The SedLine® Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The SedLine® Sedation Monitor is indicated for adult and pediatric patients (1 year of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. The system includes the Patient State Index (PSi), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The PSi is indicated for use on adults sedated with the following agents: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The PSi is not indicated for use in the pediatric population and is not displayed when using the pediatric sensors. The SedLine® is only to be used with Masimo SedLine® sensors and cables. The use of any other sensor or cable is neither supported nor recommended by Masimo and could give erroneous results.
Device Story
SedLine Sedation Monitor is a 4-channel processed EEG monitor; acquires/processes EEG signals via forehead sensors; computes Patient State Index (PSi) to reflect anesthetic effect; displays EEG waveforms, Density Spectral Array (DSA), EMG Index, Suppression Ratio (SR), and Artifact (ARTF) on host device (Masimo Root Monitoring System). Used in OR, ICU, and clinical research labs by clinicians. Pediatric expansion introduces smaller form-factor sensors for patients 1-17 years. PSi calculation is disabled for pediatric sensors. Device provides real-time brain state monitoring to assist clinicians in managing sedation levels.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical performance testing, including software verification and validation, and mechanical/environmental testing of the pediatric sensor.
Technological Characteristics
4-channel processed EEG monitor; MOC-9 interface; low voltage DC power from host. Sensors applied to forehead. Biocompatibility per ISO 10993-1. Electrical safety/EMC per IEC 60601-1 and IEC 60601-1-2. Software level of concern: moderate.
Indications for Use
Indicated for adult and pediatric patients (1 year and older) in OR, ICU, and clinical research labs for monitoring brain state via EEG. PSi parameter indicated for adults sedated with Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, or Sevoflurane. PSi not indicated for pediatric patients.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 25, 2022
Masimo Corporation Sindura Penubarthi Regulatory Affairs Manager 52 Discovery Irvine, California 92618
Re: K203113
Trade/Device Name: Masimo SedLine Sedation Monitor and Accessories Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLW, OLT, GXY, OMC, ORT Dated: December 15, 2021 Received: December 17, 2021
Dear Sindura Penubarthi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203113
### Device Name
Masimo SedLine® Sedation Monitor and Accessories
Indications for Use (Describe) SedLine Sedation Monitor
The SedLine® Sedation Monitor is intended to monitor the state of the data acquisition and processing of EEG signals. The SedLine® Sedation Monitor is indicated for adult and pediatric patients (1 year of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.
The system includes the Patient State Index (PSi), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The PSi is indicated for use on adults sedated with the following agents: Alfentanil, Desflurane, Fentany], Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The PSi is not indicated for use in the pediatric population and is not displayed when using the pediatric sensors.
The SedLine® is only to be used with Masimo SedLine® sensors and cables. The use of any other sensor or cable is neither supported nor recommended by Masimo and could give erroneous results.
### SedLine Sensor
The RD SedLine Pediatric Sensor electrodes are applied directly to the patient's skin to enable the recording of electrophysiological signals (e.g., EEG). The RD SedLine Pediatric Sensors are indicated for pediatric patients (1 to 17 years).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Masimo Corporation logo. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in bold, black letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7982<br>FAX: (949) 297-7199 |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | 10/13/2020 |
| Contact: | Sindura Penubarthi<br>Regulatory Affairs Senior Manager<br>Masimo Corporation<br>Phone: (949) 396-4041<br>spenubarthi@masimo.com |
| Trade Name: | Masimo SedLine Sedation Monitor and Accessories |
| Common Name: | Brain Function Monitor |
| Classification Regulation/<br>Product Code: | 21 CFR 882.1400, Class II/OLW |
| Additional Product Code: | OLT, OMC, ORT, GXY |
| Establishment Registration<br>Number: | 3011353843 |
| Reason for Premarket<br>Notification: | Expansion to the indications for use and introduction of Pediatric Sensor |
| Predicate Device: | K172890 - Masimo SedLine® Sedation Monitor |
| Reference Predicate: | K072286 - Aspect Medical System, Inc, BISx |
| Performance Standards | No performance standards for the above device have been promulgated<br>pursuant to Section 514 of the Food and Drug Administration<br>Modernization Act of 1997 (FDAMA) |
# 1. Device Description
The Masimo SedLine® Sedation Monitor is a patient-connected, 4-channel processed Electroencephalograph (EEG) monitor. It displays electrode status, EEG waveforms, Density Spectral Array (DSA), and Patient State Index (PSi), EMG Index, Suppression Ratio (SR) and Artifact (ARTF).
The Masimo SedLine® Sedation Monitor includes the SedLine Module, SedLine EEG Sensor, and SedLine Patient Cable. The SedLine Module includes Masimo technology that processes the signal data collected from the SedLine sensor on the Host/Backboard device which provides the user interface.
The performance specifications for SedLine are included in Table 1.1 below:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "Page 2 of 11". This indicates that the image is a page from a document. The document has a total of 11 pages. This is page 2 of that document.
Image /page/4/Picture/1 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
| TABLE 1.1 SedLine Specifications | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| FEATURE | SPECIFICATION |
| Display Range | |
| PSI | 0 to 100 |
| EMG | 0 to 100% |
| SR | 0 to 100% |
| ARTF | 0 to 100% |
| DSA Amplitude (Left and Right) | -60 to 40 dB |
| SEFL/SEFR | 0-30Hz |
| DSA Asymmetry | -100% to +100% |
| Electrode Impedance | 0 to 65 kohms |
| DSA frequency Range | 0 to 30 Hz and 0 to 40 Hz |
| Resolution | |
| PSI | 1 |
| EMG | 1% |
| SR | 2% |
| ARTF | 1% |
| DSA Amplitude (Left and Right) | ≤1db |
| SEFL/SEFR | 1 Hz |
| DSA Asymmetry | 1% |
| Electrode Impedance | 1 kohms |
| General | |
| Visual/Audible Alarm | Host/Backboard Device (Masimo Root Monitoring System) provides the audible/visual alarm in compliance with IEC60601-1-8 |
| Storage/Recording | Host/Backboard Device (Masimo Root Monitoring System) provides trend/data storage |
| Electrical | |
| DC Power | Host/Backboard Device (Masimo Root Monitoring System) provides DC power to SedLine Module |
| Interface | |
| SedLine Module Connection | MOC-9 interface with Host/Backboard device (Masimo Root Monitoring System) |
| Mechanical | |
| Module: Dimensions | 1 3/10 in (3.3 cm) x 4 in (10.2 cm) x .8 in (2.0 cm) |
| Environmental | |
| Operating Conditions | |
| Temperature | +41°F to +104°F (+5°C to +40°C) |
| Humidity | 15% to 95%, non-condensing |
| Storage Conditions | |
| Temperature | -40°F to +158°F (-40°C to +70°C) |
| Humidity | 15-95%, non-condensing |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
| Table 1.2 SedLine Pediatric Sensor Specifications | |
|---------------------------------------------------|---------------------------|
| Application site | Forehead |
| Intended patient population | 1 to 17 years |
| Mechanical | |
| Dimensions | 7" by 5.5" |
| Biocompatibility | ISO 10993-1 |
| Environmental specifications | |
| Operating Temperature | 10°C to 40°C |
| Storage Temperature | -40°C to +70°C |
| Humidity | 10% to 95% non-condensing |
#### 1.1 Intended Use
#### 1.1.1 Intended Use
The SedLine® Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The computed values are displayed on a Host/Backboard monitor such as the Masimo Root Monitoring System.
#### 1.1.2 Indications for Use
### SedLine Sedation Monitor
The SedLine® Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The SedLine® Sedation Monitor is indicated for adult and pediatric patients (1 year of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.
The system includes the Patient State Index (PSi), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The PSi is indicated for use on adults sedated with the following agents: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The PSi is not indicated for use in the pediatric population and is not displayed when using the pediatric sensors.
The SedLine® is only to be used with Masimo SedLine® sensors and cables. The use of any other sensor or cable is neither supported nor recommended by Masimo and could give erroneous results.
### SedLine Sensor
The RD SedLine Pediatric Sensor electrodes are applied directly to the patient's skin to enable the recording of electrophysiological signals (e.g., EEG). The RD SedLine Pediatric Sensor is indicated for pediatric patients (1 to 17 years).
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in bold, black letters. To the right of the logo, the text "MASIMO CORPORATION" is displayed above the address "52 Discovery, Irvine, CA 92618".
#### 1.2 Technological Characteristics
#### 1.2.1 Principle of Operation
The SedLine is based upon the principle that brain activity results in electrical activity that can create detectable potential difference at the skin surface of the forehead that increased brain activity and decrease with reduced brain activity. The Patient State Index (PSi) utilizes this relationship to establish characteristics of the EEG that can be quantified to establish multivariate combinations of quantitative electroencephalogram (QEEG) variables that are sensitive to the brain activity under changing levels of anesthesia. The PSi is the result of a proprietary computation that combines weighted quantitative values reflecting many dimensions of brain electrical activity, such as: (1) changes in power in various EEG frequency bands, (2) changes in symmetry and synchronization between critical brain regions, and (3) the inhibition of regions of the frontal cortex.
The PSi is computed continuously from monitored changes in the QEEG when the sensor is applied. The PSi values are intended to provide information on the changes in sedation with the lower values reflecting lower levels of brain activity and deeper levels of sedation.
#### 1.2.2 Mechanism of Action for Achieving the Intended Effect
The SedLine EEG Sensor is noninvasively applied to the patient's forehead and is connected to the SedLine Module using a SedLine patient cable. The SedLine Module connects to a Host/Backboard device through an electrical connection. The SedLine EEG Sensor detects the potential differences at the forehead which are transmitted to the SedLine Module, which processes those potential differences as the EEG waveform and the analysis of the signal characteristics to be displayed on the Host/Backboard device.
To accommodate the smaller foreheads of the pediatric patient population, the SedLine is provided with a smaller form factor sensor more tailored to pediatric patients.
#### 1.2.3 Summary of Technological Characteristics of Subject Device Compared to Predicate Device
The subject device, SedLine® Sedation Monitor, and the predicate device, SedLine® Sedation Monitor (K172890), have the following key similarities:
- . Both devices have the same intended use to monitor the state of the brain by real-time data acquisition and processing of EEG signals;
- Both devices rely on the same principles of operations;
- Both devices have the same environmental specifications; ●
- Both devices utilize the same sensor materials; ●
- Both devices provide similar displays of EEG features. ●
The subject device and the predicate device. SedLine® Sedation Monitor (K172890), have the following key differences:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold font. To the right of the company name is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
- . Subject device includes the indication for pediatric patients 1 to 17 years of age;
- . Subject device includes pediatric sensors that have a smaller sensor form factor.
The main difference between the subject device and the predicate device (K172890) is the expansion of the indications to support the pediation and the addition of pediatric sensors that have a smaller form factor compared to the adult sensor and a pediatric sensor that is sized equivalent to the adult sensor. To support the substantial equivalence of the subject device non-clinical testing was performed.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left and the word "Masimo" in black letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray letters, followed by the address "52 Discovery, Irvine, CA 92618" also in gray letters.
| TABLE 5 SUBSTANTIAL EQUIVALENCE | | | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| FEATURE | Masimo SedLine® Sedation Monitor | Masimo SedLine® Sedation Monitor | Comparison |
| 510(k) Number | Subject device | Predicate device<br>(K172890) | |
| General<br>Information | | | |
| Intended Use<br>/Indications for Use<br>(IFU) | The SedLine® Sedation Monitor is<br>intended to monitor the state of the brain<br>by real-time data acquisition and<br>processing of EEG signals. The system<br>includes the Patient State Index (PSi), a<br>proprietary computed EEG variable that<br>is related to the effect of anesthetic<br>agents. The SedLine® Sedation Monitor<br>is indicated for adult and pediatric<br>patients (1 year of age and older) in the<br>operating room (OR), intensive care unit<br>(ICU), and clinical research laboratory.<br><br>The PSi is indicated for use on adults<br>sedated with the following agents:<br>Alfentanil, Desflurane, Fentanyl,<br>Isoflurane, Nitrous Oxide, Propofol,<br>Remifentanil, and Sevoflurane.<br><br>The SedLine® is only to be used with<br>Masimo SedLine® sensors and cables.<br>The use of any other sensor or cable is<br>neither supported nor recommended by<br>Masimo, and could give erroneous<br>results. | The SedLine Sedation Monitor is<br>indicated for use in the operating room<br>(OR), intensive care unit (ICU), and<br>clinical research laboratory. It is intended<br>to monitor the state of the brain by real-time<br>data acquisition and processing of<br>EEG signals for adult patients (18 years<br>of age and older). The system includes<br>the Patient State Index (PSI), a<br>proprietary computed EEG variable that<br>is related to the effect of anesthetic<br>agents.<br><br>The agents include: Alfentanil,<br>Desflurane, Fentanyl, Isoflurane, Nitrous<br>Oxide, Propofol, Remifentanil, and<br>Sevoflurane. | Added the indication for pediatrics and the<br>specific use of SedLine with Masimo<br>SedLine sensors and cables. |
| TABLE 5 SUBSTANTIAL EQUIVALENCE | | | |
| FEATURE | Masimo SedLine® Sedation Monitor<br>Subject device | Masimo SedLine® Sedation Monitor<br>Predicate device<br>(K172890) | Comparison |
| 510(k) Number | SedLine Sensor<br><br>The RD SedLine Pediatric Sensor<br>electrodes are applied directly to the<br>patient's skin to enable the recording of<br>electrophysiological signals (e.g., EEG).<br>The RD SedLine Pediatric Sensors are<br>indicated for pediatric patients (1 to 17<br>years). | | |
| Monitoring<br>Parameters | Electrode Status, EEG Waveforms,<br>Patient State Index (PSi), Density<br>Spectral Array (DSA),<br>Electromyography (EMG), Artifacts<br>(ARTF), Suppression Ratio (SR), and<br>Spectral Edge Frequency (SEFL/SEFR) | Electrode Status, EEG Waveforms,<br>Patient State Index (PSi), Density<br>Spectral Array (DSA),<br>Electromyography (EMG), Artifacts<br>(ARTF), Suppression Ratio (SR), and<br>Spectral Edge Frequency (SEFL/SEFR) | Same |
| Classification<br>Regulation/ Product<br>Code | 21 CFR 882.1400,Class II/OLW | 21 CFR 882.1400,Class II/OLW | Same |
| Additional Product<br>Codes | OLT, OMC, ORT, GXT | OLT, OMC, ORT, GXT | Same |
| Principle of<br>Operation | The SedLine is based upon the principle<br>that brain activity results in electrical<br>activity that can be create detectable<br>potential difference activity at the skin<br>surface of the forehead that increase with<br>increased brain activity and decrease<br>with reduced brain activity. The Patient<br>State Index (PSi) utilizes this relationship | The SedLine is based upon the principle<br>that brain activity results in electrical<br>activity that can be create detectable<br>potential difference activity at the skin<br>surface of the forehead that increase with<br>increased brain activity and decrease<br>with reduced brain activity. The Patient<br>State Index (PSi) utilizes this relationship | Same |
| TABLE 5 SUBSTANTIAL EQUIVALENCE | | | |
| FEATURE<br>510(k) Number | Masimo SedLine® Sedation Monitor<br>Subject device | Masimo SedLine® Sedation Monitor<br>Predicate device<br>(K172890) | Comparison |
| | to establish characteristics of the EEG<br>that can be quantified to establish<br>multivariate combinations of quantitative<br>electroencephalogram (QEEG) variables<br>that are sensitive to the brain activity<br>under changing levels of anesthesia. The<br>PSi is the result of a proprietary<br>computation that combines weighted<br>quantitative values reflecting many<br>dimensions of brain electrical activity,<br>such as: (1) changes in power in various<br>EEG frequency bands, (2) changes in<br>symmetry and synchronization between<br>critical brain regions, and (3) the<br>inhibition of regions of the frontal cortex.<br>The PSi is computed continuously from<br>monitored changes in the QEEG when<br>the sensor is applied. The PSi values are<br>intended to provide information on the<br>changes in sedation with the lower values<br>reflecting lower levels of brain activity<br>and deeper levels of sedation. | to establish characteristics of the EEG<br>that can be quantified to establish<br>multivariate combinations of quantitative<br>electroencephalogram (QEEG) variables<br>that are sensitive to the brain activity<br>under changing levels of anesthesia. The<br>PSi is the result of a proprietary<br>computation that combines weighted<br>quantitative values reflecting many<br>dimensions of brain electrical activity,<br>such as: (1) changes in power in various<br>EEG frequency bands, (2) changes in<br>symmetry and synchronization between<br>critical brain regions, and (3) the<br>inhibition of regions of the frontal cortex.<br>The PSi is computed continuously from<br>monitored changes in the QEEG when<br>the sensor is applied. The PSi values are<br>intended to provide information on the<br>changes in sedation with the lower values<br>reflecting lower levels of brain activity<br>and deeper levels of sedation. | |
| Function | | | |
| EEG Data Display | EEG Waveform, DSA | EEG Waveform, DSA | Same |
| PSi Algorithm | Adult | Adult | Same |
| Electrical | | | |
| TABLE 5 SUBSTANTIAL EQUIVALENCE | | | |
| FEATURE<br>510(k) Number | Masimo SedLine® Sedation Monitor<br>Subject device | Masimo SedLine® Sedation Monitor<br>Predicate device<br>(K172890) | Comparison |
| Type of power | Operates on low voltage DC provided through connection to host device. | Operates on low voltage DC provided through connection to host device. | Same |
| Electrical Safety/<br>EMC | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | Same |
| Input/Output<br>Interface | | | |
| Type of Interface | MOC-9 | MOC-9 | Same |
| Communication<br>Type | Masimo Proprietary Protocol | Masimo Proprietary Protocol | Same |
| Mechanical |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.