← Product Code [NVE](/productcode/NVE) · K203199

# RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP Single Chamber Washer-Disinfector (K203199)

_STERIS Corporation · NVE · Dec 1, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203199

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [NVE](/productcode/NVE.md)
- **Decision Date:** Dec 1, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

## Device Story

RAS Racks are specialized loading accessories for AMSCO 7052HP/7053HP Single-Chamber Washer-Disinfectors; designed to hold up to 12 da Vinci EndoWrist instruments. Device operates via automated RAS Cycle; performs series of spray-washing and lumen-flushing steps using Prolystica Ultra Concentrate HP Enzymatic Cleaner and Neutral Detergent. Cycle includes prewash, wash phases, rinse, thermal disinfection (A0=600), and heated drying. Used in hospital sterile processing departments by technicians. Output is cleaned, disinfected, and dried instruments ready for steam sterilization. System monitors water temperature, fill volume, chemistry injection, and pump rotation. Benefits include standardized, automated reprocessing of complex robotic surgical instruments.

## Clinical Evidence

Bench testing only. No clinical data presented.

## Technological Characteristics

Stand-alone pass-through single-chamber washer-disinfector with integrated software. Features in-line self-cleaning filtration assembly. Uses temperature-controlled water, enzymatic cleaner, and neutral detergent. Thermal disinfection parameters: 194°F (90°C) for 1 minute (A0=600). Connectivity includes USB port for data download/printing. No sonication capability.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector ([K200577](/device/K200577.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2020

STERIS Corporation Jennifer Nalepka Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060

Re: K203199

Trade/Device Name: RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP Single Chamber Washer-Disinfector Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NVE Dated: October 28, 2020 Received: October 29, 2020

Dear Jennifer Nalepka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie-Williams, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

# 510(k) Number (if known)

K203199

## Device Name

RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP Single Chamber Washer-Disinfector

## Indications for Use (Describe)

The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the STERIS logo. The word STERIS is in all capital letters and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines that are blue. The STERIS logo is a registered trademark.

# 510(k) Summary For RAS Racks and RAS Cycle in AMSCO® 7052HP / 7053HP Single Chamber Washer/Disinfectors

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Contact:

Jennifer Nalepka Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198 Email: jennifer_nalepka@steris.com

Submission Date: November 16, 2020

Premarket Notification Number: K203199

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

#### Device Name 1.

| Trade Name:            | RAS 12 Rack / RAS 12 Long Rack, used in RAS<br>Cycle of AMSCO 7052HP / 7053HP Single<br>Chamber Washer-Disinfector |  |
|------------------------|--------------------------------------------------------------------------------------------------------------------|--|
| Device Class:          | Class II                                                                                                           |  |
|                        | Endoscope cleaning accessory                                                                                       |  |
| Common/usual Name:     | Endoscope washer, cleaner, automated                                                                               |  |
| Classification Name:   | 21 CFR 876.1500                                                                                                    |  |
| Classification Number: | NVE                                                                                                                |  |
| Product Code:          |                                                                                                                    |  |

#### 2. Predicate Device

RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector, K200577

#### 3. Device Description

The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector. Twelve soiled da Vinci X/Xi and S/Si Endowrist® instruments, with limited prior manual pre-cleaning, are loaded into the appropriate RAS Rack according to the provided instructions for use. The rack is placed in the AMSCO 7052HP or 7053HP Single-Chamber Washer/Disinfector, and the RAS Cycle is selected.

The RAS Cycle performs automated cleaning through a validated series of spraywashing, lumen flushing steps that use Prolystica Ultra Concentrate HP Enzymatic Cleaner alternately with Prolystica Ultra Concentrate HP Neutral Detergent in temperature-controlled solutions. When the series of automated prewash and washing stages are complete, a one-minute rinse occurs. Next the RAS Cycle completes a thermal rinse to achieve intermediate level disinfection of the load before drying it. Upon RAS Cycle completion, the devices are ready to be prepared and packed for steam sterilization.

#### 4. Indications for Use

The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level

{5}------------------------------------------------

disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

#### ട. Technological Characteristics Comparison Table

A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1.

| Feature                                                  | Proposed Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K203199)                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K200577)                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                                                                                                                                                                               |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use<br>Indications<br>for Use                | The RAS Racks are used in<br>the RAS Cycle of the<br>AMSCO 7052HP Single-<br>Chamber Washer/Disinfector<br>and the AMSCO 7053HP<br>Single-Chamber<br>Washer/Disinfector for the<br>effective cleaning, rinsing,<br>intermediate level<br>disinfection and drying of<br>reusable da Vinci® X/Xi and<br>S/Si EndoWrist®<br>instruments.                                                                                                                                                                                                   | The RAS Racks are used in<br>the RAS Cycles of the<br>AMSCO 7052HP and<br>7053HP Single-Chamber<br>Washer/Disinfectors for the<br>effective cleaning, rinsing,<br>drying and intermediate level<br>disinfection of reusable da<br>Vinci EndoWrist® X/Xi,<br>S/Si, and Single-Site<br>instruments and staplers.                                                                                                                                                                                                                                     | Identical except<br>for removal of<br>Single-Site and<br>Stapler<br>instruments and<br>re-ordering of<br>wording to<br>more<br>accurately<br>identify the<br>intended da<br>Vinci devices.                               |
| Operating<br>Principles<br>/<br>Technology               | The RAS Racks and RAS<br>Cycle of the 7052HP/7053HP<br>SC w/d provide the necessary<br>combination of cleaning<br>agents in hot water with<br>temperature and water<br>pressure control during timed<br>sequences to achieve<br>effective cleaning of complex<br>lumened da Vinci<br>instruments. Validated<br>parameters of the RAS cycle,<br>which includes prewash,<br>cleaning stages, rinsing,<br>intermediate level thermal<br>disinfection, and heated<br>drying, cannot be reduced<br>from the minimum default<br>parameters.   | The RAS Racks and RAS<br>Cycle of the 7052HP/7053HP<br>SC w/d provide the necessary<br>combination of cleaning<br>agents in hot water with<br>temperature and water<br>pressure control during timed<br>sequences to achieve<br>effective cleaning of complex<br>lumened da Vinci instruments<br>and staplers. Validated<br>parameters of the RAS cycle,<br>which includes prewash,<br>cleaning stages, rinsing,<br>intermediate level thermal<br>disinfection, and heated<br>drying, cannot be reduced<br>from the minimum default<br>parameters. | Identical except<br>for the removal<br>of Single-Site<br>and Stapler<br>instruments.                                                                                                                                     |
| Feature                                                  | Proposed Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K203199)                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K200577)                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                                                                                                                                                                               |
| Where<br>Used                                            | Hospital/medical center SPD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Hospital/medical center SPD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                |
| Design                                                   | The Washer/Disinfector is a<br>stand-alone pass-through<br>design single-chamber unit<br>with integrated software. The<br>RAS Racks position and<br>provide flow to loaded da<br>Vinci items throughout the<br>RAS Cycle while spray arms<br>assure all surfaces are cleaned<br>with validated chemistries.<br>rinsed, thermal disinfected,<br>and dried.                                                                                                                                                                               | The Washer/Disinfector is a<br>stand-alone pass-through<br>design single-chamber unit<br>with integrated software. The<br>RAS Racks position and<br>provide flow to loaded da<br>Vinci items throughout the<br>RAS Cycle while spray arms<br>assure all surfaces are cleaned<br>with validated chemistries,<br>rinsed, thermal disinfected,<br>and dried.                                                                                                                                                                                          | Identical                                                                                                                                                                                                                |
| Instrument<br>preparation                                | Reusable EndoWrist® X/Xi,<br>S/Si instruments are handled<br>at point of use in the OR, then<br>transferred to the<br>Decontamination Area of the<br>sterile processing department<br>where they are prepared for<br>automated cleaning and<br>installed in the appropriate<br>RAS Rack according to the<br>detailed instructions provided<br>in its Operator Manual.                                                                                                                                                                   | Reusable EndoWrist® X/Xi,<br>S/Si, and Single-Site<br>instruments and staplers are<br>handled at point of use in the<br>OR, then transferred to the<br>Decontamination Area of the<br>sterile processing department<br>where they are prepared for<br>automated cleaning per the da<br>Vinci Addendum for<br>Automated Reprocessing<br>Manual) and installed in the<br>appropriate RAS Rack<br>according to the detailed<br>instructions provided in its<br>Operator Manual.                                                                       | Similar;<br>removed<br>Single-Site and<br>Stapler<br>instrument<br>processing<br>instructions.<br>Incorporated<br>the manual pre-<br>cleaning steps<br>from da Vinci<br>Addendum<br>Automated<br>Reprocessing<br>Manual. |
| Critical<br>Parameters<br>for<br>Cleaning                | Minimum critical cycle<br>parameters are provided by<br>default in the RAS Cycle:<br>Dosing of validated<br>●<br>concentrated chemistries<br>at specified volume of<br>0.74 oz. (22 mL)<br>Series of washing stages:<br>●<br>> 2-minute Prewash, not<br>heated<br>> Wash phase 1 - initial 6-<br>minute stage @122°F<br>with Prolystica Ultra<br>Concentrate HP                                                                                                                                                                         | Minimum critical cycle<br>parameters are provided by<br>default in the RAS Cycle:<br>Dosing of validated<br>●<br>concentrated chemistries<br>at specified volume of<br>0.74 oz. (22 mL)<br>Series of washing stages:<br>●<br>> 2-minute Prewash, not<br>heated<br>> Wash phase 1 - initial 6-<br>minute stage @ 122°F<br>with Prolystica Ultra<br>Concentrate HP                                                                                                                                                                                   | Identical                                                                                                                                                                                                                |
| Feature                                                  | Proposed Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K203199)                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K200577)                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                                                                                                                                                                               |
|                                                          | Enzymatic Cleaner<br>followed by 6-minute<br>stage @150°F with<br>Prolystica Ultra<br>Concentrate HP<br>Detergent<br>> 2-minute rinse, not<br>heated<br>> Wash phases 3 and 4,<br>each - initial 6-minute<br>stage @122°F with<br>Prolystica Ultra<br>Concentrate HP<br>Enzymatic Cleaner<br>followed by 6-minute<br>stage @150°F with<br>Prolystica Ultra<br>Concentrate HP<br>Detergent<br>> 1-minute rinse, not<br>heated<br>Pump provides<br>continuous circulation<br>through lumens and spray<br>arms at pressure above 45<br>psi | Enzymatic Cleaner<br>followed by 6-minute<br>stage @150°F with<br>Prolystica Ultra<br>Concentrate HP<br>Detergent<br>> 2-minute rinse, not<br>heated<br>> Wash phases 3 and 4,<br>each - initial 6-minute<br>stage @122°F with<br>Prolystica Ultra<br>Concentrate HP<br>Enzymatic Cleaner<br>followed by 6-minute<br>stage @150°F with<br>Prolystica Ultra<br>Concentrate HP<br>Detergent<br>> 1-minute rinse, not<br>heated<br>Pump provides<br>continuous circulation<br>through lumens and spray<br>arms at pressure above 45<br>psi            |                                                                                                                                                                                                                          |
| Critical<br>Parameters<br>for<br>Thermal<br>Disinfection | Minimum critical cycle<br>parameters are provided by<br>default in the RAS Cycle:<br>● Temperature 194°F<br>(90°C)<br>● Time 1 minute<br>● A₀ = 600                                                                                                                                                                                                                                                                                                                                                                                     | Minimum critical cycle<br>parameters are provided by<br>default in the RAS Cycle:<br>● Temperature 194°F<br>(90°C)<br>● Time 1 minute<br>● A₀ = 600                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                                |
| Drying                                                   | ● Temperature (high =<br>setpoint 220°F)<br>● Default time 20 minutes<br>(adjustable from 2 to 30<br>minutes)                                                                                                                                                                                                                                                                                                                                                                                                                           | ● Temperature (high =<br>setpoint 220°F)<br>● Default time 20 minutes<br>(adjustable from 2 to 30<br>minutes)                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                |
| Record<br>keeping                                        | Provides printout or<br>download capability from<br>USB port or using optional<br>printer                                                                                                                                                                                                                                                                                                                                                                                                                                               | Provides printout or<br>download capability from<br>USB port or using optional<br>printer                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                |
| Water<br>Quality                                         | < 120 ppm hardness                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | < 120 ppm hardness                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                |
| Feature                                                  | Proposed Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K203199)                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K200577)                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                                                                                                                                                                               |
| Sonication                                               | No capability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | No capability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                |
| Process<br>Monitors:                                     | Monitors water<br>temperature for each<br>filling of the sump Monitors time of each<br>phase at set point<br>temperature. Monitors water fill<br>volume Monitors volume of<br>cleaner injected Monitors pump rotation                                                                                                                                                                                                                                                                                                                   | Monitors water<br>temperature for each<br>filling of the sump Monitors time of each<br>phase at set point<br>temperature. Monitors water fill<br>volume Monitors volume of<br>cleaner injected Monitors pump rotation                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                |
| Cleaner<br>dispensed                                     | Washer/disinfector<br>automatically dispenses the<br>validated volume of<br>concentrated chemistry at<br>specified points in the RAS<br>Cycle.                                                                                                                                                                                                                                                                                                                                                                                          | Washer/disinfector<br>automatically dispenses the<br>validated volume of<br>concentrated chemistry at<br>specified points in the RAS<br>Cycle.                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                |
| Cleaning<br>chemistries                                  | Uses only Prolystica Ultra<br>Concentrate HP Enzymatic<br>Cleaner and Prolystica Ultra<br>Concentrate HP Neutral<br>Detergent in RAS Cycle.                                                                                                                                                                                                                                                                                                                                                                                             | Uses only Prolystica Ultra<br>Concentrate HP Enzymatic<br>Cleaner and Prolystica Ultra<br>Concentrate HP Neutral<br>Detergent in RAS Cycle.                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                |
| Cycle time                                               | Approximately 1 hour and 15<br>- 20 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Approximately 1 hour and 15<br>- 20 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                |
| #<br>Instruments                                         | Up to 12 da Vinci<br>instruments/cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Up to 12 da Vinci<br>instruments and/or<br>staplers/cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical<br>except for the<br>removal of<br>Single-Site and<br>Stapler<br>instruments                                                                                                                                   |
| Self-<br>Disinfection<br>Cycle                           | No self-disinfection<br>cycle. A Decon Cycle using<br>AMSCO Liquid Descaler<br>is run once weekly.                                                                                                                                                                                                                                                                                                                                                                                                                                      | No self-disinfection<br>cycle. A Decon Cycle using<br>AMSCO Liquid Descaler<br>is run once weekly.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                |
| Filters                                                  | Each RAS Rack features an<br>in-line, self-cleaning filtration<br>assembly. Operator performs                                                                                                                                                                                                                                                                                                                                                                                                                                           | Each RAS Rack features an<br>in-line, self-cleaning filtration<br>assembly. Operator performs                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                |
| Feature                                                  | Proposed Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K203199)                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>RAS Racks and Cycles in<br>AMSCO® 7052HP/7053HP<br>Single-Chamber<br>Washer/Disinfectors<br>(K200577)                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                                                                                                                                                                               |
|                                                          | a manual cleaning step<br>weekly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | a manual cleaning step<br>weekly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                          |
| Accessories                                              | Prolystica Ultra Concentrate<br>HP Enzymatic Cleaner and<br>Prolystica Ultra Concentrate<br>HP Neutral Detergent                                                                                                                                                                                                                                                                                                                                                                                                                        | Prolystica Ultra Concentrate<br>HP Enzymatic Cleaner and<br>Prolystica Ultra Concentrate<br>HP Neutral Detergent                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                |

# Table 1. Technological Characteristics Comparison Table

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

#### 6. Summary of Non-Clinical Performance Testing

The purpose of this Special 510(k) is to update labeling consistent with a labeling revision being enacted by Intuitive Surgical to allow for automated cleaning and intermediate level disinfection of reusable da Vinci X/Xi and S/Si EndoWrist® instruments in the RAS Racks and RAS Cycles of the AMSCO 7052HP and 7053HP Single-Chamber Washer/Disinfectors. The RAS Racks and associated RAS Cycles were not altered in anyway.

#### 7. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K200577), Class II (21 CFR 876.1500), product code NVE.

---

**Source:** [https://fda-staging.innolitics.com/device/K203199](https://fda-staging.innolitics.com/device/K203199)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com