K203135 · Medline Industries, Inc. · BTA · May 4, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203135
Device Name
Medline Vacu-line Suction Aspirator
Applicant
Medline Industries, Inc.
Product Code
BTA · General, Plastic Surgery
Decision Date
May 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Medline Vacu-line Suction Aspirator is a portable, AC-powered device to be used on general patient population excluding neonatal due to pressure range, to supply a vacuum source adequate within the stated operating vacuum range, to aspirate fluids from respiratory airway. Intended use in home care or hospital environments.
Device Story
Portable, AC-powered suction aspirator; removes saliva, mucus, and secretions from patient airways. Components include vacuum pump, vacuum gauge, pressure regulating valve, and disposable 850cc canister. Inline hydrophobic filter (0.45 micron) prevents liquid/bacteria ingress to pump. Operated by healthcare professionals or patients/caregivers in hospitals or home environments. Device provides high vacuum/low flow suction; output monitored via gauge. Clinical benefit: airway clearance. Device is non-sterile; reusable aspirator unit with single-patient-use accessories.
Clinical Evidence
Bench testing only. Performance testing included airflow rate (24 ± 1.1 L/min), vacuum pressure range (21.15 ± 4.86 kPa to 82.84 ± 1.84 kPa), noise levels, and working current. Safety testing included IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, and EN ISO 10079-1. Cleaning validation confirmed performance stability after 150 cycles.
Technological Characteristics
AC-powered (120V, 60Hz) portable suction pump. Class II, Type BF electrical safety. High vacuum/low flow. Includes vacuum gauge, pressure regulating valve, and 0.45-micron hydrophobic inline filter. Dimensions: 293 x 192 x 245 mm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, EN ISO 10079-1.
Indications for Use
Indicated for general patient population, excluding neonates, requiring vacuum-assisted aspiration of fluids from the respiratory airway in home care or hospital settings.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
May 4, 2021
Medline Industries, Inc. Kelsey Closen Regulatory Affairs Specialist Three Lake Drive Northfield, Illinois 60093
Re: K203135
Trade/Device Name: Medline Vacu-line Suction Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: March 30, 2021 Received: April 1, 2021
Dear Kelsey Closen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equiv alence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203135
Device Name Medline Vacu-line Suction Aspirator
#### Indications for Use (Describe)
The Medline Vacu-line Suction Aspirator is a portable, AC-powered device to be used on general patient population excluding neonatal due to pressure range, to supply a vacuum source adequate within the stated operating vacum range, to aspirate fluidsfrom respiratory airway. Intended use in home care or hospital environments.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>□ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Medline. The logo consists of the word "MEDLINE" in white, stacked on top of a white abstract star shape. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.
Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093
# 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
## Submitter / 510(k) Sponsor
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592
## Contact Person
Kelsey Closen Regulatory Affairs Specialist Phone: 847-949-2283 Email: KClosen@medline.com
Summary Preparation Date May 4, 2021
Type of 510(k) Submission Traditional
## Device Name / Classification
Trade Name: Medline Vacu-line Suction Aspirator Common Name: Powered suction pump Classification Name: Pump, Portable, Aspiration (manual or powered) Product Code: BTA Classification Panel: General & Plastic Surgery Regulatory Class: II Regulation Number: 21 CFR 878.4780
## Predicate Device
EasyGoVac K140179
## Device Description
Medline Vacu-line Suction Aspirator is an AC powered medical suction aspirator that is intended to remove saliva, thick mucus and other secretions from the patient's airway or respiratory system. The device is intended to be used by healthcare facilities and in homecare. The device is small in size, lightweight and easy to carry allowing it to be portable. The main structure of the suction device
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image features the Medline logo against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a white, stylized starburst symbol. The starburst is composed of four elongated triangles converging at a central point, creating a dynamic and eye-catching design. A small gray rectangle is visible in the upper right corner of the image.
Medline Industries. Inc Three Lakes Drive Northfield. IL 60093
includes vacuum pump, vacuum gauge, pressure regulating valve, and is to be used in conjunction with accessories like suction tubing, air filters, and canisters (these accessories are outside the scope of this submission). The Medline Vacu-line Suction Aspirator is a non-sterile device and can be used on multiple patients when following the proper cleaning techniques detailed in the instruction for use, but the accessories are all single patient use and should be disposed of properly and in accordance with their respective instructions for use. The device is a high vacuum/low flow with a vacuum gauge of minimum vacuum range 21.15 ± 4.86 kPa to maximum vacuum range 82.84 ± 1.84 kPa, 2.5 grade and a flow rate of ≥14L/min. The power supply for the device is AC 120 voltage (60Hz) and can be used continuously for up to 30 minutes. The device has a sound level of 66.81 ± 1.37 dB in an open atmosphere and 68.42 ± 0.93 dB in a closed atmosphere. The canister bottles are 850cc, single patient use and disposable. This devices uses an Inline Hydrophobic filter that has a pore size of 0.45 microns; the filter blocks bacteria from entering the vacuum and it will prevent liquid from entering the vacuum pump and tubing.
## Indications for Use
The Medline Vacu-line Suction Aspirator is a portable, AC-powered device to be used on general patient population excluding neonatal due to pressure range, to supply a vacuum source adequate within the stated operating vacuum range, to aspirate fluids from respiratory airway. Intended use in home care or hospital environments.
## Summary of Technological Characteristics
| Device<br>Characteristic | Proposed<br>Device | Predicate Device | Comparison<br>Analysis |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Product Name | Medline Vacu-line<br>Suction Aspirator | EasyGoVac | N/A |
| 510(k)<br>Reference | K203135 | K140179<br>(mod<br>el PM66AC) | N/A |
| Product Owner | Medline Industries,<br>Inc. | Precision Medical | N/A |
| Product Code | BTA | BTA | Same |
| Intended Use | The Medline Vacu-line Suction<br>Aspirator is a portable, AC-<br>powered device to be used on<br>general patient population<br>excluding neonatal due to<br>pressure range, to supply a<br>vacuum source adequate within<br>the stated operating vacuum<br>range, to aspirate fluids from<br>respiratory airway. Intended use<br>in home care or hospital<br>environments. | The<br>Easy<br>Go<br>Aspirator<br>provides<br>portable, AC<br>a<br>medical vacuum<br>powered<br>source. It is intended for use in<br>the homecare / health care<br>environments. | Same |
| Regulation<br>Number | 878.4780 | 878.4780 | Same |
| Design<br>Features | -Disposable canisters (850 cc<br>float and filter lids)<br>-Lid has shut off valve to<br>protects from overfill<br>-Portable, easy carry<br>-Light Weight<br>- Inline Hydrophobic Filter<br>-Vacuum gauge | -Reusable and disposable<br>containers (800cc/1200cc)<br>-Lid has a float shut off valve that<br>protects from overfill<br>-Portable, easy carry<br>-Light Weight<br>-Filters (inline, Bacterial or<br>Hydrophobic)<br>-Vacuum gauge | Similar |
| Safety Features | -Overflow protection | -Overflow Protection | Same |
| Accessories | -Suction Tubing<br>- Disposable canisters<br>-Intermediate Tubing<br>- Inline Hydrophobic Filter<br>-Suction Catheter | -Suction tubing<br>-Inline filter (Bacterial or<br>Hydrophobic)<br>-Disposable or reusable<br>Canister (800/1200cc)<br>-800cc Inlet Adaptor Kit<br>- Carry Bag<br>-AC Power Cord | Similar |
| Device<br>Dimensions | (L) 293 x (W) 192 x<br>(H) 245 mm | With Disposable<br>bottle- 218x 160x<br>204mm<br>With Reusable bottle-<br>186x160x 187mm | Different |
| Prescription vs.<br>OTC | Rx | Rx | Same |
| Sterile vs. Non-<br>Sterile | Non- Sterile | Non-Sterile | Same |
| Disposable<br>VS.<br>Non-<br>Disposable | Aspirator is reusable<br>accessories - single<br>patient use | Aspirator is reusable<br>accessories-<br>single<br>patient use, canister is<br>also offer reusable | Same |
| Power | AC Only Model | AC only model<br>(PM66AC) | Same |
| Voltage | AC 120V, 60Hz | AC 100-240V, 50-<br>60Hz | Similar |
| Input Current | 1.5 Amps | 0.60-0.25 Amps | Different |
| Vacuum Range | 16.20~ 84.68kPa | 0~94kPa | Different |
| Canister | 850cc Disposable | 800cc and 1200cc<br>Disposable and<br>reusable | Similar<br>(for<br>disposable canister) |
| Electrical<br>Requirements | Class II, Type BF | Class II, Type BF | Same |
| Equipment<br>Type | High Vacuum/Low<br>Flow | High Vacuum/ High<br>Flow | Different |
| Working Mode | <30 continuous minutes | 15 minutes on, 15 minutes off within a 2 hour cycle | Different |
| Operating<br>Temperature | 41° ~ 104° F (5° ~ 40°C) | 0° ~ 122° F (-18° ~ 50°C) | Different |
| Operating<br>Humidity | ≤93% Relative Humidity | 95% Noncondensing | Different |
#### TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, with a stylized white cross extending from the top and bottom of the word. The background is a solid dark blue color. The logo is simple and recognizable, commonly associated with medical supplies and healthcare products.
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, with a stylized white starburst symbol behind it. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.
ledline Industries, Inc. Three Lakes Drive Northfield. IL 60093
## Summary of Non-Clinical Testing
Non-clinical verification of the Medline Vacu-line Suction Aspirator has been conducted to evaluate its safety, performance and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Full performance testing reports for can be found in Appendix D.
## Performance Testing (Bench)
The following testing was conducted on the Medline Vacu-line Suction Aspirator:
## Air Flow Rate
The aspirator displayed an average airflow rate of 24± 1.1 L/min
## Pressure Test
The aspirator displayed an minimum vacuum range of 21.15 ± 4.86 kPa and a maximum vacuum range of 82.84 ± 1.84 kPa
#### Working Current
The aspirator displayed an average working current of 1.2 ± 0.0 Amps
#### Noise Level
The aspirator displayed a noise level of 66.81 ± 1.37 dB in an open atmosphere and 68.42 ± 0.93 dB in a closed atmosphere
# Other Testing
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for Medline. The logo is set against a blue square background. The word "MEDLINE" is in white, bold, sans-serif font. A white star-like shape is behind the word "MEDLINE".
Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093
Electromagnetic Compatibility & Electrical Safety
- IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General Requirements for Basic . Safety and Essential Performance, Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11:2015 Medical Electrical Equipment -- Part 1-11: General Requirements for Basic ● Safety and Essential Performance- Collateral Standards: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-6:2010, Medical Electrical Equipment Part 1-6: General requirements for safety-Collateral Standard: Usability
- EN ISO 10079-1:2015, Medical suction equipment -- Part 1: Electrically powered suction equipment
## Cleaning Validation
A Cleaning validation was conducted by cleaning the suction aspirator 150 times with Microkill+ alcohol based wipes, without affecting the performance or appearance of the device.
## Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Vacu-line Suction Aspirator is as safe and as effective for their intended use as the predicate device, EasyGoVac K140179.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.