K203179 · Contract Medical International, GmbH · DYB · Nov 20, 2020 · Cardiovascular
Device Facts
Record ID
K203179
Device Name
Fortress Introducer Sheath System
Applicant
Contract Medical International, GmbH
Product Code
DYB · Cardiovascular
Decision Date
Nov 20, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Fortress Introducer Sheath System is intended to provide access and to facilitate percutaneous introduction of guide wires, catheters and other devices into the femoral, popliteal and infrapopliteal arteries while maintaining hemostasis during diagnostic and interventional procedures.
Device Story
Fortress Introducer Sheath System facilitates percutaneous vascular access; consists of coil-reinforced polymer sheath with hemostasis valve, side port, and tapered-tip dilator. Sheath features PTFE inner tube, Pebax outer cover, stainless-steel coil reinforcement, and distal hydrophobic coating; radiopaque platinum-iridium marker at distal end. Used in healthcare facilities by clinicians; inserted over guide wire to provide working lumen for catheters/devices. Hemostasis valve seals around instruments to minimize blood loss; side port allows flushing. System is sterile, single-use, disposable; intended for up to 24-hour placement. Provides stable access path for interventional procedures; benefits patient by enabling minimally invasive diagnostic/therapeutic interventions.
Clinical Evidence
Bench testing only. Performance verified via mechanical testing (kink resistance, pull-out, force at break, creep to break, insertion force, bond strength), usability evaluation per IEC 62366-1, accelerated aging (3 years), and biocompatibility assessment per ISO 10993 series. Sterilization validated per ISO 11135.
Technological Characteristics
Materials: PTFE, Pebax, stainless-steel coil, silicone, platinum-iridium. Design: Coil-reinforced multi-layer polymer shaft, hemostasis valve, side port with 3-way stopcock, tapered dilator. Dimensions: 7F/8F, lengths 45-90cm. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10555-1, ASTM F2394, EN 13868, ISO 10993, IEC 62366-1, ASTM F1980.
Indications for Use
Indicated for patients requiring percutaneous access to femoral, popliteal, or infrapopliteal arteries for diagnostic or interventional procedures using guide wires and catheters.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 20, 2020
Contract Medical International, GmbH Marketa Shanelova Regulatory Manager Lauensteiner Strasse 37 Dresden, Saxony 01277 Germany
Re: K203179
Trade/Device Name: Fortress Introducer Sheath System Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: October 23, 2020 Received: October 26, 2020
Dear Marketa Shanelova:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and
{1}------------------------------------------------
regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH and -Learn (https://www.fda.gov/training-and-continuing-education/odrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203179
Device Name Fortress Introducer Sheath System
### Indications for Use (Describe)
The Fortress Introducer Sheath System is intended to provide access and to facilitate percutaneous introduction of guide wires, catheters and other devices into the femoral, popliteal arteries while maintaining hemostasis during diagnostic and interventional procedures.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|--------------------------------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: underline;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY 2
#### 2.1 Submitter
| Submitter: | Contract Medical GmbH | |
|------------------------|---------------------------------------|--|
| | Lauensteiner Strasse 37 | |
| | 01277 Dresden | |
| | Germany | |
| Contact Person: | Marketa Shanelova, Regulatory Manager | |
| Phone: | +420 494 949 586 | |
| Email: | marketa.shanelova@heraeus.com | |
| Date Summary Prepared: | 23 October 2020 | |
#### 2.2 Device Device Trade Name: Fortress Introducer Sheath System Device Common Name: Introducer Sheath Classification Name: Introducer Catheter, and Dilator, Vessel, For Percutaneous Catheterization Device Class: . Product Code: DYB, and DRE
#### 2.3 Predicate Device
The predicate device is the 6F size of the Fortress Introducer Sheath System (K153197)
#### 2.4 Device Description
The Fortress Introducer Sheath System consists of an introducer sheath with hemostasis valve and side port, as well as a dilator with a tapered tip and luer lock at the proximal end. The main introducer sheath tubing is connected at the proximal end to a hemostasis valve with side port tubing that is connected to a plastic color coded 3-way stopcock valve. The side port is used for flushing the introducer sheath. The introducer sheath is introduced into the vascular system with the aid of the dilator. The hemostasis valve at the proximal end of the introducer sheath conforms and seals around guide wires and catheters to reduce blood leakage from the introducer sheath. A radiopaque marker helps identify the distal end of the introducer sheath. The introducer sheath has a hydrophobic silicone coating on the outer surface of the distal 30cm portion.
The System consists of the following components:
- A One Introducer Sheath (a) with hemostasis valve (b)
- A One dilator (c)
(a) Shaft. The coil reinforced multi-layer polymer shaft contains a tapered tip at the distal end. A continuous inner PTFE tube forms the core of the shaft and provides a circular working lumen through which devices can be passed. A single, full length polyether (Pebax) tube creates the outer cover of the shaft. A stainless-steel flat wire coil is fused between the two polymer tubes along the entire
{4}------------------------------------------------
length of the shaft. Hydrophobic coating is applied along the distal 30cm of the outer layer for increased lubricity in this area. A radiopaque marker made of platinum iridium is embedded at the distal end of the shaft. At the proximal end of the shaft, a female, winged luer hub is over-molded onto the shaft to support handling and to provide for the connection for the hemostasis valve. The hub is color coded to match the French size of the device.
(b) Hemostasis valve. A removable hemostasis valve is thread onto the luer hub at the proximal end of the shaft. Inside the valve housing, a lubricated, silicone slit disc provides a seal around devices passed through the sheath, thereby preventing blood leakage through the valve. Just distal of the valve, the valve housing is connected to a side port tube leading to three-way stopcock valve. The side port is used for flushing the introducer sheath. At the proximal end of the valve housing, the color-coded housing cap provides a snap fit connection to the hub of the dilator.
(c) Dilator. The dilator made of Pebax contains a full length round lumen to allow placement over a guide wire. The distal end of the dilator is configured as a tapered tip that extends 35+/-7mm beyond the end of the sheath when the dilator is fully inserted through the sheath. An integral, color coded luer hub that is over-molded onto the proximal end of the dilator supports handling of the dilator and provides a snap fit connection to the valve housing cap at the proximal end of the sheath introducer.
The shaft and hemostasis valve are pre-assembled and packaged together with a dilator. Each shaft and each dilator are placed in a separate protection tube, the protection tubes are clamped together with clips. Products are sealed inside an inner and outer medical grade Tyvek pouch on which a product label is placed. Devices are packaged in labeled packers containing five units and one Instruction for Use (IFU) booklet.
The Fortress Introducer Sheath System is a sterile, disposable device. It is a prescription medical device that is used only in healthcare facilities or hospitals. The device is placed in patients for up to 24 hours.
Devices are sterilized using ethylene oxide (maximum levels of remaining tested sterilant residuals of EO < 2.9mg/24h and ECH < 1mg/24h; sterilant residual limits of EO < 4mg/24h and ECH < 9mg/24h; Sterility Assurance Level, SAL 10-6).
#### 2.5 Physical Description
The schematic diagram below illustrates the range of dimensions for the new 7/8F models.
Image /page/4/Figure/9 description: This image shows a technical drawing of a medical instrument. The drawing includes labels for different parts of the instrument, such as the Luer Hub, Valve, Stopcock, Sheath, Dilator, and Marker Band. The drawing also includes measurements, such as 450-900mm, 515-965mm, and 300 mm, as well as 7Fr/8Fr.
{5}------------------------------------------------
| French Size | Length (cm) | Shape | Catalogue Number |
|-------------|-------------|-----------|------------------|
| 7F | 45 | Crossover | 452386 |
| 7F | 45 | Straight | 452382 |
| 7F | 65 | Straight | 452383 |
| 7F | 90 | Straight | 452388 |
| 8F | 45 | Crossover | 452387 |
| 8F | 45 | Straight | 452385 |
| 8F | 65 | Straight | 452384 |
| 8F | 90 | Straight | 452389 |
#### 2.6 Device Models
#### 2.7 Materials Used
The materials that make up the device are either the same or very similar to those used for the predicate. Any differences in the material do not raise any new issues of safety or effectiveness, as demonstrated by the design verification test results.
#### 2.8 Indications for Use
The Fortress Introducer Sheath System is intended to provide access and to facilitate percutaneous introduction of guide wires, catheters and other devices into the femoral, popliteal and infrapopliteal arteries while maintaining hemostasis during diagnostic and interventional procedures.
This is the same intended use as for the previously cleared 6F Fortress Introducer Sheath System (K153197).
#### 2.9 Comparison of Technological Characteristics with the Predicate Device
The 7F and the 8F size of the Fortress Introducer Sheath System is a manually operated, sterile, single patient use sheath system made predominantly of thermoplastic polymers. The sheath is reinforced with a stainless steel coil in order to provide kink resistance when passed through tortuous paths. With regard to the design, device features, method of sterilization, and mode of operation, the 7F and 8F Fortress Introducer Sheath System does not differ from the predicate device.
Materials used for manufacture of the 7F and 8F Fortress Introducer Sheath System are the same or very similar as those contained in the predicate device. The slight differences in technological characteristics between the subject device and the predicate do not raise any new concerns of safety and effectiveness, as demonstrated by the data collected.
{6}------------------------------------------------
## 2.10 Performance Data
Performance data demonstrate that the Fortress Introducer Sheath System 7F and 8F performance is substantially equivalent to the predicate. The following performance data from non-clinical tests are being provided in support of the substantial equivalence determination:
- Mechanical testing, including tests required under relevant international standards, and usability test performed to verify and validate the design.
- . Biocompatibility Risk Assessment (BRA) and performed biocompatibility device testing to demonstrate biocompatibility.
- . Sterilization information to confirm sterility of the device upon exposure to the selected sterilization cycle.
- . Accelerated aging testing to confirm product performance at the end of the shelf life.
The list of tests performed in support of determination of substantial equivalence is provided in the table below.
| No. | Verification / Validation Activity | Test Type | Applicable Standard(s) |
|-----|-----------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| 1 | Sheath/dilator fit test | Mechanical/Visual | Internal requirement |
| 2 | Sheath pull out test | Mechanical/Visual | ASTM F2394<br>ISO 10555-1:2013 |
| 3 | Sheath kink resistance test | Mechanical | EN 13868:2002 |
| 4 | Sheath force at break test | Mechanical | ISO 10555-1:2013 |
| 5 | Sheath creep to break test | Mechanical | ISO 10555-1:2013 |
| 6 | Dilator hub bond strength | Mechanical | ISO 10555-1:2013 |
| 7 | Sheath system insertion force | Mechanical | Internal requirement |
| 8 | Usability evaluation | Usability Study | IEC 62366-1:2015 |
| 9 | Simulated use test | Mechanical/Visual | ASTM F2394 |
| 10 | Accelerated age test<br>(3 years) | Mechanical/ External<br>Laboratory | ASTM F1980-16<br>Various per performed tests |
| 11 | Biocompatibility testing<br>Cytotoxicity test<br>Biological Risk Assessment | External Laboratory<br><br>Documented assessment | ISO 10993-1:2009<br>ISO 10993-4:2017<br>ISO 10993-5:2009<br>ISO 10993-10:2010<br>ISO 10993-11:2017<br>ISO 10993-12:2012 |
## List of Tests
{7}------------------------------------------------
| No. | Verification / Validation Activity | Test Type | Applicable Standard(s) |
|-----|------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------|
| | | | ISO 14971:2019<br>FDA Guidance Use of<br>International Standard ISO<br>10993-1 |
| 12 | Sterilization adoption | Documented assessment/<br>External laboratory testing | ISO 10993-7:2008<br>ISO 11737-1:2018<br>ISO 11135:2014<br>ISO 10993-1:2018 |
# 2.11 Conclusions
The results of performed testing based on risk analysis demonstrate that the 7F and 8F Fortress Introducer Sheath System performs comparably with the predicate. The 7F and 8F Fortress Introducer Sheath System is substantially equivalent to the predicate device in terms of intended use, design and materials, technological characteristics, and principle of operation. Any differences between the subject device and the predicate do not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.