Infinity Acute Care System (IACS) Monitoring System
K203088 · Draeger Medical Systems, Inc. · MHX · Feb 1, 2022 · Cardiovascular
Device Facts
Record ID
K203088
Device Name
Infinity Acute Care System (IACS) Monitoring System
Applicant
Draeger Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Feb 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patients in environments where patient care is provided by trained health care professionals. The IACS obtains the physiologic, multi-parameter data from the M540 monitor and optional medical devices and displays. The real-time physiologic and multi-parameter data collected from the IACS is made available on the Infinity network. The IACS and optional connected hardware are not intended for use in the following hospital environments: Hyperbaric chambers Environments containing MRI equipment
Device Story
Infinity Acute Care System (IACS) is a multi-parameter physiological patient monitoring solution for hospital use. System components include M540 patient monitor, M500 docking station, and Infinity Medical Cockpits (C500/C700). M540 acquires real-time vital signs (ECG, SpO2, NIBP, IBP, CO2, etc.) via sensors/cables; displays data locally; transmits data over Infinity network to bedside Cockpits. Cockpits provide centralized monitoring, alarm management, and setting control. System supports adult, pediatric, and neonatal patients. Healthcare professionals use output for continuous patient assessment and clinical decision-making regarding life-threatening conditions. Software updates in VG4.1.1/VG4.2 include IEC 60601-1-8 alarm compliance, ST algorithm improvements, and cybersecurity enhancements (DDoS resiliency, port management).
Clinical Evidence
No clinical studies were performed or required. Substantial equivalence is supported by non-clinical data, including software verification and validation, electrical safety/EMC testing (IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-25, 60601-2-27, ISO 80601-2-55), and bench testing for cardiac rhythm and ST segment measurement algorithms (EC57).
Technological Characteristics
Multi-parameter monitor system; components include M540 monitor, M500 docking station, and C500/C700 cockpits. Connectivity via Infinity network. Software-driven; Major Level of Concern. Complies with IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-25, 60601-2-27, and ISO 80601-2-55. Features include configurable QRS detection, ST algorithm decimation filter, and IEC 60601-1-8 compliant alarm management.
Indications for Use
Indicated for multi-parameter physiologic monitoring of adult, pediatric, and neonatal patients in professional healthcare settings. Parameters include heart rate, arrhythmia (adult/pediatric), 12-lead ECG (adult/pediatric), ST segment (adult/pediatric), apnea, respiratory rate, invasive/non-invasive blood pressure, temperature, cardiac output (adult/pediatric), CO2, SpO2, pulse rate, perfusion index, total hemoglobin (adult/pediatric), total oxygen content (adult/pediatric), carboxyhemoglobin (adult/pediatric), methemoglobin, and pleth variability index.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Infinity Acute Care System Monitoring Solution (K201764)
Infinity Acute Care System Monitoring Solution Version VG2.0 (K113798)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 1, 2022
Draeger Medical Systems, Inc. Deborah Herrington Senior Manager, Regulatory Affairs 6 Tech Dr. Andover, Massachusetts 01810
Re: K203088
Trade/Device Name: Infinity Acute Care System (IACS) Monitoring System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, MSX, DRT, DOA, DSK, MLD, DXN, DSF, DSA, DSB, DQE, DXG, BZO, FLS, CCK, FLL, DSI, DPS Dated: January 26, 2022 Received: January 27, 2022
Dear Deborah Herrington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203088
Device Name Infinity® Acute Care System™ (IACS)
# Indications for Use (Describe)
Intended Use
Infinity® Acute Care System™ (IACS)
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patients in environments where patient care is provided by trained health care professionals.
The IACS obtains the physiologic, multi-parameter data from the M540 monitor and optional medical devices and displays. The real-time physiologic and multi-parameter data collected from the IACS is made available on the Infinity network.
The IACS and optional connected hardware are not intended for use in the following hospital environments:
- Hyperbaric chambers
- Environments containing MRI equipment
Indications for use
The IACS, in connection with the M540 Patient Monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric patients only)
- 12-lead Rest ECG analysis (adult and pediatric patients only)
- ST segment analysis including TruST (adult and pediatric patients only)
- Apnea
- Impedance respiratory rate RRi
- Invasive blood pressure IBP
- Non-invasive blood pressure NIBP
- Temperature
- Cardiac output (adult and pediatric patients only)
- Carbon dioxide etCO2, inCO2, RRc
- Arterial oxygen saturation SpO2
- Pulse rate
- Perfusion index PI
- Total hemoglobin SpHb, (adult and pediatric patients only)
- Total oxygen content SpOC, (adult and pediatric patients only)
- Carboxyhemoglobin saturation SpCO (adult and pediatric patients only)
- Methemoglobin saturation SpMet
- Pleth variability index PVI
The IACS can connect to third-party devices to display physiologic, multi-parameter data and stored trends. It can also send information across the Infinity network.
Intended Use
The Infinity Medical Cockpits, consisting of the C500, and the C700 are monitoring and control displays for the Infinity Acute Care System (IACS).
Medical Cockpits are intended to be used to monitor waveforms, parameter information, and alarms as well as to control
{3}------------------------------------------------
#### settings.
The Infinity Series Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare professionals.
Indications for Use
As the Infinity Medical Cockpits are used in conjunction with the Infinity Acute Care System and the M540 Patient Monitors they do not have their own indication for use.
Intended Use
The Infinity M540 (M540) is intended for the monitoring of multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patient information in environments where patient care is provided by trained health care professionals.
The M540 obtains physiological data from connection to optional accessory devices. The real-time physiologic and multiparameter data collected from the M540 is made available on the Infinity network. The M540 is intended to monitor one patient at a time.
The M540 and any connected hardware are not intended for use in the following hospital environments:
- Hyperbaric chambers
- Environments containing MRI equipment
Indications for use
The M540 monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric patients only)
- 12-lead Rest ECG analysis (adult and pediatric patients only)
- ST segment analysis including TruST (adult and pediatric patients only)
- Apnea
- Impedance respiratory rate RRi
- Invasive blood pressure IBP
- Non-invasive blood pressure NIBP
- Temperature
- Cardiac output only available when the M540 is docked in an IACS configuration (adult and pediatric patients only)
- Carbon dioxide etCO2, inCO2, RRc
- Arterial oxygen saturation SpO2
- Pulse rate
- Perfusion index PI
- Total hemoglobin SpHb, (adult and pediatric patients only)
- Total oxygen content SpOC, (adult and pediatric patients only)
- Carboxyhemoglobin saturation SpCO (adult and pediatric patients only)
- Methemoglobin saturation SpMet
- Pleth variability index PVI
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The background is white.
# age
Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810-2434
# 510(k) SUMMARY
#### 1. Submitter:
Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810-2434
| Contact Person: | Deborah L. Herrington<br>Senior Manager, Regulatory Affairs |
|--------------------|-------------------------------------------------------------|
| E-mail: | Deborah.Herrington@Draeger.com |
| Tel: | (978) 379-6574 |
| Alternate Contact: | Thomas Hirte<br>Director of Regulatory Affairs |
| E-mail: | Tom.Hirte@Draeger.com |
| Tel: | (978) 379-8291 |
01-Dec-21 Date Prepared:
#### 2. Device Names / Common Names / Classification Names
Table 5.1: Device Classification / Pro Code / Common Name
| Device Trade Name: | Infinity® Acute Care System™ Monitoring Solution |
|----------------------|-----------------------------------------------------------------------|
| Classification Name: | Monitor, Physiological, Patient (with Arrhythmia detection or alarms) |
| Common Name: | Multi-Parameter Patient Monitor |
| Classifications: | See Table 5.1 below |
| Classification | Pro Code | Common Name |
|---------------------------------------|----------|-----------------------------------------------------------------------|
| <b>Primary Product Classification</b> | | |
| §870.1025 | MHX | Monitor, physiological, Patient (with Arrhythmia Detection or Alarms) |
| <b>Subsequent Product Codes</b> | | |
| §870.2300 | MSX | System, Network and Communication, Physiological Monitors |
| §870.2300 | DRT | Monitor, Cardiac (incl. Cardio-tachometer & Rate Alarm) |
| §870.2700 | DQA | Oximeter |
| §870.1110 | DSK | Computer, Blood Pressure |
| §870.1025 | MLD | Monitor, ST Segment with Alarm |
| §870.1130 | DXN | System, Measurement, Blood-Pressure, Non-Invasive |
# K203088 Page 1 of 9
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue and are slightly larger than the rest of the letters. The word is centered and takes up most of the image.
| Table 5.1: Device Classification / Pro Code / Common Name | | |
|-----------------------------------------------------------|----------|-------------------------------------------------------|
| Classification | Pro Code | Common Name |
| §870.2810 | DSF | Recorder, Paper Chart |
| §870.2900 | DSA | Cable, Transducer and Electrode, Patient |
| §820.2770 | DSB | Plethysmograph, Impedance |
| §870.1230 | DQE | Catheter, Oximeter, Fiber-Optic |
| §870.1435 | DXG | Computer, Diagnostic, Pre-Programmed, Single-Function |
| §868.2375 | BZQ | Monitor, Breathing Frequency |
| §868.2377 | FLS | Monitor, Apnea, Facility Use |
| §868.1400 | CCK | Analyzer Gas, Carbon-Dioxide, Gaseous-Phase |
| §880.2910 | FLL | Thermometer, electronic, clinical |
| §870.1025 | DSI | detector and alarm, arrhythmia |
| §870.2340 | DPS | Electrocardiograph |
#### 3. Predicate Device:
Primary: K201764 - Infinity® Acute Care System™ Monitoring Solution Version
Secondary: K113798 - Infinity® Acute Care System™ Monitoring Solution Version VG2.0
The predicate devices have not been the subject of a design-related recall.
#### 4. Device Description:
The Infinity® Acute Care System™ (IACS) Monitoring Solution is a multi-parameter physiological patient monitoring system that acquires and displays patient data at the bedside. The IACS is a combination of the following devices: Infinity M540 Patient Monitor (Model #MS20401) with the Infinity M500 Docking Station (Model #MS20407) integrated with the Infinity Medical Cockpit (Model #MK31501 - 17" screen or Model MK31701 - 21" screen) and respective software, powered by the Infinity P2500 power supply (Model #MS22277).
The physiological patient data acquired from the interconnected components is displayed on the Infinity M540 Patient Monitor and is transmitted via network to the patient bedside Infinity
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color, and the background is white. The two dots above the "a" are also dark blue.
Medical Cockpit. The Infinity M540 Patient Monitor is a lightweight hand-held patient monitor that displays real-time vital signs, provides continuous trending and produces visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and transmits the data over the network. The Infinity M540 Patient Monitor can be used as a stand-alone medical device without integration with the Infinity Medical Cockpits.
Associated accessories may include:
- •Infinity MCable Masimo SET Model #MS20667
- •Infinity MCable Masimo SET Rainbow Model #MS27900 / MS27003
•Infinity MCable – Nellcor OxiMax – Model #MS20668
- · Infinity MPod Quad Hemo Model #MS20725
- · Infinity MCable Dual Hemo Model #MS20783
- ·Infinity MCable Mainstream CO2 Model #6871950
•Infinity MCable - Analog/Sync - Model #MS20662
•Infinity MCable – Nurse Call – Model #8417370
·PS120 Desktop Power Supply - Model #2606270
- · Y-Adapter (for PS120) Model #MS29702
The IACS, Monitoring Solution is a software driven device that is supplied and used non-sterile in a hospital environment by trained personnel. The IACS, Monitoring Solution is capital equipment.
#### 5. Intended Use/Indications for Use:
## Infinity® Acute Care System™ (IACS) Intended Use
The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patients in environments where patient care is provided by trained health care professionals.
The IACS obtains the physiologic, multi-parameter data from the connection to the M540 monitor and optional medical devices and displays. The real-time physiologic and multi-parameter data collected from the IACS is made available on the Infinity network.
The IACS and optional connected hardware are not intended for use in the following hospital environments:
- Hyperbaric chambers
- Environments containing MRI equipment
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The color of the text is a dark blue. The two dots above the "a" are also dark blue and are positioned closely together.
# 510(k) Summary
#### Indications for Use
The IACS, in connection with the M540 Patient Monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric patients only)
- 12-lead Rest ECG analysis (adult and pediatric patients only)
- ST segment analysis including TruST (adult and pediatric patients only)
- Apnea
- Impedance respiratory rate RRi
- Invasive blood pressure IBP
- Non-invasive blood pressure NIBP
- Temperature
- Cardiac output (adult and pediatric patients only)
- Carbon dioxide etCO2, inCO2, RRc
- Arterial oxygen saturation SpO2
- Pulse rate
- Perfusion index PI
- Total hemoglobin SpHb, (adult and pediatric patients only)
- Total oxygen content SpOC, (adult and pediatric patients only)
- Carboxyhemoglobin saturation SpCO (adult and pediatric patients only)
- Methemoglobin saturation SpMet
- Pleth variability index PVI
The IACS can connect to third-party devices to display physiologic, multi-parameter data and stored trends. It can also send information across the Infinity network.
# Infinity Medical Cockpits
#### Intended Use
The Infinity Medical Cockpits, consisting of the C500, and the C700 are monitoring and control displays for the Infinity Acute Care System (IACS).
Medical Cockpits are intended to be used to monitor waveforms, parameter information, and alarms as well as to control settings.
The Infinity Series Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare professionals.
#### Indications for Use
As the Infinity Medical Cockpits are used in conjunction with the Infinity Acute Care System and the M540 Patient Monitors they do not have their own indication for use.
## Infinity M540 Patient Monitor
#### Intended Use
The Infinity M540 (M540) is intended for the monitoring of multi-parameter, physiologic patient monitoring of adult, pediatric, and neonatal patient information in environments where patient care is provided by trained health care professionals.
The M540 obtains physiological data from connection to optional accessory devices. The real-time
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the two dots above the "a" are also blue. The background is white.
510(k) Summary
physiologic and multi-parameter data collected from the M540 is made available on the Infinity network.
The M540 is intended to monitor one patient at a time.
The M540 and any connected hardware are not intended for use in the following hospital environments:
- Hyperbaric chambers
- Environments containing MRI equipment
## Indications for Use
The M540 monitors the following parameters:
- Heart rate
- Arrhythmia (adult and pediatric patients only)
- 12-lead Rest ECG analysis (adult and pediatric patients only)
- ST segment analysis including TruST (adult and pediatric patients only)
- Apnea
- Impedance respiratory rate RRi
- Invasive blood pressure IBP
- Non-invasive blood pressure NIBP
- Temperature
- Cardiac output - only available when the M540 is docked in an IACS configuration (adult - and pediatric patients only)
- Carbon dioxide etCO2, inCO2, RRc
- Arterial oxygen saturation SpO2
- Pulse rate
- Perfusion index PI
- Total hemoglobin SpHb, (adult and pediatric patients only)
- Total oxygen content SpOC, (adult and pediatric patients only)
- Carboxyhemoglobin saturation SpCO (adult and pediatric patients only)
- Methemoglobin saturation SpMet
- Pleth variability index PVI
#### 6. Comparison of technological characteristics with the predicate device:
Both the subject and predicate devices use the same basic fundamental scientific technology and are
comprised of the same components which include the:
- Infinity C500 or C700 Medical Cockpit, -
- Infinity Communications Hub (power supply), -
- Infinity M500 Docking Station, -
- Infinity M540 Patient Monitor, and -
- various accessories.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The background is white.
The subject and predicate devices have the same technological characteristics pertaining to:
- Intended use: ●
- Patient categories O
- O Intended hospital environments
- Physiologic patient information obtained from connected hardware o
- Device connection to an R50N recorder, either directly or via the Infinity Network. O
- Indications for use: ●
- Measured parameters: O
- Heart Rate, Arrhythmia, ■
- ST Segment Analysis,
- 12-Lead ST Segment Analysis, ■
- Apnea,
- . Respiration Rate,
- Invasive Blood Pressure,
- Non-Invasive Blood Pressure,
- Temperature.
- Cardiac Output,
- Arterial Oxygen Saturation,
- . Pulse Rate,
- . Perfusion Index.
- 트 Mainstream End Tidal CO2,
- Total Hemoglobin,
- Total Oxygen Content,
- . Carboxyhemoglobin Saturation,
- Methemoglobin saturation,
- . Pleth Varibility Index
- Physical characteristics: ●
I
- Cockpit & M540 display, weight, dimensions, cooling, power, alarm indicator light O
- Battery indicator, AC Power LED, Navigation O
- User Interface: ●
- O Waveform area, parameter boxes-displays of physiological parameters
- ECG heartrate leads & measuring range O
- Respiration leads, measuring method and measuring range, detection threshold O
- O Arrhythmia types
- Invasive blood pressure measuring method and range, display, frequency O
- Non-invasive blood pressure measuring method, display O
- Temperature measuring range O
- QRS detection O
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The background is white.
The following characteristics were added to the subject device in VG4.1.1:
Software changes:
- Improvements: O
■
■
- Added header bar touch functionality, graphical volume indicator and reset . tone to the Medical Cockpit
- 트 Configurable QRS detection threshold
- Connectivity to certain 3rd party ventilators and anesthesia machines
- ST algorithm improvements
- ST algorithm modified decimation filter - M540
- New System Settings & Patient Profile Settings Medical Cockpit
- . Target bed automatic discharge before network transfer complete - Medical Cockpit
- Configurable buttons & increased options - Medical Cockpit
- Alarm for units of measure mismatch when M540 is undocked from the Medical Cockpit
- Battery alarms recorded in alarm history - Medical Cockpit
- Multicast configurable option - Medical Cockpit
- Configurable clinical password - M540
- . M540 alarm volume escalation in stand-alone configuration, docked & not connected to a Central Station ■
- Storage of arrhythmia events
- IEC 60601-1-8 Third edition related changes: O
- Audible alarms speaker volume increase on Cockpit & M540 ■
- Audio alarms set during audio pause
- Configurable Alarms
- Updated banner colors & alarms
- . New IEC tones
- Transducer failure detection
- Quiet mode on/off
- Data storage - Alarm Volume and Store Events configured for both the Masimo & Nellcor Sensor messages
- Configurable low battery alarm - M540
- Alarm volume controls - M540
- Defect corrections: O
- Non-invasive blood pressure intervals adjusted to actual time and setting of 5 ■ minutes or more
- . Real-time clock error handling - Medical Cockpit
- . Display of ECG Artifact Message w/ESU Artifacts
- Y-Adapter Cable for M540 in stand-alone configuration O
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue. The word is the logo for the Dräger company.
# The following technological differences exist between the subject and predicate devices:
Software changes:
- Defect corrections o
- Wrong data output scale for the ECG network is sent to the Infinity Network O
- o Waveform template for the ECG algorithm
- New instructions for correct ECG pacer sensing O
- Change of the etCO2 adapter during sensor warm up O
- Duplicate IP function and alarms in different network use cases o
- M540 not joining the network while roaming O
- O Production of unusual amount of Infinity protocol traffic
- Network data package handling on big loads O
- Incorrect detection of internal hardware failure O
- Synchronization between system components in specific workflows O
- Device handling of malformed Infinity network messages O
- Export protocol disconnections O
- Device resiliency for internal error reading and transmitting files in the memory of o the device
- O Improvements:
- M540 Distributed Denial of Service packet storm resiliency о
- Added security for the FTP connection between components of the system O
- Closing unused ports and services on network connection O
#### 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility:
The Infinity® Acute Care System™ Monitoring Solution and its components are not intended to
directly contact the patient during use. If patient contact is made it is transient with intact skin.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on devices equivalent to the IACS VG4.0.3 and M540 VG4.1.1 devices. The system complies with:
- IEC 60601-1 ●
- IEC 60601-1-2 ●
- IEC 60601-1-8
- IEC 606001-2-25
- IEC 60601-2-27
- ISO 80601-2-55 .
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is a dark blue color. The two dots above the "a" are also dark blue.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDAs Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Major Level of Concern", since the device provides diagnostic information that directly drives a decision regarding treatment or therapy, such that if misapplied it could result in serious injury or death. Additionally, the device provides vital signs monitoring and alarms for potentially life-threatening situations in which medical intervention is necessary.
## Performance Bench Testing
Verification testing was conducted to confirm the new requirements that were added to the IACS system, Medical Cockpit or M540 Patient Monitor based on the differences between the subject and predicate devices as listed above. EC57 testing was conducted for performance of cardiac rhythm and ST segment measurement algorithms.
Human Factors/Usability Engineering Validation testing was conducted regarding the changes made to the Instructions for Use relevant to the device modifications since the IACS VG2 devices were cleared.
#### Animal Studies
Animal performance testing was not performed or required for the device modifications subject of this 510(k).
## Clinical Studies
Clinical studies were not performed or required for the device modifications subject of this 510(k).
#### 8. Conclusion
The predicate devices were cleared based on results which did not include clinical studies. The non-clinical data supports the substantial equivalence of the device and the hardware and software verification and validation demonstrate that the subject device, IACS VG4.2, should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.