K203144 · Eclipse Medcorp, LLC · QAI · May 25, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203144
Device Name
MicroPen EVO
Applicant
Eclipse Medcorp, LLC
Product Code
QAI · General, Plastic Surgery
Decision Date
May 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Eclipse MicroPen EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
Device Story
MicroPen EVO is a minimally invasive, powered microneedling device for aesthetic use. It consists of a reusable pen body, rechargeable Li-ion battery, battery charger, and sterile, single-use 14-needle cartridge. The device uses a rotary motor to drive needles in a reciprocating motion, creating microscopic punctures in the epidermis and dermis. It is operated by licensed healthcare practitioners or individuals directed by them in a clinical setting. The user adjusts needle penetration depth (0-2.0 mm) via the device controls. A disposable MicroSleeve sheath is used to prevent cross-contamination. The device is intended to improve facial acne scar appearance. Performance is maintained through consistent puncture rates (105-128 stamps/second) and needle depth accuracy. The device is cleaned and disinfected between uses. Benefits include mechanical skin remodeling via controlled micro-injury.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included motor speed/puncture rate consistency, needle penetration depth accuracy, needle retention force, battery life (>4 hours), use life (2000 hours), cartridge reliability, microbial ingress protection, and suction prevention. Biocompatibility testing (ISO 10993) confirmed the device is non-cytotoxic, non-sensitizing, and non-pyrogenic. Cleaning and disinfection validation demonstrated 6-log reduction of vegetative organisms.
Indicated for the treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
Regulatory Classification
Identification
A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.
Special Controls
The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 25, 2021
Eclipse MedCorp, LLC Julie Summerville Senior Director of Product Management 5916 Stone Creek Drive The Colony, Texas 75056
Re: K203144
Trade/Device Name: MicroPen EVOTM Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: QAI Dated: March 1, 2021 Received: March 8, 2021
Dear Julie Summerville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lixin Liu, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203144
Device Name MicroPen EVO™
Indications for Use (Describe)
The Eclipse MicroPen EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
| 510(k) Summary | |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| [This 510(k) Summary is submitted in accordance with 21 CFR Part 807.92(c)] | |
| Submitted by: | Eclipse Medcorp, LLC.<br>5916 Stone Creek Drive<br>The Colony, TX 75056 |
| | <b>Contact Person:</b> Julie Summerville<br>Senior Dir of Product Management<br>972-380-2911 x 2405<br>jsummerville@eclipsemed.com |
| Date Prepared: | May 19, 2021 |
| Trade Name: | MicroPen EVO™ |
| Common Name: | Powered Microneedling Device |
| Product Code;<br>Regulation Name & No: | QAI Microneedling device for aesthetic use 21 CFR §878.4430 |
| Device Classification: | II |
# Device Description:
The Eclipse MicroPen EVO™ is a minimally invasive microneedling device that mechanically creates microscopic punctures in the epidermal and dermal layers of the skin by means of micro-needles in a reciprocating cartidge head. The MicroPen EVO is comprised of a reusable pen body, a sterile, single use microneedling cartridge, a rechargeable battery pack, a battery charger with power supply, and a disposible MicroSleeve sheath. The microneedling cartridge is attached to the pen body and activated with an On/Off button. The depth of needle penetration can be adjusted by the user depending on the skin being treated. Charging is accomplished by placing the MicroPen EVO pen body or the battery pack on the Charger base.
## Indications for Use:
The Eclipse MicroPen EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.
## Predicate Device:
The Eclipse MicroPen EVO is predicated against the SkinPen Precision System by Bellus Medical (DEN160029).
## Technological Characteristics and Comparison to Predicate
The Eclipse MicroPen EVO has the identical intended use and indications for use as the predicate device and the same or equivalent technological features. A comparison of the subject and predicate device's technological features is presented in the following table.
| Technological Characteristics Comparison Chart | | | | |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Subject device<br>Eclipse MicroPen EVO | Predicate (Primary)<br>SkinPen Precision System | Comparison | |
| 510(k) | TBD | DEN160029 | NA | |
| Manufacturer | Eclipse MedCorp LLC<br>The Colony, TX, U.S.A. | Bellus Medical, LLC<br>Addison, TX, U.S.A. | NA | |
| Trade Name | MicroPen EVO™ | SkinPen® Precision System | NA | |
| Product Code | QAI | QAI | Same | |
| Regulation # | 21 CFR Part 878.4430 | 21 CFR Part 878.4430 | Same | |
| Reg Name | Microneedling device for aesthetic<br>use | Microneedling device for aesthetic<br>use | Same | |
| Device Class | Class II | Class II | Same | |
| Indication<br>for Use /<br>Intended Use | The Eclipse MicroPen EVO™ is a<br>microneedling device and<br>accessories intended to be used as<br>a treatment to improve the<br>appearance of facial acne scars in<br>adults aged 22 years and older. | SkinPen® Precision System is a<br>microneedling device and<br>accessories intended to be used as a<br>treatment to improve the appearance<br>of facial acne scars in adults aged 22<br>years and older. | Same | |
| Intended Users | Rx Only: Licensed healthcare<br>practitioners or individuals directed<br>by practitioners | Rx Only: licensed healthcare<br>practitioners or individuals directed<br>by practitioners | Same | |
| Technological Characteristics Comparison Chart - continued | | | | |
| Use Location | Face | Face | Same | |
| Power Source<br>(Pen Body) | Rechargeable Li-ion battery | Rechargeable Li-ion battery | Same | |
| Power Source<br>(Battery Charger) | AC Powered | AC Powered | Same | |
| Control<br>Mechanism | Microprocessor;<br>embedded software | Microprocessor; embedded software | Same | |
| Operating<br>Principle | Rotary | Rotary | Same | |
| Single Speed<br>(RPM) | 6300-7700 | 6300 - 7700 | Same | |
| Puncture Rate | 105-128 stamps/second | 105 -128 stamps/second | Same | |
| Microneedling<br>Cartridge | Sterile, Single Use | Sterile, Single Use | Same | |
| No. of Needles | 14 | 14 | Same | |
| Needle Gauge | 32 Ga | 32 Ga | Same | |
| Needle Material | Stainless Steel | Stainless Steel | Same | |
| Needle Shape<br>Geometry | Straight, cylindrical body with a<br>conical tapered, sharp point | Straight, cylindrical body with a<br>conical tapered, sharp point | Same | |
| Arrangement | Needles radially arranged | Needles radially arranged | Same<br>Equivalent:<br>subject device<br>has a slightly<br>smaller total<br>surface area of<br>the hub. No<br>affect to<br>geometry,<br>puncture<br>pattern, needle<br>stamp. | |
| Needle Spacing | 2 mm spacing/3.48 mm² per needle | 2 mm spacing/3.54 mm² per needle | Equivalent:<br>subject device<br>has a slightly<br>smaller total<br> | |
| Penetration<br>Depth | 1.5 mm<br>(Recommended) | 1.5 mm<br>(Recommended) | Same | |
| Max Needle<br>Depth Setting | 2.0 mm | 2.5 mm | Different: Not<br>significant;<br>treatment depth<br>is 1.5 mm for<br>both devices. | |
| Penetration<br>Depth Selection | 9 depth settings; 0 mm to 2.0 mm<br>in 0.25 mm increments | 11 depth settings; 0 mm to 2.5 mm in<br>0.25 mm increments | treatment depth<br>is 1.5 mm for<br>both devices. | |
| Sterility<br>(cartridge) | Ethylene Oxide | Ethylene Oxide | Same | |
| Shelf Life<br>(Cartridge) | 2 years | 2 years (min) | Same | |
| Barrier: Cross-<br>Contamination | MicroSleeve Sheath<br>(Disposable) | BioSheath<br>(Disposable) | Same | |
{4}------------------------------------------------
# Performance Data
In combination with the general controls of the FD&C Act, the Eclipse MicroPen EVO™ microneedling device for aesthetic use has been subjected to performance testing and adheres to the following special controls and standards:
{5}------------------------------------------------
## Motor Speed: Puncture Rate and Needle Penetration Depth Testing
In order to determine if the MicroPen EVO can perform as intended for the recommended 30 minute treatment sessions, maintain accuracy of needle penetration, and resist dislodgement and serious deformation of the needles, Eclipse conducted testing under simulated worst-case skin conditions. Worst-case skin was simulated by the use of SynDaver SynTissue plate 10N puncture grade synthetic skin. Normal adult skin exhibits a mean penetration resistance of approximately 2N in areas such as the face and forearm. SynDaver SynTissue with a puncture resistance of 6-8N is representative of callused skin or scar tissue. The 10N SynDaver substrate used in the testing has a penetration force beyond that of thick, calloused human skin or scar tissue, and represents a worst-case load for the device.
## Puncture Rate Testing
Eclipse conducted testing to determine if the MicroPen EVO motor can achieve consistent speed and the intended puncture rate within the established tolerance of 7000 RPM ± 10% under worst case skin condition (load) over the duration of the intended operating time. The MicroPen EVO handpieces with attached needling cartridges and battery pack were run continually for 30 minutes (i.e., recommended operating time) under a representative worst-case skin substrate, the SynDaver SynTissue plate 10N penetration grade. The predicate device, SkinPen Precision was also tested so that a side-by-side comparison could be made. Measurements were taken before and after load application, and also at defined intervals while under load to verify the motor is operating at RPMs within the pre-defined specification range of 6300 – 7700 RPMs (7000 ± 10%).
All test articles met the established criteria demonstrating that the MicroPen EVO motor can achieve the intended puncture rate within the established tolerance range of 6300-7700 RPM over the duration of the typical, intended 30 minute operating time. The puncture rate ranged from 6513 to 7164. The results show that the MicroPen EVO motor can perform its intended purpose, even under worst case load, and is substantially equivalent to the predicate device.
## Needle Penetration Depth and Extension Accuracy
Eclipse conducted penetration depth testing of the MicroPen EVO microneedling cartridge to determine if, under worst-case conditions, the needle settings meet the specified acceptance criteria by remaining within +/- 0.25 mm of the selected needle depths below 1.50 mm and +0mm/-0.50 mm at selected needles depths of 1.50 mm and above. Worst-case conditions were created by utilizing the synthetic tissue substrate, SynDaver SynTissue plate 10N. This testing evaluated extension accuracy to demonstrate if the maximum possible needle extension does not exceed what has been demonstrated safe in clinical data and literature. The predicate device, the SkinPen Precision, was included for side-byside comparative purposes.
All MicroPen EVO test articles passed the needle penetration testing coming within +/- 0.25 mm for depth settings up to 1.25 mm: and within +0.00/-0.50 for depth setting between 1.5mm -2.0 mm and no test articles exceeded 2.0 mm Further, none of the needles were dislodged or showed serious deformation at the conclusion of each test. These results demonstrate that the MicroPen EVO can meet the established performance criteria and are substantially equivalent to the predicate device.
## Needle Retention
In order to determine if aged MicroPen EVO microneedling cartridges demonstrate adequate retention when subjected to pullout forces.
## Needle Retention for Aged Samples Eclipse
To determine if the MicroPen EVO microneedles demonstrate adequate retention under force, nine cartridges and 35 needles were subjected to pullout forces. The microneedles were considered to have passed if the needle retention force is greater than or equal to 110 g (1.08 N) which is the retention force of the SkinPen Precision (predicate).
{6}------------------------------------------------
The needle retention of the all MicroPen EVO test needles (n=35) exceeded the minimum force of 110 g (1.08 N). The results of this test demonstrate that the MicroPen EVO microneedling cartridge needles will perform as intended. The results also demonstrate that MicroPen EVO microneedling cartires better than the predicate device with regard to needle retention and is considered substantially equivalent.
## Battery Life
In order to determine if the MicroPen EVO will perform as intended and provide >4 hours of operation under normal use conditions on a single full battery charge, cyclic testing of EVO battery packs was performed under simulated worst-case condition. The 10N SynDaver substrate used in the testing has a penetration force beyond that of thick, calloused human skin or scar tissue, and represents a worst-case load for the device.
## Battery Life Testing
In order to establish that the MicroPen EVO will perform as intended and provide >4 hours of operation under normal use conditions on a single full battery charge, cyclic testing was performed under simulated worst case clinical condition using Syn Tissue Plate 10N. Each simulated operation cycle consisted of:
- 1. Installation of a cartridge on the hand piece;
2. 30 minutes of operation against a 10N SynDaver SynTissue plate;
3. 5 minute rest following each cycle to simulate cleaning, disinfection and set-up for next patient).
The activity cycle above was repeated until battery shutdown occurs.
Thirty trials were completed achieving a >95% confidence (95/90). All test articles passed the pre-established acceptance criteria of "at least 8 cycles (four hours) on a single charge" and "no degradation in battery longevity after multiple charges". The results of the Battery Life Test demonstrated than under worst case clinical conditions, battery run times are significantly higher than required (i.e., Battery Trials 1 and 2: avg 9.38 hours; Battery Trial 3: avg 10.32 hours). These results establish that the Eclipse MicroPen EVO can perform its intended use and supports the substantial equivalence to the predicate device which has similar battery performance requirements.
## Use Life
In order to determine if the MicroPen EVO handpiece, charger and battery can perform its intended use for the expected duration of its identified use life of 2000 hours, cyclic testing was performed under simulated worst-case clinical conditions. The 10N SynDaver substrate used in the testing has a penetration force beyond that of thick, calloused human skin or scar tissue, and represents a worst-case load for the device.
# Use Life Testing
In order to establish that the MicroPen EVO handbiece, charger, and battery will perform as intended over its identified useful life, test articles were subjected to simulated procedure cycles to determine the number of cycles each can sustain before degraded device function is observed. The test articles were considered to have passed if it survived at least 2000 hours without physical, mechanical, or visual degradation to the components.
{7}------------------------------------------------
Each simulated operation cycle consisted of:
- 1. Installation of cartridge assembly onto the MicroPen handpiece:
- 2. 30 minutes of continuous operation against a 10N SynDaver SynTissue plate (clinical worst case):
- 3. Following each 30 minute cycle, the handpiece, battery and charger were cleaned and
disinfected by the recommended method (i.e., CaviWipes).
The activity cycle above was repeated every 30 minutes for the duration of each day until a total time of 2000 hours of simulated use was achieved. (Cartridge assemblies and batteries were replaced as needed.)
The results of the testing demonstrated that the handbieces, batteries and chargers continued to function for the 2000 hour test and had no observable degradation to finish, seals, mechanical function or labels. All batteries continued to operate beyond 4 hours for the test. The MicroPen EVO test articles survived the continuous, repeated, simulated use including cleaning and disinfection without degradation, the device is considered to meet the identified use life of 2000 hours.
## Cartridge Life
Testing was conducted to determine if the MicroPen EVO microneedling cartridge can continuously perform its intended purpose for the recommended 30 minute treatment sessions. To make this determination, the cartridge was tested under simulated worst-case clinical use to verify if the device can perform consistently under extreme use conditions and show no signs of damage to component form or function. The worst-case scenario was created by the use of aged cartridges (2 year equivalent), 60 minute continuous run times (i.e., twice the recommended) and the use of SynDaver SynTissue plate 10N puncture grade synthetic skin.
## EVO Cartridge Reliability Testing
The aged. EVO microneedling cartridges were attached to the hand piece with the needles extended to 2.0 mm and operated at 6300-7700 RPMs aqainst the 10N synthetic tissue for 60 minutes. Test articles were considered to have met acceptance criteria if they completed 60 minutes of continuous use with the applied load and showed no evidence of wear or other damage to the components. The Eclipse MicroPen EVO cartridges met the established acceptance criteria with all nine, test articles operating continuously under load for the entire 60 minutes and were still functional. There were no signs of unacceptable wear, and no observable damage to needles including burrs, hooks, bending, breakage or loss.
The results from the EVO Cartridge Reliability Test demonstrates the ability of the device to perform consistently under extreme conditions of use. This establishes that the MicroPen EVO can perform its intended use under recommended use conditions. Further, these results demonstrate the cartridge's functionality after undergoing to 2 year accelerated shelf-life aging.
This worst-case study demonstrates that the device can perform consistently under extreme conditions of use and its intended purpose throughout the life of the device. These results support the substantial equivalence to the predicate device for intended use performance.
## Microbial Ingress Testing
Eclipse has conducted a microbial ingress study utilizing worst case microorganisms including a motile microorganism to determine if the microneedling cartridge and MicroSleeve sheath prevent the ingress of these worst case microorganisms under simulated worst case use. A robust, quantitative method of recovery, incubation and enumeration by colony forming units (CFUs) was employed.
{8}------------------------------------------------
Validation of Ingress Protection of a Microneedling Device Eclipse conducted ingress testing utilizing worst-case motile microorganisms and simulated worst case use. The microorganisms used are as follows: •Staphylococcus aureus •Klebsiella pneumoniae
•Pseudomonas aeruginosa ·Escherichia coli (motile)
Fully assembled MicroPen EVO test units, with MicroSleeve sheaths and aged (2 year equivalent) microneedling cartridges were operated in the presence of the worst-case microorganisms identified above. Aged cartridges were used to demonstrate consistent performance throughout the labeled 2 year shelf life. The test article handpieces (pen bodies) were handled with a gloved hand to which a 10.0 µL aliquot of inoculum had been applied. The test units were turned on, and the cartridge tips were moved over the inoculum placed in a sterile petri dish. To ensure coverage and the greatest ingress challenge, the cartridge tip was rotated and maneuvered over multiple angles to allow the inoculum to drip into the cartridge. The test articles continued to run for 60 minutes – a time twice the recommended treatment session.
A robust, quantitative method of recovery, incubation and enumeration of CFUs was employed to determine the effectiveness of the ingress protection. The results of this study demonstrate the MicroSleeve Protective Sheath and microneedling cartridge for the MicroPen EVO provide effective protection against the ingress of fluid, and aid in the mitigation of cross contamination. These results also support substantial equivalency to the predicate device in performance.
## Suction Testing
Testing was performed to determine if the MicroPen EVO would create suction on human skin receiving treatment during device operation. To make this determination, Eclipse performed simulated use on human skin to observe if there were visible signs of suction (e.g., red marks) after 1 minute settings representing the minimum and maximum piston extension limits.
## Suction Prevention Testing
Eclipse conducted simulated use of the MicroPen EVO with a modified cartridge assembly on an anterior human forearm. The microneedling cartridge test articles had the needles removed to render 'dummy pistons' that would simulate potential suction without needle penetration. The test cartridges were attached to an EVO handpiece and set to both 0.0 and 2.0mm, representing the minimum and maximum extension limits of the piston respectively. The tip of the cartridge was placed on the anterior forearm while the device was turned on and held in place for 1 minute. This was repeated at both 0.0 and 2.0 settings with all nine test cartridges. The results showed no evidence of suction on the skin after 1 minute of simulated use at minimum and maximum extension limits. The results of this testing demonstrate the needle cartridge design prevents suction during normal use and supports substantial equivalent to the predicate device subjected to similar performance testing.
# Cleaning and Disinfection
The handpiece test units were soiled by placing a clean glove onto a dish with artificial test soil (ATS) with 20% defibrinated bovine blood. The entire surface of the palm and fingers were coated. The device was then placed in the gloved hand and soiled by aggressively handling the device, moving around to ensure all surfaces of the device are contacted. A cotton swab was dipped in the ATS and used to force the test soil into all mated areas, cracks, crevasses and seams. This included the saddle and gear, switch seal, and belly seal. The battery pack test units and charger test units were independently soiled by placing a clean glove onto a dish with artificial test soil (ATS) with 20% defibrinated bovine blood. The entire surface of the palm and fingers were coated. The battery and charger were then placed in the gloved hand and soiled by aggressively handling the device, moving around to ensure all surfaces of the battery and the charger were contacted including the battery seal. The test soil represents a worst-case challenge for cleaning and simulates clinical blood constituents including its viscosity and drying characteristics on critical areas and without the use of the provided MicroSleeve Protective Sheath. To further simulate worst-case cleaning extremes, testing was performed on freshly soiled (wet) devices and also on soiled devices allowed to dry for 24 hours. This is to simulate worst-case extremes on a device
{9}------------------------------------------------
that has been cleaned immediately after use and a device that has been left uncleaned for 24 hours. This combination of worst-case soiling and worst-case simulated-use soiling presents the appropriate challenge, given a typical procedure should only produce a small amount of pin-point bleeding.
For the handpiece, intermediate-level disinfection validation shows a 6-log reduction of common vegetative microorganisms (list below and tested individually), as well as a 3-log reduction of mycobacterium. The battery pack and charger were soiled, cleaned and disinfected with the same procedure as the handbiece, however they were validated for low-level disinfection (6-log reduction of common vegetative organisms but did not include mycobacterium.) This new testing includes the following organisms.
| Gram Negative (-) | Gram Positive(+) | Mycobacterium |
|------------------------|-----------------------|----------------------|
| Pseudomonas aeruginosa | Staphylococcus aureus | Mycobacterium terrae |
| Escherichia coli | | |
| Klebsiella pneumoniae | | |
# Sterilization and Shelf Life
- · Ethylene oxide sterilization per ISO 11135-2014; ISO11737-1:2018; ISO 11737-2: 2009; ISO 10993-7:2008
- · Sterilization, Shelf Life/Package Integrity in accordance with the following standards: ASTM-F1980 Std Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices; ASTM-F1886-2016 Std Test Method for Determining Integrity of Seals for Medical Packaging by Visual Inspection; ASTM-F1929-2015 Std Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration; ASTM-F88 Std Test Method for Seal Strength of Flexible Barrier Materials; ANSI/AAMI/ISO 11607-1: 2006(R) 2010/ A1-2014, Packaging for Terminally Sterilized Medical Devices Part 1: Requirements for Materials, Sterile Barrier System and Packaging Systems; ASTM F2096-11 (2019) Std Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test); ASTM F88/F88M-15 -Std Test Method for Seal Strength of Flexible Barrier Materials. Real time aging studies are being conducted at the extremes of storage limits, ambient temperature, and cycling between the 3 conditioning temperatures. Visual sampling and verification is conducted at 6 month intervals, while seal strength, dye migration and performance testing and cartridge reliability test are performed at 1 year intervals up to the 2 year stated shelf life.
# Biocompatibility.
- · The following tests were performed on the final, finished microneedling cartridge : (1) Cytotoxicity (ISO 10993-5:2009; (2) Sensitization and (3) Irritation/Intracutaneous Reactivity (ISO 10993-10: 2010); (4) Acute Systemic Toxicity (ISO 10993-11:2017); (5) Material Mediated Pyrogenicity (ISO 10993-11:2017, USP 41 NF 36:2018. <151> Pvrogen Test. The results of these tests demonstrated the device to be biocompatible with no evidence of material mediated pyrogenicity.
## Electrical Safety and Electromagnetic Compatibility:
- IEC- 60601-1:2005 + A1: 2012 Medical electrical equipment-Part 1: General Requirements for Basic Safety and Essential Performance; EN/JEC 60601-1-2: 2015 /IEC 60601-1-2: 2014-Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirement and tests.
- Software: ISO 62304:2006
Substantial Equivalence: The MicroPen EVO is substantially equivalent to the Bellus Medical SkinPen® Precision System predicate devices are under the same product code (QAI), both have the same intended use/ indication for use, same number of needles, gauge, shape and arrangement, same
{10}------------------------------------------------
material, recommended penetration depth, speed, puncture rate and sterilization method. The only technological differences are the maximum needle depth setting: the MicroPen EVO is 0.5 mm shorter (2.0 mm) than the predicate (2.5 mm); and the penetration the MicroPen EVO has 9 depth settings (0-2.0 mm in 0.25 mm increments) and the predicate device has 11 depth settings (0-2.5 mm in 0.25 mm increments). These differences are minor and are not significant since both devices recommend the same treatment depth (1.5 mm). There is a small and insignificant difference between the subject and predicate in total surface area of the hub, however the needle spacing is the same (2 mm) and there is no effect on geometry, puncture pattern, needle stamp. These minor different questions of safety and effectiveness. Further, the results of performance testing support substantial equivalence of the Eclipse MicroPen EVO to the predicate device.
## Conclusion
The Eclipse MicroPen EVO is considered to be substantially equivalent to the predicate device based on the intended use, technological characteristics, and the results of device testing submitted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.