Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, US, NM, PT, and XA, etc. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Device Story
Synapse 3D Base Tools is medical imaging software for Windows-based PCs; used by trained medical professionals in clinical settings. It retrieves DICOM-compliant images (CT, MR, CR, US, NM, PT, XA) from PACS or imaging devices via network. Software provides tools for 2D/3D viewing, volume rendering, MPR (orthogonal/oblique/curved), intensity projections (MIP/MinIP/RaySum), image fusion, subtraction, and segmentation. It includes modality-specific tools like virtual endoscopy (CT), delayed enhancement viewing (MR), and diffusion-weighted analysis (MR). Clinicians use these outputs to assist in image interpretation, reporting, and treatment planning. The device facilitates anatomical segmentation, path definition, and boundary detection to support clinical decision-making. It integrates with other Synapse 3D clinical modules and PACS systems.
Clinical Evidence
No clinical studies were conducted. Evidence consists of bench testing, including system-level functionality, component, integration, usability, and installation/upgrade testing. Benchmark performance testing using actual clinical images was performed to validate the accuracy of semi-automatic and automatic segmentation, detection, and registration functions against design specifications. All tests passed.
Technological Characteristics
Software-based medical imaging application; runs on Windows-compatible PCs. Connectivity via DICOM (SCP/SCU) for image retrieval/storage. Segmentation utilizes 'graph cuts' and 'Fully Convolutional Network' (deep learning) algorithms. Complies with DICOM (PS 3.1-3.20), AAMI/ANSI/IEC 62304:2015 (software lifecycle), and ISO 14971:2007 (risk management).
Indications for Use
Indicated for trained medical professionals to aid in reading, interpreting, reporting, and treatment planning using DICOM-compliant medical images (CT, MR, CR, US, NM, PT, XA). Not for primary diagnostic interpretation of mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 9, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fujifilm Corporation % Mr. Jeffrey Wan Senior Regulatory Affairs Specialist Fujifilm Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 LEXINGTON MA 02421
Re: K203103
Trade/Device Name: Synapse 3D, Synapse 3D Base Tools v6.1 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 14, 2020 Received: December 16, 2020
Dear Mr. Wan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K203103
Device Name Synapse 3D, Synapse 3D Base Tools v6.1
#### Indications for Use (Describe)
Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, NM, PT, and XA, etc.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Synapse 3D Base Tools provides several levels of tools to the user:
Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing, orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Average (RaySum) and Minimum (MinIP) Intensity Projection. 4D volume viewing, image subtraction, surface rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and displaying multiple MPR images along an object, time-density distribution, basic image processing, CINE, measurements, annotations, reporting, storing, distribution, and general image management and administration tools, etc.
-Tools for regional segmentation of anatomical structures within the image data, path definition through vascular and other tubular structures, and boundary detection.
-Image viewing tools for modality specific images, including CT PET fusion and ADC image viewing for MR studies. -Imaging tools for CT images including virtual endoscopic viewing.
-Imaging tools for MR images including delayed enhancement image viewing, diffusion-weighted MRI data analysis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5
# 510(K) Summary
This Section contains:
510(k) Summary (per 21 CFR 807.92)
{4}------------------------------------------------
# 510(k) Summary
K203103
| Date Prepared: | Aug. 15, 2020 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Information: | FUJIFILM Corporation<br>26-30,NISHIAZABU 2-CHOME MINATO-KU<br>TOKYO 106-8620<br>Telephone: (781) 323-5315<br>Contact: Tommy San |
| Device Trade Name: | Synapse 3D, Synapse 3D Base Tools v6.1 |
| Device Common Names: | Picture Archiving and Communications System (PACS) |
| Device Clasification Name: | System, Image Processing, Radiological |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Device Class: | Class II |
| Panel: | Radiology |
| Predicate Devices: | Synapse 3D Base Tools (K120361)<br>FUJIFILM Medical Systems U.S.A., Inc. |
{5}------------------------------------------------
# 1. Description of the Device
Synapse 3D Base Tools (V6.1) (this submission) is updated software of previously-cleared Synapse 3D Base Tools (V3.0) (cleared by CDRH via K120361 on 04/06/2012).
The 3D image analysis software Synapse 3D Base Tools (V6.1) is medical application software running on Windows server/client configuration installed on commercial general-purpose Windows-compatible computers. It offers software tools which can be used by trained professionals to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.
Synapse 3D Base Tools is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communications based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools (V6.1), and the associated image-related information of the image data is registered in its database and is used for display, image processing, analysis, etc. Images newly created by Synapse 3D Base Tools (V6.1) not only can be displayed on a display, but also can be printed on a hardcopy using a DICOM printer or a Windows printer.
Synapse 3D Base Tools (V6.1) is a basic software module that works with other cleared clinical applications, including Synapse 3D Cardiac Tools (K200973), Synapse 3D Perfusion Analysis (K162287), Synapse 3D Lung and Abdomen Analysis (K130542), Synapse 3D Liver and Kidney Analysis (K142521), Synapse 3D Nodule Analysis (K120679), Synapse 3D Colon Analysis (K123566), Synapse 3D Tensor Analysis (K141514) and Synapse 3D Blood Flow Analysis (K191544). All these software modules consist of the Synapse 3D product family.
Synapse 3D Base Tools can be integrated with Fujifilm's Synapse PACS, and can be used as a part of a Synapse system. Synapse 3D Base Tools also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.
{6}------------------------------------------------
# 2. Indications for Use
Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, US, NM, PT, and XA, etc.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Synapse 3D Base Tools provides several levels of tools to the user:
Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing, orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Maximum (MIP), Average (RaySum) and Minimum (MinIP) Intensity Projection, 4D volume viewing, image fusion, image subtraction, surface rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and displaying multiple MPR images along an object, time-density distribution, basic image processing, CINE, measurements, annotations, reporting, printing, storing, distribution, and general image management and administration tools, etc.
- > Tools for regional segmentation of anatomical structures within the image data, path definition through vascular and other tubular structures, and boundary detection.
- > Image viewing tools for modality specific images, including CT PET fusion and ADC image viewing for MR studies.
- > Imaging tools for CT images including virtual endoscopic viewing.
- > Imaging tools for MR images including delayed enhancement image viewing, diffusion-weighted MRI data analysis.
# 3. Substantial Equivalence Comparison
Synapse 3D Base Tools has the same intended use, similar labeling, and clinical application tools as those of the cleared predicate device Synapse 3D Base Tools (K120361). The device features and technical characteristics comparison with predicates is shown as Table 1 Device Features and Technical Characteristics Comparison Matrix.
{7}------------------------------------------------
| Device Parameters | Synapse 3D Base Tools<br>(V6.1)<br>(This submission) | Synapse 3D Base<br>Tools(V3.0)<br>(K120361)<br>(Primary predicate<br>device) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Basic features and basic imaging tools | | |
| 2D Viewing | Yes | Yes |
| Image Storing (DICOM<br>SCP) | Yes | Yes |
| Image Communication<br>(DICOM SCU) | Yes | Yes |
| DICOM Interface<br>(SCP/SCU) | Yes | Yes |
| Printing (DICOM SCU) | Yes | Yes |
| Measurements (2D and<br>3D) | Yes | Yes |
| Annotations -<br>Standardized and Free<br>Text | Yes | Yes |
| Reporting | Yes | Yes |
| Cine | Yes | Yes |
| Volume Rendering and<br>3D Viewing | Yes | Yes |
| MPR<br>• orthogonal / oblique /<br>curved Multi-Planar<br>Reconstructions (MPR),<br>• Sector and rectangular<br>shape MPR image<br>viewing<br>• MPR for dental images<br>• Multiple MPR images<br>along an object (Slicer) | Yes | Yes |
| Maximum, Average,<br>Minimum Intensity | Yes | Yes |
| Device Parameters | Synapse 3D Base Tools (V6.1) (This submission) | Synapse 3D Base Tools(V3.0) (K120361) (Primary predicate device) |
| Projection | | |
| 4D viewing | Yes | Yes |
| Image fusion | Yes | Yes |
| Surface rendering | Yes | Yes |
| Image subtraction (3D) | Yes | Yes |
| Time-density distribution | Yes | Yes |
| General image data management and administration tools | Yes | Yes |
| In-depth analysis tools | | |
| Segmentation | Yes (Some segmentation applications are implemented using a deep learning method called as “Fully Convolutional Network”) | Yes |
| Path definition | Yes | Yes |
| Boundary detection | Yes | Yes |
| Modality specific imaging tools | | |
| CT PET fusion | Yes | Yes |
| ADC image viewing (MRI) | Yes | Yes |
| Virtual Endoscopic Simulator | Yes | N/A |
| Diffusion-weighted MRI Data Analysis | Yes | N/A |
| Delayed Enhancement Image Viewing | Yes | N/A |
| Product characteristics | | |
| Device Parameters | Synapse 3D Base Tools<br>(V6.1)<br>(This submission) | Synapse 3D Base<br>Tools(V3.0)<br><span style="text-decoration:underline"><b>(K120361)</b></span><br>(Primary predicate<br>device) |
| Product Availability | Software Product | Software Product |
| Hardware Platform | Windows PC | Windows PC |
Table 1 Device Features and Technical Characteristics Comparison Matrix
{8}------------------------------------------------
{9}------------------------------------------------
The main differences as in-depth analysis tools are shown as follows.
- 1. The new segmentation features implemented using an algorithm "graph cuts" same as the algorithm used in our previously cleared Synapse 3D Basic Tools (K120361).
- 2. The old segmentation features that had existed in our previously cleared Synapse 3D Basic Tools (K120361) are implemented using a new algorithm called as "Fully Convolutional Network" - a new deep learning-based algorithm.
- 3. The new segmentation features are implemented using a new algorithm called as "Fully Convolutional Network" - a new deep learning-based algorithm.
The main differences presented as new features of modality specific imaging tools are shown as follows.
- 1. Virtual Endoscopic Simulator (CT).
- 2. Diffusion-weighted MRI Data Analysis (MRI).
- 3. Delayed Enhancement Image Viewing (MRI).
## 4. Safety Information
Synapse 3D Base Tools introduces no new safety or efficacy issues other than those already indentified with the predicate devices. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse 3D Base Tools labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.
# 5. Testing and Performance Information
### Nonclinical testing result:
The purpose of Software Development Process for Synapse 3D Bae Tools is to carry out the activities relating to the establishment of the software development plan (or plans) for definitely conducting software hazard analysis, risk management, requirement analysis, architectural design, the design specification, unit implementation and verification, software integration and integration testing, software system test, software release, software maintenance. The main activities in software development process are described as follows.
{10}------------------------------------------------
- Software development plan ●
- Software hazard analysis and risk management
- Software requirements analysis/specification ●
- Software architectural design ●
- Software detailed design specification
- Software unit module implementation and verification ●
- Software integration and system testing ●
## Clinical tests:
The subject of this 510(k) notification, Synapse 3D Base Tools does not require clinical studies to support safety and effectiveness of the software.
## Verification and Validation:
Testing for verification and validation involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, benchmark performance testing was conducted using actual clinical images to help demonstrate that the semi-automatic or automatic segmentation, detection, and registration functions implemented in Synapse 3D Base Tools achieved the expected accuracy performance. Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests passed successfully according to the design specifications. All of the different components of the Synapse 3D Base Tools software have been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices.
## Cybersecurity:
The confidentiality, integrity and availability are maintained by Synapse 3D Base Tools in accordance with Section 6 of the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices; Guidance for Industry and Food and Drug Administration Staff (October 2, 2014).
Synapse 3D Base Tools is connected through DICOM standard to medical devices and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communication based on the DICOM standard. Therefore Synapse 3D Base Tools assures an adequate degree of protection for cybersecurity.
## Performance standards:
- Digital Imaging and Communications in Medicine (DICOM) Set (PS 3.1 3.20) (2016). .
- AAMI / ANSI / IEC 62304:2015, Medical Device Software Software Life Cycle Processes.
- • ISO 14971 Second Edition 2007-03-01, Medical Devices Application of Risk Management to Medical Devices.
{11}------------------------------------------------
# 6. Conclusion
Performance tests were conducted to test the functionality of the subject device, Synapse 3D Base Tools. Results of all conducted testing were acceptable in supporting the claim of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.