K203019 · Daesung Maref Co., Ltd. · IRP · Jun 22, 2021 · Physical Medicine
Device Facts
Record ID
K203019
Device Name
LF900
Applicant
Daesung Maref Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Jun 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.
Device Story
Intermittent pneumatic compression system; improves blood circulation. Device delivers air to 4-chamber sleeve; pressurizes chambers sequentially (1st to 4th). Used in professional healthcare and home environments; operated by medical professionals or patients. Output consists of sequential pneumatic compression. Healthcare providers use output to treat edema and venous conditions; benefits include improved circulation and symptom management. Device must be used over clothing/socks; never direct skin contact.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), and performance verification (pressure, time, sleeve durability).
Indicated for patients with primary lymphedema, edema following trauma/sports injuries, post-immobilization edema, and venous insufficiencies. Intended for use by medical professionals or patients at home under medical supervision.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Compressible Limb Therapy System (Power-Q1000 Premium)(K113275)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
June 22, 2021
DaeSung Maref Co., Ltd So Hyeon Assistant Researcher 298-24. Gongdan-ro Gunpo-si, 15809 Korea
Re: K203019
Trade/Device Name: LF900 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: March 8, 2021 Received: March 29, 2021
Dear So Hyeon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203019
Device Name LF900
Indications for Use (Describe)
LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|--|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 1. Data Prepared [21 CFR 807.92(a)(a)]
March 8, 2021
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer :
- DAESUNG MAREF CO., LTD.
- Address :
298-24, Gongdan-ro Gunpo-si, Gyeonggido Republic of Korea
- Contact Name :
- Su Hyeon, So
- Telephone No. :
- 82-31-459-7211
- Fax No. :
- 82-31-459-7215
- Email Adress :
rndra@dsmaref.com
- Registration No. : 3004116008
## 3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
| Trade / Device Name | LF900 |
|---------------------|-----------------------------------|
| Regulation Number | 21 CFR 890.5650 |
| Regulation Name | Massager, Powered Inflatable Tube |
| Regulation Class | II |
| Product Code | IRP |
## 1
/
{4}------------------------------------------------
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
Predicate Device
- 510(k) Number :
- K113275
- Applicant :
BSR KOREA Corp.(K113275)
- Trade / Device Name :
Compressible Limb Therapy System (Power-Q1000 Premium)(K113275)
- Regulation Number :
21 CFR 890.5650
- Regulation Name :
Massager, Powered Inflatable Tube
- Regulation Class:
II
- Product Code:
- IRP
Predicate device has not been subject to a design-related recall.
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
This product is an Intermittent Pneumatic Compression system to improve the blood circulation of the human body.
The operating principle is that the air from the device will be delivered to the sleeve with 4 air chambers and the air pressurizes sequentially the chambers from 1st to 4th.
### 6. Indications For Use [21 CFR 807(a)(5)]
LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.
> 5 2 /
{5}------------------------------------------------
## 7. Determination of Substantial Equivalence
Summary of technological characteristics of the device compared to the predicate device. [21CFR 807.92(a)(6)]
The LF900 is substantially equivalent to legally marketed predicate device (POWER-Q1000 Premium) with respect to indications for use and technology characteristics. The table below presents comparisons for device :
| | Proposed Device | Predicate Device |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Name | LF900 | POWER-Q1000 Premium |
| 510(k) Number | K203019 | K113275 |
| Manufacturer | DAESUNG MAREF CO., LTD. | WONJIN MULSAN Co., Ltd. |
| Product Code | IRP | IRP |
| Device Class | II | II |
| Regulation<br>Number | 21 CFR 890.5650 | 21 CFR 890.5650 |
| Regulation<br>Name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube |
| Indications For<br>Use | LF900 is intended for use by<br>medical professionals and<br>patients at home, who are under<br>medical supervision, in treating<br>many conditions, such as:<br>Primary lymphedema, Edema<br>following trauma and sport<br>injuries, Postimmobilization<br>edema, Venous insufficiencies,<br>Lymphedema. | The device is indicated for use by<br>medical professionals and patient at<br>home, who are under medical<br>supervision, in treating many<br>conditions, such as: Primary<br>lymphedema, edema following<br>trauma and sport injures, Post-<br>immobilization edema, Venous<br>insufficiencies, Lymphedema |
| Intended Use<br>environment | Professional healthcare environment<br>&<br>Home environment | Professional healthcare environment<br>&<br>Home environment |
| OTC or Rx only | Rx only | Rx only |
### [Table 1. Comparison of Proposed Device to Predicate Device]
5 /
3
{6}------------------------------------------------
| | Proposed Device | Predicate Device |
|----------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Accessories | Leg sleeves<br>Extension zipper<br>Single tubing<br>Double tubing | Leg Cuffs<br>Single connectinig hose<br>Double connectinig hose<br>Extension zipper for the leg cuffs<br>(option) |
| Specifications | | |
| Power Source | 100-127V~, 50/60Hz | AC 120V, 50/60Hz |
| Time | 5-90 min | 0-99 min |
| Pressure | $10-180mmHg \pm 20 mmHg$ | $0-240mmHg \pm 20 %$ |
| Operation Mode | A,B,C,D | A,B,C,D,E |
| Number of<br>chamber | 4 | 4 |
The table also provides rationale for a little difference in support of substantial equivalence to the Predicate devices.
#### [Table 2. Little differnence with Predicate Device]
#### Justification to Support Substantial Equivalence
Proposed device (LF900) and Predicate device (POWER-Q1000 Premium) can be considered to be almost equivalent except for pressure range, time range, operation mode.
The time range and pressure range of the proposed device are included within the time range of the predicate device.
The operation mode simply differs in the seqeuence of pressurization, the principle of operation and usage are the same. And the pressure is also lower than that of the predicate device & Reference device.
Therefore, the differences in technological characteristics do not raise different questions of safety and effectiveness.
ব
{7}------------------------------------------------
#### Non-Clinical Test Summary
The LF900 complies with voluntary standards for electrical safety, electromagnetic compatibility, use in the home healthcare environment and usability.
The following data were provided in support of the substantial equivalence determination :
- 1) Electrical Safety, Electromagnetic Compatibility and Performance
The LF900 complies with the electrical safety and electromagnetic compatibility requirements established by the standards.
• IEC 60601 : 2005/A1:2012, Medical Electrical Equipment:Part 1: General Requirements for Basic Safety and Essential Performance
- · IEC 60601 60601-1-2:2014, Medical Electrical Equipment Part 1 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests
- IEC 60601 60601-1-11:2015, Medical Electrical Equipment Part 1 1-11: General Requirements for Basic Safety and Essential Per formance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
#### 2) Device Performance test
The proposed device is at a safer level in terms of pressure and time compared to equivalent devices, Safety and effectiveness were confirmed by conducting verification tests for these essential performances (pressure, time, and sleeve durability).
Therefore, it can be confirmed that the proposed device (K203019) is substantially equivalent to the equivalent device (K113275).
#### Clinical Test Summary
Clinical testing was not required to demonstrate the substantial equivalence of the LF900 to its predicate device.
The device should only be used over full clothing and socks (because the cuffs cover the feet), never on direct skin.
The biocompatibility of the materials has not been verified by the FDA and contact of the cuffs/accessories to direct skin may lead to skin irritation, skin sensitization and/or cytotoxicity.
#### 8. Conclusion [21 CFR 807.92(b)(3)]
The LF900 has same intended use and technical characteristics to the predicate device.
Based on that information, we conclude that the differences between the proposed device and predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness.
> / ર
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.