The Quantum Perfusion Arterial Cannula Graft is designed to the patient through an arterial vessel during cardiopulmonary bypass procedures for periods of maximum 6 hours. The device is equipped with an additional valved access which allows accessing arterial circulation.
Device Story
Standalone single-use arterial cannula graft; used during cardiopulmonary bypass procedures; sutured to arterial vessel to infuse blood from extracorporeal circuit into patient vascular district. Features 3/8" barbed connection and side branch with hemostatic valve; allows additional access for interventional catheterization; enables single-site access; reduces invasiveness. Operated by clinicians in medical settings. Benefits patient by facilitating extracorporeal life support and simplifying interventional access.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing, including biocompatibility, mechanical integrity, pressure drop, valve leakage, suture retention, kinking resistance, and sterilization validation.
Technological Characteristics
Single-use arterial cannula graft with 3/8" barbed connection and hemostatic valve. Sterilized by ethylene oxide. Complies with ISO/DIS 18193, ISO 7198:2016, ISO 10555-1:2013, ISO 11135:2014, ISO 11607-1:2019, and ISO 10993-1:2018.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures requiring arterial vessel access for a maximum duration of 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2021
Qura S.r.1. Raffaella Tommasini QA & RA Director Via di Mezzo 23 Mirandola, Modena 41037 Italy
Re: K203057
Trade/Device Name: Quantum Perfusion Arterial Cannula Graft Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: June 1, 2021 Received: June 3, 2021
Dear Raffaella Tommasini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Nicole Gillette Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203057
Device Name Quantum Perfusion Arterial Cannula Graft
### Indications for Use (Describe)
The Quantum Perfusion Arterial Cannula Graft is designed to the patient through an arterial vessel during cardiopulmonary bypass procedures for periods of maximum 6 hours. The device is equipped with an additional valved access which allows accessing arterial circulation.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a circular design. To the right of the figure, the word "QURA" is written in a bold, sans-serif font. Below "QURA", the words "advanced medical solutions" are written in a smaller, lighter font.
Traditional 510(k) Quantum Perfusion Arterial Cannula Graft 510(k) Summary
### 510(K) SUMMARY Section 5.0
### l. SUBMITTER
| Submitter Name: | Qura S.r.l. |
|------------------------|---------------------------------------------------|
| Submitter Address: | Via di Mezzo, 23 41037 Mirandola (MO) Italy |
| Contact Person: | Raffaella Tommasini, QA&RA Director – Qura s.r.l. |
| Phone: | +39 0535 1803050 |
| e-mail: | raffaella.tommasini@quramed.com |
| Fax: | +39 0535 1803051 |
| Date Summary Prepared: | June 1, 2021 |
#### II. DEVICE
| Proprietary Name: | Quantum Perfusion Arterial Cannula Graft |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Arterial Cannula Graft |
| Classification Name: | Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass |
| Regulatory Class: | II |
| Product Code: | DWF |
| Panel: | Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of<br>Health Technology 2 B (Circulatory Support, Structural and Vascular Devices) |
#### PREDICATE DEVICE lll.
| Trade Name: | VASCUTEK CANNULA GRAFT |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registered Establishment Name: | VASCUTEK LTD. |
| Classification Name: | Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass |
| Regulatory Class: | II |
| Product Code: | DWF |
| Panel: | Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of<br>Health Technology 2 B (Circulatory Support, Structural and Vascular Devices) |
| 510(k) Clearance: | K081560 |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image contains the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a circular design. To the right of the figure, the word "QURA" is written in bold, blue letters, with the words "advanced medical solutions" written in a smaller font underneath.
Traditional 510(k) Quantum Perfusion Arterial Cannula Graft 510(k) Summary
#### DEVICE DESCRIPTION IV.
The Quantum Perfusion Arterial Cannula Graft is a standalone single use device intended to be used in medical procedures requiring extracorporeal life support. It is designed to infuse blood coming from extracorporeal circuit into the patient's vascular district while being sutured to an arterial vessel.
The device also consists of a 3/8″ barbed connection with a side branch containing a hemostatic valve. Due to this valve, the Quantum Perfusion Arterial Cannula Graft can also act as an additional access for interventional catheterization. This feature allows the clinician to perform a single site access during medical procedures, avoiding multiple site access and thus reducing invasiveness.
The application duration of the Quantum Perfusion Arterial Cannula Graft is 6 hours.
The device is non-toxic, non-pyrogenic, sterilized by ethylene oxide and packaged in a double pouch.
#### v. INTENDED USE / INDICATIONS FOR USE
The Quantum Perfusion Arterial Cannula Graft is designed to the patient through an arterial vessel during cardiopulmonary bypass procedures for periods of maximum 6 hours. The device is equipped with an additional valved access which allows accessing arterial circulation.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Quantum Perfusion Arterial Cannula Graft has the same intended use, technological characteristics and operating principle as the predicate device (VASCUTEK CANNULA GRAFT, K081560).
In-vitro performance tests have been performed in order to support claimed substantial equivalence determining that proposed devices do not raise any new issues in terms of product's safety or effectiveness if compared to currently cleared predicate products.
Quantum Perfusion Arterial Cannula Graft can be therefore considered as substantially equivalent to predicate device, according to FDA's Guidance "Evaluating Substantial Equivalence in Premarket Notifications [510(k)]," issued July 28, 2014.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a partial circle. To the right of the figure is the word "QURA" in bold, blue letters. Below "QURA" are the words "advanced medical solutions" in a smaller, lighter font.
Traditional 510(k) Quantum Perfusion Arterial Cannula Graft 510(k) Summary
#### VII. PERFORMANCE DATA
### NON-CLINICAL TESTING
In-vitro testing was performed to demonstrate product substantial equivalence with the predicate device and also to comply with user needs and safety and effectiveness requirements. Testing supplied in the 510(k) Notification includes biocompatibility evaluation, mechanical and performance verification, labeling and Instructions for Use (IFU), verification and validation tests.
All testing passed by meeting the established requirements set for the use of Quantum Perfusion Arterial Cannula Graft. The following data were provided:
- Evaluation of devices' performances: -
- O Device pressure drop;
- o Valve backflow and leakage;
- o Graft Water Entry Pressure;
- O Suture retention strength;
- O Dynamic blood damage;
- O Kinking resistance;
- o Mechanical integrity;
and mainly performed according to ISO/DIS 18193, ISO 7198:2016 [Recognition Nr.: 3-144] and ISO 10555-1:2013 + AMD 2017 [Recognition Nr.: 6-408];
- -Evaluation of product shelf life, including product's sterility according to EP/UPS requirements;
- -Validation of the EtO Sterilization process, according to ISO 11135:2014 [Recognition Nr.: 14-452],
- Packaging Validation tests according to ISO 11607-1:2019 [Recognition Nr.: 14-530]; -
- -Biocompatibility of the finished product, according to International Standard ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
### Animal Study
No animal studies have been performed except for mandatory biocompatibility tests according to International Standard ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
### CLINICAL TESTING
No clinical data have been included in the current Traditional 510(k) submission to support substantial equivalence to legally marketed predicate device.
{6}------------------------------------------------
K203057/S001
Quantum Perfusion Arterial Cannula Graft
Image /page/6/Picture/2 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a circular shape. To the right of the figure is the word "QURA" in a bold, sans-serif font, and below that is the text "advanced medical solutions" in a smaller font.
Traditional 510(k) Quantum Perfusion Arterial Cannula Graft 510(k) Summary
#### CONCLUSIONS VIII.
Based on the indications for use, technological characteristics and results of non-clinical testing, the Quantum Perfusion Arterial Cannula Graft has been demonstrated to its intended use and is considered substantially equivalent to claimed predicate VASCUTEK CANNULA GRAFT, cleared by K081560.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.