← Product Code [BZD](/productcode/BZD) · K203126

# S10 Kirra (K203126)

_Resmed Pty , Ltd. · BZD · Dec 18, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203126

## Device Facts

- **Applicant:** Resmed Pty , Ltd.
- **Product Code:** [BZD](/productcode/BZD.md)
- **Decision Date:** Dec 18, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The S10 Kirra is indicated to provide CPAP and Bi-level therapy for the treatment of obstructive sleep apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg). ASV and ASVAuto modes are also indicated for the treatment of central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use.

## Device Story

S10 Kirra is a prescription-only positive airway pressure (PAP) ventilator for sleep apnea treatment. Device uses a microprocessor-controlled brushless centrifugal blower, pressure sensors, and flow sensors to regulate air delivery. It provides CPAP, Bi-level, ASV, and ASVAuto therapy modes. Optional integrated heated humidification is available. Used in home and hospital settings; operated by patients or clinicians. Device provides therapeutic and technical data via a user interface. Bluetooth, cellular, and SD card connectivity enable data transfer. Output affects clinical decision-making by providing therapy and monitoring patient compliance/faults. Benefits include airway splinting to treat apnea events.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including system verification and comparative side-by-side testing against predicates for pressure performance, breath event detection (flow limitations, snore, apneas), response to periodic breathing, and humidification output.

## Technological Characteristics

Microprocessor-controlled brushless centrifugal blower; pressure/flow sensors. Materials: polymers, plastics, stainless steel. Connectivity: SD card, cellular, Bluetooth. Integrated thermostatically controlled heated humidifier. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-70, ISO 80601-2-74, ISO 10993-1, ISO 18562-1.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- S9 Greenhills ([K140279](/device/K140279.md))
- S9 Wanda ([K140159](/device/K140159.md))
- S9 Elouera ([K140124](/device/K140124.md))

## Reference Devices

- VPAP Adapt ([K133801](/device/K133801.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

December 18, 2020

ResMed Pty Ltd % Sheila Bruschi Director, Regulatory Affairs ResMed Corp 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K203126

Trade/Device Name: S10 Kirra Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: October 16, 2020 Received: October 19, 2020

Dear Sheila Bruschi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K203126

Device Name S10 Kirra

Indications for Use (Describe)

The S10 Kirra is indicated to provide CPAP and Bi-level therapy for the treatment of obstructive sleep apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg). ASV and ASVAuto modes are also indicated for the treatment of central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use.

| Type of Use (Select one or both, as applicable)                |
|----------------------------------------------------------------|
| <div>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> |
| <div>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div>  |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the ResMed logo. The logo features a curved line made up of a series of colored dots, transitioning from blue to red. Below the line is the word "ResMed" in a bold, sans-serif font.

# 510(k) Summary

|                            | [As required by 21 CFR 807.92(c)]                                                                                                                                                                     |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared:             | 16 December 2020                                                                                                                                                                                      |
| Company Name/Owner:        | ResMed Pty Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW, 2153 Australia                                                                                                                     |
| Prepared and Submitted by: | Mr Peter Jennings<br>Senior Regulatory Affairs Manager<br>Tel: +612 8884 1000<br>Fax: +612 8884 2004<br>peter.jennings@resmed.com.au                                                                  |
| Official Contact:          | Ms Sheila Bruschi<br>Director, Regulatory Affairs<br>ResMed Corp.<br>9001 Spectrum Center Blvd<br>San Diego CA 92123 USA<br>Tel: +1 858 836 5934<br>Fax: +1 858 836 5519<br>sheila.bruschi@resmed.com |
| Device Trade Name:         | S10 Kirra                                                                                                                                                                                             |
| Device Common Name:        | Non continuous ventilator (IPPB)                                                                                                                                                                      |
| Classification:            | 21 CFR 868.5905, BZD (Class II)                                                                                                                                                                       |
| Product Code:              | BZD                                                                                                                                                                                                   |
| Predicate Device(s):       | S9 Greenhills (K140279)<br>S9 Wanda (K140159)<br>S9 Elouera (K140124)                                                                                                                                 |
| Reference Device(s):       | VPAP Adapt (K133801)                                                                                                                                                                                  |
| Reason for Submission:     | New Device                                                                                                                                                                                            |

# Device Description:

The S10 Kirra is a prescription only Positive Airway Pressure (PAP) ventilator device intended to treat individuals that are diagnosed with sleep apnea conditions. The S10 Kirra uses a micro-processor controlled blower, along with pressure and flow sensors, to achieve pressure, flow and time regulation of air delivery. It includes optional humidification, with air delivery to the patient via

{4}------------------------------------------------

S10 Kirra Traditional 510(k)

Image /page/4/Picture/1 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

heated or non-heated breathing circuits. The device provides both therapeutic (e.g. tidal volume) and technical data (e.g. system fault), and a user interface allowing adjustment of device parameters. The device uses an external AC power supply, and allows the addition of low flow supplemental oxygen.

#### Indications for Use:

The S10 Kirra is indicated to provide CPAP and Bi-level therapy for the treatment of obstructive sleep apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg). ASV and ASVAuto modes are also indicated for the treatment of central and/or mixed apneas, or periodic breathing.

It is intended for home and hospital use.

#### Substantial Equivalence:

The subject and predicate devices have the same intended use and the following similarities:

- . Similar Indications for Use
- . Same operating principle
- . Similar technological characteristics

The S10 Kirra combines the PAP therapy modes of the predicate devices S9 Greenhills (K140279), S9 Elouera (K140124) and S9 Wanda (K140159) in a new flow generator system. There are only minor differences between the S10 Kirra and the predicate devices including new materials and the addition of Bluetooth technology.

| Characteristic                      | Predicate device:<br>S9 Greenhills<br>K140279                                                                                                                                                                                                                      | Predicate device:<br>S9 Wanda<br>K140159                                                                                                                                                                         | Predicate device:<br>S9 Elouera<br>K140124                                                                                                                                                                                                                        | Subject device:<br>S10 Kirra                                                                                                                                                                                                                                    | Substantially<br>Equivalent?                                                                                                                                                                    |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                        | Positive airway<br>pressure support<br>for sleep apnea<br>Patients >66 lb<br>(>30kg)<br>Home and Hospital                                                                                                                                                          | Positive airway<br>pressure support<br>for sleep apnea<br>Patients >66 lb<br>(>30kg)<br>Home and Hospital                                                                                                        | Positive airway<br>pressure support<br>for sleep apnea<br>Patients >66 lb<br>(>30kg)<br>Home and Hospital                                                                                                                                                         | Positive airway<br>pressure support<br>for sleep apnea<br>Patients >66 lb<br>(>30kg)<br>Home and Hospital                                                                                                                                                       | YES                                                                                                                                                                                             |
| Indications for<br>Use<br>Statement | The S9 Greenhills is<br>indicated for the<br>treatment of<br>patients weighing<br>more than 66 lb (30<br>kg) with obstructive<br>sleep apnea (OSA),<br>central and/or<br>mixed apneas, or<br>periodic breathing.<br>It is intended for<br>home and hospital<br>use | The S9 WANDA<br>VPAP ST is indicated<br>for the treatment of<br>Obstructive Sleep<br>Apnea (OSA) in<br>patients weighing<br>more than 66 lb (30<br>kg).<br>It is intended for<br>use in the hospital<br>and home | The S9 Elouera self-<br>adjusting device is<br>indicated for the<br>treatment of<br>Obstructive Sleep<br>Apnea (OSA) in<br>patients (female<br>patients with mild<br>to moderate OSA<br>when using AfH<br>treatment mode)<br>weighing more than<br>66 lb (30 kg). | The S10 Kirra is<br>indicated to provide<br>CPAP and Bi-level<br>therapy for the<br>treatment of<br>obstructive sleep<br>apnea (OSA) in<br>patients (female<br>patients with mild<br>to moderate OSA<br>when using AfH<br>treatment mode)<br>weighing more than | YES<br>S10 Kirra<br>includes a<br>combined IFU<br>statement<br>within the<br>Intended Use of<br>the predicate<br>devices.<br>Disease states<br>treated by each<br>therapy mode<br>are unchanged |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for ResMed. The logo consists of the word "ResMed" in a bold, sans-serif font. Above the word, there is a curved line made up of a series of dots that transition in color from blue to red. The dots are arranged in a wave-like pattern.

| Characteristic                | Predicate device:<br>S9 Greenhills<br>K140279                                                                                                        | Predicate device:<br>S9 Wanda<br>K140159                                     | Predicate device:<br>S9 Elouera<br>K140124                               | Subject device:<br>S10 Kirra                                                                                                                                                                             | Substantially<br>Equivalent?                                                                                                                                                   |  |  |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|                               | The humidifier is<br>intended for single<br>patient use in the<br>home environment<br>and re-use in a<br>hospital /<br>institutional<br>environment. |                                                                              | It is intended for<br>home and hospital<br>use                           | 66 lb (30 kg). ASV<br>and ASVAuto<br>modes are also<br>indicated for the<br>treatment of<br>central and/or<br>mixed apneas, or<br>periodic breathing.<br>It is intended for<br>home and hospital<br>use. | from the<br>predicates.                                                                                                                                                        |  |  |
| Environment<br>of Use         | Home Healthcare<br>Environment,<br>Professional<br>Healthcare Facilities                                                                             | Home Healthcare<br>Environment,<br>Professional<br>Healthcare Facilities     | Home Healthcare<br>Environment,<br>Professional<br>Healthcare Facilities | Home Healthcare<br>Environment<br>(including aircraft),<br>Professional<br>Healthcare Facilities                                                                                                         | YES<br>S10 Kirra and<br>predicate<br>devices for use<br>in same<br>environments.<br>Specific use on<br>aircraft based<br>on reference<br>device K133801<br>testing<br>methods. |  |  |
| Therapies                     |                                                                                                                                                      |                                                                              |                                                                          |                                                                                                                                                                                                          |                                                                                                                                                                                |  |  |
| Therapy<br>Modes              | CPAP<br>ASV<br>ASVAuto                                                                                                                               | CPAP<br>-<br>-<br>-<br>VAuto<br>S, ST, T                                     | CPAP<br>AutoSet<br>AutoSet for Her<br>(AfH)                              | CPAP<br>ASV<br>ASVAuto<br>AutoSet<br>AutoSet for Her<br>(AfH)<br>VAuto<br>S, ST, T                                                                                                                       | YES<br>S10 Kirra has<br>same therapy<br>modes as<br>predicates                                                                                                                 |  |  |
| Pressure<br>Range             | 4-20 cm H2O<br>(CPAP)<br>3-25 cm H2O (ASV,<br>ASVAuto)                                                                                               | 4-20 cm H2O<br>(CPAP)<br>4-25 cm H2O<br>(VAuto)<br>3-25 cm H2O (S, ST,<br>T) | 4-20 cm H2O (CPAP,<br>AutoSet, AutoSet<br>for Her)                       | 4-20 cm H2O (CPAP,<br>AutoSet, AutoSet<br>for Her)<br>4-25 cm H2O<br>(VAuto)<br>3-25 cm H2O (S, ST,<br>T, ASV, ASVAuto)                                                                                  | YES<br>S10 Kirra has<br>identical<br>pressure range<br>for<br>corresponding<br>modes.                                                                                          |  |  |
| Comfort<br>Features           |                                                                                                                                                      | Expiratory Pressure<br>Relief (EPR)<br>Easybreathe                           | Expiratory Pressure<br>Relief (EPR)                                      | Expiratory Pressure<br>Relief (EPR)<br>Easybreathe                                                                                                                                                       | YES                                                                                                                                                                            |  |  |
| Technology & Design<br>System |                                                                                                                                                      |                                                                              |                                                                          |                                                                                                                                                                                                          |                                                                                                                                                                                |  |  |
| Components                    | Flow Generator<br>Integrated<br>Humidifier<br>Air Tubing<br>Mask                                                                                     | Flow Generator<br>Integrated<br>Humidifier<br>Air Tubing<br>Mask             | Flow Generator<br>Integrated<br>Humidifier<br>Air Tubing<br>Mask         | Flow Generator<br>Integrated<br>Humidifier<br>Air Tubing<br>Mask                                                                                                                                         | YES                                                                                                                                                                            |  |  |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.

| Characteristic                         | Predicate device:<br>S9 Greenhills<br>K140279                                                                                                                                                                                | Predicate device:<br>S9 Wanda<br>K140159                                                                                                                                                                                     | Predicate device:<br>S9 Elouera<br>K140124                                                                                                                                                                                   | Subject device:<br>S10 Kirra                                                                                                                                                                                                 | Substantially<br>Equivalent?                                                                                                                       |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating<br>Principle                 | Micro-processor<br>controlled brush-<br>less centrifugal<br>blower as air source<br>to provide splinting<br>of patient airway                                                                                                | Micro-processor<br>controlled brush-<br>less centrifugal<br>blower as air source<br>to provide splinting<br>of patient airway                                                                                                | Micro-processor<br>controlled brush-<br>less centrifugal<br>blower as air source<br>to provide splinting<br>of patient airway                                                                                                | Micro-processor<br>controlled brush-<br>less centrifugal<br>blower as air source<br>to provide splinting<br>of patient airway                                                                                                | YES                                                                                                                                                |
| Materials                              | Various materials,<br>including:<br>Polymers<br>Plastics<br>Stainless steel                                                                                                                                                  | Various materials,<br>including:<br>Polymers<br>Plastics<br>Stainless steel                                                                                                                                                  | Various materials,<br>including:<br>Polymers<br>Plastics<br>Stainless steel                                                                                                                                                  | Various materials,<br>including:<br>Polymers<br>Plastics<br>Stainless steel                                                                                                                                                  | YES<br>Substantially<br>equivalent<br>materials<br>verified for<br>reprocessing<br>and BioC.<br>Same base<br>material.                             |
| Data<br>Connectivity                   | SD card, Cellular<br>Wireless                                                                                                                                                                                                | SD card, Cellular<br>Wireless                                                                                                                                                                                                | SD card, Cellular<br>Wireless                                                                                                                                                                                                | SD card, Cellular<br>Wireless, Bluetooth<br>Wireless                                                                                                                                                                         | YES<br>Bluetooth is an<br>alternate<br>wireless<br>technology<br>option to<br>provide the<br>same data<br>transfer<br>capabilities as<br>cellular. |
| Humidification                         |                                                                                                                                                                                                                              |                                                                                                                                                                                                                              |                                                                                                                                                                                                                              |                                                                                                                                                                                                                              |                                                                                                                                                    |
| Humidifier                             | Integrated<br>thermostatically<br>controlled heated<br>humidifier with<br>detachable water<br>chamber                                                                                                                        | Integrated<br>thermostatically<br>controlled heated<br>humidifier with<br>detachable water<br>chamber                                                                                                                        | Integrated<br>thermostatically<br>controlled heated<br>humidifier with<br>detachable water<br>chamber                                                                                                                        | Integrated<br>thermostatically<br>controlled heated<br>humidifier with<br>detachable water<br>chamber                                                                                                                        | YES                                                                                                                                                |
| Operating<br>Principle                 | Water is heated by<br>a heater plate to<br>create warm and<br>moist air within the<br>water chamber.<br>Airflow from the<br>device is passed<br>through the heated<br>water chamber,<br>which moistens and<br>warms the air. | Water is heated by<br>a heater plate to<br>create warm and<br>moist air within the<br>water chamber.<br>Airflow from the<br>device is passed<br>through the heated<br>water chamber,<br>which moistens and<br>warms the air. | Water is heated by<br>a heater plate to<br>create warm and<br>moist air within the<br>water chamber.<br>Airflow from the<br>device is passed<br>through the heated<br>water chamber,<br>which moistens and<br>warms the air. | Water is heated by<br>a heater plate to<br>create warm and<br>moist air within the<br>water chamber.<br>Airflow from the<br>device is passed<br>through the heated<br>water chamber,<br>which moistens and<br>warms the air. | YES                                                                                                                                                |
| Humidifier<br>output                   | 12.0mg/L @ 20cm<br>H2O (50 L/min)                                                                                                                                                                                            | 12.0mg/L @ 20cm<br>H2O (50 L/min)                                                                                                                                                                                            | 12.0mg/L @ 20cm<br>H2O (50 L/min)                                                                                                                                                                                            | 12.6mg/L @ 20cm<br>H2O (50 L/min)                                                                                                                                                                                            | YES                                                                                                                                                |
| Heated Tube<br>temperature<br>settings | 60-86 °F<br>(16-30 °C)                                                                                                                                                                                                       | 60-86 °F<br>(16-30 °C)                                                                                                                                                                                                       | 60-86 °F<br>(16-30 °C)                                                                                                                                                                                                       | 60-86 °F<br>(16-30 °C)                                                                                                                                                                                                       | YES                                                                                                                                                |
| Characteristic                         | Predicate device:<br>S9 Greenhills<br>K140279                                                                                                                                                                                | Predicate device:<br>S9 Wanda<br>K140159                                                                                                                                                                                     | Predicate device:<br>S9 Elouera<br>K140124                                                                                                                                                                                   | Subject device:<br>S10 Kirra                                                                                                                                                                                                 | Substantially<br>Equivalent?                                                                                                                       |
| Heated Tube<br>temperature<br>cutout   | 106 °F (41 °C)                                                                                                                                                                                                               | 106 °F (41 °C)                                                                                                                                                                                                               | 106 °F (41 °C)                                                                                                                                                                                                               | 106 °F (41 °C)                                                                                                                                                                                                               | YES                                                                                                                                                |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transitions from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

#### Non-Clinical Data:

- . Verification bench testing for S10 Kirra comprises system verification and comparative sideby-side predicate testing. Verification confirmed the S10 Kirra met the predetermined acceptance criteria as defined in the relevant compliance standards and as defined in the system verification protocols.
- . Comparative predicate testing supports the determination that the S10 Kirra is substantially equivalent to the predicate devices (S9 Greenhills (K140279), S9 Wanda (K140159), S9 Elouera (K140124)). Verification bench testing included testing the performance of the therapy modes and therapy functions including:
	- o Pressure performance
	- Breath events including flow limitations, snore and apneas o
	- Response to periodic breathing O
	- Humidification O

The S10 Kirra was designed and tested in accordance with the applicable requirements in relevant FDA consensus standards including:

- . IEC 60601-1:2005+AMD1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- . IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-70, Medical Electrical Equipment – Part 2-70: Particular requirements for basic safety and essential performance of sleep apnoea breathing therapy equipment
- ISO 80601-2-74, Medical Electrical Equipment – Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
- . ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 18562-1:2017 – Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for ResMed. The logo consists of a curved line of colored dots that transition from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.

#### Clinical Data:

Clinical performance data is not required as the subject device uses established therapeutic technology and bench testing is sufficient to demonstrate substantial equivalence.

# Substantial Equivalence Conclusion:

The S10 Kirra has the same intended use and similar indications and technological characteristics as the predicate devices. The differences in the technological characteristics between the predicate devices and subject device do not impact or raise new questions of safety or efficacy. Non-clinical performance data supports the determination that the subject device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K203126](https://fda-staging.innolitics.com/device/K203126)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).