine Navigation System, and adherence to all acceptance criteria.
Technological Characteristics
Manual, non-sterile, reusable surgical instruments. Materials are well-characterized and biocompatible per applicable standards. Connectivity via Rotational Navigation Adapter and trackers for use with the Stryker Spine Navigation System.
Indications for Use
Indicated for patients requiring spinal fixation surgery where computer-assisted planning and navigation are appropriate; used for intraoperative guidance via rigid anatomical reference structures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
March 23, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Stryker Spine Megan Callanan Senior Regulatory Affairs Specialist 600 Hope Parkway SE Leesburg, Virginia 20175
Re: K203205
Trade/Device Name: Navigated Spine Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 21, 2020 Received: December 23, 2020
### Dear Megan Callanan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203205
Device Name Navigated Spine Instruments
#### Indications for Use (Describe)
Navigated Everest Instruments
The Navigated Everest Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.
The Navigated Everest Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The Navigated Everest Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spinal Fixation System using the Stryker Spine Navigation System.
#### Navigated Mesa Instruments
The Navigated Mesa Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.
The Navigated Mesa Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The Navigated Mesa Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spinal Fixation System using the Stryker Spine Navigation System.
#### Navigated Yukon Instruments
The Navigated Yukon Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.
The Navigated Yukon Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The Navigated Yukon Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spinal Fixation System using the Stryker Spine Navigation System.
#### Navigated Xia 4.5- Xia CT Taps
The Navigated Xia 4.5- Xia CT Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.
The Navigated Xia 4.5- Xia CT Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
{3}------------------------------------------------
The Navigated Xia 4.5- Xia CT Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Xia 4.5- Xia CT Spinal Fixation System using the Stryker Spine Navigation System.
#### Navigated Xia 3- Serrato Deformity Screwdriver
The Navigated Xia 3- Serrato Deformity Screwdriver is intended to be used as an accessory to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.
The Navigated Xia 3-Serrato Deformity Screwdriver may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a regid anatomical structure can be identified.
The Navigated Xia 3- Serrato Deformity Screwdriver is intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of the Stryker Spine XIA® 3 and ES2® Spinal Fixation Systems using the Stryker Spine Navigation System.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 12pt;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12pt;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
K203205
{4}------------------------------------------------
# K203205
| 510(k) Summary: Navigated Spine Instruments | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Stryker Spine<br>600 Hope Parkway SE<br>Leesburg, VA 20175 |
| Contact Person : | Name: Megan Callanan<br>Phone: (551)262-2429<br>Email: megan.callanan@stryker.com |
| Date Prepared: | 10/29/2020 |
| Trade Name: | Navigated Spine Instruments |
| Common Name: | Navigation Instruments |
| Proposed Class: | Class II |
| Classification Name: | Orthopedic / Orthopedic Stereotaxic Instruments |
| Regulation Number | 21 CFR 882.4560 |
| Product Code: | OLO |
| Predicate Devices: | Primary Predicate: K172034 Spine Map 3D 3.1<br>Additional Predicate: K192911 Brainlab Compatible K2M Navigation Instruments<br>Reference Predicate: K183196 SpineMap Go |
| Device Description: | Navigated Spine Instruments are nonsterile, reusable surgical instruments (screw<br>inserters and taps) intended to be used when implanting previously cleared<br>components of Mesa, Everest, Yukon, Xia 3, Xia 4.5, and ES2. These instruments are<br>intended to be used with the Stryker Spine Navigation System. |
| Indications for Use: | Navigated Everest Instruments |
| | The Navigated Everest Screw Inserters and Taps are intended to be used as<br>accessories to the Stryker Spine Navigation System. They are manual surgical<br>instruments used to facilitate placement of Stryker Spine implants. |
| | The Navigated Everest Screw Inserters and Taps may be used as part of the Stryker<br>Spine Navigation System, which is indicated for any medical condition in which the<br>use of computer assisted planning and surgery may be appropriate. The System can<br>be used for intraoperative guidance where a reference to a rigid anatomical structure<br>can be identified. |
| | The Navigated Everest Screw Inserters and Taps are intended for use with Rotational<br>Navigation Adapter and associated trackers to facilitate the placement of screws of<br>the Stryker Spine Everest Spinal Fixation System using the Stryker Spine Navigation<br>System. |
| 510(k) Summary: Navigated Spine Instruments | |
| Navigated Mesa Instruments<br>The Navigated Mesa Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.<br>The Navigated Mesa Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.<br>The Navigated Mesa Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Mesa Spinal Fixation System using the Stryker Spine Navigation System. | |
| Navigated Yukon Instruments<br>The Navigated Yukon Screw Inserters and Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.<br>The Navigated Yukon Screw Inserters and Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.<br>The Navigated Yukon Screw Inserters and Taps are intended for use with Rotational Navigation Adapter and associated trackers to facilitate the placement of screws of the Stryker Spine Yukon Spinal Fixation System using the Stryker Spine Navigation System. | |
| Navigated Xia 4.5- Xia CT Taps<br>The Navigated Xia 4.5- Xia CT Taps are intended to be used as accessories to the Stryker Spine Navigation System. They are manual surgical instruments used to facilitate placement of Stryker Spine implants.<br>The Navigated Xia 4.5- Xia CT Taps may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The System can be used | |
{5}------------------------------------------------
{6}------------------------------------------------
| 510(k) Summary: Navigated Spine Instruments | |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | for intraoperative guidance where a reference to a rigid anatomical structure can be<br>identified. |
| | The Navigated Xia 4.5- Xia CT Taps are intended for use with Rotational Navigation<br>Adapter and associated trackers to facilitate the placement of screws of the Stryker<br>Spine Xia 4.5- Xia CT Spinal Fixation System using the Stryker Spine Navigation<br>System. |
| | Navigated Xia 3- Serrato Deformity Screwdriver |
| | The Navigated Xia 3- Serrato Deformity Screwdriver is intended to be used as an<br>accessory to the Stryker Spine Navigation System. They are manual surgical<br>instruments used to facilitate placement of Stryker Spine implants. |
| | The Navigated Xia 3- Serrato Deformity Screwdriver may be used as part of the<br>Stryker Spine Navigation System, which is indicated for any medical condition in<br>which the use of computer assisted planning and surgery may be appropriate. The<br>System can be used for intraoperative guidance where a reference to a rigid<br>anatomical structure can be identified. |
| | The Navigated Xia 3- Serrato Deformity Screwdriver is intended for use with<br>Rotational Navigation Adapter and associated trackers to facilitate the placement of<br>screws of the Stryker Spine XIA® 3 and ES2® Spinal Fixation Systems using the<br>Stryker Spine Navigation System. |
| Summary of the<br>Technological<br>Characteristics | The subject Navigated Spine Instruments have the same technological characteristics<br>as the predicate device including design, material composition, intended use, function,<br>and range of sizes. Biocompatibility of patient-contacting materials was demonstrated<br>by using well characterized materials that meet applicable standards. |
| Summary of the<br>Performance Data | Performance testing was conducted to ensure that the Navigated Spine Instruments<br>are acceptable for their intended use, to ensure functionality and compatibility with<br>the Stryker Spine Navigation System using the Rotational Navigation Adapter and<br>associated trackers, and to demonstrate substantial equivalence to the predicate<br>device. The performance of Navigated Spine Instruments has met all acceptance<br>criteria. |
| Conclusion | Stryker Navigated Spine Instruments are substantially equivalent to<br>the noted predicate device as they have the same intended use, technological<br>characteristics, safety and efficacy profile, and performance specifications.<br>The content of this submission supports the determination of substantial<br>equivalence. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.